The US SA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. The abutment is intended for use with a dental implant fixture to provide support for prosthetic restorations such as bridges, or overdenture.
Device Story
US SA Implant System comprises titanium dental implants and abutments for surgical placement in jaw bone; supports prosthetic restorations (bridges, overdentures). Fixtures feature external hex-connection, submerged design, straight or tapered body, and self-tapping 3-sided cutting edges. Abutments (Multi Angled) allow path adjustment for multi-unit restorations. Used by dental professionals in clinical settings. System provides structural support for dental prosthetics; benefits patients by restoring function in edentulous areas. Device is supplied sterile; requires surgical implantation.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including fatigue testing per FDA guidance for root-form endosseous dental implants and sterilization validation per ISO 11137 and ISO 17665.
Technological Characteristics
Materials: Pure Titanium (ASTM F67) for fixtures, Titanium Alloy (ASTM F136) for abutments. Design: External hex-connected, submerged, straight/tapered body, self-tapping. Surface: Sandblasted and Acid-etched (SA). Sterilization: Radiation (fixtures) and Steam. Dimensions: Diameters 3.5mm-5.13mm, Lengths 7mm-18mm.
Indications for Use
Indicated for partially or fully edentulous patients (mandibles and maxillae) requiring single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, and fixed bridgework support. Intended for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 2, 2017
OSSTEM IMPLANT Co., Ltd. c/o Mr. David Kim Manager HiOSSEN Inc. 85 Ben Fairless Dr. Fairless Hills. Pennsylvania 19030
Re: K161103
Trade/Device Name: US SA Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: January 6, 2017 Received: January 6, 2017
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo includes the company name and the word "IMPLANT" in orange. The address is listed as 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea, and the telephone number is +82 51 850-2500, the fax number is +82 51 850-4341, and the website is www.osstem.com.
# Indications for Use Statement
# Indications for Use
510(k) Number K 161103
Device Name : US SA Implant System
Indication for use : The US SA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. The abutment is intended for use with a dental implant fixture to provide support for prosthetic restorations such as bridges, or overdenture.
Prescription Use (Per 21CFR801 Subpart D) OR Over-The-Counter Use (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
QS-QI-505-2(Rev.0)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Osstem Implant. The logo is orange and gray. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray. The logo is simple and modern.
STEM Implant Co., Ltd.
6-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
# 510(k) Summary
Date: February 1, 2017
1. Company and Correspondent making the submission:
| - Submitter's Name : | OSSTEM Implant Co., Ltd. |
|-------------------------|-------------------------------------------------------------------------|
| - Address : | 66-16, Bansong-ro 513beon-gil, Haeundae-gu,<br>Busan, Republic of Korea |
| - Contact : | Mr. Hee Kwon Son |
| - Phone: | +82 51 850 2575 |
| - Correspondent's Name: | HIOSSEN Inc. |
| - Address: | 85 Ben Fairless Dr. Fairless Hills, PA 19030 |
| - Contact: | DAVID KIM |
| - Phone: | 267 759 7031 |
- 2. Device :
| Trade or (Proprietary) Name : | US SA Implant System |
|-------------------------------|---------------------------|
| Common or usual name : | Dental Implant |
| Classification Name : | Endosseous Dental Implant |
| Regulation Number: | 21CFR872.3640 |
| Regulatory Class: | Class II |
| Product Code: | DZE, NHA |
- 3. Predicate Device :
- Primary predicate HU II / HS II FIXTURE SYSTEM, HiOSSEN Inc, K080387
- Reference predicate
Multi Angled Abutment, OSSTEM Implant Co., Ltd., K123755 TS Fixture System, OSSTEM Implant Co., Ltd., K121995 US.SS.GS System, OSSTEM Implant Co., Ltd., K073247 3i OSSEOTITE Dental Implants, Implant Innovations, Inc, K063286 HIOSSEN Implant System, HIOSSEN Inc., K140934
# 4. Description :
The US SA Implant System is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches.
QS-QI-505-3(Rev.0)
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, gray, sans-serif font. The logo is simple and modern.
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
Fixture is made of pure titanium metal and supplied sterile. The surface is SA, Sandblasting and Acid etching, treated.
The US SA Implant System is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics.
