THE FIAGON NAVIGATION - EXTENDED INSTRUMENT SET ENT, REGISTRATIONPOINTER, VENTERAPOINTER, GUIDEWIRE, POINTERSHELL
Applicant
Fiagon GmbH
Product Code
PGW · Neurology
Decision Date
Oct 8, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Fiagon Navigation – Extended Instrument Set ENT is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation. The Fiagon Navigation - Extended Instrument Set ENT is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to: ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. Skull base procedures for ENT access
Device Story
Extended Instrument Set ENT consists of reusable, electromagnetically navigated surgical instruments (suction tubes, pointers, registration probes, and adapters) for use with the Fiagon Navigation system. Instruments feature integrated sensors at the tip, tracked within a low-energy magnetic field generated by the navigation system. The system overlays instrument position onto preoperative CT/MRI scans. Used in ENT procedures (e.g., sinus surgery, skull base access) by surgeons to precisely locate anatomical structures. Malleable suction instruments allow for simultaneous suction and navigation. Instrument adapters enable tracking of standard surgical tools. Output is visual guidance on the navigation display, assisting surgeons in navigating complex anatomy, potentially improving surgical precision and reducing risk of injury to critical structures.
Clinical Evidence
Bench testing only. Evaluated device precision and accuracy. Mean bench accuracy measured at 0.7 mm - 1.2 mm (SD 0.29 mm - 0.42 mm). 95% confidence levels were < 2 mm, comparing favorably to predicate values (0.9 mm mean, < 3 mm 95% confidence).
Technological Characteristics
Reusable, electromagnetically tracked surgical instruments. Includes suction tubes, pointers, and adapters. Sensors integrated at instrument tips. Operates via low-energy magnetic field generator. Compatible with CT/MRI-based navigation software. Materials are user-sterilizable. Dimensions vary by component type.
Indications for Use
Indicated for patients undergoing ENT procedures where stereotactic surgery is appropriate and rigid anatomical structures (e.g., paranasal sinuses, mastoid anatomy) can be referenced to CT or MR imaging. Applicable for open or percutaneous procedures including transphenoidal, intranasal, sinus, and skull base surgeries.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 8, 2014
Fiagon GmbH % Dr. Dirk Mucha Manager, Regulatory Affairs Neuendorfstrasse 23b 16761 Hennigsdorf, Germany
Re: K141456
Trade/Device Name: Fiagon Navigation - Extended Instrument Set ENT Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: September 5, 2014 Received: September 10, 2014
Dear Dr. Mucha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known): K141456
Device Name: Fiagon Navigation – Extended Instrument Set ENT
Indications for Use:
The Fiagon Navigation – Extended Instrument Set ENT is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation. The Fiagon Navigation - Extended Instrument Set ENT is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. Skull base procedures for ENT access
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
And / Or
Over-The-Counter-Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) Summary
### May 28, 2014
#### 1. Submitter Information
| Submitter: | Fiagon GmbH |
|------------|----------------------------|
| Address: | Neuendorfstrasse 23b |
| | 16761 Hennigsdorf, Germany |
| <i>Telephone:</i> | +49 3302 201 21 10 |
|-------------------|--------------------|
| <i>Telefax:</i> | +49 3302 201 21 15 |
Contact: Mr. Dirk Mucha, Manager Regulatory Affairs
#### Device Information 2.
| Trade Name: | Fiagon Navigation - Extended Instrument Set ENT |
|-----------------|-------------------------------------------------|
| Common Name: | Image guided surgery system |
| Classification: | Class II per 21 CFR 882.4560 |
| Device: | Ear, Nose, and Throat Stereotaxic Instrument |
| Product Code: | PGW |
#### 3. Purpose of Submission
The purpose of this submission is to gain clearance for a new set of instruments for the Image Guided Surgery System, Fiagon Navigation system (K133573).
