Closure screws, healing caps, and healing abutments are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process. Customizable healing abutments made of PEEK are for use for up to six months.
Device Story
Straumann Healing Abutments, Healing Caps, and Closure Screws are dental prosthetic accessories used with Straumann Tissue Level and Bone Level Dental Implant Systems. Devices protect implant internal configurations and stabilize/form soft tissue during post-surgical healing. Available in various diameters and heights; constructed from titanium or PEEK with titanium alloy inlays. Used by dental clinicians in clinical settings. PEEK variants are indicated for temporary use up to six months. Devices are provided sterile to ensure patient safety during the healing phase.
Clinical Evidence
Bench testing only. Sterilization validation performed per ISO 11137-1:2006 and ISO 11137-2:2006. Shelf-life validation performed per ASTM F1980. All testing met predetermined acceptance criteria.
Technological Characteristics
Materials: Titanium and PEEK with Titanium Alloy inlay. Design: Various diameters and heights for dental implant interface. Sterilization: Radiation (ISO 11137-1/2). Shelf-life: Accelerated aging (ASTM F1980).
Indications for Use
Indicated for patients requiring dental implant procedures using the Straumann Dental Implant System to protect implant inner configurations and manage soft tissue healing. Customizable PEEK abutments are limited to six months of use.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Straumann Narrow Neck CrossFit (NNC) 03.3mm Dental Implant System (K111357)
BoneFit Implant System Accessories and Instruments (K894844)
Submission Summary (Full Text)
{0}------------------------------------------------
K130808
# 510(K) Summary
#### Applicant's Name and Address 1.
Straumann USA, LLC (on behalf of Institut Straumann AG) 60 Minuteman Rd. Andover, MA 01810 Telephone Number: 800-448-8168, ext. 2513 Fax Number: 978-747-0023 Contact Person: Elaine Alan Regulatory Project Manager
- 2. Date of Submission: March 22, 2013
- 3. Name of the Device
Trade Name:
Straumann Healing Abutments Straumann Healing Caps Straumann Closure Screws
Common Name: Healing Abutment Healing Caps Closure Screws
Product Code: NHA
Classification Name: Endosseous Dental Implant Abutments
Regulation Number: §872.3630
### 4. Legally Marketed Devices to which Equivalence is Claimed (Predicate Devices)
K111357, Straumann Narrow Neck CrossFit (NNC) 03.3mm Dental Implant System
- K072679, P.004 Abutments
K071585, P.004 Healing Abutments and Closure Screws
- K070478, P.004 RC Temporary Abutment, P.004 Temporary Healing Abutment (Cap)
K062129, P.004 Implants
K013798, Prosthetic Accessories to the ITI Dental Implant System
K960634, Titanium Healing Caps
K894844, BoneFit Implant System Accessories and Instruments
## 5. Description of the Device
The Straumann Healing Abutments, Healing Caps, and Closure Screws are
intended for use with the Straumann Tissue Level and Bone Level Dental
Implant System to protect the inner configuration of the implant and to
maintain, stabilize and form the soft tissue during the healing process. The
JUN 2 8 2013
{1}------------------------------------------------
devices are available in various diameters and heights, and the material is Titanium for all Healing Abutments, Healing Caps and Closure Screws; the customizable Healing Abutments are made of polyetheretherketone (PEEK) with a Titanium Alloy inlay.
#### ర్. Intended Use of the Device
Closure screws, healing caps, and healing abutments are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process.
Customizable healing abutments made of PEEK are for use for up to six months.
#### 7. Technological Characteristics
The proposed Sterile Healing Abutments, Healing Caps, and Closure Screws intended use, material, fundamental operating principles and overall design are identical to the predicate devices. The modification of the proposed healing components versus the currently cleared devices is from non-sterile to sterile devices. The non-significant design change of the proposed RC Conical Healing Abutments versus the currently cleared devices is to increase the diameter by 0.5mm.
#### 8. Performance Testing
Verification and validation testing were performed to ensure that the Straumann Sterile Healing Abutments, Healing Caps, and Closure Screws function as intended and that the modifications do not impact the performance of the devices. Testing included:
- i. Sterilization validation in accordance with ISO 11137-1:2006: Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of sterilization process for medical devices, and ISO 11137-2:2006: Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose
{2}------------------------------------------------
- 1) Requirements were met; there were not deviations to the applicable standards.
- ii. Shelf Life validation in accordance ASTM F1980, Standard Guide for Accelerated Aging of Sterile Medical Device Packages
- 1) Requirements were met; there were not deviations to the applicable standard.
#### Conclusion ல்
The results from the testing conducted demonstrated that the Straumann Sterile Healing Abutments, Healing Caps, and Closure Screws function as intended and met the predetermined acceptance criteria.
The results of the verification/validation testing and risk analysis indicate that the Straumann Sterile Healing Abutments, Healing Caps, and Closure Screws perform as intended, are substantially equivalent to the named predicate devices and do not raise new issues of safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle. The text is in all capital letters and is black.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Go(b) Silver Spring, MI) 20993-0002
June 28, 2013
Ms. Elaine Alan Regulatory Project Manager Straumann USA 60 Minuteman Road ANDOVER MA 01810
Re: K130808
Trade/Device Name: Straumann Healing Abutments, Healing Caps, and Closure Screws Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: May 30, 2013 Received: May 31, 2013
Dear Ms. Alan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kwame O. Ulmer -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{5}------------------------------------------------
# KISOBOR
## Indications for Use Statement
510(k) Number (if known): K130808
Device Name: Straumann Healing Abutments, Healing Caps, and Closure Screws Indications for Use:
Closure Screws, healing caps, and healing abutments, are intended for use with the Straumann Dental Implant System (SDIS) to protect the inner configuration of the implant and maintain, stabilize and form the soft tissue during the healing process.
Customizable healing abutments made of PEEK are for up to six months.
Prescription Use X (Part 21 CFR 801 Subpart D)
> ప ﺎ
AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Sheena A. Green -S 2013.06.28 11:36:22 -04'00'
for Susan Runner, DDS, MA
Division Sign-Off) Jivision of Anesthestology, General Hospital ifection Control, Dental Davices
510(k) Number: K130808
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.