The Symphony breast pump is a powered breast pump to be used by lactating women to express and collect milk from their breast.
Device Story
Symphony breast pump is a multi-user, powered device for expressing and collecting breast milk; used by lactating women in hospital or home settings. Device features a belt-driven vacuum pump with two diaphragms for single or double pumping. User adjusts vacuum levels and cycle speeds via control knob; LCD provides status. 2-Phase Expression Technology mimics natural nursing: Stimulation Phase (fast cycles, low vacuum) initiates flow; Expression Phase (slower cycles, higher vacuum) extracts milk. Preemie+ program card provides specialized suction patterns mimicking irregular sucking of preterm infants to assist initiation. Protective membrane prevents milk overflow into pump mechanism. Device operates on AC power or rechargeable lead-acid batteries. Healthcare providers or patients use the device to support lactation maintenance or initiation; output is collected breast milk. Benefits include efficient, gentle milk expression and support for lactation initiation in clinical or home environments.
Clinical Evidence
No clinical testing was required. Substantial equivalence is supported by bench testing, including performance verification of vacuum levels and cycle rates across various power sources and motor types, cleaning validation, and software verification/validation.
Technological Characteristics
Powered diaphragm pump; 100-240 Vac or rechargeable Pb battery power; Polyamide housing; 16 adjustable vacuum levels (-50 to -250 mmHg); 35-120 cycles/minute. Features 2-Phase Expression technology. Software-controlled suction patterns via program cards. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and EN ISO 10079-1. Media separation via protective membrane.
Indications for Use
Indicated for lactating women to express and collect breast milk. Suitable for use in hospital and home environments.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract symbol that resembles three human profiles facing to the right. The symbol is composed of three curved lines that overlap each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 5, 2015
Medela AG % Adrienne Lenz Member Pathway Regulatory Consulting, LLC W324 S3649 County Road E Dousman, WI 53118
Re: K151632
> Trade/Device Name: Symphony Breastpump, Symphony Reconditioned, Symphony Preemie+ Breastpump, Symphony Plus Breastpump (battery Version), Symphony Consignment, Preemie Consignment Pump, Symphony Rental, Symphony Plus Rental, Symphony Preemie+ Rental Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: October 2, 2015 Received: October 5 17, 2015
Dear Adrienne Lenz,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151632
#### Device Name
Symphony breast pump, Models: Symphony Preeme+ Breastpump, Symphony Plus Breastpump (battery version), Symphony Consignment, Symphony Preemie Consignment Pump, Symphony Plus Rental, Symphony Preemie+ Rental, Symphony Reconditioned
#### Indications for Use (Describe)
The Symphony breast pump is a powered breast pump to be used by lactating women to express and collect milk from their breast.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# MEDELA AG SYMPHONY BREAST PUMP
{4}------------------------------------------------
#### 510(k) Summary
#### K151632
In accordance with 21 CFR 807.92 the following summary of information is provided:
DATE: October 30, 2015
#### SUBMITTER:
Medela AG Lättichstrasse 4b 6341 Baar / Switzerland Phone +41 (0)41 769 51 51 Fax + 41 (0)41 769 51 00
PRIMARY CONTACT PERSON:
Adrienne Lenz, RAC Member Pathway Regulatory Consulting, LLC T 262-290-0023
SECONDARY CONTACT PERSON:
Orlando Antunes Vice President Regulatory Affairs Medela AG
#### DEVICE:
TRADE NAME: Symphony breast pump, Models: Symphony Breastpump, Symphony Preemie+ Breastpump, Symphony Plus Breastpump (battery version), Symphony Consignment, Symphony Preemie Consignment Pump, Symphony Rental, Symphony Plus Rental, Symphony Preemie+ Rental, Symphony Reconditioned
COMMON/USUAL NAME: Powered Breast Pump
DEVICE CLASS: Class II
CLASSIFICATION NAMES: 21 CFR 884.516022 Pump, Breast, Powered
PRODUCT CODE: HGX
{5}------------------------------------------------
#### PREDICATE DEVICE(S):
#### K020518 Medela Symphony Breast Pump Model 024
#### DEVICE DESCRIPTION:
The Symphony breast pump is intended to be used by lactating women to express and collect milk from their breasts. The pumping can be performed on one breast (single pumping) or on both breasts at the same time (double pumping). The Symphony breast pump is a multi-user breast pump and therefore, designed for a safe and hygienic use by different mothers.
