CELTRA Press is an all-ceramic system for the creation of Occlusal veneers Thin veneers Veneers Inlays Onlays Crowns in the anterior and posterior region 3-unit bridges in the anterior and posterior region 3-unit bridges in the premolar region up to the second premolar as the terminal abutment Crown, splinted crown or 3 unit bridge up to the second premolar placed on top of an implant abutment.
Device Story
CELTRA Press is a high-strength zirconia-reinforced lithium silicate glass ceramic material used by dental laboratory technicians to fabricate all-ceramic dental restorations. The device consists of industrially produced, homogeneous ingots available in varying translucencies. Technicians use a hot-pressing technique within a dental furnace and press investment materials to form the restorations. The process avoids the formation of a surface reaction layer. Finished restorations may be veneered, stained, or glazed using compatible CELTRA Press veneering porcelains. The final product provides aesthetic, durable tooth-colored restorations for patients, including crowns and 3-unit bridges. The device is intended for professional use in a dental laboratory setting.
Clinical Evidence
No clinical data included. Substantial equivalence supported by non-clinical bench testing, including flexural strength, chemical solubility, coefficient of thermal expansion, and radioactivity testing per ISO 6872:2015. Biocompatibility established via testing per ISO 10993-1, 10993-5, and 10993-18.
Technological Characteristics
Zirconia-reinforced lithium silicate glass ceramic. Classified as Type II, Class 4a/4b per ISO 6872:2015. Requires minimum mean flexural strength of 500 MPa. Form factor: industrially produced ingots. Processing: hot-pressing technique in dental furnaces. Biocompatibility: tested per ISO 10993.
Indications for Use
Indicated for patients requiring dental restorations including occlusal veneers, thin veneers, veneers, inlays, onlays, crowns, and 3-unit bridges in anterior, posterior, and premolar regions, or implant-supported restorations.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
Predicate Devices
IPS e.max Press and IPS e.max Press Multi (K120134)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 16, 2016
Dentsply Sirona Helen Lewis Director Corporate Regulatory Affairs 221 West Philadelphia Street, Suite 60 York, Pennsylvania 17404
Re: K161269
Trade/Device Name: CELTRA Press Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: Class II Product Code: EIH Dated: July 5, 2016 Received: July 6, 2016
Dear Ms. Helen Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Susan Runno DDS, MA
Tina Kiang, Ph.D. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name CELTRA Press
Indications for Use (Describe) CELTRA Press is an all-ceramic system for the creation of Occlusal veneers Thin veneers Veneers Inlays Onlays Crowns in the anterior and posterior region 3-unit bridges in the anterior and posterior region 3-unit bridges in the premolar region up to the second premolar as the terminal abutment Crown, splinted crown or 3 unit bridge up to the second premolar placed on top of an implant abutment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Dentsplv Sirona 221 West Philadelphia Street Suite 60 York, PA 17404
Image /page/3/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized leaf-like shape on the left, followed by the words "Dentsply" and "Sirona" stacked on top of each other on the right. The text and the leaf are in a gray color.
## SECTION 5. 510(k) SUMMARY for CELTRA Press
#### Submitter Information:
Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404
| Contact Person: | Helen Lewis |
|-------------------|--------------|
| Telephone Number: | 717-487-1332 |
| Fax Number: | 717-849-4343 |
Date Prepared: 24 June 2016
#### Device Name:
| ● | Proprietary Name: | CELTRA Press |
|---|----------------------|-----------------------------------|
| ● | Common Name: | All-ceramic system |
| ● | Classification Name: | Porcelain powder for clinical use |
| ● | CFR Number: | 21 C.F.R. 872.6660 |
| ● | Device Class: | II |
| ● | Product Code: | EIH |
#### Predicate Device:
| Predicate Device Name | 510(k) | Company Name |
|-------------------------------------------------------------------|---------|----------------------|
| IPS e.max Press and IPS e.max Press Multi | K120134 | Ivoclar Vivadent AG |
| Reference Device for Biocompatibility<br>Quattro Porcelain System | K091706 | DENTSPLY Prosthetics |
## Description of Device:
CELTRA Press is a high-strength zirconia-reinforced lithium silicate glass ceramic material that can be used for the fabrication of highly aesthetic all-ceramic restorations by using the hot-pressing technique in dental labs. The homogeneous, industrially produced ingots are available in both high and low translucency for full-contour application. They are pressing furnaces using press investments material which do not form surface reaction layer to obtain tooth-colored, highly aesthetic, all-ceramic restorations. Subsequently, CELTRA Press may be veneered, stained and/or glazed with compatible CELTRA Press veneering porcelains.
{4}------------------------------------------------
## Indications for Use:
CELTRA Press is an all-ceramic system for the creation of Occlusal veneers Thin veneers Veneers Inlays Onlays Crowns in the anterior and posterior region 3-unit bridges in the anterior and posterior region 3-unit bridges in the premolar region up to the second premolar as the terminal abutment Crown, splinted crown or 3 unit bridge up to the second premolar placed on top of an implant abutment.
