Gibralt® Spine System and Gibralt® Occipital Spine System
K160697 · Exactech, Inc. · NKG · Jun 28, 2016 · Orthopedic
Device Facts
Record ID
K160697
Device Name
Gibralt® Spine System and Gibralt® Occipital Spine System
Applicant
Exactech, Inc.
Product Code
NKG · Orthopedic
Decision Date
Jun 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3075
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Gibralt® Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1-C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Gibralt® Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In order to achieve additional levels of fixation, the Gibralt® Spine System may be connected to the Gibralt® Occipital Spine System with rod-to-rod connectors. The Gibralt® Spine System may also be connected to the Exactech® Proliant® System, Exactech® Silverbolt® and Mainframe® Spinal Screw Systems, or Exactech® Hydralok® Exactech, using rod-to-rod connectors and transitional rods. Refer to the specific system package inserts for a list of their indications for use.
Device Story
Spinal fixation system comprising rods, screws, and connectors; provides immobilization and stabilization of cervical/thoracic spinal segments; used as adjunct to fusion or for temporary stabilization in advanced tumor cases. Components manufactured from titanium and cobalt chromium. System allows for modular connectivity to other Exactech spinal systems via rod-to-rod connectors. Operated by surgeons in clinical/OR settings to restore spinal column integrity or stabilize segments. Output is mechanical fixation of the spine; assists in clinical decision-making by providing structural support for fusion or palliative stabilization. Benefits include reduced instability and pain management for patients with degenerative or traumatic spinal conditions.
Clinical Evidence
Bench testing only. Performed static and dynamic axial compression and compression bending tests per ASTM F1717-14 to validate mechanical strength and equivalence to predicate devices.
Technological Characteristics
Components constructed from titanium (ASTM F136) and cobalt chromium (ASTM F1537). Mechanical fixation system; includes rods, screws, and connectors. No software or energy source.
Indications for Use
Indicated for patients with acute/chronic instabilities of craniocervical junction, cervical spine (C1-C7), and thoracic spine (T1-T3) requiring immobilization/stabilization as adjunct to fusion. Conditions include traumatic fractures/dislocations, instability/deformity, pseudarthrosis, tumors, and degenerative disease (radiculopathy, myelopathy, discogenic pain, facet instability). Also indicated for temporary spinal column integrity restoration in patients with advanced cervical tumors with limited life expectancy. Occipital bone screws limited to occipital fixation.
Regulatory Classification
Identification
Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
Special Controls
*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.
Predicate Devices
Synapse Occipital-Cervical-Thoracic System (K142838)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 28, 2016
Exactech, Incorporated % Mr. Kenneth C. Maxwell II Regulatory and Quality Specialist Empirical Testing Corporation 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K160697
Trade/Device Name: Gibralt® Spine System and Gibralt® Occipital Spine System Regulatory Class: Unclassified Product Code: NKG, KWP Dated: May 31, 2016 Received: June 1, 2016
Dear Mr. Kenneth C. Maxwell II:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 | | |
|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Food and Drug Administration | Expiration Date: January 31, 2017 | | |
| Indications for Use | See PRA Statement on last page. | | |
| 510(k) Number (if known) | K160697 | | |
| Device Name | Gibralt® Spine System | | |
| Indications for Use (Describe) | The Gibralt® Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1-C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The GibraltⓇ Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. | | |
| | In order to achieve additional levels of fixation, the Gibralt® Spine System may be connected to the Gibralt® Occipital Spine System with rod-to-rod connectors. The Gibralt® Spine System may also be connected to the Exactech® Proliant® System, Exactech® Silverbolt® and Mainframe® Spinal Screw Systems, or Exactech® Hydralok® Exactech, using rod-to-rod connectors and transitional rods. Refer to the specific system package inserts for a list of their indications for use. | | |
| Type of Use (Select one or both, as applicable) | <table style="border:none;"><tr><td><span style="border: 1px solid black;"><b>[X]</b></span> Prescription Use (Part 21 CFR 801 Subpart D)</td><td><span style="border: 1px solid black;">[ ]</span> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <span style="border: 1px solid black;"><b>[X]</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="border: 1px solid black;">[ ]</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <span style="border: 1px solid black;"><b>[X]</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="border: 1px solid black;">[ ]</span> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
| PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | |
| FOR FDA USE ONLY | | | |
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | | | |
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement on last page. |
|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
**Indications for Use**
| 510(k) Number (if known) | K160697 |
|--------------------------------|---------------------------------|
| Device Name | Gibralt® Occipital Spine System |
| Indications for Use (Describe) | |
When used with the Gibralt® Spine System, the Gibralt® Occipital Spine System is intended
to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the
following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-
C7), and the thoracic spine from T1 - T3: traumatic spinal fractures and/or traumatic
dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors
involving the cervical/thoracic spine; and degenerative disease, including intractable
radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by
radiographic studies, and degenerative disease of the facets with instability. The GibraltⓇ
Occipital Spine System is also intended to restore the integrity of the spinal column even in the
absence of fusion for a limited time period in patients with advanced stage tumors involving
the cervical spine in whom life expectancy is of insufficient duration to permit achievement of
fusion.
