K161433 · Voluntis Sa. · NDC · Nov 9, 2016 · Anesthesiology
Device Facts
Record ID
K161433
Device Name
Insulia Diabetes Management Companion
Applicant
Voluntis Sa.
Product Code
NDC · Anesthesiology
Decision Date
Nov 9, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1890
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic
Indications for Use
Insulia Diabetes Management Companion is indicated for use by healthcare professionals (HCPs) and their type 2 adult diabetes patients treated with long-acting insulin analog. Insulia Diabetes Management Companion is intended to provide secure capture, storage and transmission of diabetesrelated healthcare information, to enhance data management, to display reports and graphs, and to aid the patient in the review, and evaluation of patient data in order to support effective diabetes management. Insulia Diabetes Management Companion includes a basal calculator intended to provide direction to the patient in response to blood glucose and health events, within the scope of a pre-planned treatment program from a healthcare professional for insulin adjustments, similar to the directions provided to patients as a part of routine clinical practice. Insulia Diabetes Management Companion intended for use on commercially available mobile platforms, personal computers, in the home or in professional healthcare settings, and uses generally available networks and communication protocols. Insulia Diabetes Management Companion is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.
Device Story
System for type 2 diabetes management; comprises mobile app (iOS/Android), web-based app, and secure cloud database. Inputs: blood glucose measurements, health events. Transforms inputs via basal calculator; provides insulin dose adjustments based on pre-planned, physician-specified treatment programs; generates educational coaching messages. Used in home or professional settings by patients and HCPs. Output: reports, graphs, insulin dose guidance. Benefits: supports effective diabetes management by aiding patient/HCP in data review and adherence to treatment plans.
Clinical Evidence
Bench testing only. Evidence includes human factors study results and software verification and validation conducted in accordance with FDA guidance. No clinical trial data presented.
Technological Characteristics
Software-only system; mobile and web-based applications. Operates on commercially available smartphones, tablets, and PCs. Uses generally available networks and communication protocols. Secure database hosted in private cloud. Basal calculator algorithm provides insulin dose adjustments based on pre-planned treatment programs.
Indications for Use
Indicated for type 2 adult diabetes patients treated with long-acting insulin analog. Contraindicated for patients using NPH or premixed insulin, patients on basal-plus or basal-bolus regimens (multiple mealtime injections or pump therapy), pregnant patients, and non-adult patients.
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
WellDoc BlueStar (WellDoc DiabetesManager® System and DiabetesManager®-Rx System) (K141273)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 9, 2016
Voluntis S.A. Raffi Krikorian Vice President, Quality Assurance & Regulatory Affairs 2, Rue Des Bourets 92150 Suresnes FRANCE
Re: K161433
Trade/Device Name: Insulia Diabetes Management Companion Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive Pulmonary-Function Value Calculator Regulatory Class: II Product Code: NDC Dated: October 18, 2016 Received: October 19, 2016
Dear Raffi Krikorian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
# 510(k) Number (if known) K161433
### Device Name
Insulia Diabetes Management Companion
### Indications for Use (Describe)
Insulia Diabetes Management Companion is indicated for use by healthcare professionals (HCPs) and their type 2 adult diabetes patients treated with long-acting insulin analog.
Insulia Diabetes Management Companion is intended to provide secure capture, storage and transmission of diabetesrelated healthcare information, to enhance data management, to display reports and graphs, and to aid the patient in the review, and evaluation of patient data in order to support effective diabetes management.
Insulia Diabetes Management Companion includes a basal calculator intended to provide direction to the patient in response to blood glucose and health events, within the scope of a pre-planned treatment program from a healthcare professional for insulin adjustments, similar to the directions provided to patients as a part of routine clinical practice.
Insulia Diabetes Management Companion intended for use on commercially available mobile platforms, personal computers, in the home or in professional healthcare settings, and uses generally available networks and communication protocols.
Insulia Diabetes Management Companion is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.
Type of Use (Select one or both, as applicable)
| <span></span> | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|--------------------------------------------------------------------------------------|
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Voluntis Connected Therapeutics. The logo features a blue circle with a white bird-like shape inside, along with a small yellow dot. To the right of the circle, the word "VOLUNTIS" is written in blue, with the words "Connected Therapeutics" in a smaller, lighter font below it.
