K150910 · Roche Diabetes Care, Inc. · NDC · Jun 3, 2015 · Anesthesiology
Device Facts
Record ID
K150910
Device Name
ACCU-CHEK Connect Diabetes Management App
Applicant
Roche Diabetes Care, Inc.
Product Code
NDC · Anesthesiology
Decision Date
Jun 3, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.1890
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic
Indications for Use
The ACCU-CHEK Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format. The ACCU-CHEK Bolus Advisor, as a component of the ACCU-CHEK Connect Diabetes Management App, is indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data. Before its use, a physician or healthcare professional must activate the bolus calculator and provide the patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the software.
Device Story
Mobile application for diabetes management; collects, stores, and displays blood glucose data; receives data via Bluetooth Low Energy (BLE) from ACCU-CHEK Aviva Connect meter; allows manual data entry. Features Bolus Advisor for insulin dosing calculations; requires physician activation and parameter entry (target blood glucose, insulin-to-carbohydrate ratio, insulin sensitivity). Used by patients with MDI therapy to optimize glycemic control; reduces fear of hypoglycemia; improves confidence. Data can be shared with healthcare providers or caregivers via online management system. App provides reports and graphs for clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by design verification bench testing and human factors expert evaluation comparing iOS and Android versions. Risk management performed per ISO 14971 using FMEA.
Technological Characteristics
Mobile software application (iOS platform). Connectivity via Bluetooth Low Energy (BLE) to blood glucose meters. Bolus calculator algorithm is rule-based, requiring physician-programmed parameters. Software level of concern: Major.
Indications for Use
Indicated for diabetes patients treated with multiple daily insulin injection (MDI) therapy. Bolus calculator requires physician activation and programming of patient-specific parameters (target blood glucose, insulin-to-carbohydrate ratio, insulin sensitivity).
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of a human figure, represented by three overlapping profiles facing to the right. The profiles are stylized and appear to be connected, creating a sense of unity and collaboration.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3, 2015
Roche Diabetes Care Inc. Ms. Chunhong Tao Regulatory Affairs Specialist 9115 Hague Road Indianapolis, IN 46250
Re: K150910
Trade/Device Name: ACCU-CHEK Connect Diabetes Management App Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive pulmonary-function value calculator Regulatory Class: II Product Code: NDC, LZG, LFR, JQP Dated: May 6, 2015 Received: May 7, 2015
Dear Ms. Tao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K150910
Device Name
ACCU-CHEK Connect Diabetes Management App
Indications for Use (Describe)
The ACCU-CHEK Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### Device Name
ACCU-CHEK Connect Diabetes Management App
#### Indications for Use (Describe)
The ACCU-CHEK Bolus Advisor, as a component of the ACCU-CHEK Connect Diabetes Management App, is indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data. Before its use, a physician or healthcare professional must activate the bolus calculator and provide the patient-specific target blood glucose. insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the software.
Type of Use (Select one or both, as applicable)
| <span style="font-size: large;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: large;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Roche logo. The logo consists of the word "Roche" in black, centered within a black hexagon. The hexagon is outlined in black and has a simple, clean design.
# 510(k) Summary
Introduction
According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
| 1) Submitter name, address, contact | Roche Diabetes Care Inc.<br>9115 Hague Rd.<br>Indianapolis, IN 46250<br>(317) 521-7227<br>Contact Person: Chunhong (Emma) Tao |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| | Date Prepared: April 2nd, 2015 |
| 2) Device name | Proprietary Name: ACCU-CHEK Connect Diabetes Management App<br>Common Name: diabetes management software |
|----------------|----------------------------------------------------------------------------------------------------------|
|----------------|----------------------------------------------------------------------------------------------------------|
Primary
Classification Name: predictive pulmonary-function value calculator
Classification Regulation: 21 C.F.R. § 868.1890; Class II
Product Code: NDC
Secondary
Classification Regulations: 21 C.F.R. § 880.5725, 862.1345, 862.2100; Class II
Product Code: LZG, LFR, JQP
| 3) Predicate device | ACCU-CHEK Connect Diabetes Management App (K141929), concurrence received on March 16, 2015. |
|---------------------|----------------------------------------------------------------------------------------------|
|---------------------|----------------------------------------------------------------------------------------------|
Continued on next page
{5}------------------------------------------------
| 4) Device<br>Description | The ACCU-CHEK Connect Diabetes Management App is designed to<br>facilitate efficient collecting, transmitting, and analyzing of blood glucose<br>results and other diabetes management data. The App helps:<br><br>Wireless transfer of data from ACCU-CHEK Aviva Connect Blood<br>Glucose Meter. Assist in general diabetes management through logging of contextual<br>data. ACCU-CHEK Bolus Advisor support of mealtime insulin dosing<br>calculations. Perform structured testing. Wireless transfer of data from mobile devices to ACCU-CHEK<br>Connect Online Diabetes Management System and optionally share<br>this data with healthcare provider (HCP) or caregiver.<br>The insulin bolus calculations provided by the app are meant for patients<br>undergoing multiple daily injection therapy. Bolus calculators, such as the<br>ACCU-CHEK Bolus Advisor, have been demonstrated to facilitate the<br>optimization of glycemic control in patients who are trained in multiple daily<br>insulin injection therapy and under the supervision of healthcare professional |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
experienced in managing insulin-treated patients. Such calculators have also been shown to reduce patient fear of hypoglycemia and improve patient confidence in diabetes management.