In order to achieve additional levels of fixation, the Gibralt® Occipital Spine System may be
connected to the Gibralt® Spine System, using rod-to-rod connectors. Refer to the Gibralt®
Spine System package insert for a list of the Gibralt® Spine System indications of use.
The Occipital Bone Screws are limited to occipital fixation only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| | <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
| FORM FDA 3881 (9/13)<br>Page 1 of 2 | PSC Publishing Services (301) 443-6740 EF |
|-------------------------------------|-------------------------------------------|
|-------------------------------------|-------------------------------------------|
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
# 5.510(K) SUMMARY
| Submitter's Name: | Exactech, Inc. |
|----------------------------|----------------------------------------------------------------------------------------|
| Submitter's Address: | 2320 NW 66th Court<br>Gainesville, FL 32658 |
| Submitter's Telephone: | 352.377.1140 |
| Contact Person: | Kenneth C. Maxwell II<br>Empirical Testing Corp.<br>719.291.6874 |
| Date Summary was Prepared: | 23 June 2016 |
| Trade or Proprietary Name: | Gibralt® Spine System<br>Gibralt® Occipital Spine System |
| Common or Usual Name: | Orthosis, Cervical Spinal Pedicle Fixation<br>Appliance, Fixation, Spinal Interlaminal |
| Classification: | Unclassified, Class II per 21 CFR §888.3050 |
| Product Code: | NKG, KWP |
| Classification Panel: | Orthopedic Devices |
### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The line extension and device modifications for the Gibralt® Spine System are intended to provide immobilization and stabilization of spinal segments in the upper thoracic, cervical, and occipital spine. The system components are manufactured from titanium and cobalt chromium. The modifications included in the scope of this submission are: expanded indications for use for both systems, incorporation of design changes, and addition of new components.
#### INDICATIONS FOR USE
#### Gibralt® Spine System:
The Gibralt® Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1-C7) and the thoracic spine from T1-T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Gibralt® Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the Gibralt® Spine System may be connected to the Gibralt® Occipital Spine System with rod-to-rod connectors. The Gibralt®
{7}------------------------------------------------
Spine System may also be connected to the Exactech® Proliant® System. Exactech® Silverbolt® and Mainframe® Spinal Screw Systems, or Exactech® Hydralok® Exactech, using rod-to-rod connectors and transitional rods. Refer to the specific system package inserts for a list of their indications for use.
Gibralt® Occipital Spine System:
When used with the Gibralt® Spine System, the Gibralt® Occipital Spine System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1-C7), and the thoracic spine from T1 - T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The Gibralt® Occipital Spine System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the Gibralt® Occipital Spine System may be connected to the Gibralt® Spine System, using rod-to-rod connectors. Refer to the Gibralt® Spine System package insert for a list of the Gibralt® Spine System indications of use.
The Occipital Bone Screws are limited to occipital fixation only.
The indications for use for the Gibralt® Spine System and the Gibralt® Occipital Spine System are similar to that of the predicate devices in Table 5-1.
## TECHNOLOGICAL CHARACTERISTICS
Gibralt® Spine System components are made from materials that conform to titanium per ASTM F136 and cobalt chromium per ASTM F1537. The Gibralt® Occipital Spine System components are made from materials that conform to titanium per ASTM F136. The subject and predicate devices have similar technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are similar between the subject and predicates:
- Indications for Use
- Materials of Manufacture ●
{8}------------------------------------------------
| 510k Number | Trade or Proprietary or<br>Model Name | Manufacturer | Type |
|-------------|------------------------------------------------|-----------------|----------------------|
| K142838 | Synapse Occipital-Cervical-<br>Thoracic System | Synthes USA | Primary<br>Predicate |
| K083071 | VERTEXTM Reconstruction<br>System | Medtronic | Reference |
| K050979 | S4 Spinal System | Aesculap®, Inc. | Reference |
Table 5-1 Predicate Devices
#### PERFORMANCE DATA
The proposed line extensions and modified Gibralt® Spine System have been tested in the following test modes:
- Static axial compression per ASTM F1717-14
- Static axial compression bending per ASTM F1717-14 ●
- Dynamic axial compression per ASTM F1717-14 ●
- o Dynamic axial compression bending per ASTM F1717-14
The results of this non-clinical testing show that the strength of the Gibralt® Spine System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
#### CONCLUSION
The overall technology characteristics and mechanical performance data lead to the conclusion that the Gibralt® Spine System and Gibralt® Occipital Spine System are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.