Hiossen Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
Device Story
Hiossen Prosthetic System comprises dental abutments, overdenture components, and abutment screws; used to support prosthetic restorations (crowns, bridges, over-dentures) on dental implants. Components include ET Custom Healing Abutments (PEEK), NP-Cast Abutments (Co-Cr-Mo alloy/POM), O-Ring Retainer sets (Titanium/Polymer), and Esthetic Low Gold Cylinders (Gold alloy/POM). Used by dental professionals in clinical settings to facilitate soft tissue shaping or prosthetic attachment. Device functions as mechanical interface between implant fixture and final restoration. Benefits include restoration of dental function and aesthetics. No software or AI components involved.
Clinical Evidence
No clinical studies submitted. Substantial equivalence supported by non-clinical bench testing, including biocompatibility and retention testing for modified O-ring materials.
Technological Characteristics
Components: PEEK, Co-Cr-Mo alloy, Gold alloy, Titanium, POM, and Acrylonitrile-Butadiene polymer. Mechanical interface for dental implants. Sterilization via blister/Tyvek packaging (validated for 8-year shelf life). No software, electronics, or energy sources.
Indications for Use
Indicated for patients requiring prosthetic restoration (crowns, bridges, or over-dentures) supported by dental implant fixtures.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure, with three overlapping profiles facing to the right. The profiles are rendered in a dark color, contrasting with the white background of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Auqust 20, 2014
OSSTEM IMPLANT Company, Ltd. C/O Mr. Patrick Lim Manager 85 Ben Fairless Dr Fairless Hills, PA 19030
Re: K140507
Trade/Device Name: Hiossen Prosthetic System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous dental implant abutment Regulatory Class: Class II Product Code: NHA Dated: July 17, 2014 Received: July 23, 2014
Dear Mr. Lim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Lim
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mary S. Runner -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The address is #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea. The telephone number is +82 51 850-2500, and the fax number is +82 51 850-4341. The website address is www.osstem.com.
## Indications for Use Statement
### Indications for Use
510(k) Number K _140507
Device Name : Hiossen Prosthetic system
Indication for use : Hiossen Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
Prescription Use X (Per 21CFR801 Subpart D) OR Over-The-Counter Use (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
QS-QI-505-2(Rev.0)
{3}------------------------------------------------
Section 010
### OSSTEM Implant Co., Ltd
#507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: December 13, 2013
1. Company and Correspondent making the submission:
- Submitter's Name :
- Address :
- Contact :
- Phone:
- Correspondent's Name:
- Address:
- Contact:
- Phone:
. 2. Device :
Trade or (Proprietary) Name : Common or usual name :
Classification Name :
OSSTEM Implant Co., Ltd. #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804, Republic of Korea Mr. Hee Kwon Son +82 51 850 2575
HIOSSEN Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 Patrick Lim 888 678 0001
Hiossen Prosthetic system Dental Abutment Endosseous dental implant abutment 21CFR872.3630 Class II NHA
#### 3. Predicate Device:
HU.HS.HG Prosthetic System, OSSTEM Implant Co., Ltd., K081575 TS implant system, OSSTEM Implant Co., Ltd., K121585 NP-cast abutment system, OSSTEM Implant Co., Ltd., K121843 Prosthetic system, OSSTEM Implant Co., Ltd., K110308 US.SS.GS System, OSSTEM Implant Co., Ltd., K073247 TS Fixture system, OSSTEM Implant Co., Ltd., K121995
#### 4. Description :
The Hiossen Prosthetic system is intended for use as an aid in prosthetic restoration. It consists of Abutments, overdenture components and Abutment Screws
The Hiossen Prosthetic system is similar to other commercially available products based on
QS-QI-505-3(Rev.0)
{4}------------------------------------------------
### OSSTEM Implant Co., Ltd.
STEM® #507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
the intended use, the technology used, the claims, the material composition employed and performance characteristics.
The Hiossen Prosthetic system is substantially equivalent in design, function and intended use to the HU.HS.HG Prosthetic System (K081575), TS implant system (K121585), NP-cast abutment system (K121843), US.SS.GS System, (K073247) and Prosthetic system (K110308)
These proposed abutments are compatible with the HIOSSEN Implant fixture only. HTIII SA Fixture (K101096)
| Item | Content |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ET Custom Healing<br>Abutment | Description<br>Used after successful osseointegration and removing cover screw to make a soft tissue shape before loading prosthetics. |
| | Material<br>PEEK |
| | Diameter<br>7.05mm/ 5.0mm (Oval in shape) |
| | Height<br>7mm, 9mm, 11mm, 7.5mm, 9.5mm, 11.5mm |
| NP-Cast Abutment | Description<br>NP-Cast Abutment is used for cases with path and aesthetic and spatial constraints<br>After customization, be sure to use only dental non-precious metal alloy for casting to make the prosthesis |
| | Material<br>Co-Cr-Mo Alloy + POM |
| | Diameter<br>4.0mm, 4.5mm |
| | Height<br>10.5mm, 15.5mm, 13.71mm, 15.71mm, 13mm, 15mm |
| O-Ring | Description<br>Use for making stud-type overdenture |
| | Material<br>Polymer |
| | Diameter<br>4.6mm |
| O-Ring Retainer Cap set | O-Ring Retainer Cap set consist of Retainer and O-ring<br>Retainer Cap is cleared K81575 |
| O-Ring Retainer set | O-Ring Retainer set consist of Retainer and O-ring<br>Retainer is cleared K81575 |
| Esthetic Low Gold<br>Cylinder | Description<br>Use for making screw-retained prosthesis<br>After customization, be sure to use only dental gold alloy for casting to make the prosthesis |
| | Material<br>Gold Alloy + POM |
| | Diameter<br>4.8mm |
| | Height<br>12mm |
QS-QI-505-3(Rev.0)