The Symphony breast pump employs a control knob for the user to adjust the applied vacuum. The suction patterns are programmed with either constant or variable cycles (pump speed). The Symphony breast pump is capable of providing vacuum levels from -50 to -250 mmHg with cycling rates up to 120 cycles per minute.
The Symphony® breast pump provides the following user features:
- Belt-driven vacuum pump design, for quiet operation.
- . Portable or trolley-mounted transport/operation options.
- LCD display, for user assistance/device status.
- Control knob, for user adjustment of vacuum level/pump speed.
- . 2-Phase Expression® Technology designed to mimic a baby's natural nursing rhythm:
- . Stimulation Phase (phase 1): Suction pattern with fast cycles and low vacuum to start milk flowing
- . Expression Phase (phase 2): Suction pattern with slower cycles and higher vacuum to express more milk gently and efficiently.
- Symphony program card with the initiation program (Preemie+) software. The Preemie+ program contains stimulation and expression phases with breast pump suction patterns which assist pump users in the initiation of milk production. The Preemie+ program mimics the irregular sucking pattern of healthy term infants right after birth.
- Symphony program card with the "Standard" pumping program software. The Standard program contains stimulation and expression phases with breast pump suction patterns which assist users in coming to volume, and/or maintenance of lactation. The Standard program imitates a breastfeeding baby during established lactation.
- "Let-down" control button to change between stimulation phase and expression phase.
- Option of either single or double breast pumping.
{6}------------------------------------------------
- Disposable protective membrane, designed to isolate the pump mechanism from the breast milk collection apparatus (prevent milk overflow into the pump). No vacuum can be created if the protective membrane is missing or damaged.
- User-friendly pump exterior, designed for ease of cleaning.
- A variety of accessories and breast pump kits.
## INDICATIONS FOR USE:
The Symphony breast pump is a powered breast pump to be used by lactating women to express and collect milk from their breast.
## DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
The Symphony breast pump has identical indications for use compared to the previously cleared version (K020518). A number of modifications have been made to the Symphony Breast Pump since the last 510(k) clearance (K020518). These are types of changes expected during years of production, including changes to meet new or modified standards and non-US regulations, manufacturing improvements to increase yield, improve stability and precision of components and prevent damage during use. Changes to the software have also been made, including the introduction of the Preemie+ initiation program.
The table below identifies key similarities and differences of the proposed Symphony breast pump to the legally marketed predicate version of this device (K020518).
| | Symphony (Current Version) | Symphony (Predicate Version) |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| | | K020518 |
| Indications for<br>Use | The Symphony breast pump is a<br>powered breast pump to be used by<br>lactating women to express and collect<br>milk from their breast. | The Symphony breast pump is a<br>powered breast pump to be used by<br>lactating women to express and collect<br>milk from their breast. |
| Environment of<br>Use | Hospital, Home | Hospital, Home |
| | TABLE 5.1 COMPARISON OF SYMPHONY TO PREDICATE DEVICE | |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| | Symphony (Current Version) | Symphony (Predicate Version) |
| | | K020518 |
| User Interface | Hardware interfaces | Hardware interfaces |
| User Control | State-dependent controls:<br>On-Off membrane switch<br>Let-Down Button (membrane switch),<br>also used to select the Preemie+<br>initiation program<br>Value-adjustment control:<br>Vacuum/Cycle-adjustment control | State-dependent controls:<br>On-Off membrane switch<br>Let-Down Button (membrane switch)<br>Value-adjustment control:<br>Vacuum/Cycle-adjustment control |
| Visual Indicator | Liquid-crystal display | Liquid-crystal display |
| Single/Double<br>Pumping | Single and double pumping | Single and double pumping |
| Accessories | Breast shield Breast shield inserts Valve Membrane Tubing Bottles, with disks lids and caps Trolly Vehicle adapter Power cord Container Stand Cooler Colostrum and Collection Containers Pump and Save Bags | Breast shield Breast shield inserts Valve Membrane Tubing Bottles, with disks lids and caps Trolly Vehicle adapter Power cord |
