K161622 · Optics Laboratory, Inc. · LPN · Aug 11, 2016 · Ophthalmic
Device Facts
Record ID
K161622
Device Name
LensGiene Sterile Saline Solution
Applicant
Optics Laboratory, Inc.
Product Code
LPN · Ophthalmic
Decision Date
Aug 11, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
The LensGiene™ Sterile Saline Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The LensGiene™ Sterile Saline Solution is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the cav to rinse contact lenses.
Device Story
LensGiene™ Sterile Saline Solution is a preservative-free, aqueous solution of sodium chloride, boric acid, and sodium borate. Provided in single-use 10 ml plastic containers. Used by contact lens wearers to rinse debris and cleaning solution from soft, rigid gas permeable, or hard lenses prior to insertion. Also functions as an insertion solution for scleral lenses and for rinsing lens cases. The solution is applied topically to the lens; it provides a natural environment for the cornea compared to preserved multipurpose solutions. The device is intended for over-the-counter use.
Clinical Evidence
No clinical data. Clinical studies were deemed unnecessary as the solution was previously cleared for rinsing soft contact lenses under K110221.
Technological Characteristics
Preservative-free aqueous solution; ingredients: sodium chloride, boric acid, sodium borate. pH: 7.2-7.8; tonicity: 290-320 mOsm/kg. Form factor: 10 ml single-use plastic resin container with twist-off cap. Sterility: Sterile.
Indications for Use
Indicated for rinsing soft (hydrophilic), rigid gas permeable, and hard contact lenses following disinfection; also for use as an insertion solution for scleral lenses and for rinsing contact lens cases.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Optics Laboratory, Inc. EyeCept™ Sterile Saline Solution (K110221)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping heads or faces.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 11, 2016
Optics Laboratory, Inc. % Mr. Paul Kramsky President Rockin' Regulatory, Inc. 21831 Tumbleweed Circle Lake Forest, CA 92630
Re: K161622
Trade/Device Name: LensGiene™ Sterile Saline Solution Regulation Number: 21 CFR 886.5918 Regulation Name: Rigid Gas Permeable Contact Lens Care Products Regulatory Class: Class II Product Code: MRC, LPN Dated: June 10, 2016 Received: June 13, 2016
Dear Mr. Kramsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161622
Device Name LensGiene(TM) Sterile Saline Solution
#### Indications for Use (Describe)
The LensGiene™ Sterile Saline Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The LensGiene™ Sterile Saline Solution is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the cav to rinse contact lenses.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------|--|--|
| | | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary for the LensGiene™ Sterile Saline Solution
### "Submission Sponsor:
Optics Laboratory, Inc. 9480 Telstar Ave Ste 3 El Monte, CA 91731-2988 Telephone: 626-350-1926 Fax: 626-350-1906 FDA Establishment Registration #: 3001918419
# Submission Correspondent:
Rockin' Regulatory, Inc. 21831 Tumbleweed Circle Lake Forest, CA 92630 Telephone: 949-636-1464 Contact: Paul Kramsky, President Email: pkramsky@cox.net
# Date Prepared:
June 10, 2016
### Device Name:
| Trade/Proprietary Name: | LensGiene™ Sterile Saline Solution |
|-------------------------|------------------------------------------------------------------|
| Classification Name: | Contact Lens Care Products |
| Product Code: | MRC and LPN |
| Regulation Number: | 21 CFR 886.5918 (Rigid Gas Permeable Contact Lens Care Products) |
| | 21 CFR 886.5928 (Soft (hydrophilic) Contact Lens Care Products) |
| Device Classification: | II |
| Review Panel: | Ophthalmic |
### Predicate Devices:
The LensGiene™ Sterile Saline Solution is substantially equivalent to the Menicon Co., Ltd. Saline Rinse Solution (K151768) and the Optics Laboratory, Inc. EyeCept™ Sterile Saline Solution (K110221).