#### 4. Predicate Device Information
The Fiagon Navigation System described in this submission is substantially equivalent to the following predicates:
| | Predicate Device | Manufacturer | 510(k) No. |
|---|--------------------------------|-----------------------------------|------------|
| 1 | Fiagon Navigation System | Fiagon GmbH | K133573 |
| 2 | Malleable Suction™ Instruments | Medtronic<br>Navigation, Inc. | K133665 |
| 3 | Instatrak® | Visualization<br>Technology, Inc. | K960330 |
{4}------------------------------------------------
### 5. Device Description
The Fiagon Navigation - Extended Instrument Set ENT are reusable instruments intended to be used with the Fiagon Navigation system. The instruments in the Set are electromagnetically navigated devices that are
- Navigated suction instruments (malleable, designed to be bendable by hand, sensor a. within the tip, attachable to standard surgical vacuum suction systems)
- b. Navigated pointing devices (flexible, sensor within the tip)
- c. Registration probe (designed for non.sterile patient registration)
- d. Instrument adaptors (designed for mechanical connection to cylindric schaped surgical instruments of different diameters)
Each device incorporates a sensor device which is tracked by the navigation system with in the low-energy magnetic field of a field generator (part of the navigation system)
The navigation software (part of the navigation system) displays the position of the instruments in preoperative scans (e.g., CT, MRI, fluoroscopy)
| Components | Grouping | Sterility State | Patient contact |
|--------------------------------|-------------------------------------------------------------------|----------------------------------------------------------|-------------------------------------------------------------------|
| FlexTube 3 mm | Suction and navigation<br>Instrument | user sterilized, reusable, delivered in non steril state | limited duration (<24h) External communicating Tissue/bone/dentin |
| FlexTube 4 mm | Suction and navigation<br>Instrument | user sterilized, reusable, delivered in non steril state | limited duration (<24h) External communicating Tissue/bone/dentin |
| PointerTube<br>Straight | Suction and navigation<br>Instrument | user sterilized, reusable delivered in non steril state | limited duration External communicating Tissue/bone/dentin |
| PointerTube<br>Stammberger | Suction and navigation<br>Instrument | user sterilized, reusable delivered in non steril state | limited duration External communicating Tissue/bone/dentin |
| PointerTube Sinus<br>Frontalis | Suction and navigation<br>Instrument | user sterilized, reusable delivered in non steril state | limited duration External communicating Tissue/bone/dentin |
| RegistrationPointer | Pointing Instrument<br>for patient registration<br>only, ball tip | reusable not to be sterilized outside sterile field | limited duration Surface communicating Skin |
### List of components
{5}------------------------------------------------
| Components | Grouping | Sterility State | Patient contact |
|-------------------|---------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| VenteraPointer | Pointing Instrument | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>(< 24h)<br>• External communicating<br>• Tissue/bone/dentin |
| GuideWire | Pointing Instrument | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>(< 24h)<br>• External communicating<br>• Tissue/bone/dentin |
| PointerShell 4 mm | Instrument adapter | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>• Surface communicating<br>• Skin |
| PointerShell 5 mm | Instrument adapter | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>• Surface communicating<br>• Skin |
| PointerShell 3 mm | Instrument adapter | • user sterilized,<br>• reusable,<br>• delivered in non steril<br>state | • limited duration<br>• Surface communicating<br>• Skin |
### 6. Intended Use
The Fiagon Navigation - Extended Instrument Set ENT is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation. The Fiagon Navigation - Extended Instrument Set ENT is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. Skull base procedures for ENT access.
{6}------------------------------------------------
### 7. Comparison of Technological Characteristics
The substantial equivalence of the Fiagon Navigation – Extended Instrument Set ENT to the predicates is shown by similarity in intended use, indications for use, materials, and performance. The Fiagon Navigation System and its predicates utilize:
- Electromagnetic tracking technology for navigation attached or included in surgical ● instruments (all predicates)
- Malleable suction instruments enabled for image guidance (Predicate 2)
- Adapters for Instruments for enabling image guidance (Predicate 3)
- . Instruments with tracking sensor at the instrument tip, precalibrated (Predicate 1 and 2)
The primary difference between the Fiagon Navigation - Extended Instrument Set ENT and it predicates Fiagon Navigation System is that the Fiagon Navigation - Extended Instrument Set ENT provides electromagnetic tracking of the same kind in instruments of different geometry. This different does not raise new issues of safety and effectiveness, with respect to the following:
Cannulated instruments have been demonstrated by predicate 2. Therefore, the suction instruments of the Extended Instrument Set ENT do not raise new issues of safety and effectiveness.
Instrument adapters have been demonstrated by predicate 3. Therefore, the instrument adapters of the Extended Instrument Set ENT do not raise new issues of safety and effectiveness
### 8. Performance Data
Testing was performed in order to determine device precision and accuracy. The following nonclinical tests were performed to determine substantial equivalence:
Bench testing was conducted to determine the device accuracy and the performance of the electromagnetic field distortion mechanism.
A mean bench accuracy of 0.7 mm - 1.2 mm (Standard deviation 0.29 mm - 0.42 mm) was measured for the different instrunents. All 95% confidence levels were < 2mm which compares to the values 0.9 mm (mean) resp. < 3 mm (95% confidence) reported for the predicate devices.
The results supports the claim of substantial equivalence to the predicate devices.
Testing was completed to ensure functionality and compatibility with the Fiagon Navigation system.
{7}------------------------------------------------
### 9. Conclusion
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicates, the Fiagon Navigation – Extended Instrument Set ENT has been shown to be substantially equivalent to the predicate devices identified in this submission, and does not present any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.