Implants: The ASTRA TECH Implant System - OsseoSpeed Profile EV implants are intended for both one- and two-stage surgical procedures in the following situations and with the following clinical protocols: · replacing missing teeth in single or multiple unit applications in the mandible or maxilla. · immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge · especially indicated for use in soft bone applications where implants with other implant surface treatments may be less effective · immediate and early loading for all indications · together with immediate loading protocol in all indications, except in single tooth situations in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate · only together with Profile EV components, Implant Driver Profile EV, Radiographic Implant Guides Profile EV and non-Indexed prosthetic components Abutments: ASTRA TECH Implant System™ - OsseoSpeed™ EV abutments are intended to be used in conjunction with ASTRA TECH Implant System™ - OsseoSpeed™ EV in fully edentulous or partially edentulous maxillary and/or mandibular arches. The ATLANTIS™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. The ATLANTIS™ Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in partially or completely edentulous patients. The prosthesis is screw retained. The abutment screw is intended to secure the crown abutment to the endosseous implant. The ATLANTIS™ Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS™ Abutment to the endosseous implant. ATLANTIS™ Abutment, ATLANTIS™ Crown Abutment and ATLANTIS™ Conus Abutment are compatible with ASTRA TECH Implant System Profile EV.
Device Story
Endosseous dental implant system; includes implants and abutments for tooth replacement. Implants feature coronal contour designed for anatomical bone alignment; available in straight or tapered neck designs. Abutments support prosthetic restorations (cemented, screw-retained, or friction-fit). Used by dental clinicians in surgical/restorative settings. System facilitates immediate/early loading protocols. Benefits include improved esthetic restoration and stability in soft bone. Performance verified via static/dynamic compression-bending testing.
Clinical Evidence
Bench testing only. Static and dynamic compression-bending testing performed in accordance with ISO 14801 to demonstrate mechanical integrity and substantial equivalence.
Technological Characteristics
Endosseous dental implants and abutments. Materials: Titanium alloy and zirconia. Design: Straight or tapered neck; diameters 3.0-4.8mm; lengths 8-17mm. Surface treatment: OsseoSpeed. Mechanical testing: ISO 14801 static/dynamic compression-bending.
Indications for Use
Indicated for patients with missing teeth (single/multiple) in mandible or maxilla, including edentulous or partially edentulous arches. Suitable for immediate placement in extraction sites or healed ridges; indicated for soft bone applications. Contraindicated for immediate loading in single-tooth soft bone (type IV) cases where stability is insufficient.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
510(k) Summary K130999
# 510(k) Summary DENTSPLY Implants OsseoSpeed™ Profile EV K130999
September 23, 2013
# OCT 2 5 2013
### ADMINISTRATIVE INFORMATION
| Manufacturer Name | DENTSPLY Implants |
|---------------------------|---------------------------------|
| | Aminogatan 1, P.O. Box 14 |
| | Mölndal, Sweden SE-431-21 |
| Telephone: | +46 31 776 30 00 |
| Fax: | +46 31 776 30 10 |
| Official Contact | Christina Lewing |
| | Head of Regulatory Affairs |
| Representative/Consultant | Linda K. Schulz, BSDH, RDH |
| | Kevin A. Thomas, PhD |
| | PaxMed International, LLC |
| | 12264 El Camino Real, Suite 400 |
| | San Diego, CA 92130 |
| Telephone: | +1 (858) 792-1235 |
| Fax: | +1 (858) 792-1236 |
| email: | lschulz@paxmed.com |
| | kthomas@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
Trade/Proprietary Name Common Name
Classification Name Classification Regulations
Product Code Classification Panel Reviewing Branch
OsseoSpeed™ Profile EV Dental implant and abutment
Endosseous dental implant Class II, 21 CFR 872.3640
DZE, NHA Dental Products Panel Dental Devices Branch
{1}------------------------------------------------
#### INTENDED USE
#### Implants:
The ASTRA TECH Implant System - OsseoSpeed Profile EV implants are intended for both one- and two-stage surgical procedures in the following situations and with the following clinical protocols: · replacing missing teeth in single or multiple unit applications in the mandible or maxilla.
- · immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge
- · especially indicated for use in soft bone applications where implants with other implant surface treatments may be less effective
- · immediate and early loading for all indications
- · together with immediate loading protocol in all indications, except in single tooth situations in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate
- · only together with Profile EV components, Implant Driver Profile EV, Radiographic Implant Guides Profile EV and non-Indexed prosthetic components
#### Abutments
ASTRA TECH Implant System™ - OsseoSpeed™ EV abutments are intended to be used in conjunction with ASTRA TECH Implant System™ - OsseoSpeed™ EV in fully edentulous or partially edentulous maxillary and/or mandibular arches.
The ATLANTIS™ Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.
The ATLANTIS™ Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in partially or completely edentulous patients. The prosthesis is screw retained. The abutment screw is intended to secure the crown abutment to the endosseous implant.
The ATLANTIS™ Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment. The abutment screw is intended to secure the ATLANTIS™ Abutment to the endosseous implant.
ATLANTIS™ Abutment, ATLANTIS™ Crown Abutment and ATLANTIS™ Conus Abutment are compatible with ASTRA TECH Implant System Profile EV.
