The ATLANTIS™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the ATLANTIS Abutment to the endosseous implant. The ATLANTIS™ Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS Crown Abutment to the endosseous implant. The ATLANTIS™ Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment screw is intended to secure the ATLANTIS Conus Abutment to the endosseous implant.
Device Story
Patient-specific endosseous dental implant abutments; fabricated via CAD/CAM technology at Dentsply sites based on clinician prescription. Inputs: clinician-provided anatomical/restorative data. Device transforms inputs into custom-designed titanium alloy abutments (straight or up to 30° angulation). Used in dental clinics by dentists/specialists to support crowns, bridges, or overdentures. Output: physical abutment secured to implant via screw; serves as substructure for final restoration or as final restoration itself (Crown Abutment). Benefits: provides patient-specific prosthetic interface for dental implants, restoring function and aesthetics in edentulous patients.
Clinical Evidence
Bench testing only. Evidence includes mechanical design analysis, dimensional analysis, and static/dynamic compression-bending testing per ISO 14801. Compatibility analysis confirms fit with CONELOG implant platforms (3.3, 3.8, 4.3, 5.0 mm). No clinical data provided.
Technological Characteristics
Materials: Titanium alloy (ASTM F136). Design: Patient-specific CAD/CAM fabricated abutments. Interface: Internal conical connection with indexing (three grooves). Dimensions: 3.3-13 mm diameter, 4-15 mm height, 0-30° angulation. Connectivity: None. Sterilization: Same cycle as predicate.
Indications for Use
Indicated for partially or completely edentulous patients requiring prosthetic support via endosseous dental implants. Supports single or multiple tooth prostheses (cemented, screw-retained, or friction-fit) in the mandible or maxilla.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image. The logo is simple and recognizable, conveying the department's authority and purpose.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 26, 2016
Dentsply Sirona Ms. Helen Lewis Director Corporate Regulatory Affairs 221 West Philadelphia Street, Suite 60 York, Pennsylvania 17404
Re: K161030
Trade/Device Name: ATLANTISTM Abutment for CONELOG Implant Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: July 27, 2016 Received: July 29, 2016
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161030
Device Name
ATLANTISTM Abutment for CONELOG implant
## Indications for Use (Describe)
The ATLANTIS™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment screw is intended to secure the ATLANTIS Abutment to the endosseous implant.
The ATLANTIS™ Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS Crown Abutment to the endosseous implant.
The ATLANTIS™ Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment screw is intended to secure the ATLANTIS Conus Abutment to the endosseous implant.
ATLANTISTM products are compatible with the implants shown in the table below.
| Implant Manufacturer | Trade Name | Implant Diameter | Abutment Platform Diameter |
|----------------------|--------------------------------|------------------------|----------------------------|
| Altatec GmbH | CONELOG SCREW-<br>LINE Implant | Ø3.3, 3.8, 4.3, 5.0 mm | Ø3.3, 3.8, 4.3, 5.0 mm |
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404
Image /page/3/Picture/1 description: The image shows the logo for Dentsply Sirona. The logo consists of a stylized leaf-like shape in gray on the left, with a white stripe running through it. To the right of the leaf shape are the words "Dentsply" and "Sirona" stacked vertically, also in gray. The font is sans-serif and modern.
# SECTION 5. 510(k) SUMMARY
for
# ATLANTISTM Abutment for CONELOG implant
- 1. Submitter Information:
Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404
| Contact Person: | Helen Lewis |
|-------------------|--------------|
| Telephone Number: | 717-487-1332 |
| Fax Number: | 717-849-4343 |
August 18, 2016 Date Prepared:
- 2. Device Name:
- Proprietary Name:
- Classification Name: ●
- CFR Number: ●
- Device Class: .
- Product Code: NHA .
