K142519 · Implant Direct Sybron Manufacturing, LLC · KCT · Apr 15, 2015 · General Hospital
Device Facts
Record ID
K142519
Device Name
InterActive Complete Surgical Tray
Applicant
Implant Direct Sybron Manufacturing, LLC
Product Code
KCT · General Hospital
Decision Date
Apr 15, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The InterActive Complete Surgical Tray is designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. The tray is to be enclosed in an FDA cleared sterilizable wrap and sterilized in an FDA cleared sterilizer for one of the following cycles: (1) Prevaccum Steam At 132℃ for 4 minutes with a 20 minutes dry time. (2) Gravity Steam At 132% for 15 minutes with a 30 minutes dry time. The trays are intended for sterilization of non-porous loads. The trays are recommended not to be stacked during sterilization. The Complete Surgical Tray represents the worst case validated load due to the number of components) and the weight of 608.05 grams. Implant Direct Sybron Manufacturing LLC does not make any lumen claims for the Interactive Complete Surgical Tray.
Device Story
Reusable rigid sterilization container system; holds dental surgical drills/tools during transport, steam sterilization, and storage. Composed of perforated base tray, lid, and internal insert tray; made of Radel R-5000 polymer resin with medical-grade silicone grommets/brackets for instrument retention. Used in healthcare facilities; operated by clinical staff. Requires enclosure in FDA-cleared sterilization wrap before processing in standard autoclaves (prevacuum or gravity steam cycles). Protects instruments during handling; ensures effective sterilant penetration via perforated design. Benefits include organized instrument management and maintenance of sterility post-processing.
Clinical Evidence
No clinical data. Bench testing only: steam sterilization validation (AAMI/ANSI/ISO 17665, AAMI TIR 12), cytotoxicity (ISO 10993-5/12), cleaning validation (AAMI TIR 30/12), and distribution studies (ASTM D4169).
Technological Characteristics
Materials: Radel R-5000 polymer resin (lid, base, insert), medical-grade silicone (grommets/brackets per FDA CFR 177.2600). Design: Perforated multi-piece rigid container with locking lid. Dimensions: 7.2" x 5.6" x 2.4". Sterilization: Steam (Gravity/Pre-vacuum). Connectivity: None. Software: None.
Indications for Use
Indicated for holding, organizing, steam sterilizing, and transporting dental surgical drills and tools in healthcare facilities. Intended for non-porous loads only.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Polyvac Inc. Surgical Instrument Delivery Tray (K012105)
Tuttnauer USA Co Ltd. Cassette Container (K990761)
C/T Medical Systems Sterilization Cassettes (K980065)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 15, 2015
Implant Direct Sybron Manufacturing LLC Debleena Sinha, M.S. Regulatory Affairs Specialist 3050 East Hillcrest Drive Thousand Oaks, CA 91362
Re: K142519
Trade/Device Name: InterActive Complete Surgical Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: March 20, 2015 Received: March 23, 2015
Dear Ms. Sinha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Debleena Sinha,M.S.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K142519
Device Name
InterActive Complete Surgical Tray
#### Indications for Use (Describe)
The InterActive Complete Surgical Tray is designed to hold various dental surgical drills and tools in order to organize. steam sterilize, and transport the instruments between uses. The tray is to be enclosed in an FDA cleared sterilizable wrap and sterilized in an FDA cleared sterilizer for one of the following cycles:
- (1) Prevaccum Steam At 132℃ for 4 minutes with a 20 minutes dry time.
- (2) Gravity Steam At 132% for 15 minutes with a 30 minutes dry time.
- The trays are intended for sterilization of non-porous loads.
- The trays are recommended not to be stacked during sterilization.
- The Complete Surgical Tray represents the worst case validated load due to the number of components) and the weight of 608.05 grams.
- Implant Direct Sybron Manufacturing LLC does not make any lumen claims for the Interactive Complete Surgical Tray.
- The tray will be marketed in following variations.