The ACCU-CHEK Connect Diabetes Management App is not intended to serve as an accessory to an insulin pump.
Continued on next page
{6}------------------------------------------------
5) Intended For Over-the-Counter Use: use The ACCU-CHEK Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format. For Prescription Use: The ACCU-CHEK Bolus Advisor, as a component of the ACCU-CHEK Connect Diabetes Management App, is indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on userentered data. Before its use, a physician or healthcare professional must activate the bolus calculator and provide the patient-specific target blood glucose, insulin-to-carbohydrate ratio, and insulin sensitivity parameters to be programmed into the software. 6) Substantial The ACCU-CHEK Connect Diabetes Management App utilizes the equivalence processing power of a mobile phone operating system platform to transfer data from a blood glucose meter, transfer data to an online diabetes management system, facilitate structured testing data collection, assist in general diabetes management through logging of contextual data, and calculate insulin bolus in response to blood glucose, health events, and carbohydrate input. The App is compatible with glucose meters with BLE transmitter. The insulin bolus calculator algorithm is unchanged as compared to the predicate device. The modification of the ACCU-CHEK Connect Diabetes Management App to work on iOS platform does not change the App's intended use or the interface to the ACCU-CHEK Aviva Connect meter or the ACCU-CHEK Connect Online diabetes management system. The modification of ACCU-CHEK Connect Diabetes Management App is substantially equivalent to ACCU-CHEK Connect Diabetes Management App (K141929).
Continued on next page
{7}------------------------------------------------
Modifying the ACCU-CHEK Connect Diabetes Management App to work on 7) Data demonstrating iOS platform does not change the App's intended use or its place within the substantial ACCU-CHEK Connect Diabetes Management System. The modification to equivalence the ACCU-CHEK Connect Diabetes Management App does not change the data collection, tracking, trending, or transmission functions, the bolus calculator algorithm, the bolus calculator activation prescription control process, the bolus calculator activation and patient training materials, or the bolus calculator user interface screens.
> The risk assessment for the iPhone version relied heavily on the risk assessment performed for the Android OS version of the App. The project management staff assessed potential faulty conditions and monitored ACCU-CHEK Connect Diabetes Management App post launch (Android OS version) to identify any possible faulty conditions that might lead to possible hazards for the patient. A risk analysis according to the "Risk Reduction Principle" laid down in the harmonized ISO standard "14971 Medical Devices -Application of risk management to medical devices" was carried out for the ACCU-CHEK Connect App. Possible hazards and consequences were systematically identified and evaluated by using the "Failure Mode Effect and Criticality Analysis" technique. Where appropriate, adequate protection measures relating to the risk that cannot be eliminated have been implemented, as well as safety information described in the instructions for use.
Design verification bench testing on the modification of ACCU-CHEK Connect Diabetes Management App demonstrated that the device meets the performance requirements for its intended use. The data demonstrate that the new device is substantially equivalent to the predicate device.
A Human Factors expert evaluation was performed to show that the predicate design validation can be used to support iPhone version's design validation. Three types of changes were reviewed using side-by-side comparison of screenshots from the Android version and the iPhone version:
- 1. Changes based on the inherent differences between the iOS and Android operating systems (OS) user interface standards
- 2. Enhancements that were made based upon results of the Android version summative and iPhone version formative human factors study
- 3. Changes to Validation study tasks
All of the changes are attributed to user interface standard differences between iOS and Android operating systems or to enhancements to the user experience. No new use-related hazard was identified during the expert evaluation. This expert evaluation supports design validation of the product.