The US SA Implant System is substantially equivalent in design, function and intended use to the HU II / HS II Fixture System ( K080387) of Osstem Implant Co., Ltd., the TS Fixture System (K121995) of OSSTEM Implant Co., Ltd., the Multi Angled Abutment (K123755) of OSSTEM Implant Co., Ltd. and HIOSSEN Implant System (K140934) of HIOSSEN Inc.
The US Multi Angled Abutment is device made of titanium alloy intended for use as an aid in prosthetic restoration.
The US Multi Angled Abutment is used with Multi Ti Cylinder or US Esthetic Low Cylinder that is predicated by 510(K), K140507 Hiossen Prosthetic system and connected to USII SA Fixture and USIII SA Fixture. The US II SA and US III SA fixtures are only compatible with the straight Cemented Abutment and straight UCLA Temporary Abutment from K073247 and the US Multi Angled Abutment included in this submission.
| Item | Content | | |
|---------------------|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| US II SA Fixture | Description | External Hex-connected<br>Submerged Fixture<br>Straight body shape<br>3 side cutting edge with self-tapping | |
| | Material | Titanium (ASTM F67) | |
| | Diameter and Length | 3.5mm X 8.5, 10, 11.5, 13, 15, 18 mm<br>4.1mm X 7, 8.5, 10, 11.5, 13, 15, 18 mm<br>4.45mm X 7, 8.5, 10, 11.5, 13, 15, 18 mm<br>4.9mm X 7, 8.5, 10, 11.5, 13, 15, 18 mm (wide)<br>4.9mm X 7, 8.5, 10, 11.5, 13, 15, 18 mm (wide ps) | |
| | US III SA Fixture | Description | External Hex-connected<br>Submerged Fixture<br>Taper body shape<br>3 sided cutting edge with self-tapping |
| | | Material | Titanium (ASTM F67) |
| Diameter and Length | | 3.75mm X 8.5, 10, 11.5, 13 mm<br>4.23mm X 8.5, 10, 11.5 mm<br>4.27mm X 7 mm<br>4.65 mm X7, 8.5, 10, 11.5, 13 mm<br>5.05mm X 10, 11.5 mm (wide)<br>5.1mm X 7, 13 mm (wide)<br>5.13mm X 8.5 mm (wide)<br>5.05mm X 10, 11.5 mm (wide ps)<br>5.1mm X 7, 13 mm (wide ps)<br>5.13mm X 8.5 mm (wide ps) | |
QS-QI-505-3(Rev.0)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image features the logo for Osstem Implant. The logo is in orange and consists of the word "OSSTEM" in a bold, sans-serif font, with the superscript symbol. Below "OSSTEM" is the word "IMPLANT" in a smaller, sans-serif font. The logo is simple and modern, and the orange color is eye-catching.
**OSSTEM Implant Co., Ltd.**
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeunda-gu, Busan, Republic of Korea
Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| Cover Screw | Description | To be used to protect the exposed platform of the implant<br>during healing period | |
|-----------------------------|-------------|----------------------------------------------------------------------------------------------|--|
| | Material | Titanium (ASTM F67) | |
| | Diameter | 6.4mm, 7.7mm | |
| | Length | 5.9mm | |
| US Multi Angled<br>Abutment | Description | US Multi Angle Abutment is used to adjust path of<br>prosthesis for multi-unit restorations. | |
| | Material | Titanium Alloy (ASTM F 136) | |
| | Component | US Multi Angled Abutment Screw. | |
| | Diameter | 3.4mm, 4.0mm, 4.1mm | |
| | Angle | 17°, 30° | |
| Multi Ti Cylinder | Description | Multi Ti Cylinder is used to make final prosthesis using<br>US Multi Angled Abutment | |
| | Material | Titanium Alloy (ASTM F 136) | |
| | Component | Consists of Multi Ti Cylinder Screw. | |
| | Diameter | 4.8mm | |
| | Length | 7.0mm | |
# - Substantial Equivalence Matrix
| | | Predicate devices | | |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Primary<br>predicate | Reference<br>predicate | Reference<br>predicate |
| | US SA Implant<br>System<br>USII SA Fixture<br>USIII SA Fixture | HUII/ HSII<br>FIXTURE<br>SYSTEM<br>HUII Fixture<br>(K080387) | TS Fixture<br>System<br>TSII/III SA<br>Fixture<br>(K121995) | ET II SA<br>Fixture<br>(K140934) |
| Design | Image: USII and USIII | Image: HUII | Image: TSII and TSIII | Image: ETII |