| Device Model Name | Max no. of Instruments | Weight of each tray<br>with instruments (gms.) |
|-------------------|------------------------|------------------------------------------------|
| CST | 45 | 608.05 |
| ICST | 34 | 467.26 |
| IBST | 18 | 453.59 |
| IST | 0 | 404.37 |
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(K) Submission
# 510 (K) Summary (21CFR 807.92(a)(1))
### 1. Submitter Information
| 510 K Sponsor | Implant Direct Sybron Manufacturing LLC |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 3050 East Hillcrest Drive<br>Thousand Oaks, CA 91362 |
| Registration Number | 3001617766 |
| Correspondence Contact | Debleena Sinha, M.S<br>Specialist, Regulatory Affairs<br>Ph: 818-444-3306<br>Fax: 818-444-3406<br>Email: debleena.sinha@implantdirect.com |
| Date Summary Prepared | April 6, 2015 |
## 2. Device Name and Classification
| Trade Name | InterActive Complete Surgical Tray |
|---------------------|------------------------------------------------------------------------|
| Classification Name | Sterilization Wrap Containers, Trays, Cassettes, and other Accessories |
| Common Name | Instrument Sterilization Tray |
| Regulation Number | 21 CFR 880.6850 |
| Product Code | KCT |
| Regulatory Class | II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Implant Direct, a dental company. The logo is set on an orange background and features the company name in blue, with a blue curved line above the name. Below the company name, the words "simply smarter." are written in a smaller, white font. At the bottom of the logo, the words "Sybron Dental Specialties" are written in a smaller, blue font.
Traditional 510(K) Submission
#### 3. Predicate Devices
Conmed Linvatech Instrument Sterilization Tray (K052992) Polyvac Inc. Surgical Instrument Delivery Tray (K012105) Tuttnauer USA Co Ltd. Cassette Container (K990761) C/T Medical Systems Sterilization Cassettes (K980065)
#### 4. Description of Device
The InterActive Complete Surgical Tray is a reusable rigid sterilization container or organizing tray intended for use in health care facilities for the purpose of containing reusable medical devices for sterilization. It is composed of multiple pieces, designed to be integrated into a single unit which contains and protects the interior components during sterilization. Each tray consists of three components - a base tray, a lid, and an internal individualized insert tray. All of the three components are perforated for steam penetration. The internal insert tray has the ability to hold individualized pieces and accessories which include dental tools, drills, and ratchet. The tray is available in only one size of approximately 7.2 inches X 5.6 inches X 2.4 inches and will be marketed in four different variations – (1) InterActive Complete Surgical Tray (ICST) containing 34 instruments, (2) InterActive Basic Surgical Tray (IBST) containing almost half the number of instruments as ICST, (3) InterActive
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Implant Direct, a Sybron Dental Specialties company. The logo is set against an orange background. The words "Implant Direct" are in blue, with a blue arc above them. Below that, the words "simply smarter" are in white, and below that, the words "Sybron Dental Specialties" are in gray.
Surgical Tray (IST), empty tray containing no instruments, and (4) Complete Surgical Tray (CST), containing 45 instruments.
The rigid multi-piece tray holds dental device apparatus and accessories before, during, and after the sterilization process. The tray set has a locking lid to contain the products. The trays are designed to fit any standard autoclave, which allows it to be effective for sterilization and be able to withstand the environment of repeated steam sterilization cycles in normal operating room. Since the trays are perforated, an FDA cleared sterilization wrap must be used for sterilization purposes to maintain the sterility of the contents.
The trays are reusable and the tray material allows repeated sterilization cycles. The lid, base, and insert are made of Radel R-5000. This material is a polymer resin produced by Solvay Advanced Polymers, LLC, and is identical to the Radel used in predicate device cleared under K052992 and K012105. Radel R 5000 CL 301 is used for the lid, and Radel R 5100 NT15 is used for base and insert tray where all the drills and tools are mounted. The small circular brackets (grommets) throughout the insert tray which are used to contain the drills and the tool holder with cradle on the base tray are made of medical grade silicone material which has been manufactured to meet FDA CFR 177.2600. These brackets and holders are used to secure the
{6}------------------------------------------------
instruments during transport, sterilization, and storage. Although these trays are reusable they will not be serviced or repaired.
The instruments to be sterilized in the proposed tray are all non-porous devices and include dental surgical drills and tools. The tools are class I exempt and the drills have Class II pre-market notification clearance.
### 5. Intended Use
The InterActive Complete Surgical Tray is designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. The tray is to be enclosed in an FDA cleared steam sterilizable wrap and sterilized in an FDA cleared sterilizer for one of the following cycles:
(1) Prevaccum Steam – At 132ºC for 4 minutes with a 20 minutes dry time.
(2) Gravity Steam - At 132ºC for 15 minutes with a 30 minutes dry time.
- -The trays are intended for sterilization of non-porous loads.
- -The trays are recommended not to be stacked during sterilization.
- -The Complete Surgical Tray represents the worst case validated load due to the number of components (45 instruments) and the weight of 608.05 grams.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for Implant Direct, a dental company. The logo is set on an orange background and features the company name in blue, with a blue arc above the name. Below the company name is the tagline "simply smarter." in a smaller font. At the bottom of the logo is the text "Sybron Dental Specialties".
- Implant Direct Sybron Manufacturing LLC does not make any lumen claims for the -
InterActive Complete Surgical Tray.