#### 510(K) SUMMARY 5
[per 21 CFR 807.92]
#### Submitter Information 5.1
| Name: | Voluntis S.A. |
|---------------|-----------------------------------------------|
| Address: | 58, avenue de Wagram<br>75017 Paris<br>FRANCE |
| Phone: | +33 141 383 920 |
| Fax: | +33 141 383 926 |
| Contact Name: | Raffi Krikorian |
Date of summary: November 8, 2016
| <b>5.2 Subject Device</b> | |
|---------------------------|------------------------------------------------|
| Trade Name: | Insulia Diabetes Management Companion |
| Common Name: | Diabetes Management Software |
| Regulation Number: | 21 CFR 868.1890 |
| Regulation Name: | Predictive pulmonary-function value calculator |
| Regulatory Class: | II |
| Product Code: | NDC |
| Classification Panel: | General Hospital |
#### 5.3 Predicate
| Devices | Primary | |
|---------|-----------------------|------------------------------------------------|
| | Trade Name: | ACCU-CHEK Connect Diabetes Management App |
| | 510(k) Reference: | K150910 |
| | Common Name: | Diabetes Management Software |
| | Regulation Number: | 21 CFR 868.1890 |
| | Regulation Name: | Predictive pulmonary-function value calculator |
| | Regulatory Class: | II |
| | Product Code: | NDC |
| | Classification Panel: | General Hospital |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Voluntis logo. The logo features a blue and white circular design with a yellow dot in the upper left corner. To the right of the design is the word "VOLUNTIS" in blue, with the words "Connected Therapeutics" in a smaller font below.
| Secondary | |
|-----------------------|-----------------------------------------------------------------------------------------|
| Trade Name: | WellDoc BlueStar (WellDoc<br>DiabetesManager® System and<br>DiabetesManager®-Rx System) |
| 510(k) Reference: | K141273 |
| Common Name: | Medical computers and software<br>Infusion pump accessories |
| Regulation Number: | 21 CFR 880.5725 |
| Regulation Name: | Infusion Pump |
| Regulatory Class: | II |
| Product Code: | MRZ, LNX |
| Classification Panel: | General Hospital |
#### Device Description 5.4
Insulia Diabetes Management Companion is a mobile and web based diabetes management system for type 2 adult diabetes patients and their healthcare team. Insulia Diabetes Management Companion includes three components:
- I A mobile medical application for use by patients on commercially available smartphones (iPhones and Android phones) and tablets.
- . A web-based application for use by patients in their home on their personal computer and on their mobile device, or by HCPs in professional healthcare settings through a compatible web browser on a personal computer.
- A secure database hosted in a private cloud environment and used to securely store patient data.
Insulia Diabetes Management Companion provides secure capture, storage and transmission of blood glucose data and other diabetes-related healthcare information to enhance data management, to display reports and graphs, and to aid the HCP and the patient in the review. analysis, and evaluation of patient data in order to support effective diabetes management.
Insulia Diabetes Management Companion includes a Basal Calculator intended to provide directions to the patient in response to blood glucose measurements and other diabetesrelated events, within the scope of a pre-planned treatment program from a healthcare professional for insulin adjustments. The guidance is similar to the directions provided to patients as a part of routine clinical practice.
Insulia Diabetes Management Companion provides educational coaching messages based on blood glucose values.
Insulia Diabetes Management Companion is not to be used in conjunction with a type of long-acting insulin analog different from insulin glargine (Lantus®) or insulin detemir (Levemir®).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Voluntis logo. The logo consists of a blue and white circular graphic on the left, with a stylized bird in flight. To the right of the graphic is the word "VOLUNTIS" in a bold, blue sans-serif font. Below "VOLUNTIS" is the phrase "Connected Therapeutics" in a smaller, lighter blue font.
Insulia Diabetes Management Companion is not to be used by patients using NPH (Neutral Protamine Hagedorn) insulin and premixed insulin.
Insulia Diabetes Management Companion is not to be used by patients treated with a basalplus or a basal-bolus regimen (i.e. including multiple mealtime regular or rapid-acting insulin injections per day or insulin pump therapy).
Insulia Diabetes Management Companion is not to be used during pregnancy nor by nonadult patients.
#### ર્સ્ટ ર્ Indications for Use
Insulia Diabetes Management Companion is indicated for use by healthcare professionals (HCPs) and their type 2 adult diabetes patients treated with long-acting insulin analog.
Insulia Diabetes Management Companion is intended to provide secure capture, storage and transmission of diabetes-related healthcare information, to enhance data management, to display reports and graphs, and to aid the HCP and the patient in the review, analysis, and evaluation of patient data in order to support effective diabetes management.
Insulia Diabetes Management Companion includes a basal calculator intended to provide direction to the patient in response to blood glucose and health events, within the scope of a pre-planned treatment program from a healthcare professional for insulin adjustments, similar to the directions provided to patients as a part of routine clinical practice.
Insulia Diabetes Management Companion includes software intended for use on commercially available mobile platforms, personal computers, in the home or in professional healthcare settings, and uses generally available networks and communication protocols.
Insulia Diabetes Management Companion is not intended to replace the care provided by a licensed healthcare professional, including prescriptions, diagnosis, or treatment.