| Intended use | The US SA Implant<br>System is indicated<br>for use in partially or<br>fully edentulous<br>mandibles and<br>maxillae, in support<br>of single or multiple-<br>unit restorations<br>including: cemented | The HUII / HSII<br>Fixture System is<br>indicated for use in<br>partially or fully<br>edentulous<br>mandibles and<br>maxillae, in support<br>of single or multiple-<br>unit restorations | The TS Fixture<br>System is indicated<br>for use in partially or<br>fully edentulous<br>mandibles and<br>maxillae, in support<br>of single or multiple-<br>unit restorations<br>including : cemented | The ETII Fixture<br>System is indicated<br>for use in partially or<br>fully edentulous<br>mandibles and<br>maxillae, in support<br>of single or multiple-<br>unit restorations<br>including: cemented |
| | retained, screw<br>retained, or<br>overdenture<br>restorations, and final<br>or temporary<br>abutment<br>support for fixed<br>bridgework. It is<br>intended for delayed<br>loading.<br>The abutment is<br>intended for use with<br>a dental implant<br>fixture to provide<br>support for prosthetic | including;<br>cemented retained,<br>screw retained, or<br>overdenture<br>restorations, and<br>terminal or<br>intermediate<br>abutment support for<br>fixed bridgework.<br>The HU II / HS II<br>Fixture System is for<br>one and two stage<br>surgical procedures.<br>It is not for<br>immediate load. | retained, screw<br>retained, or<br>overdenture<br>restorations, and<br>final or temporary<br>abutment<br>support for fixed<br>bridgework. It is<br>intended for delayed<br>loading | retained, screw<br>retained, or<br>overdenture<br>restorations, and<br>final or temporary<br>abutment support for<br>fixed bridgework. It<br>is intended for<br>delayed loading.<br>Ultra wide Fixture<br>System is intended to<br>be used in the molar<br>region. |
| | restorations such as<br>bridges, or<br>overdenture. | | | |
| Surgery type | One and two stage<br>Surgery | One and two stage<br>Surgery | One and two stage<br>Surgery | One and two stage<br>Surgery |
| Structure | - External Hex-<br>connected<br>- Submerged Fixture<br>- Straight / Taper<br>body shape<br>- 3 sided cutting edge<br>with self-tapping | -Submerged Fixture<br>-Body type : Straight<br>Body<br>-Self tapping<br>-Connection :<br>External hex-<br>connected | - Internal Hex-<br>connected<br>- Submerged Fixture<br>- Straight / Taper<br>body shape<br>- 3 sided cutting edge<br>with self-tapping | - Internal Hex-<br>connected<br>- Submerged Fixture<br>- Straight body<br>shape<br>- 3 sided cutting<br>edge with self-<br>tapping |
| Platform<br>(D) | 3.6, 4.2, 5.1, 5.2 | 3.5~5.1 | N/A | N/A |
| Body<br>Diameter<br>(D) | 3.5, 4.1, 4.45, 4.9<br>3.75, 4.23, 4.27,<br>4.65, 5.05, 5.1, 5.13 | 3.5~5.5 | TS II SA: 3.5, 4.2,<br>4.4, 4.9<br>TS III SA: 3.77,<br>3.75, 4.25, 4.2, 4.65,<br>4.63, 4.6, 5.1, 5.08,<br>5.05, | 3.5, 4.2, 4.45, 4.9,<br>5.0 |
| Length (mm) | 7, 8.5, 10, 11.5, 13,<br>15, 18 | 7.0~15.0 | 7.0~15 | 6.2 ~ 18.2 |
| Material of<br>Fixture | Pure Titanium Grade<br>4<br>(ASTM F67) | Pure Titanium Grade<br>4<br>(ASTM F67) | Pure Titanium Grade<br>4<br>(ASTM F67) | Pure Titanium Grade<br>4<br>(ASTM F67) |
| Surface | SA | RBM | SA | SA |
| Sterilization | Radiation Sterile | Radiation Sterile | Radiation Sterile | Radiation Sterile |
| Shelf life | 8years | 3 years | 8 years | 8 years |
| SE | USII SA Fixture and USIII SA Fixture have the same material, indication for use,<br>connection structure as the HUII Fixture in the HU II / HS II Fixture System<br>(K080387) except surface treatment.<br>surface treatment of USII SA Fixture is different from Primary predicate, HUII<br>Fixture but the surface treatment of USII SA Fixture and USIII SA Fixture is the<br>same with surface treatment of TS Fixture System (K121995) | | | |
{6}------------------------------------------------
K161103
Image /page/6/Picture/1 description: The image contains the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a bold, sans-serif font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, gray, sans-serif font. There is a small orange circle with a degree symbol in the upper right corner of the logo.