- -The tray will be marketed in following variations.
| Device Model<br>Name | Max no. of Instruments | Weight of each tray<br>with instruments (gms.) |
|----------------------|------------------------|------------------------------------------------|
| CST | 45 | 608.05 |
| ICST | 34 | 467.26 |
| IBST | 18 | 453.59 |
| IST | 0 | 404.37 |
### 6. Substantial Equivalence
The tables below shows that the InterActive Complete Surgical Tray compared against 510(K) cleared products. The comparison analysis consists of the products' intended use, technological characteristics and performance testing to support the substantial equivalency to their corresponding predicate devices.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for Implant Direct, a dental company. The logo is set on an orange background and features a blue swoosh above the company name. Below the company name is the tagline "simply smarter" in a smaller, white font. The logo also includes the text "Sybron Dental Specialties" in a smaller font.
## TABLE1: Substantial Equivalence: Comparison of Proposed Device with Predicates on Technological Considerations
| Manufacturer | Implant Direct | ConMed<br>Linvatech | PolyVac Inc. | Tuttnauer<br>USA Co Ltd | C/TMedical<br>Systems |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| | Proposed Device | Predicate | Predicate | Predicate | Predicate |
| 510 (K) | K142519 | K052992 | K012105 | K990761 | K980065 |
| Trade Name | InterActive<br>Complete Surgical<br>Tray | Instrument<br>Sterilization<br>Tray | Surgical<br>Instrument<br>Delivery<br>System | Mini<br>Container<br>Cassette;<br>Standard<br>Cassette<br>Container | Sterilization<br>cassettes |
| Product Code | KCT | FRG | KCT | FRG | KCT |
| Intended Use | Perforated<br>instrument<br>cassette system<br>to hold dental<br>instruments in<br>place during<br>transport, steam<br>sterilization, and<br>storage | Perforated trays<br>with lids to hold<br>surgical<br>instruments in<br>place during<br>transport,<br>steam<br>sterilization,<br>and storage | Perforated<br>trays with lids<br>to hold surgical<br>instruments in<br>place during<br>transport,<br>steam<br>sterilization,<br>and storage | Dental<br>instrument<br>cassette<br>system to<br>hold<br>instruments<br>in place<br>during<br>transport,<br>steam<br>sterilization,<br>and storage | Dental<br>instrument<br>cassette<br>system to<br>hold<br>instruments<br>in place<br>during<br>transport,<br>steam<br>sterilization,<br>and storage |
| Material<br>Composition | Polymer Resin<br>Radel-5000,<br>biomedical grade<br>silicone | Polymer Resin<br>Radel-5000,<br>biomedical<br>grade silicone | Radel-5000<br>plastic,<br>Biomedical<br>grade silicone,<br>and Aluminum,<br>300 series<br>stainless steel | Plastic<br>Polymer,<br>biomedical<br>grade silicone | Plastic,<br>Metal, and<br>Stainless<br>Steel<br>container |
| Manufacturer | Implant Direct | ConMed<br>Linvatech | PolyVac Inc. | Tuttnauer<br>USA Co Ltd | C/TMedical<br>Systems |
| Design | Plastic tray with<br>locking lid and<br>silicone<br>containment<br>brackets | Plastic tray with<br>stainless steel<br>latch and<br>silicone<br>containment<br>brackets | Tray made of<br>plastic or metal<br>base with lid | Small to<br>medium tray<br>with silicone<br>mat inserts | Trays and<br>cassettes<br>made of<br>metal and<br>plastic. Latch<br>made of<br>stainless<br>steel. |
| Sterilization by | | | | | |
| (1) Gravity<br>Steam | Yes | Yes | Yes | Yes | Yes |
| (2) Pre Vaccum<br>Steam | Yes | Yes | Yes | Yes | Yes |
| Air Permeance | Yes | Yes | Yes | Yes | Assume |
| Locking system to<br>hold lid in place | Yes | Yes | Yes | Yes | Yes |
| Reusable | Yes | Yes | Yes | Yes | Yes |
| Material<br>compatibility<br>with sterilization<br>process | Yes | Yes | Yes | Yes | Yes |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for Implant Direct, a Sybron Dental Specialties company. The logo is set against an orange background. Above the company name is a blue curved line. Below the company name is the tagline "simply smarter."
Traditional 510(K) Submission
#### 7. Non-Clinical Performance Testing
The performance testing of the proposed device complies with a special control: 21 CFR 880.6850 - Class II Special Controls Guidance Document for sterilization wrap (pack, sterilization wrapper, bag, and accessories. The performance characteristics of the device are supported by Steam Sterilization validations, Cytotoxicity tests,
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image is a logo for Implant Direct, a dental company. The logo is set on an orange background. The words "Implant Direct" are written in blue, with a blue arc above the words. Below the company name, the words "simply smarter" are written in white, and below that, the words "Sybron Dental Specialties" are written in gray.