| Cleaning | Breastpump/case - wipe with clean, damp cloth Tubing - wash or sanitize only if milk or condensation in tubing Breastpump kit and bottles - wash and sanitize | Tubing - wash or sanitize weekly or if milk or condensation in tubing Breastpump kit and bottles - wash and sanitize (home) or autoclave sterilization (hospital) |
| Specifications | | |
| Power Source | 100-240 Vac | 100-240 Vac |
| | 50/60Hz | 50/60Hz |
| | 0.3A | 0.3A |
| | Symphony (Current Version) | Symphony (Predicate Version) |
| | | K020518 |
| Batteries | 2 x 6V/1.2Ah<br>Rechargeable Pb Batteries | 2 x 6V/1.2Ah<br>Rechargeable Pb Batteries |
| Battery charge/discharge | Operation time: approx. 1 hour<br>Charging time: approx.. 4 hours (80%)<br>approx.. 8 hours (100%) | Operation time: approx. 1 hour<br>Charging time: approx.. 4 hours (80%)<br>approx.. 8 hours (100%) |
| Housing Material | Polyamide | PS coated |
| Vacuum aggregate type | Diaphragm pump<br>Two diaphragms - for left and right breast<br>Linear movement by driving bars | Diaphragm pump<br>Two diaphragms - for left and right breast<br>Linear movement by driving bars |
| Vacuum Regulation type | Electronic measuring of the displacement of the membranes which corresponds to the generated vacuum. | Electronic measuring of the displacement of the membranes which corresponds to the generated vacuum. |
| Maximum vacuum mmHg/kPa | Approx. -290 mmHg / -40 kPa | Approx. -290 mmHg / -40 kPa |
| Vacuum range - double mmHg/kPa | -50 to -250mmHg ± 20 mmHg /<br>-7 to -33 kPa ± 3 kPa<br>(with Standard Program and 24 mm / size M PersonalFit breastshield)<br><br>-70 to -250mmHg ± 20 mmHg /<br>-7 to -33 kPa ± 3 kPa<br>(with Preemie+ Program and 24 mm / size M PersonalFit breastshield) | -50 to -250mmHg ± 20 mmHg /<br>-7 to -33 kPa ± 3 kPa<br>(with Standard Program and 24 mm / size M PersonalFit breastshield) |
| Vacuum range - single mmHg/kPa | -50 to -250mmHg ± 20 mmHg /<br>-7 to -33 kPa ± 3 kPa<br>(with Standard Program and 24 mm / size M PersonalFit breastshield)<br><br>-70 to -250mmHg ± 20 mmHg /<br>-7 to -33 kPa ± 3 kPa<br>(with Preemie+ Program and 24 mm / size M PersonalFit breastshield) | -50 to -250mmHg ± 20 mmHg /<br>-7 to -33 kPa ± 3 kPa<br>(with Standard Program and 24 mm / size M PersonalFit breastshield) |
| Adjustable Suction Levels | 16 levels | 16 levels |
| Cycling Control Mechanism | Electronic measuring of the Motor speed and position. | Electronic measuring of the Motor speed and position. |
| | Symphony (Current Version) | Symphony (Predicate Version) |
| | | K020518 |
| Cycle Speed | 54 – 120 Cycles/Minute<br>(with Standard Program)<br><br>35 – 120 Cycles/Minute<br>(with Preemie+ Program) | 54 – 120 Cycles/Minute<br>(with Standard Program) |
| Overflow<br>protection | A Protection Membrane on the pump<br>side acts as a media separation and<br>prevents milk from going into the<br>pump in case of a milk overflow into<br>the vacuum tubes. | A Protection Membrane on the pump<br>side acts as a media separation and<br>prevents milk from going into the<br>pump in case of a milk overflow into<br>the vacuum tubes. |
| Software Cards | Standard 2.0 or Preemie+ program<br>cards, each featuring stimulation and<br>expression phases to assist for the<br>initiation, coming to volume, and/or<br>maintenance of lactation.<br><br>Vacuum range and cycle speed are<br>dependent on the breast pump suction<br>patterns.<br><br>The Standard 2.0 program consists of<br>two different curves.<br><br>The Preemie+ initiation program<br>consists of three different curves:<br><br>Standard 2.0 program card: Standard<br>program only.<br><br>Preemie+ program card: Standard 2.0<br>and Preemie+ programs. | Standard program card featuring<br>stimulation and expression phases to<br>assist for the coming to volume,<br>and/or maintenance of lactation.<br><br>Vacuum range and cycle speed are<br>dependent on the breast pump suction<br>patterns.<br><br>The Standard program consists of two<br>different curves.<br><br>The Standard program card includes<br>the Standard program only |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
SUMMARY OF NON-CLINICAL TESTS:
The Symphony breast pump complies with voluntary standards for electrical safety, electromagnetic compatibility, use in the home healthcare environment and powered suction pumps. The following data were provided in support of the substantial equivalence determination:
- . Risk Analysis developed in accordance with ISO 14971: 2007.