{8}------------------------------------------------
| 8)<br>Similarities/<br>Differences | The following is a listing of the key similarities and differences between the<br>modified ACCU-CHEK Connect Diabetes Management App and the<br>predicate device. | |
|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Feature /Claim | Modified ACCU-CHEK Connect<br>Diabetes Management App | ACCU-CHEK Connect Diabetes<br>Management App (#k141929) |
| Intended Use | The ACCU-CHEK Connect<br>Diabetes Management App is<br>indicated as an aid in the treatment<br>of diabetes. The software provides<br>for electronic download of blood<br>glucose meters, manual data entry,<br>storage, display, transfer, and self-<br>managing of blood glucose and<br>other related health indicators which<br>can be shown in report and<br>graphical format.<br><br>The ACCU-CHEK Bolus Advisor,<br>as a component of the ACCU-<br>CHEK Connect Diabetes<br>Management App, is indicated for<br>the management of diabetes by<br>calculating an insulin dose or<br>carbohydrate intake based on user-<br>entered data. Before its use, a<br>physician or healthcare professional<br>must activate the bolus calculator<br>and provide the patient-specific<br>target blood glucose, insulin-to-<br>carbohydrate ratio, and insulin<br>sensitivity parameters to be<br>programmed into the software. | Same |
| Prescription /<br>over-the-counter<br>use? | Over-the-counter distribution of<br>App, with bolus calculator function<br>disabled until activated by physician | Same |
| User Group | Diabetes patients treated with<br>multiple daily insulin injection<br>(MDI) therapy | Same |
| Accessory to<br>Blood Glucose<br>Meters? | ACCU-CHEK Aviva Connect blood<br>glucose meter | Same |
| Feature /Claim | Modified ACCU-CHEK Connect<br>Diabetes Management App | ACCU-CHEK Connect Diabetes<br>Management App (#k141929) |
| Bolus Calculator<br>Set-up stage | During the modified ACCU-CHEK<br>Connect Diabetes Management<br>App's bolus calculator set-up stage,<br>the following information must be<br>entered.<br>• Time Blocks<br>• Target Range<br>• Carb Ratio<br>• Insulin Sensitivity<br>• Health Event Percentages<br>• Meal Rise<br>• Snack Size<br>• Acting Time<br>• Offset Time | Same |
| Bolus Calculator<br>inputs | Once the set-up stage has been<br>completed, the patient may begin<br>using the ACCU-CHEK Bolus<br>Advisor in the modified ACCU-<br>CHEK Connect Diabetes<br>Management App. Bolus<br>recommendations can be triggered<br>by inputting the following<br>information prior to each meal:<br>• Measure blood glucose<br>• Enter carbohydrates<br>• Enter Health Event | Same |
| Communicate<br>with insulin<br>pumps? | No | No |
| Software Level of<br>Concern | Major | Major |
| Connectivity to<br>Meter | Bluetooth Low Energy (BLE) | Same |
| Control or affect<br>the blood glucose<br>meter's<br>measurements? | No | No |
| Reports, graphs,<br>and Electronic<br>Log Book | Yes | Yes |
| Feature /Claim | Modified ACCU-CHEK Connect Diabetes Management App | ACCU-CHEK Connect Diabetes Management App (#k141929) |
| Structured Testing | Facilitate the collection of<br>Structured Testing data | Same |
| Carbohydrate Calculator | Calculate carbohydrate intake based<br>on user-entered data | Same |
| Bolus Calculator | Yes | Yes |
| Bolus Calculator Security | The patients will be instructed to<br>request an activation code from<br>their healthcare providers; the<br>healthcare providers will decide<br>whether any additional multiple<br>daily injection therapy training<br>should be provided to the patient<br>prior to activating the app's bolus<br>calculator function. | Same |
| Allow manual entry? | Yes | Yes |
| Back-calculation prevents insulin stacking? | Yes | Yes |
| Operating platform | iOS platform | Android platform |
| UI Standards | iOS standards | Android standards |
| Size and Spacing | iOS phones with smaller viewable<br>screens | Android phones with larger<br>viewable screens |
| Meter Pairing Code | iOS standard Bluetooth pairing code | Android standard Bluetooth pairing<br>code |
{9}------------------------------------------------
{10}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.