#### Substantial Equivalence:
The proposed CELTRA Press has the similar indications for use as the predicate IPS e.max Press and IPS e.max Press Multi (K120134) with the exception of an indication for 3-unit bridges in the posterior region. The proposed CELTRA Press meets the requirement of flexural strength for this indication in accordance with ISO 6872:2015 (Dentistry - Ceramic materials).
| Proposed Device | Predicate Device | Differences |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CELTRA Press system | IPS e.max Press and IPS<br>e.max Press Multi, K120134 | |
| Indications for Use:<br>CELTRA Press is an all-<br>ceramic system for the creation<br>of<br>Occlusal veneers<br>Thin veneers<br>Veneers<br>Inlays<br>Onlays<br>Crowns in the anterior and<br>posterior region<br>3-unit bridges in the anterior<br>and posterior region<br>3-unit bridges in the premolar<br>region up to the second<br>premolar as the terminal<br>abutment<br>Crown, splinted crown or 3 unit<br>bridge up to the second<br>premolar placed on top of an<br>implant abutment. | Indications for Use:<br>IPS e.max Press and IPS e.max<br>Press Multi is an all-ceramic<br>system for the creation of<br>Occlusal veneers<br>Thin Veneers<br>Veneers<br>Inlays<br>Onlays<br>Crowns in the anterior and<br>posterior region<br>3-unit bridges in the anterior<br>region<br>3-unit bridges in the premolar<br>region up to the second<br>premolar as the terminal<br>abutment<br>Crown, splinted crown or 3 unit<br>bridge up to the second<br>premolar placed on top of an<br>implant abutment. | The proposed device<br>CELTRA Press is indicated<br>for 3-unit posterior bridges<br>while the predicate IPS<br>e.max Press and IPS<br>e.max Press Multi,<br>K120134 is not indicated<br>for this use based on its<br>ISO 6872 standard<br>classification. |
## Technological Characteristics:
The proposed CELTRA Press was classified and tested in accordance with ISO 6872:2015 (Dentistry – Ceramic materials). CELTRA Press classification and clinical usage is noted below: Type II: All other forms of ceramic products
Class 4a: Monolithic ceramic for three-unit prostheses involving molar restoration Class 4b: Partially or fully covered substructure for three-unit prostheses involving molar restoration
{5}------------------------------------------------
| Physical<br>Properties<br>ISO 6872 | Proposed Device<br>CELTRA Press | Predicate<br>Device<br>IPS e.max<br>Press and IPS<br>e.max Press<br>Multi, K120134 | Differences |
|----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type and<br>Classification<br>Flexural Strength<br>Chemical Solubility<br>Coefficient of<br>Thermal Expansion,<br>CTE<br>Radioactivity | Conforms to ISO<br>6872 | Conforms to<br>ISO 6872 | The proposed device CELTRA<br>Press is a Type II class 4 device<br>which requires a minimum mean<br>flexural strength of 500MPa<br>while the predicate IPS e.max<br>Press and IPS e.max Press<br>Multi, K120134 is a Type II class<br>3 devices based on its ISO 6872<br>standard classification. |
The strength of the proposed CELTRA Press meets the reguirement for Type II Class 4 (materials with a minimum mean flexural strength of 500 MPa), which by definition includes three unit prostheses.
# Non-Clinical Performance Data:
## Physical Properties:
In accordance with ISO 6872:2015 (Dentistry - Ceramic materials), the product must meet the requirements for flexural strength, chemical solubility, coefficient of thermal expansion, and radioactivity. Both CELTRA Press and predicate IPS e.max Press and IPS e.max Press Multi (K120134) meet the requirements of the applicable Class, as detailed in Table under Technological The performance of CELTRA Press meets the requirements of the non-clinical Characteristics. bench testing conducted to support substantial equivalence.
## Toxicological Testing;
Testing according to ISO 10993-1, 10993-5 and 10993-18 (Biological evaluation of medical devices) was completed to characterize the biocompatibility of CELTRA Press. Results of the testing conducted in accordance with ISO standard support the biocompatibility of CELTRA Press. CELTRA Press veneering porcelains (accessory) were compared to reference device, Quattro Porcelain System (K091706) for biocompatibility purposes.
## Clinical Performance Data.
Not applicable. No data from human clinical studies has been included to support the substantial equivalence of the CELTRA Press.
## Conclusion Regarding Substantial Equivalence
CELTRA Press supports substantial equivalence to the predicate IPS e.max Press and IPS e.max Press Multi cleared under premarket notification K120134 because:
- It has the similar intended use as the predicate ●
- . Has similar technological characteristics as the predicate, with the exception of flexural strength. CELTRA Press meets the requirement of flexural strength for Type II, Class 4 materials.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.