K161697 · Acclarent, Inc. · PGW · Oct 24, 2016 · Neurology
Device Facts
Record ID
K161697
Device Name
Acclarent NavWire Sinus Navigation Guidewire
Applicant
Acclarent, Inc.
Product Code
PGW · Neurology
Decision Date
Oct 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Acclarent Nav Wire is intended for use as a navigable guidewire to provide access to nasal and sinus spaces, and confirmation of placement in the accessed anatomy. The Acclarent NavWire is designed for use during procedures where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, can be identified relative to a CT-based model of the anatomy.
Device Story
Electromagnetically (EM) trackable guidewire; provides access to nasal and paranasal sinus spaces during endoscopic/image-guided surgery. Device integrates with CARTO ENT System; tracks distal tip position within EM field. System displays sensor position relative to patient's pre-operative CT scan. Used by surgeons in clinical settings; provides real-time navigation and confirmation of placement in tortuous anatomy. Enhances surgical precision by correlating instrument location with patient-specific CT models; facilitates navigation of complex sinus structures.
Clinical Evidence
Bench testing only. Testing included nonclinical accuracy and performance testing with the CARTO ENT System, functionality testing after accelerated aging and user conditions, simulated use testing, and sterilization verification (EtO, SAL 10^-6). Device met all pre-defined design and performance specifications.
Technological Characteristics
Electromagnetically (EM) trackable guidewire; long flexible form factor with distal EM sensor. Compatible with CARTO ENT System for electromagnetic field tracking. Sterilization via Ethylene Oxide (EtO) to SAL 10^-6. Connectivity via integration with surgical navigation system software.
Indications for Use
Indicated for use as a navigable guidewire to provide access to nasal and sinus spaces and confirm placement in patients undergoing ENT surgery where anatomical structures can be referenced against a CT-based model.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
Fiagon Navigation Extended Instrument Set ENT (K141456)
Relieva Luma Sentry Sinus Illumination System (cleared as part of K111875)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a human figure, represented by three overlapping profiles facing to the right. The profiles are stylized and appear to be made of flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 24, 2016
Acclarent, Inc. Ms. Anna Hwang Regulatory Affairs Specialist 33 Technology Drive Irvine, CA 92618
Re: K161697
Trade/Device Name: Acclarent Navwire Sinus Navigation Guidewire Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: September 21, 2016 Received: September 22, 2016
Dear Ms. Hwang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K161697
Device Name
Acclarent® Nav Wire Sinus Navigation Guidewire
Indications for Use (Describe)
The Acclarent Nav Wire is intended for use as a navigable guidewire to provide access to nasal and sinus spaces, and confirmation of placement in the accessed anatomy. The Acclarent NavWire is designed for use during procedures where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, can be identified relative to a CT-based model of the anatomy.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Acclarent" in a stylized font. The first part of the word, "Accla," is in black, while the second part, "rent," is in purple. A curved line extends from the top of the "r" in "rent" over the rest of the word, giving it a distinctive appearance.
Acclarent® NavWire Sinus Navigation Guidewire
Traditional 510(k) Premarket Notification CONFIDENTIAL
# APPENDIX A: 510(K) SUMMARY
| Sponsor/Submitter: | Acclarent, Inc.<br>33 Technology Drive<br>Irvine, CA 92618 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Anna Hwang<br>Regulatory Affairs Specialist<br>Email: ahwang@its.jnj.com<br>Telephone: 949-923-4134<br>Mobile: 310-780-8792 |
| Date of Submission: | June 17, 2016 |
| Device Trade Name: | Acclarent® NavWire Sinus Navigation Guidewire |
| Common Name: | Image Guided Surgery System |
| Device Classification: | Class II |
| Regulation Number: | 21 CFR 882.4560 |
| Classification Name: | Ear, Nose, and Throat Stereotaxic Instrument |
| Product Code: | PGW |
| Predicate Device: | The Fiagon Navigation Extended Instrument Set ENT (K141456)<br>Relieva Spin Sinus Dilation System (K111875) |
| Device Description: | The Acclarent® NavWire Sinus Navigation Guidewire an<br>electromagnetically (EM) trackable guidewire that provides surgeons a<br>means to access the intranasal and paranasal sinus spaces in endoscopic<br>and image guided sinus surgeries. It is compatible with the CARTO®<br>ENT System to track the NavWire within an electromagnetic field,<br>which displays the magnetic sensor position relative to a patient's pre-<br>operative CT scan. |
| Indications for Use: | The Acclarent® NavWire Sinus Navigation Guidewire is intended for<br>use as a navigable guidewire to provide access to nasal and sinus<br>spaces, and confirmation of placement in the accessed anatomy. The<br>Acclarent® NavWire Sinus Navigation Guidewire is designed for use<br>during procedures where reference to a rigid anatomical structure in the<br>field of ENT surgery, such as the paranasal sinuses, can be identified<br>relative to a CT-based model of the anatomy. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Acclarent. The word "Acclarent" is written in a sans-serif font, with the first part of the word in black and the "ent" in purple. There is a purple swoosh above the "ent" portion of the word.
Acclarent® NavWire Sinus Navigation Guidewire
Traditional 510(k) Premarket Notification CONFIDENTIAL
conjunction with a compatible surgical navigation system which consists of computer-aided software, CT-imaging, patient tracker, registration probe, and various instruments used in sinus surgery. Acclarent® NavWire Sinus Navigation Guidewire provides real-time tracking at the distal tip of the guidewire in the nasal anatomy. Performance Data: Bench testing was performed in order to determine device accuracy and precision. The following nonclinical tests were performed to determine substantial equivalence: Nonclinical accuracy testing of the Acclarent® NavWire Sinus . Navigation Guidewire with the CARTO® ENT System Nonclinical performance testing of the Acclarent® NavWire ● Sinus Navigation Guidewire with the CARTO® ENT System Functionality of the Acclarent® NavWire Sinus Navigation ● Guidewire after accelerated aging and user conditions . Simulated use testing of the NavWire with the CARTO® ENT System according to user's needs and intended use . Sterilization verification to effectively EtO sterilize the NavWire (SAL 106) and EtO residuals are within defined limits General requirements and performance testing of Acclarent® ● NavWire Sinus Navigation Guidewire to assure conformance with identified design and performance specifications The Acclarent NavWire passed all intended criteria in accordance with appropriate test criteria and standards. The Acclarent® NavWire Sinus Navigation Guidewire is substantially Summary of Substantial Equivalence: equivalent to the predicate device Fiagon Navigation Extended Instrument Set ENT (K141456) and reference device Relieva Luma Sentry Sinus Illumination System, cleared as part of the Relieva Spin Sinus Dilation System (K1118755). Like the predicate device, the Acclarent® NavWire Sinus Navigation Guidewire is an electromagnetically trackable guidewire which can be used to navigate tortuous sinus anatomy when used in conjunction with an ENT software application for computer assisted surgery. NavWire, like the predicate and reference devices, consists of a long flexible guidewire with a distal tip used to provide access to nasal and sinus spaces, and confirmation of placement in the accessed anatomy. The proposed navigation wire is similar in connectivity, technological characteristics, intended use, and design as the predicate device. The primary differences between the Acclarent® NavWire Sinus Navigation Guidewire and predicate device do not raise new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.