K091706 · Dentsply International, Inc. · EIH · Aug 28, 2009 · Dental
Device Facts
Record ID
K091706
Device Name
QUATTRO PORCELAIN SYSTEM
Applicant
Dentsply International, Inc.
Product Code
EIH · Dental
Decision Date
Aug 28, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Indications for Use
Quattro Porcelain System is indicated for: 1. PFM Applications: Single and multiple unit porcelain fused-to-metal fixed prosthodontic restorations. 2. PTM Applications: Single and multiple unit porcelain pressed-to-metal fixed prosthodontic restorations. 3. All-Ceramic Applications: Single unit anterior and posterior premolar metal-free fixed prosthodontic restorations, laminate veneers, inlays and onlays.
Device Story
Quattro Porcelain System is an integrated dental porcelain system used by dental professionals for fabricating restorations. It utilizes three fabrication techniques: porcelain-fused-to-metal (PFM), pressed-to-metal (PTM), and pressed all-ceramic (AC). The system consists of dentin and ingot porcelain components. These materials are processed by dental technicians to create crowns, bridges, inlays, onlays, and veneers. The final restorations are placed by dentists to restore tooth structure and function. The system relies on established porcelain fabrication methods; it does not involve digital inputs, software, or automated algorithms. Biocompatibility testing confirms safety for oral use.
Clinical Evidence
Bench testing only. Performance data compared the structural components (dentin and ingot porcelain) to legally marketed Ceramco porcelains. Biocompatibility testing was performed, including cytotoxicity, irritation, systemic toxicity, sensitization, oral irritation, and genotoxicity, with all results passing.
Technological Characteristics
Integrated porcelain system comprising dentin and ingot porcelain components. Materials are consistent with those used in legally marketed dental porcelains. The system supports PFM, PTM, and all-ceramic fabrication techniques. No energy source, software, or connectivity features. Biocompatible for oral use.
Indications for Use
Indicated for patients requiring fixed prosthodontic restorations including single/multiple unit PFM and PTM crowns/bridges, and single unit anterior/posterior premolar metal-free crowns, laminate veneers, inlays, and onlays.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
K091706
Image /page/0/Picture/24 description: The image shows the word "DENTSPLY" in all capital letters. The font is bold and sans-serif. The letters are tightly spaced together. The image is in black and white.
1.
DENTSPLY International
World Headquarters Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405-0872 (800) 877-0020 Fax (717) 849-4343 www.dentsply.com
AUG 2 8 2009
# 510(k) SUMMARY for Quattro Porcelain System
- Submitter Information: DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405
| Contact Person: | Helen Lewis |
|-------------------|--------------|
| Telephone Number: | 717-849-4229 |
| Fax Number: | 717-849-4343 |
June 8, 2009 Date Prepared:
#### 2. Device Name:
Proprietary Name: ◆
- Classification Name: .
- CFR Number: ●
- Device Class: .
- Product Code: EIH .
#### Predicate Device: 3.
- DENTSPLY International Company: ● Finesses Low Fusing Porcelain System Device: ●
- K954761 510(K) No.: .
- Date Cleared: 11/02/1995 .
#### Description of Device: 4.
The Quattro Porcelain System is an integrated porcelain system used for the fabrication of dental restorations using three major porcelain fabrication techniques: porcelain-fused-to-metal (PFM), pressed-to-metal (PTM), and pressed all-ceramic (AC). Quattro PFM and PTM are indicated for both anterior and posterior crowns and bridges. Quattro AC is indicated for anterior to pre-molar crowns, inlays and onlays, and veneers.
Quattro Porcelain System
Powder, Porcelain
872.6660
II
000009
{1}------------------------------------------------
#### Indications for Use: 5.
Quattro Porcelain System is indicated for:
- 1. PFM Applications: Single and multiple unit porcelain fused-to-metal fixed prosthodontic restorations.
- 2. PTM Applications: Single and multiple unit porcelain pressed-to-metal fixed prosthodontic restorations.
- 3. All-Ceramic Applications: Single unit anterior and posterior premolar metal-free fixed prosthodontic restorations, laminate veneers, inlays and onlays.
### Description of Safety and Substantial Equivalence: 6.
## Technological Characteristics
The Quattro Porcelain System is an integrated system that can be used for porcelain fused to metal (PFM), pressed to metal (PTM), and all-ceramic (AC) restorations. All of the components found in the Quattro Porcelain System have been used in legally marketed devices and/or were found safe for dental use. The Ouattro veneering has been evaluated and passed biocompatibility testing for cvtotoxicity, irritation, systemic toxicity, sensitization, oral irritation, and genotoxicity.
### Non-Clinical Performance Data
The performance of the Quattro Porcelain System's two structural components, dentin and ingot porcelain are comparable to other legally marketed Ceramco porcelains commercially used and successful for PFM and PTM restorations.
### Conclusion as to Substantial Equivalence
We believe that the prior use of the components of the Quattro Porcelain System in legally marketed devices, the performance data provided, and the biocompatibility data provided support the safety and effectiveness of the Quattro Porcelain System for the indicated uses.
We believe that the test data and the predicate comparisons provided in the Quattro Porcelain System 510(k) application demonstrate that the device is as safe and as effective as the predicate device.
000010
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
# AUG 2 8 2009
Ms. Helen Lewis
Director of Corporate Compliance and Regulatory Affairs DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street, Suite 60 York, Pennsylvania 17404-0872
Re: K091706
Trade/Device Name: Quattro Porcelain System Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: June 8, 2009 Received: June 10, 2009
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Ms. Lewis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Runner
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
109 706
510(k) Number (if known):
Device Name: Quattro Porcelain System
### Indications for Use:
Quattro Porcelain System is indicated for:
- 1. PFM Applications: Single and multiple unit porcelain fused-to-metal fixed prosthodontic restorations.
- 2. PTM Applications: Single and multiple unit porcelain pressed-to-metal fixed prosthodontic restorations.
- 3. All-Ceramic Applications: Single unit anterior and posterior premolar metal-free fixed prosthodontic restorations, laminate veneers, inlays and onlays.
Prescription Use X · (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Reim Marley for MSR
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number: K091706Quattro Porcelain System
000008
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.