K161677 · Institut Straumann AG · NHA · Oct 7, 2016 · Dental
Device Facts
Record ID
K161677
Device Name
Straumann Sterile Healing Solution
Applicant
Institut Straumann AG
Product Code
NHA · Dental
Decision Date
Oct 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Healing abutments, healing caps and closure screws are intended for use with the implants of the Straumann Dental Implant System to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Fixation caps are used to stabilize bone grafts in cases where bone augmentation is being performed in conjunction with implant placement.
Device Story
Straumann Sterile Healing Solution consists of healing abutments, healing caps, closure screws, and fixation caps; used with Straumann Dental Implant System. Devices protect implant internal configuration; maintain/form soft tissue during healing; stabilize bone grafts during augmentation. Operated by dental professionals in clinical settings. Devices are commercially pure grade 4 titanium; provided sterile via gamma irradiation. Subject devices include one-piece solid and two-piece assemblies. Not intended for immediate loading protocols; not for use with granular bone substitutes or dental membranes. Clinical benefit involves facilitating successful implant integration and soft tissue healing.
Clinical Evidence
Bench testing only. No animal or human clinical studies conducted. Evidence includes sterilization validation (ISO 11137), biocompatibility assessment (ISO 10993), and shelf-life validation (ASTM F1980).
Technological Characteristics
Commercially pure grade 4 Titanium per ISO 5832-2. One-piece solid and two-piece assemblies. Provided sterile via gamma irradiation (SAL 10^-6) per ISO 11137-2. Packaging: PETG blister tray with Tyvek lid stock. Shelf life validated per ASTM F1980.
Indications for Use
Indicated for patients requiring dental implant healing components to protect implant internal features, manage soft tissue during healing, or stabilize bone grafts during augmentation procedures in conjunction with Straumann Dental Implant System implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and clean, and it is easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 7, 2016
Institut Straumann AG % Jennifer Jackson Director Of Regulatory Affairs And Quality Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K161677
Trade/Device Name: Straumann Sterile Healing Solution Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: September 5, 2016 Received: September 9, 2016
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161677
Device Name
Straumann® Sterile Healing Solution
Indications for Use (Describe)
Healing abutments, healing caps and closure screws are intended for use with the implants of the Straumann Dental Implant System to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Fixation caps are used to stabilize bone grafts in cases where bone augmentation is being performed in conjunction with implant placement.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 5. | 510(k) Summary | K161677 |
|----|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810 |
| | Contact Person: | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs and Quality<br>+1 (978) 747-2509 |
| | Prepared By: | Christopher Klaczyk<br>Head of North American Regulatory Affairs<br>Institut Straumann AG<br>+41 61 965 1260 |
| | Date Prepared: | October 7, 2016 |
| | Product Code(s): | NHA (21 CFR 872.3630) |
| | Device Class: | II (21 CFR 872.3630) |
| | Classification Panel: | Dental |
| | Classification Name: | Endosseous dental implant abutment (21 CFR 872.3630) |
| | Proprietary Name: | Straumann® Sterile Healing Solution |
| | Predicate Device:: | K130808, Straumann® Sterile Healing Solution (Institut<br>Straumann) |
| | Reference Device(s): | K062129, Bone Level Implants (Institut Straumann)<br>K120951, SMARTbuilder System (Osstem Implant Co.) |
| | Device Description: | The Straumann Healing Abutments, Healing Caps, and Closure<br>Screws are intended for use with the implants of the Straumann<br>Dental Implant System (SDIS) to protect the inner features of<br>the implant and to maintain, stabilize and form the soft tissue<br>during the healing process. The devices are available in various<br>diameters and heights. They are also not intended for use in<br>immediate loading protocols; none of the devices of the<br>Straumann Sterile Healing Solution are indicated for use in<br>immediate loading protocols as they are intended to protect the<br>internal features of the implant during the healing phase.<br>When the implant has sufficient primary stability, the closure |
| | | caps and fixation caps are used as a means to retain a bone graft |
{4}------------------------------------------------
used to augment the bone in the vicinity of the implant. These devices are not intended to be used with granular bone substitute materials or dental membranes.
The subject devices are identified in the table below.
