K160511 · Naya Health, Inc. · HGX · Jun 24, 2016 · Obstetrics/Gynecology
Device Facts
Record ID
K160511
Device Name
Naya Breast Pump System
Applicant
Naya Health, Inc.
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Jun 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Naya Breast Pump System is a powered breast pump to be used by lactating women in the hospital or home setting to express and collect milk from their breasts.
Device Story
Personal electric breast pump; single or double pumping; reciprocating diaphragm vacuum pump driven by microprocessor; vacuum levels up to 260 mmHg. Powered by rechargeable Li-ion battery or mains. User interface includes power button, mode button, and touch wheel; mobile app (iOS 8.0+) replicates physical interface. Used by lactating women in home or hospital. Stimulation mode (fast cycles, low vacuum) initiates flow; flow mode (slower cycles, higher vacuum) expresses milk. Healthcare provider/user adjusts settings via pump or app to optimize milk expression; benefits include efficient, gentle milk collection.
Reciprocating diaphragm vacuum pump; microprocessor-controlled. Powered by 14.8V 3200 mAh Li-ion battery or 100-240 VAC mains. Connectivity via mobile app (iOS). Materials meet ISO 10993-1 for surface contact/intact skin/limited duration and 21 CFR 174-179 food-contact standards. No natural latex. Electrical safety per ES 60601-1, IEC 60601-1-2, IEC 60601-1-11.
Indications for Use
Indicated for lactating women in hospital or home settings to express and collect breast milk. No contraindications.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 24, 2016
Naya Health Inc. % Paul Dryden Consultant 390 Bridge Pkwy. Suite C Redwood City, CA 94065
Re: K160511 Trade/Device Name: Naya Breast Pump System Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: Class II Product Code: HGX Dated: May 20, 2016 Received: Mav 23, 2016
Dear Paul Dryden,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Expiration Date: January 31, 2017
See PRA Statement on last page.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
## K160511
Device Name
# Nava Breast Pump System
Indications for Use (Describe)
The Naya Breast Pump System is a powered breast pump to be used by lactating women in the hospital or home setting to express and collect milk from their breasts.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D)
[X] Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "
FORM FDA 3881 (8/14
Page 1 of 1
SC Publishing Services (301) 443-6740
{3}------------------------------------------------
#### 510(k) Summary Page 1 of 4
| Date of Preparation: | 21-Jun-16 |
|-------------------------------------------------------------------------|-------------------------------------------------------------|
| | |
| Naya Health, Inc.<br>390 Bridge Pkwy, Suite C<br>Redwood City, CA 94065 | Tel - 650-346-6271<br>Fax - 847-680-6269 |
| Official Contact: | Janica B. Alvarez - Co-founder and CEO |
| Proprietary or Trade Name: | Naya Breast Pump System |
| Common/Usual Name: | Powered breast pump |
| Classification Name: | Powered breast pump<br>HGX - CFR 884.5760<br>Class II - OTC |
| Predicate Devices: | K151632 - Medela - Symphony |
#### Device Description:
The Naya Breast Pump System is a personal use electric breast pump capable of single or double pumping. The device is electrically powered from either an internal rechargeable battery or an external supply. The external supply also charges the battery. The device can also connect to an iPhone (iOS 8.0+) application (Mobile App) that replicates the front panel controls and indicators on the pump.
The Naya Breast Pump System may be used in the hospital or home setting. The Pump may be used by different users, but one person at a time. The Distal Breast Assembly (DBA) is intended as a single user component.
A reciprocating diaphragm vacuum pump, driven by a microprocessor, generates the suction to extract the milk at vacuum levels up to 260 mmHg. The Naya Breast Pump System is powered by a rechargeable battery or from mains.
The User Interface with the pump is:
- Power Button
- . Mode Button
- . Touch Wheel
The Mobile App looks identical to the pump user interface in that the same three interface points are present.
The device has two modes of operation:
- Stimulation Mode: Suction pattern with fast cycles and low vacuum to start milk ● flowing
- . Flow Mode: Suction pattern with slower cycles and higher vacuum to express more milk gently and efficiently.
{4}------------------------------------------------
## 510(k) Summary Page 2 of 4 21-Jun-16
#### Indications for Use:
The Naya Breast Pump System is a powered breast pump to be used by lactating women in the hospital or home setting to express and collect milk from their breasts.
| Patient Population: | Lactating women |
|---------------------|---------------------------------|
| Environment of use: | Home and / or hospital settings |
| Contraindications: | None |
Table 1 below presents a comparison between the proposed device and the predicate.
## Substantial Equivalence Discussion of Comparison to Predicates
The Naya Breast Pump System is viewed as substantially equivalent to the predicate device because:
#### Indications -
Indications for use are to express milk of lactating women ● Discussion - These are the identical indications for use of the predicate - Medela -Symphony - K151632.