Cleaning validations, and Distribution studies conducted on the Complete Surgical
Tray taking worst case scenarios into account.
## TABLE 2: Substantial Equivalence: Comparison with Predicates on Performance Testing
| Manufacturer | Implant Direct | ConMed<br>Linvatech | PolyVac Inc. | Tuttnauer<br>USA Co Ltd | C/T Medical<br>Systems |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------|
| Proposed Device | Predicate | Predicate | Predicate | Predicate | Predicate |
| Manufacturer | Implant Direct | ConMed<br>Linvatech | PolyVac Inc. | Tuttnauer<br>USA Co Ltd | C/T Medical<br>Systems |
| 510 (K) | K142519 | K052992 | K012105 | K990761 | K980065 |
| Sterilant<br>Penetration<br>Studies | Yes<br>Steam<br>Sterilization<br>Validations<br>conducted as per<br>AAMI/ANSI/ISO<br>17665;<br>AAMI TIR 12 | Yes | Yes | Yes | Yes |
| Toxicological<br>Properties<br>(biocompatibilty) | Yes<br>Cytotoxicity<br>Tests conducted<br>as per<br>ISO 10993 - 5;<br>ISO 10993 -12 | Yes<br>Brackets and<br>containment<br>mats made<br>from<br>Biomedical<br>grade silicone | Yes<br>Biocomapatibilty<br>tested to USP<br>requirements | Yes<br>Cassette<br>inserts<br>made from<br>biomedical<br>grade<br>silicone | Not stated<br>in summary |
| Transportation<br>studies<br>(Packaging<br>Integrity) | Yes<br>Distribution<br>Studies<br>conducted as per<br>ASTM D4169 | Not stated in<br>summary | Not stated in<br>summary | Not stated in<br>summary | Not stated<br>in summary |
| Cleaning<br>Instructions for<br>Reusable devices | Yes<br>Cleaning<br>Validations<br>conducted as per<br>AAMI TIR 30;<br>AAMI TIR 12 | Not stated in<br>summary | Not stated in<br>summary | Not stated in<br>summary | Yes |
| Material<br>Compatibility<br>(repeat<br>sterilization) | Yes<br>Radel Technical<br>Data Sheet | Yes<br>Radel | Yes<br>Radel | Not stated<br>in summary | Not stated<br>in summary |
| Manufacturer | Implant Direct | ConMed<br>Linvatech | PolyVac Inc. | Tuttnauer<br>USA Co Ltd | C/T Medical<br>Systems |
| Drying Time | Yes<br>Validated in<br>Steam<br>Sterilization<br>report<br>AAMI/ANSI/ISO<br>17665;<br>AAMI TIR 12 | Not stated in<br>summary | Not stated in<br>summary | Not stated<br>in summary | Yes<br>Drying time<br>in labeling |
| Vent/Volume<br>Ratio | All proposed<br>trays are<br>$0.184 in^2/in^3$ | Not stated in<br>summary | $0.02276 in^2/in^3$ | Not stated<br>in summary | Not stated<br>in summary |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image is a logo for Implant Direct, a dental company. The logo features the company name in blue, with the words "simply smarter" in a smaller font below. The text "Sybron Dental Specialties" is at the bottom of the logo. The background is orange.
Traditional 510(K) Submission
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image is a logo for Implant Direct. The logo is set on an orange background. The words "Implant Direct" are in blue, with a blue arc above them. Below that, in a smaller font, it says "simply smarter" and "Sybron Dental Specialties".
Traditional 510(K) Submission
The proposed devices have high vent to volume ratio, allowing easier sterilant penetration than the predicate device.
The Complete Surgical Tray (CST) container represents the worst case vent to volume
ratio for the Surgical Tray System making this container the most challenging
container to achieve sterilization.
### 8. Clinical Testing
No clinical tests were conducted.
#### 9. Summary and Conclusion
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image is a logo for Implant Direct, a dental company. The logo is set on an orange background and features a blue arc above the company name. Below the company name, in smaller letters, is the tagline "simply smarter." The text "Sybron Dental Specialties" is at the bottom of the logo.
Traditional 510(K) Submission
A comparison of the intended use, technological characteristics, and performance studies conducted on the InterActive Complete Surgical Tray demonstrate that the device is substantially equivalent to its predicate devices K052992 (ConMed Linvatech Instrument Sterilization Tray), K012105 (Polyvac Surgical Instrument Delivery System), K990761 (Tuttnauer USA Co Ltd. Cassette Container), and K980065 (C/T Medical Systems Sterilization Cassettes).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.