| Article<br>Number | Description |
|-------------------|-----------------------------------------|
| 024.0000S | RC Healing Abutment, conical, Ø 4.5, H2 |
| 024.0001S | RC Healing Abutment, conical, Ø 4.5, H4 |
| 024.0002S | RC Healing Abutment, conical, Ø 4.5, H6 |
| 024.0003S | RC Healing Abutment, conical, Ø 6.0, H2 |
| 024.0004S | RC Healing Abutment, conical, Ø 6.0, H4 |
| 024.0005S | RC Healing Abutment, conical, Ø 6.0, H6 |
| 024.2100S | NC Closure Screw, H 0 |
| 024.2100S-04 | NC Closure Screw, H 0 (4-Pack) |
| 024.2105S | NC Closure Screw, H 0.5 |
| 024.2105S-04 | NC Closure Screw, H 0.5 (4-Pack) |
| 024.2220S | NC Closure & Fixation Cap, Ø 5.5mm, Ti |
| 024.4100S | RC Closure Screw, H 0 |
| 024.4100S-04 | RC Closure Screw, H 0 (4-Pack) |
| 024.4105S | RC Closure Screw, H 0.5 |
| 024.4105S-04 | RC Closure Screw, H 0.5 (4-Pack) |
| 024.4220S | RC Closure & Fixation Cap, Ø 5.5mm, Ti |
| 048.324S | NNC Closure Screw, H 0 |
| 048.324SV4 | NNC Closure Screw, H 0 (4-Pack) |
| 048.325S | NNC Closure Screw, H 0.5 |
| 048.325SV4 | NNC Closure Screw, H 0.5 (4-Pack) |
| 048.371SV4 | RN Closure Cap, H 0, Ti (4-Pack) |
| 048.373SV4 | RN Closure Cap, H 1.5 mm, Ti (4-Pack) |
Indications For Use: Healing abutments, healing caps and closure screws are intended for use with the implants of the Straumann Dental Implant System to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Fixation caps are used to stabilize bone grafts in cases where bone augmentation is being performed in conjunction with implant placement. Materials: The material for all Healing Abutments, Closure Caps and Closure Screws is commercially pure Titanium (grade 4)
conforming to ISO 5832-2.
{5}------------------------------------------------
## Technological Characteristics:
A comparison of the relevant technological characteristics between the subject and primary predicate devices is provided in the table that follows.
| Feature | Primary Predicate Device<br>Straumann Sterile Healing Solution | Subject Devices<br>Straumann Sterile Healing Solution |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications For Use<br>K130808 | Closure screws, healing caps, and healing<br>abutments, are intended for use with the<br>straumann dental implant system (sdis) to<br>protect the inner configuration of the<br>implant and maintain, stabilize and form<br>the soft tissue during the healing process.<br>Customizable healing abutments made of<br>peek are for use for up to six months. | Healing abutments, healing caps and<br>closure screws are intended for use with the<br>implants of the Straumann Dental Implant<br>System to protect the inner configuration of<br>the implant and maintain, stabilize and<br>form the soft tissue during the healing<br>process. Fixation caps are used to stabilize<br>bone grafts in cases where bone<br>augmentation is being performed in<br>conjunction with implant placement. |
| Indications For Use<br>K120951 | SMARTbuilder System is a metal (Non-<br>resorbable membrane) device intended for<br>use with a dental implant to stabilize and<br>support of bone graft in dento-alveolar<br>bony defect sites. | The Indications for Use differ with the<br>removal of the PEEK abutments. The<br>subject devices do not include PEEK<br>abutments and only include Titanium<br>abutments; therefore the modification of<br>the Indications for Use is acceptable as the<br>PEEK abutment limitation for 6 months<br>does not apply to the subject device |
| Compatible Implants | Straumann Bone Level implants having the<br>NC and RC implant-to-abutment interface<br>geometries.<br>Straumann Tissue Level implants having<br>the NNC, RN and WN implant-to-abutment<br>interface geometries. | Straumann Bone Level implants having the<br>NC and RC implant-to-abutment interface<br>geometries.<br>Straumann Tissue Level implants having<br>the NNC and RN implant-to-abutment<br>interface geometries. |
| Material | Commercially pure grade 4 Titanium<br>per ISO 5832-2 | Commercially pure grade 4 Titanium<br>per ISO 5832-2 |
| Construction | One-piece solid device. | One-piece solid devices and two-piece<br>device assemblies. Two-piece devices are<br>supported via comparison with components<br>of the Osstem SMARTbuilder System<br>(K120951) |
| Packaging | Primary barrier package consists of a<br>PETG blister tray with a PETG insert and a<br>Tyvek 1073B lid stock. The primary<br>package is contained within a paperboard<br>shelf box. | Primary barrier package consists of a<br>PETG blister tray with a PETG insert and a<br>Tyvek 1073B lid stock. The primary<br>package is contained within a paperboard<br>shelf box. |
| Sterility | Provided sterile via gamma irradiation to a<br>Sterility Assurance Level of 10-6.<br>Cycle validated per ISO 11137-2. | Provided sterile via gamma irradiation to a<br>Sterility Assurance Level of 10-6.<br>Cycle validated per ISO 11137-2. |
{6}------------------------------------------------
| Performance Data: | Test data to support the evaluation of the subject closure cap<br>and fixation cap devices has been included directly or by<br>reference as follows:<br>Sterilization validation in accordance with ISO 11137 series of standards. Biocompatibility assessment per the ISO 10993 series of standards. Shelf Life validation in accordance ASTM F1980 |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | No animal or human clinical studies were conducted. |
| Conclusions: | Based upon our assessment of the design and applicable<br>performance data, the subject devices have been determined to<br>be substantially equivalent to the identified predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.