- 3. Predicate Device:
| Predicate Device Name | 510(k) | Company Name |
|-------------------------------------------------------------------------|---------|-----------------------------------------|
| ATLANTIS™ Abutment for<br>NobelActive 3.0 (Primary<br>Predicate Device) | K151039 | DENTSPLY IMPLANTS<br>MANUFACTURING GMBH |
| Osseospeed™ Profile EV<br>(Reference Predicate Device) | K130999 | DENTSPLY IMPLANTS<br>MANUFACTURING GMBH |
| CONELOG Implant System<br>(Reference Predicate Device) | K113779 | Altatec GmbH |
21 CFR 872.3630
II
ATLANTISTM Abutment for CONELOG implant
Endosseous dental implant abutment
- 4. Description of Device:
The proposed ATLANTIS™ Abutment for CONELOG implant is an endosseous dental implant abutment. The subject device is provided for implant diameter (Ø3.3, 3.8, 4.3 and 5.0 mm) and three designs: ATLANTIS™ Abutment for CONELOG implant, ATLANTIS™ Crown Abutment for CONELOG implant and ATLANTIS™ Conus Abutment for CONELOG implant, see table 5-1. All are patient-specific abutments fabricated using CAD/CAM technology at DENTSPLY Implant sites. Each abutment is designed according to prescription instructions from the clinician to support a screwretained, cement-retained or friction fit prosthesis.
{4}------------------------------------------------
Table 5-1: Compatibility table (The ATLANTIS Abutment Titanium and Gold-shaded are compatible with CONELOG implant interface (Ø3.3, 3.8, 4.3 and 5.0 mm)
| Implant<br>Manufacturer | Interface | ATLANTIS<br>TM<br>Abutment for<br>CONELOG<br>implant | ATLANTISTM<br>Crown<br>Abutment<br>for CONELOG<br>implant | ATLANTISTM<br>Conus<br>Abutment for<br>CONELOG<br>implant<br>(Custom) | ATLANTISTM<br>Conus Abutment<br>for CONELOG<br>implant<br>(Overdenture) |
|-------------------------|------------------------------------------------------------------------------------|------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------------------|-------------------------------------------------------------------------|
| Altatec GmbH | CONELOG<br>Implant<br>System<br>(SCREW-<br>LINE) (Ø3.3,<br>3.8, 4.3 and<br>5.0 mm) | Titanium,<br>Gold-shaded<br>titanium<br>(Gold-Hue) | Titanium | Titanium | Titanium |
The coronal portion of the ATLANTIS™ Abutment can be fabricated as a conventional abutment for prosthesis attachment (ATLANTIS™ Abutment or ATLANTIS™ Conus Abutment) or fabricated as a single tooth final restoration onto which porcelain is added (ATLANTIS™ Crown Abutment). The ATLANTIS™ abutment interface is compatible with the CONELOG implants from the CONELOG Implant System (K113779).
The CONELOG implant interface is an internal connection with indexing feature (three grooves) and provided for implant platform diameter (3.3, 3.8, 4.3 and 5.0 mm). The abutment diameter ranges from 3.3 to 13 mm, the maximum abutment height is 15 mm above implant interface and the minimum abutment height is 4 mm above the transmucosal collar. The abutment is provided straight and up to 30° of angulation.
- న. Indications for Use:
The ATLANTIS™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in mandible or maxilla. The prosthesis can be cemented or screw retained to the abutment. The abutment screw is intended to secure the ATLANTIS Abutment to the endosseous implant.
The ATLANTIS™ Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in a partially or completely edentulous patient. The abutment screw is intended to secure the ATLANTIS Crown Abutment to the endosseous implant.
The ATLANTIS™ Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS Conus Abutment to the endosseous implant.
{5}------------------------------------------------
ATLANTIS™ products are compatible with the implants shown in the table below.
| Implant<br>Manufacturer | Trade Name | Implant Diameter | Abutment Platform Diameter |
|-------------------------|--------------------------------|------------------------|----------------------------|
| Altatec GmbH | CONELOG SCREW-<br>LINE Implant | Ø3.3, 3.8, 4.3, 5.0 mm | Ø3.3, 3.8, 4.3, 5.0 mm |
- 6. Substantial Equivalence:
Technological Characteristics
ATLANTIS™ Abutment for CONELOG implant is a patient specific restorative device designed under the control of DENTSPLY and manufactured by DENTSPLY using CAD/CAM technology.
Table 5-2 and 5-3 below summarizes the differences and similarities of the subject and predicate devices.
