The instrument kits are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and protect the instruments that are sterilized by healthcare provider. The cassette is to be enclosed in an FDA cleared steam sterilizable pouch. The cassettes are not intended on their own to maintain sterility. The cycle of sterilization is: Pre-vacuum steam: 134 °C during 3 minutes with 16 minutes drying time Anthogyr does not make any lumen claims for the instrument kits. The Axiom range surgery kit represents the worst case validated load due to the number of components (34 instruments). The cassettes are not intended to be stacked during sterilization process.
Device Story
Instrument kits are reusable cassettes designed to hold dental surgical drills and tools during steam sterilization and storage. Cassettes are constructed from Radel R-5000/R-5100 and medical-grade silicone. The device is used in clinical settings by healthcare providers. The cassette is loaded with instruments, placed inside an FDA-cleared steam sterilizable pouch, and subjected to a pre-vacuum steam sterilization cycle (134°C for 3 minutes, 16 minutes drying). The device organizes instruments and facilitates sterilization; it does not maintain sterility on its own. The Axiom range surgery kit (34 instruments) serves as the worst-case validated load. The output is a sterilized, organized set of dental instruments ready for surgical use. The device benefits patients by ensuring dental tools are properly sterilized and organized for procedures.
Clinical Evidence
No clinical data. Bench testing only. Validations included cleaning (AAMI TIR 30, AAMI TIR 12, ISO 17665), steam sterilization (ISO 17665) using worst-case loads, drying time verification, and cytotoxicity testing (ISO 10993-5).
Technological Characteristics
Materials: Radel R-5000, Radel R-5100, medical-grade silicone, stainless steel. Form factor: Single size (129x156x47.5 mm) tray with locking lid and silicone brackets. Sterilization: Pre-vacuum steam (134°C). Connectivity: None. Software: None.
Indications for Use
Indicated for use by healthcare providers to organize, steam sterilize, and protect dental surgical drills and tools. Not intended to maintain sterility independently; must be used with an FDA-cleared steam sterilizable pouch.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 27, 2017
Anthogyr Therese Candau Regulatory Affairs Manager 2237 Avenue Andre Lasquin Sallanches. 74700 FR
Re: K160730
Trade/Device Name: Instrument Kits Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: March 20, 2017 Received: March 24, 2017
Dear Therese Candau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160730
Device Name Instrument kits
Indications for Use (Describe)
The instrument kits are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and protect the instruments that are sterilized by healthcare provider. The cassette is to be enclosed in an FDA cleared steam sterilizable pouch. The cassettes are not intended on their own to maintain sterility.
The cycle of sterilization is:
Pre-vacuum steam: 134 °C during 3 minutes with 16 minutes drying time
Anthogyr does not make any lumen claims for the instrument kits.
The Axiom range surgery kit represents the worst case validated load due to the number of components (34 instruments).
The cassettes are not intended to be stacked during sterilization process.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY K160730
#### SUBMITTER'S CONTACT INFORMATION 5.1
### ANTHOGYR SAS
2237, avenue André Lasquin SALLANCHES, FRANCE 74700 Registration Number: 8020776 Phone number: (33) (0)4 50 58 02 37 Fax number: (33) (0)4 50 93 78 60 Contact person: Thérèse CANDAU (m.candau@anthogyr.com) Preparation date: 04/12/2017
#### DEVICE 5.2
Trade name: Instrument kits Classification name: STERILIZATION CASSETTE Class II Product Code: KCT CFR Section: 21CFR 880.6850 Device panel: GENERAL HOSPITAL
#### 5.3 PREDICATE DEVICE
InterActive Complete Surgical Tray (K142519), manufactured by Implant Direct
#### DEVICE DESCRIPTION 5.4
Instrument kits are cassettes used to enclose and hold surgical instruments and accessories in an organized manner during the sterilization process and subsequent storage. The cassettes by themselves do not maintain sterility. They are to be used with an FDA cleared steam sterilizable pouch.
The cassette is available in only one size (129x156x47.5 mm) marketed in different variations, including different number of tools.
The tools included in the kits are class I exempt or already have class II pre-market notification clearance.
The cassettes are reusable and made of the same materials as the predicate device (Radel R-5000 and R-5100 for tray and lid and medical grade silicone for circular brackets holding the tools).