#### DEVICE DESCRIPTION
The purpose of this submission is to expand the current OsseoSpeed EV Implant System (previously named OsseoSpeed Plus) to include the OsseoSpeed Profile EV implants. The OsseoSpeed EV Implant line is a part of the ASTRA TECH Implant System. The coronal contour of the OsseoSpeed Profile EV implant is designed to follow the anatomical contour of the bone and facilitate a more esthetic restoration.
{2}------------------------------------------------
Implants are available in a straight (cylindrical) design or with a taper at the implant neck. Each design is provided in two diameters (4.2 and 4.8 mm) and six lengths (8, 9, 11, 13, 15, and 17 mm).
Abutments are provided in titanium allov and zirconia materials. Abutments have platform diameters of 3.0, 4.2, and 4.8 mm, prosthetic diameters from 3.3 to 7.0 mm, and gingival heights ranging from 2.0 to 6.5 mm. Abutments are provided straight and with up to 30° of angulation.
#### EQUIVALENCE TO MARKETED DEVICE
DENTSPLY Implants submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices OsseoSpeed™ Profile EV is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
Astra Tech AB, OsseoSpeed™ Plus - K120414 Astra Tech AB, OsseoSpeed™ Profile, System - K080156 Astra Tech, Inc., Atlantis™ Abutment in Zirconia for OsseoSpeed TX Profile Implant - K 103759 Astra Tech, Inc., Atlantis™ Abutment in Zirconia for Astra Tech Osseospeed Plus Implant - K 112138 Astra Tech AB, Astra Tech Implant System, New Component - K072624 Astra Tech AB, Astra Tech Implant System - K 101732 Atlantis Components, Inc. AtlantisTM Abutment for Astra Implant - K051652 FRIADENT GmbH, ANK YLOS® Dental Implant System - K041509
The subject device and the predicate devices have intended use and have the same technological characteristics. The subject device and the predicate devices encompass the same range of physical dimensions and characteristics, including diameter, length, angulation and surface treatment. Any differences in the technological characteristics between the subject device and the predicate devices do not raise new issues of safety or efficacy.
Performance testing is provided to demonstrate substantial equivalence and includes static and dynamic compression-bending testing according to ISO 14801. The data included in this submission demonstrates substantial equivalence to the predicate devices listed above.
Overall. OsseoSpeed™ Profile EV has the following similarities to the predicate devices:
- . has the same intended use,
- . uses the same operating principle,
- . incorporates the same basic design,
- incorporates the same or very similar materials, and
- has similar packaging and is sterilized using the same materials and processes .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird figure. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 25, 2013
DENTSPLY Implants C/O Linda K. Schulz, BSDH, RDH PaxMed International, LLC 12264 El Camino Real, Suite 400 SAN DIEGO CA 92130
Re: KI30999
Trade/Device Name: OsseoSpeed™ Profile EV Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: September 23, 2013 Received: September 24, 2013
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Schulz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/6 description: The image shows the name "Mary Si Bunner -S" in a simple, bold font. The text is arranged horizontally, with the name "Mary" appearing first, followed by "Si Bunner" and then "-S". The letters are uniformly sized and spaced, creating a clean and legible presentation of the name. The overall impression is straightforward and professional.
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
K130999 510(k) Number:
OsseoSpeed™ Profile EV Device Name:
Implants:
The ASTRA TECH Implant System - OsseoSpeed Profile EV implants are intended for both one- and two-stage surgical procedures in the following situations and with the following clinical protocols:
- · replacing missing teeth in single or multiple unit applications in the mandible or maxilla.
- · immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge
- · especially indicated for use in soft bone applications where implants with other implant surface treatments may be less effective
- · immediate and early loading for all indications
- · together with immediate loading protocol in all indications, except in single tooth situations in soft bone (type IV) where implant stability may be difficult to obtain and immediate loading may not be appropriate
- · only together with Profile EV components, Implant Driver Profile EV, Radiographic Implant Guides Profile EV and non-Indexed prosthetic components
Abutments:
ASTRA TECH Implant System™ - OsseoSpeed™ EV abutments are intended to be used in conjunction with ASTRA TECH Implant System™ - OsseoSpeed™ EV in fully edentulous or partially edentulous maxillary and/or mandibular arches.
The ATLANTISTM Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutnent. The abutment screw is intended to secure the abutment to the endosseous implant.
The ATLANTIS™ Crown Abutment is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in partially or completely edentulous patients. The prosthesis is screw retained. The abutment screw is intended to secure the crown abutment to the endosseous implant.
The ATLANTIS™ Conus Abutment is intended for use with an endosseous implant to support a prosthetic device in partially or completely edentulous patients. It is intended for use to support a removable multiple tooth prosthesis, in the mandible or maxilla. The prosthesis is attachment-retained by friction fit to the abutment screw is intended to secure the ATLANTIS™ Abutment to the endosseous implant.
A TLANTIS™ Abutment, ATLANTIS™ Crown Abutment and ATLANTIS™ Conus Abutment are compatible with ASTRA TECH Implant System Profile EV.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|---|
|-------------------------------------------------|---|
AND/OR
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
|------------------------------------------------|--|
|------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Susan Runner
DOSPA 2013.10.25 08:23:13
Mary S. Runner -S
04'00"
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.