{6}------------------------------------------------
Table 5-2: Indications for use for the proposed and the predicate devices
| ATLANTISTM Abutment for<br>CONELOG implant (Proposed | ATLANTISTM Abutment<br>for NobelActive 3.0 | OsseoSpeed Profile EV<br>(K130999) | CONELOG Implant System<br>(K113779) | Summary of differences in<br>the indications for use |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device) | (K151039)<br>(Primary Predicate) | (Reference Predicate) | (Reference Predicate) | |
| The ATLANTISTM Abutment is<br>intended for use with an endosseous<br>implant to support a prosthetic device<br>in a partially or completely<br>edentulous patient. It is intended for<br>use to support single and multiple<br>tooth prosthesis, in mandible or<br>maxilla. The prosthesis can be<br>cemented or screw retained to the<br>abutment. The abutment screw is<br>intended to secure the ATLANTIS<br>Abutment to the endosseous implant.<br>The ATLANTISTM Crown<br>Abutment is intended for use with an<br>endosseous implant to function as a<br>substructure that also serves as the<br>final restoration, in a partially or<br>completely edentulous patient. The<br>abutment screw is intended to secure<br>the ATLANTIS Crown Abutment to<br>the endosseous implant.<br>The ATLANTISTM Conus<br>Abutment is intended for use with an<br>endosseous implant to support a<br>prosthetic device in partially or<br>completely edentulous patients. It is<br>intended for use to support a<br>removable multiple tooth prosthesis,<br>in the mandible or maxilla. The<br>prosthesis is attachment-retained by<br>friction fit to the abutment. The<br>abutment screw is intended to secure<br>the ATLANTIS Conus Abutment to<br>the endosseous implant. | The ATLANTISTM<br>Abutment is intended for<br>use with an endosseous<br>implant to support a<br>prosthetic device in a<br>partially or completely<br>edentulous patient. It is<br>intended for use to support<br>single and multiple tooth<br>prosthesis, in the mandible<br>or maxilla. The prosthesis<br>can be cemented or screw<br>retained to the abutment.<br>The abutment screw is<br>intended to secure the<br>ATLANTISTM<br>Abutment to the endosseous<br>implant.<br>The ATLANTISTM Crown<br>Abutment is intended for<br>use with an endosseous<br>implant to function as a<br>substructure that also serves<br>as the final restoration, in a<br>partially or completely<br>edentulous patient. The<br>abutment screw is intended<br>to secure the ATLANTISTM<br>Crown<br>Abutment to the endosseous<br>implant.<br>The ATLANTISTM Conus<br>Abutment is intended for<br>use with an endosseous<br>implant to support a<br>prosthetic device in partially | Implants:<br>The ASTRA TECH Implant<br>System – OsseoSpeed Profile<br>EV implants are intended for<br>both one- and two-stage<br>surgical procedures in the<br>following situations and with<br>the following clinical<br>protocols:<br>Replacing missing teeth<br>in single or multiple unit<br>applications in the<br>mandible or maxilla. Immediate placement in<br>extraction sites and in<br>situations with a partially<br>or completely healed<br>alveolar ridge Especially indicated for<br>us in soft bone<br>applications where<br>implants with other<br>implant surface<br>treatment may be less<br>effective Immediate and early<br>loading for all<br>indications Together with immediate<br>loading protocol in all<br>indications, except in<br>single tooth situations in<br>soft bone (type IV)<br>where implant stability<br>may be difficult to<br>obtain and immediate<br>loading may not be | CONELOG® Implant System<br>Implants are intended for immediate<br>or delayed placement in the bone of<br>the maxillary or mandibular arch.<br>CONELOG® Implant System<br>Abutments are intended for use as<br>support for crowns, bridges or<br>overdentures. When a one-stage<br>surgical approach is applied, the<br>implant may be immediately loaded<br>when good primary stability is<br>achieved and the functional load is<br>appropriate.<br>CONELOG® Implants with Ø3.3<br>mm diameter have the following<br>additional specific indications:<br>These are an alternative in cases<br>where the alveolar ridge width is<br>only 5 - 6 mm. Because of their<br>lower mechanical strength<br>compared with larger diameter<br>implants, they should only be used<br>under the following conditions:<br>As single implants, they<br>should be used only to replace<br>mandibular incisors and/or<br>maxillary lateral incisors. An edentulous arch can only<br>be restored with a bar retained<br>superstructure with at least<br>four implants of Ø3.3 mm<br>diameter without distal<br>extensions. Implants of Ø3.3 mm are<br>suitable for a partially<br>edentulous arch when | The indications for use of the<br>proposed device are identical<br>to the indications for use of<br>the primary predicate device.<br>The indications for use of the<br>proposed device are similar to<br>the indications for use of the<br>reference predicate device<br>(K130999). The difference is<br>that the reference predicate<br>device (K130999) also<br>includes the indications for<br>use of the implants, which is<br>not the subject of this<br>submission.<br>The indications for use of the<br>second reference predicate<br>device cover the CONELOG<br>implant system (K113779).<br>The indications for use are<br>similar. The difference<br>between the proposed device<br>and the reference predicate<br>device (K113779) is that the<br>prosthesis, in addition to<br>screw-retained restoration or<br>cement-retained restoration,<br>can be attachment-retained<br>(by friction-fit) to the<br>proposed device. |
{7}------------------------------------------------
ATLANTIS™ products are compatible with the implants shown in the table below.