#### ર્ડ.6 Comparison to Predicate Devices
A comparison of Insulia Diabetes Management Companion to the predicate devices is provided in Table 5.1. This table lists the key similarities and differences between Insulia Diabetes Management Companion and the predicate devices.
| Feature | Insulia Diabetes<br>Management<br>Companion | ACCU-CHEK Connect<br>Diabetes Management<br>App (K150910) | WellDoc BlueStar<br>DiabetesManager-Rx<br>System (K141273) |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| General Characteristics | | | |
| Regulation<br>No. | 21 CFR 868.1890 | 21 CFR 868.1890 | 21 CFR 880.5725 |
| Device Class | Class II | Class II | Class II |
| Product Code | NDC | NDC, JQP, LFR, LZG | MRZ, LNX |
| Environment<br>of Use | Home and Professional<br>Healthcare settings | Home and Professional<br>Healthcare settings | Home and Professional<br>Healthcare settings |
| Intended User | HCPs and their adult type<br>2 diabetes patients | HCPs and their patients<br>with diabetes | HCPs and their adult type<br>2 diabetes patients |
| Prescription<br>Use | Yes | OTC distribution of the<br>app, but the bolus<br>calculator function can be<br>activated only upon<br>prescription by a<br>physician | Yes |
| Technological Characteristics | | | |
| Components | Software only, patient<br>mobile based application,<br>patient and HCP web-<br>based application | Software only, patient<br>mobile based application,<br>patient and HCP web-<br>based application when<br>used in conjunction with<br>ACCU-CHEK Connect<br>Online | Software only, patient<br>mobile based application,<br>patient and HCP web-<br>based application |
| Treatment<br>Guidance | Adjustments to insulin<br>doses within the scope of<br>a pre-planned, physician-<br>specified treatment<br>program similar to<br>routine clinical practice | Adjustments to insulin<br>doses within the scope of<br>a pre-planned, physician-<br>specified treatment<br>program and calculation<br>of carbohydrate intake | Insulin dose reminder<br>and schedule |
| Type of<br>Calculated<br>Insulin | Basal Insulin | Bolus Insulin | N/A |
| Manual Data<br>Entry | Yes | Yes | Yes |
| Logbook | Yes | Yes | Yes |
| Personal<br>Health Record | Yes | Yes | Yes |
| Reports &<br>Statistics | Yes | Yes | Yes |
| Coaching<br>Messages | Yes | No | Yes |
| Secure<br>Database | On computer media | On computer media | On computer media |
| Data Transfer | Public Internet | Public Internet | Public Internet |
Table 5.1: List of the key similarities and differences between Insulia Diabetes Management Companion and the predicate devices
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Voluntis, a company focused on connected therapeutics. The logo features a stylized circular icon with a white bird-like shape against a blue background. To the right of the icon, the company name "VOLUNTIS" is displayed in a bold, sans-serif font, with the tagline "Connected Therapeutics" written in a smaller, lighter font beneath it.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Voluntis Connected Therapeutics. The logo features a blue circle with a white bird-like shape inside, along with the company name "VOLUNTIS" in blue, bold letters. Below the company name, the words "Connected Therapeutics" are written in a smaller, lighter blue font.
Insulia Diabetes Management Companion has the same intended use as the predicate devices. Indeed, all are software applications intended for use by healthcare professionals and their diabetes patients as an aid in the management of diabetes. Furthermore, the proposed indications for use for the subject device are similar to and consistent with those of the predicate devices, and do not raise different questions of safety or effectiveness.
Insulia Diabetes Management Companion's technological characteristics are similar to and consistent with those of the predicate devices, i.e. all include software applications that provide secure capture, storage, transmission and display of blood glucose data as well as other diabetes related healthcare information.
Insulia Diabetes Management Companion and the predicate devices provide directions which are similar to directions that physicians provide to patients as part of routine clinical practice. In addition, Insulia Diabetes Management Companion and ACCU-CHEK Bolus Advisor (which is a component of the ACCU-CHEK Connect Diabetes Management App) both provide directions within the scope of a pre-planned treatment program for adjustments to prescribed insulin, similar to the directions physicians provide to patients as a part of routine clinical practice.
Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Therefore, Insulia Diabetes Management Companion is substantially equivalent to the predicate devices.
#### 5.7 Performance Data Demonstrating Substantial Equivalence
Comprehensive performance testing was conducted on the Insulia Diabetes Management Companion to support a determination of substantial equivalence.
The results of the performance testing, including the human factors study results and software verification and validation documented in accordance with FDA's Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005), were provided and demonstrate that the device meets the performance requirements for its intended use, and supports substantial equivalence.
#### 5.8 Conclusion
Insulia Diabetes Management Companion has similar indications for use, the same intended use, and similar technological characteristics as those of the predicate devices. The
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Voluntis logo. The logo features a blue circle with a white abstract design inside, along with the word "VOLUNTIS" in blue, and the words "Connected Therapeutics" in a smaller font below.
differences in technological characteristics have been analysed and addressed through performance testing, which demonstrate that Insulia Diabetes Management Companion meets its intended use. Any technological differences between Insulia Diabetes Management Companion and the predicate devices do not raise any new issues of safety or effectiveness. Furthermore, performance testing has demonstrated that the Insulia Diabetes Management Companion performs as intended and is substantially equivalent to the predicate devices.
In summary, Insulia Diabetes Management Companion is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.