When intended to promote fusion of the cervical spine and the thoracic spine (C1 -T3), the S4 Spinal System is intended for the following; - DDD (Neck pain pf discogenic origin with degeneration of the disc as . confirmed by patient history and radiographic studies) - Spondvlolisthesis . - Spinal stenosis . - Fracture/dislocation . - Failed previous fusion . - Tumors . The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1 -- T3) spine. The use of the polyaxial screws is limited to placement in T1 - T3 in treating thoracic conditions only. Screws are not intended to be placed in the cervical spine.
Device Story
S4 Spinal System is a posterior spinal fixation system used by surgeons to stabilize the cervical and upper thoracic spine (C1-T3) to promote fusion. Components include 3.5mm rods, lamina hooks, and a cross connector. Polyaxial screws are utilized specifically for thoracic (T1-T3) stabilization. The system is implanted during surgical procedures to address degenerative disc disease, fractures, dislocations, tumors, or failed prior fusions. By providing rigid fixation, the device maintains spinal alignment, facilitating bone fusion and patient recovery.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Constructed from Titanium forged alloy (Ti6Al4V). System includes 3.5mm diameter rods, lamina hooks, polyaxial screws, and cross connectors. Mechanical fixation device; no energy source or software components.
Indications for Use
Indicated for patients requiring spinal fusion or stabilization in the cervical and upper thoracic spine (C1-T3) due to DDD, spondylolisthesis, spinal stenosis, fracture/dislocation, failed previous fusion, or tumors. Polyaxial screws restricted to T1-T3 thoracic use only; contraindicated for cervical spine placement.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
Koso979
S4 Spinal System
## JUL 27 2005
Page 1 of 2
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS C.
[in Accordance with SMDA of 1990]
### S4 Spinal System
April 18, 2005
Aesculap®, Inc. COMPANY: 3773 Corporate Parkway Center Valley, PA 18034
Establishment Registration Number: 2916714
Joyce Kilroy CONTACT: 800/258-1946 x 5074 (phone) 610/791-6882 (fax)
SA TRADE NAME:
COMMON NAME: S4 Spinal System
DEVICE CLASS: Class II
PRODUCT CODE: MNI, KWP
CLASSIFICATION: 888.3070 - Pedicle screw spinal system
888.3050 - Spinal interlaminal fixation orthosis
REVIEW PANEL: Orthopedics
## INDICATIONS FOR USE
When intended to promote fusion of the cervical spine and the thoracic spine (C1 -T3), the S4 Spinal System is intended for the following;
- DDD (Neck pain pf discogenic origin with degeneration of the disc as . confirmed by patient history and radiographic studies)
- Spondvlolisthesis .
- Spinal stenosis .
- Fracture/dislocation .
- Failed previous fusion .
- Tumors .
The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1 -- T3) spine.
The use of the polyaxial screws is limited to placement in T1 - T3 in treating thoracic conditions only. Screws are not intended to be placed in the cervical spine.
## DEVICE DESCRIPTION
The S4 Spinal System consists of 3.5mm rods in 6 lengths, thin and thick lamina hooks, 3.5 The 64 Opinal System of various lengths and a cross connector. The S4 Spinal System is manufactured from Titanium forged alloy Ti6Al4V.
{1}------------------------------------------------
# PERFORMANCE DATA
PERFORMANCE DATA
All required testing per "Draft Guidance for the Preparation of Premarket Notifications where All required testing per Drait Ouldance Tor Roll Clements" were done where
(510(k)s) Applications for Orthopedic Devices-The Basic Elements "who (310(K)S) Applications for Orthopodio Dotholo Riosem 510(K)s" was completed where applicable.
# SUBSTANTIAL EQUIVALENCE
Aesculap believes that the new S4 Spinal System is substantially equivalent in design to:
- Nex- Link Spinal System (K031985) .
- Vertex Spinal System (K042789, 042524) .
- Summit Spinal System (K013222, K030103, K041203) .
ﻢ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three horizontal lines representing the branches of government. The caduceus is surrounded by a circle of text that reads "U.S. Department of Health and Human Services, USA."
JUL & 7 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Joyce Kilroy Director of Regulatory Affairs/Quality Assurance Aesculap, Inc. 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K050979
Trade/Device Name: S4 Spinal System Regulation Number: 21 CFR 888.3050, 21 CFR 888.3070 Regulation Name: 21 CF CF 866558888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888888 Regulatory Class: II Product Code: KWP, MNI Dated: July 20, 2005 Received: July 21, 2005
Dear Ms. Kilroy:
We have reviewed your Section 510(k) premarket notification of intent to market the device wf nave reviewed your becaren of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated to togensy to tegains.
commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to conniner of the ridge 2011-12-11 in accordance with the provisions of the Federal Food, Drug, de nees mat have been that do not require approval of a premarket approval application (PMA). and Cosmotic Ace (110) that the device, subject to the general controls provisions of the Act. The I ou may, mercrore, mains of the Act include requirements for annual registration, listing of general controll pro resering practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rryour device is elassified (600 a00 rols. Existing major regulations affecting your device can thay or subject to observed Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean r rouse of acrease a determination that your device complies with other requirements of the Act that I Dr Has Intacted and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of to rate 077) accesses (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Ms. Joyce Kilroy
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter witi anow you to oghi maxing of substantial equivalence of your device to a legally premarket notheation: "The PDF Intains of cour device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your acon) 276-0120. Also, please note the regulation entitled, Connact the Office of Compraiset notification" (21CFR Part 807.97). You may obtain Misoranding of reference to premail.org/w/index.php?title=Ret from the Division of Small other general information on your response at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A. Millman
Mark N. Melkerson Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT B.
Page 1 of 1
510(k) Number: K050979
S4 Spinal System Device Name:
### Indication for Use:
When intended to promote fusion of the cervical spine and the thoracic spine (C1 - T3), the S4 Spinal System is intended for the following;
- DDD (Neck pain pf discogenic origin with degeneration of the disc as . confirmed by patient history and radiographic studies)
- Spondylolisthesis .
- Spinal stenosis ●
- Fracture/dislocation .
- Failed previous fusion .
- Tumors .
The hooks and rods are also intended to provide stabilization to provide fusion The Trooks 'and Tous-The 'are 'interest - trauma in the cervical/upper thoracic (C1 -T3) spine.
The use of the polyaxial screws is limited to placement in T1 – T3 in treating thoracic conditions only. Screws are not intended to be placed in the cervical spine.
Mark W. Millenn
(Division Sign-Off) Division of General, Restorative and Neurological Device
510(k) Number_
Prescription Use X or Over-the-Counter Use __
(per 21 CFR 801.109)
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
1 +1 ﮈﯾﭩﺎ ﺳﮯ ﻣﻄﺎﺑﻘﺖ ﻣﯿﮟ ﻭﺍﻗﻊ ﮨﮯ۔
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.