{5}------------------------------------------------
## OSSTEM Implant Co., Ltd.
OSSTEM®
#507-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea
Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
#### - Substantial Equivalence Matrix
### < ET Custom Healing abutment>
| | ET Custom Healing<br>abutment | Predicate devices | |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Part Name | | Healing Abutment | Quick Temporary<br>Abutment |
| 510K | Proposed | K081575 | K121585 |
| Material | PEEK | Titanium | PEEK |
| Manufacturer | OSSTEM Implant Co.,<br>Ltd. | OSSTEM Implant Co.,<br>Ltd. | OSSTEM Implant Co.,<br>Ltd. |
| Description<br>(Intended for<br>use) | Used to make a soft<br>tissue shape before<br>setting up prosthetics and<br>removing cover screw<br>after osseointegration. | Used to make a soft<br>tissue shape before<br>setting up prosthetics and<br>removing cover screw<br>after osseointegration | It is used temporary until<br>final prosthesis is made<br>to maintain esthetic<br>Apperance |
| Indication for<br>use | Hiossen Prosthetic<br>system is intended for<br>use with a dental implant<br>to provide support for<br>prosthetic restorations<br>such as crowns, bridges,<br>or overdentures. | HU/HS/HG Prosthetic<br>System is intended for<br>use as an aid in<br>prosthetic<br>restoration. | The abutment is intended<br>for use with a dental<br>implant fixture to<br>provide support for<br>prosthetic restorations<br>such as crowns, bridges,<br>or overdenture. |
| Design | Image: ET Custom Healing abutment design | Image: Healing Abutment design | Image: Quick Temporary Abutment design |
| SE | ET Custom Healing abutment has tha same "Inatended for use" with predicate<br>device, Healing Abutment (K081575) but material is difference<br>Materier of ET Custom Healing abutment is the same with predicate device,<br>Quick Temporary Abutment (K121585) | | |
#### < Substantial equivalence of ET Custom Healing Abutment Package>
Package for ET Custom Healing Abutment to keep sterile is Blister & Tyvek package Package of ET Custom Healing Abutment, has the same material and contact area as package for TS Fixture System that has already been cleared by 510(K), K121995 with shelf life of 8years. Therefore, shelf life of 8years for TS Custom Healing Abutment was submitted
QS-QI-505-3(Rev.0)
{6}------------------------------------------------
### STEM Implant Co.,
OSTEON
IMPLANT #50
07-8 Geoje3-Dong Yeonje-Gu Busan, 611-804 Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| Part Name | O-Ring Retainer Cap Set<br>O-Ring Retainer Set<br>O-ring | Predicate devices<br>O-Ring Retainer Cap Set<br>O-Ring Retainer Set<br>O-ring |
|--------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| 510K | Proposed | K081575 |
| Material | Retainer: Titanium<br>O-Ring : Acrylonitrile & Butadiene<br>Polymer | Retainer: Titanium<br>O-Ring : Silicone |
| Manufacturer | OSSTEM Implant Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Intended use | Use for marking stud-type overdenture | Use for marking stud-type overdenture |
| Design | Image: O-Ring Retainer Cap Set Design | Image: O-Ring Retainer Cap Set Design |
| SE | O-Ring Retainer Cap, O-Ring Retainer and O-Ring have a cleared with 510(K),<br>K81575 but material of O-ring is changed | |
#### 5. Indication for use :
Hiossen prosthetic system is intended for use with a dental implant fixture to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
- 6. Review :
Hiossen prosthetic system has same indication for use and technological characteristics as the predicate device.
Hiossen prosthetic system has been subjected to safety, performance, and product validations prior to release. Safety tests including biocompatibility have been considered to ensure the devices comply with the applicable International and US regulations.
- 7. Summary of nonclinical testing
Proposed devices in this submission are substantially equivalent to the predicate devices Therefore we didn't consider conducting additional test but biocompatibility test and retention test of O-Ring are conducted because material of O-Ring is changed
- 8. Summary of clinical testing No clinical studies are submitted
- 9. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Osstem Implant Co., Ltd. concludes that the Hiossen prosthetic system is substantially equivalent to the predicate devices as described herein.
QS-QI-505-3(Rev.0)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.