| Implant<br>Manufacturer | Trade<br>Name | Implant<br>Diameter | Abutment<br>Platform<br>Diameter |
|-------------------------|--------------------------------------|------------------------------|----------------------------------|
| Altatec GmbH | CONELOG<br>SCREW-<br>LINE<br>Implant | Ø3.3, 3.8,<br>4.3, 5.0<br>mm | Ø3.3, 3.8,<br>4.3, 5.0<br>mm |
or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS™ Conus Abutment to the endosseous implant.
ATLANTIS™ Abutment for NobelActive 3.0 is compatible with the NobelActive 3.0 implant.
appropriate
●
- Only together with Profile EV components, Implant Driver Profile EV, Radiographic Implant Guides Profile EV and non-Indexed prosthetic components
### Abutments:
ASTRA TECH Implant System™-OsseoSpeed EV abutments are intended to be used in conjunction with ASTRA TECH Implant System™-OsseoSpeed EV in fully edentulous or partially edentulous maxillary and/or mandibular arches.
## The ATLANTISTM
Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.
The ATLANTIS™ Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in partially or completely edentulous patients. The prosthesis is
combined with implants of larger diameter for splinted superstructures. However, the limited strength of the implants with Ø3.3 mm must be taken into account.
- . Avoid excessive mechanical stressing of the implants when using ball abutments in combination with Ø3.3 mm implants.
- . The healing time for Ø3.3 mm implants is at least 12 weeks.
CONELOG® Implants with 7 mm length have the following additional specific indications:
### CONELOG® SCREW-LINE
Implants should only be used when there is not enough space for a longer implant. Delayed loading in single tooth replacement is indicated with these implants. If the ratio of crown length to implant length is unfavorable the biomechanical risk factors have to be considered and appropriate measures have to be taken by the dental professional.
{8}------------------------------------------------
| | screw retained. The abutment<br>screw is intended to secure the<br>ATLANTIS™ Abutment to<br>the endosseous implant.<br>The ATLANTIS Conus<br>Abutment is intended for use<br>with an endosseous implant to<br>support a prosthetic device in<br>partially or completely<br>edentulous patients. It is<br>intended for use to support a<br>removable multiple tooth<br>prosthesis, in the mandible or<br>maxilla. The prosthesis is<br>attachment-retained by<br>friction fit to the abutment.<br>The abutment screw is<br>intended to secure the<br>ATLANTIS Abutment to the<br>endosseous implant.<br><br>ATLANTISTM Abutment,<br>ATLANTIS™ Crown<br>Abutment and ATLANTISTM<br>Conus Abutment are<br>compatible with ASTRA<br>TECH Implant System Profile<br>EV | |
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
{9}------------------------------------------------
Table 5-3: Similarities and differences between the proposed and the predicate devices
| Element | Proposed Device<br>ATLANTISTM<br>Abutment for<br>CONELOG implant | Primary Predicate<br>Device<br>ATLANTISTM<br>Abutment for<br>NobelActive 3.0 | Reference Predicate<br>Device<br>OsseoSpeed Profile<br>EV | Reference Predicate<br>Device<br>CONELOG Implant<br>System | Summary of differences |
|-------------------------------|------------------------------------------------------------------|------------------------------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | To be assigned | K151039 | K130999 | K113779 | - |
| Prosthesis<br>Attachment | Screw-retained<br>Cement-retained<br>Friction Fit | Screw-retained<br>Cement-retained<br>Friction Fit | Screw-retained<br>Cement-retained<br>Friction Fit | Screw-retained<br>Cement-retained | No difference between the<br>proposed device and the primary<br>predicate and the reference device<br>(K130999). |
| Restoration | Single or<br>Multi-unit | Single or<br>Multi-unit | Single or | Single or<br>Multi-unit | No difference. |
| Abutment Platform<br>Diameter | 3.3, 3.8, 4.3, 5.0 | 3.0 | 3.0, 4.2, 4,8 | 3.3, 3.8, 4.3, 5.0 | The proposed abutment platform<br>is designed to fit the reference<br>predicate device (K113779).<br>The proposed device<br>encompasses the same range of<br>physical dimensions; diameter,<br>height and angle of abutments for<br>the primary predicate as for the<br>reference predicate (K113779).<br>The subject abutment platform is<br>not designed to fit the primary<br>predicate device (K151039) or the<br>reference predicate device<br>(K130999). |