{4}------------------------------------------------
### 5.5 INTENDED USE
The instrument kits are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and protect the instruments that are sterilized by healthcare provider. The cassette is to be enclosed in an FDA cleared steam sterilizable pouch. The cassettes are not intended on their own to maintain sterility. The cycle of sterilization is:
Pre-vacuum steam: 134 °C during 3 minutes with 16 minutes drying time Anthogyr does not make any lumen claims for the instrument kits.
The Axiom range surgery kit represents the worst case validated load due to the number of components (34 instruments).
The cassettes are not intended to be stacked during sterilization process.
| | ANTHOGYR (K160730)<br>Surgical kits | Implant Direct<br>(K142519)<br>InterActive Complete<br>Surgical Tray | Comparison |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| 21CFR | 880.6850 | | Same part section |
| Indications<br>for use | The instrument kits are<br>designed to hold various<br>dental surgical drills and<br>tools in order to organize,<br>steam sterilize, and protect<br>the instruments that are<br>sterilized by healthcare<br>provider. The cassette is to<br>be enclosed in an FDA<br>cleared steam sterilizable<br>pouch. The cassettes are<br>not intended on their own<br>to maintain sterility. | The InterActive<br>Complete Surgical Tray<br>is designed to hold<br>various dental surgical<br>drills and tools in order<br>to organize, steam<br>sterilize, and transport<br>the instruments between<br>uses. The tray is to be<br>enclosed in an FDA<br>cleared steam<br>sterilizable wrap and<br>sterilized in an FDA<br>cleared sterilizer. | Same indications for<br>use |
| Raw material | Radel R5000<br>Radel R5100<br>Medical grade silicone<br>Stainless steel | Polymer resin Radel-<br>5000<br>Biomedical grade silicone | Some materials are<br>common.<br>Radel R5100 meets<br>USP class VI<br>requirements and<br>compatibility testing<br>demonstrated safety<br>and efficacy |
| Design | Plastic tray with locking lid and silicone containments<br>brackets | | Same design |
| Sterilization<br>method | Pre-vacuum steam | | Same sterilization<br>method |
| Reusable | Yes | | Both are reusable<br>devices |
#### TECHNOLOGICAL CHARACTERISTICS 5.6
{5}------------------------------------------------
| Material<br>compatibility<br>with<br>sterilization<br>process | Yes | | Both devices<br>demonstrate material<br>compatibility with<br>sterilization process |
|---------------------------------------------------------------|-------------------------------------------------------------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Microbial<br>barrier<br>properties | To be used in conjunction with an approved<br>sterilization pouch | | Both are used with an<br>approved sterilization<br>pouch to maintain<br>sterility |
| Vent/volume<br>ratio | $0.049 \frac{in^2}{in^3}$ | $0.184 \frac{in^2}{in^3}$ | The vent to volume<br>ratio is lower for the<br>candidate device, but<br>sterilant penetration<br>testing demonstrates<br>safety and efficacy |
#### 5.7 PERFORMANCE TESTING
Performance characteristics of the device are supported by cleaning validation, sterilization validation and cytotoxicity tests.
No clinical tests were conducted.
### Cleaning instructions for reusable devices
Cleaning validations were conducted as per AAMI TIR 30, AAMI TIR 12, ISO 17665 for both manual and automatic cleaning.
### Sterilization validation
Sterilization validation was performed with a legally marketed pouch on the most complete instrument kit, taking into account worst case scenarios. Steam sterilization validations were conducted as per ISO 17665
### Drying time
Validated in steam sterilization report according to ISO 17665
### Toxicological properties
Cytotoxicity tests were conducted as per ISO 10993-5
### Material compatibility
Radel technical data sheet
{6}------------------------------------------------
#### 5.8 CONCLUSION
ANTHOGYR INSTRUMENT KITS are substantially equivalent to their predicate devices in terms of intended use, material, design, mechanical properties and function.
All the kits have the same vent to volume ratio, so the most challenging container to achieve sterilization is the most loaded one.
Non clinical performance testing (sterilization and cleaning validation, cytotoxicity test) according to special control demonstrate that ANTHOGYR INSTRUMENT KITS are as safe, as effective, and perform as safely and effectively as their predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.