# Device Description:
The LensGiene™ Sterile Saline Solution is a sterile, preservative-free, aqueous solution of sodium chloride, boric acid, and sodium borate provided in single does, carry-on size 10 ml bottles. The sterile saline solution has a pH 7.2-7.8 and a tonicity of 290-320 mOsm/kg. It rinses loose debris and cleaning solution off soft (hydrophilic), rigid gas permeable (RGP) and hard contact lenses prior to insertion. The rinsing solution removes
{4}------------------------------------------------
### Optics Laboratory, Inc. LensGiene™ Sterile Saline Solution Traditional 510(k) Premarket Notification
debris and bacteria following proper disinfection as recommended by the eye care practitioner. The sterile saline solution can be used to rinse contact lens cases, rinse lenses as needed throughout the day and to fill the concave posterior surface of scleral lenses prior to insertion to provide a more natural environment than currently approved multipurpose solutions which contain preservatives and osmolarity agents that are not designed to be held against the cornea for extended periods of time.
The LensGiene™ Sterile Saline Solution is a clear solution and should not be used if cloudy or discolored. The individual 10 ml bottle should be discarded after opening and use and should not be used after its labeled expiry date.
### Indications for Use:
The LensGiene™ Sterile Saline Solution is indicated for use following proper lens disinfection as recommended by the eye care practitioner. The LensGiene™ Sterile Saline Solution is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.
# Performance Data:
Non-Clinical Data
Non-clinical tests were unnecessary for this application because this testing was previously submitted for this solution under cleared 510(k) K110221.
# Clinical Data
Clinical studies were unnecessary for this application because this solution was previously cleared for rinsing soft (hydrophilic) contact lenses under K110221.
# Conclusion
Based upon the composition of the saline solution and previous test data submitted in K110221, the LensGiene™ Sterile Saline Solution is as safe, as effective and performs as well as the predicate devices. A comparison of the LensGiene™ Sterile Saline Solution and the predicate devices is presented in Table 1.
# Substantial Equivalence:
The claim of substantial equivalence to the previously cleared Menicon Co., Ltd. Saline Rinse Solution (K151768) and the Optics Laboratory, Inc. EyeCept™ Sterile Saline Solution (K110221) is supported by the following Comparison of Characteristics in Table 1. The LensGiene™ Sterile Saline Solution and the predicates are similar in composition and intended use. Therefore, the LensGiene™ Sterile Saline Solution is substantially equivalent to the predicate devices.
{5}------------------------------------------------
| Table 1 Comparison of Characteristics | | | |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| | LensGiene™ Sterile Saline Solution | Menicon Saline Rinse Solution | EyeCept™ Sterile Saline Solution |
| Device Name | Contact Lens Care Product | SAME | SAME |
| Trade Name | LensGiene™ Sterile Saline Solution | Menicon Saline Rinse Solution | EyeCept™ Sterile Saline Solution |
| Document Number | | K151768 | K110221 |
| Classification | Ophthalmic | SAME | SAME |
| Product Code | MRC and LPN | SAME | LPN |
| Regulation Number | 21 CFR 886.5918<br>21 CFR 886.5928 | SAME | 21 CFR 886.5928 |
| Class | II | II | II |
| Intended Use | For use following proper lens<br>disinfection as recommended<br>by the eye care practitioner.<br>The LensGiene™ Sterile<br>Saline Solution is for rinsing<br>soft (hydrophilic), rigid gas<br>permeable and hard contact<br>lenses prior to lens insertion.<br>This solution may also be used<br>as an insertion solution for<br>large diameter (scleral) contact<br>lenses, as a rinse for contact<br>lens cases, and may be used as<br>needed throughout the day to<br>rinse contact lenses. | SAME | For rinsing soft<br>contact lenses after<br>cleaning and for<br>wetting soft contact<br>lenses after<br>disinfection before<br>use. |
| Volume | 10 ml | 5 ml | 10 ml |
| Preservative Free | Yes | SAME | SAME |
| Container Usage | Single use | SAME | SAME |
| Sterility | Sterile | SAME | SAME |
| Materials | Plastic resin container with twist<br>off cap | SAME | SAME |
| Biocompatibility | Evaluated in accordance with<br>FDA Guidance for Contact Lens<br>Care Products | SAME | SAME |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.