| Abutment angle | Straight, up to 30° | Straight, up to 30° | Straight, up to 30° | Straight, up to 30° | No difference. |
| Element | Proposed Device<br>ATLANTISTM<br>Abutment for<br>CONELOG implant | Primary Predicate<br>Device<br>ATLANTISTM<br>Abutment for<br>NobelActive 3.0 | Reference Predicate<br>Device<br>OsseoSpeed Profile<br>EV | Reference Predicate<br>Device<br>CONELOG Implant<br>System | Summary of differences |
| Connection | Internal conical<br>connection | Internal conical connection | Internal conical<br>connection | Internal conical<br>connection | The internal conical connection is<br>similar for the proposed device<br>when compared to the primary<br>predicate device (K151037) or the<br>reference predicate device<br>(K130999).<br>The connection of the proposed<br>device is designed to fit the<br>reference predicate (K113779). |
| Material Implant | NA | NA | Titanium | Titanium | The proposed device and primary<br>predicate is an abutment. |
| Material: Abutment | Titanium alloy | Titanium alloy | Titanium alloy | Titanium alloy | No difference. |
| Material: Screw | Titanium alloy | Titanium alloy | Titanium alloy | Titanium alloy | No difference. |
{10}------------------------------------------------
{11}------------------------------------------------
#### 7. Biocompatibility
The results of biocompatibility testing conducted for the primary predicate device, ATLANTIS™ Abutment for NobelActive 3.0 (K151039), are valid because the material composition of the proposed device is the same when compared to the primary predicate device, ATLANTISTM Abutment for NobelActive 3.0 (K151039). Atlantis abutments are made of titanium allov conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4Vandadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). As per our biological evaluations on the titanium materials used for the ATLANTIS Abutment (RP-00054490), the following tests were performed: Cytotoxicity, Sensitization, Irritation of Intracutaneous reactivity, Acute systemic toxicity, Subchronic toxicity, Genotoxicity - AMES, and Hemolysis. Since the material composition of the proposed devices is identical to the predicate devices, the result of referenced biocompatibility testing is still valid. Therefore, no additional biocompatibility testing has been performed.
#### Sterility 8.
The sterility testing conducted for the primary predicate device, ATLANTIS™ Abutment for NobelActive 3.0 (K151039), was conducted with the same materials and same sterilization cycle. Therefore, no additional sterility testing was required for the proposed device, ATLANTIS™ Abutment for CONELOG implant.
#### 9. Non-Clinical Performance Data
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence includes mechanical design analysis, dimensional analysis, and static and dynamic compression-bending testing according to ISO 14801 Dental-implants Dynamic Fatigue Test for Endosseous Dental Implants.
Compatibility analysis shows that the ATLANTIS™ Abutment for CONELOG implant is compatible with the (3.3, 3.8, 4.3 and 5.0 mm) CONELOG implants. Mechanical testing results show that the ATLANTIS™ Abutment for CONELOG implant has sufficient strength for its intended use.
# 9. Conclusion Regarding Substantial Equivalence
The ATLANTIS™ Abutment for CONELOG implant is an endosseous dental implant abutment which is intended to support a prosthetic device in a partially or completely edentulous patient. The ATLANTIS™ Abutment for CONELOG implant has the same intended use, incorporates the same fundamental technology, and has similar indications for use as the predicates ATLANTIS™ Abutment for NobelActive 3.0 (K151039), Osseospeed Profile EV (K130999) and CONELOG Implant System (K113779).
Test data to verify the performance of the ATLANTIS™ Abutment for CONELOG implant has been provided including: mechanical testing and compatibility analysis. The results of this testing, combined with the design, and intended use comparison with the predicate devices, support substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.