#### Technology -
- The technology of a reciprocating diaphragm vacuum pump to express milk and the ● flange and collection bottle system are similar to the predicate. The power source, user controls, ability to adjust vacuum level and pump mode are similar to the predicate.
Discussion - The technology and performance specifications for the proposed device are similar to the predicate - Medela - Symphony - K151632.
#### Materials -
- The materials which are in contact with the user and the expressed milk are considered by ● ISO 10993-1 as having the following classification:
- . Surface contact, Intact skin, Limited duration (< 24 hours) tests performed included: Cytotoxicity, Intracutaneous reactivity, and Sensitization.
Discussion - The materials are similar and have been found to meet ISO 10993 requirements and thus can be substantially equivalent for safety as the predicate - Medela - Symphony -K151632.
#### Environment of Use -
The environment of use, home and / or hospital is identical to the predicate. ● Discussion - The environment of use is identical to the predicate - Medela - Symphony -K151632.
{5}------------------------------------------------
## 510(k) Summary Page 3 of 4 21-Jun-16
## Patient Population -
The patient population is lactating women is identical to the predicate. ● Discussion - The user population is identical to the predicate - Medela - Symphony -K151632.
# Non-Clinical Performance Testing Summary
We have performed a number of bench tests to demonstrate the Naya Breast Pump System performs within its specifications. These tests included:
- Breast Shield Assembly (BSA) Concentrated Cleaning
- Biocompatibility Test
- Software Verification .
- ES 60601-1, IEC 60601-1-2, IEC 60601-1-11 Test
- Pump Drive Systems (PDS) ●
- O User Interface pre- and post- conditioning
- O Battery pre- and post- aging
- . Breast Shield Assembly (BSA)
- Fluid Isolation & Drainage o
- Functional pre- and post- aging O
- Conditioning Test o
- o Lifetime Cycling
## Materials -
The materials were tested as Surface contact, Intact skin, and Limited duration (< 24 hours): Cytotoxicity, Intracutaneous reactivity, and Sensitization. In addition the material contains no natural latex and are certified for food-contact in accordance with 21 CFR 174-179.
## Discussion of Differences
The basic design, performance and features of the Naya Breast Pump System are similar to the predicate. The notable differences are:
- . The ability to also control the device via a mobile App.
- o We have evaluated this App and it is configured to look and perform exactly like the interface on the pump.
These differences do not raise any new safety concerns and supports substantially equivalence.
{6}------------------------------------------------
## 510(k) Summary Page 4 of 4 21-Jun-16
| Attribute | Proposed<br>Naya Breast Pump System | Medela<br>Symphony<br>K151632 |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Naya Breast Pump System is<br>a powered breast pump to be used<br>by lactating women in the<br>hospital or home setting to<br>express and collect milk from<br>their breasts. | The Symphony breast pump is a<br>powered breast pump to be used by<br>lactating women to express and<br>collect milk from their breast. |
| Patient population | Lactating women | Lactating women |
| Environment of use | Home and Hospital | Home and Hospital |
| OTC | Yes | Yes |
| Power Source | Input: 100-240 VAC, 50/60Hz,<br>2.1A<br>1 x 14.8V / 3200 mAh<br>Rechargeable Li-ion Battery | Input: 100-240 VAC, 50/60Hz<br>0.3A<br>2 x 6V / 1.2 Ah<br>Rechargeable Pb batteries |
| Pump Style | Reciprocating Pump | Diaphragm pump |
| Single/double Pumping | Both | Both |
| Adjustable Suction Levels | 10 levels | 16 levels |
| Cycle Speed | 34 - 120 | 45 - 120 |
| Overflow Protection | Yes (diaphragm) | Yes (diaphragm) |
| Vacuum range - double (mmHg) | 50-250 | 75-200 |
| Vacuum range - single (mmHg) | 50-260 | 75-270 |
| Cycling/Suction Control Mechanism | Microprocessor | Microprocessor |
| Accessories | Flange (Breast Shield Assembly),<br>including tubing, check valve and<br>puck diaphragm<br>Bottle (Collection Cup) | Flange / Soft Flange<br>Tubing<br>Check valve<br>Diaphragm<br>Bottle |
| Software | Yes | Yes |
| Cleaning method for Accessories | Soap and warm water | Soap and warm water<br>Boiling water |
| Materials in contact with user and<br>expressed milk tested per ISO 10993-1 | Cytotoxicity<br>Sensitization<br>Intracutaneous | |
| Electrical Safety | ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11 home use | UL 1431 |
| Flange offered in multiple sizes | Yes | Yes |
# Table 1 – Comaprison of the Proposed Device ot the Predicate
## Substantial Equivalence Conclusion
The Naya Breast Pump System is substantially equivalent to the above listed predicate and we have determined that there are no significant differences which would affect safety and efficacy for the patient population. This has been demonstrated through performance testing, design, and features, and non-clinical testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.