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeunda-gu, Busan, Republic of Korea
Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The logo is orange and gray. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray. The logo is simple and modern.
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
And Diameter size and length of US SA Implant System are included dimension range of predicate, HUII Fixture in the HU II / HS II Fixture System (K080387) and ET II SA Fixture (K140934) Therefore we state that US SA Implant System is substantially equivalent to the predicate devices as described herein.
| | US Multi Angled<br>Abutment | Predicate devices | |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Primary predicate<br>Multi Angled Abutment | Reference predicate<br>Cement Abutment |
| Manufacturer | Osstem Implant Co., Ltd | Osstem Implant Co., Ltd | Osstem Implant Co., Ltd |
| 510(k) Number | New | K123755 | K073247 |
| Design | Image: US Multi Angled Abutment Design | Image: Multi Angled Abutment Design | Image: Cement Abutment Design |
| Intended use | US Multi Angled<br>Abutment is intended for<br>use with a dental implant<br>to provide support for<br>prosthetic restorations<br>such as bridges, or<br>overdentures. | Multi Angled Abutment<br>is intended for use with a<br>dental implant to provide<br>support for prosthetic<br>restorations such as<br>bridges, or overdentures. | Cement Abutment is<br>intended for use with a<br>dental implant to provide<br>support for prosthetic<br>restorations such as<br>crowns, bridges, or<br>overdentures. |
| Abutment<br>Angle(°) | 17, 30 | 17, 30 | 0 |
| platform(Ø) | 4.9 | 4.8 | - |
| connection | External Hex-connected<br>Compatible with USII<br>SA Fixture | Internal Hex-connected<br>Compatible with TSII SA<br>Fixture | External Hex-connected<br>Compatible with USII<br>SA Fixture |
QS-QI-505-3(Rev.0)
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, with the letters slightly overlapping each other. Below "OSSTEM" is the word "IMPLANT" in gray. To the right of the word "OSSTEM" is a small orange circle.
STEM Implant Co., Ltd
Bansong-ro 513beon-gil. Haeundae-gu. Busan, Republic of Korea 1: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| Connection | US Multi Angled Abutment has the same material, indication for use, function and<br>similar design as the Multi Angled Abutment (K123755) except connection<br>structure.<br>But connection structure is exactly same with Cement Abutment, K073247<br>In accordance with difference of connection and shape, we conducted fatigue test<br>to confirm proposed device strength |
|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### 5. Indication for use :
The US SA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including ; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading.
The abutment is intended for use with a dental implant fixture to provide support for prosthetic restorations such as bridges, or overdenture.
# 6. Review :
The US SA Implant System has same material and indication for use and similar design and technological characteristics as the predicate device.
### 7. Summary of nonclinical testing
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence:
Fatigue testing was conducted with subject devices according to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment".
Gamma Sterilization Validation was performed with Fixture System according to ISO 11137-1, ISO 11137-2 and ISO 11137-3. Steam sterilization validation was provided according to ISO 17665-1 and ISO 17665-2.
Subject devices are made of the same materials, manufacturing process, chemical composition, and body contact with the predicate devices, the TS Fixture System, OSSTEM Implant Co., Ltd., K121995 and Multi Angled Abutment, OSSTEM Implant Co., Ltd., K123755
therefore US SA Implant System is substantially equivalent to the predicate devices in biocompatibility, surface coating characterization
# 8. Summary of clinical testing
No clinical studies are submitted
- 9. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Osstem Implant Co., Ltd. concludes that the US SA Implant System is substantially equivalent to the predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.