- . Sterilization information in accordance with FDA's guidance document "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".
{10}------------------------------------------------
- . Cleaning Validation. Validation of cleaning methods, both manual cleaning with dish soap and cleaning in the dishwasher, and validation of the sanitation process with boiling water and the quick clean™ micro-steam™ bags were conducted. A selection of reusable breast pump kit components that contact breast milk representative of all Medela breast pump components, including those specified for use with Symphony, were challenged with a human breast milk solution containing organisms, cleaned, or sanitized according to the Instructions for Use and evaluated for the presence of surviving organisms. The validation protocol was based upon AAMI TIR30-2003, A Compendium of Processes, Materials, Test Methods, and Acceptance Criteria for Cleaning Reusable Medical Devices, 8 October 2003.
- . For sterile items, accelerated aging tests to support the labeled shelf life were conducted according to ASTM F1980 (07) Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Testing was also conducted to verify drive unit life time in support of the breast pump service life specification.
- . Biocompatibility Evaluation was completed according to the FDA guidance "Use of International Standard ISO- 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and concluded that no new testing was required as all patient contacting materials are identical to those used in the predicate Symphony Breast Pump.
- . Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since prior to mitigations of hazards, failure of the software could lead to minor injury, such as pain or engorgement.
- . Electrical safety and electromagnetic compatibility testing per IEC 60601-1:2005 (3rd Edition) with US deviations per AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 standard and IEC 60601-1-2: 2007 standards, respectively
- . Safety testing for use in the home per IEC 60601-1-11: 2010 standard
- . Performance testing demonstrating compliance with EN ISO 10079-1: 2009 Particular requirements for the safety of electrically powered suction equipment
- . Performance testing at minimum and maximum vacuum settings to determine the minimum and maximum vacuum levels of the pump as compared to its specifications, which are identical to those of the predicate version of the Symphony pump. Tests were conducted over 20 minutes with the Standard 2.0 program card or 15 minutes with the Preemie+ program card to simulate a typical pumping session using size medium (M) breast shields. Tests began in stimulation mode and continued with expression mode after the first two minutes. There are two approved suppliers of the motor for the Symphony® breast pump. Testing of six pumps was completed with the Standard 2.0 program card, using three pumps with each motor type. One of each pump type was used for testing with the Preemie+ program card. Vacuum performance during normal use was first evaluated and demonstrated that, independent of the pump
{11}------------------------------------------------
motor type, the specification for vacuum level and cycle rate is met. Additional test cases evaluated performance of the pump when running on battery, performance with varying input voltages and with the other size breast shields. Finally, testing of performance for single pumping was completed. Results demonstrated that independent of the pump motor type, the specification for battery operation time is met, that the specification for vacuum level and cycle rate is met in single pumping mode as well as with all power source and with varying input voltage. With different breast shield sizes, the specification for the maximum vacuum level is met.
#### SUMMARY OF CLINICAL TESTS:
Clinical testing was not required to demonstrate the substantial equivalence of the Symphony breast pump to its predicate device. However, published research studies are referenced to support marketing claims.
#### CONCLUSION:
The differences between the Symphony breast pump and its predicate device do not introduce a new intended use and do not raise new issues of safety and effectiveness. Verification and Validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended.
From the results of nonclinical testing described, Medela AG concludes that the Symphony breast pump is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.