K160488 · Cutera, Inc. · GEX · Oct 28, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160488
Device Name
Cutera enlighten III Laser System
Applicant
Cutera, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. 1064 nm: The 1064 nm wavelength of the enlighten III laser system is indicated for: - treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI) - tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI) 532 nm: The 532 nm wavelength of the enlighten III laser system is indicated for: - treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III - tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III 670 nm: The 670 nm wavelength of the enlighten III laser system is indicated for treatment of benign pigments with Fitzpatrick skin types I-III.
Device Story
Multi-wavelength, pulsed laser system; console-housed laser, power supplies, control electronics; touchscreen interface; footswitch-activated; articulated arm with rotational mirror knuckles; optical zoom handpiece; aiming beam diode laser. Used in dermatology, general, and plastic surgery clinics by physicians. Laser energy delivered to skin; spot size adjusted via handpiece. Enables treatment of benign pigmented lesions and tattoo removal. Benefits include safe, effective lesion clearance with minimal discomfort/adverse effects.
Clinical Evidence
Prospective IRB-approved clinical study; 48 subjects (Fitzpatrick I-III) with benign pigmented lesions (BPL). Single 670 nm picosecond treatment. Primary endpoint: blinded assessment of BPL clearing. Results: 66-74% of treated areas showed >50% lightening; median improvement scores statistically significant (p<0.0001 and p<0.001). High subject satisfaction (85-96%). Adverse events: minor erythema, edema, frosting, crusting; all resolved without intervention. No serious adverse events.
Technological Characteristics
Multi-wavelength pulsed laser (1064 nm, 532 nm, 670 nm). Q-switched Nd:YAG laser media. Pulse durations: 660 ps to 2 ns. Repetition rates: 1-10 Hz. Articulated arm delivery with optical zoom handpiece. Touchscreen control. Electrical power source. Complies with IEC 60601-1, IEC 60601-1-6, IEC 62366, IEC 60601-2-22, IEC 60825-1, and AAMI/ANSI ES60601-1.
Indications for Use
Indicated for treatment of benign pigmented lesions and tattoo removal. 1064 nm: all skin types (Fitzpatrick I-VI). 532 nm and 670 nm: Fitzpatrick skin types I-III.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol representing the human form.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 28, 2016
Cutera Incorporated Mr. Bradley Renton Vice President, Regulatory Affairs and Medical Affairs, & Compliance Officer 3240 Bayshore Boulevard Brisbane, California 94005
Re: K160488
Trade/Device Name: enlighten III Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: October 5, 2016 Received: October 6, 2016
Dear Mr. Renton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160488
Device Name enlighten III Laser System
#### Indications for Use (Describe)
The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.
1064 nm:
The 1064 nm wavelength of the enlighten III laser system is indicated for:
- · treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
• tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)
532 nm:
The 532 nm wavelength of the enlighten III laser system is indicated for:
- · treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
- tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III
670 nm:
The 670 nm wavelength of the enlighten III laser system is indicated for treatment of benign pigments with Fitzpatrick skin types I-III.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5 510(K) Summary
This 510(K) Summary of safety and effectiveness for the enlighten III laser system is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.
| Applicant: | Cutera, Inc. | | | |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Address: | 3240 Bayshore Blvd., Brisbane, CA 94005 | | | |
| Contact Person: | Bradley Renton | | | |
| | 415-657-5568 – phone<br>415-715-3568 – fax<br>brenton@cutera.com | | | |
| Preparation Date: | October 25, 2016 | | | |
| Device Trade Name: | enlighten III Laser System | | | |
| Common Name: | Dermatology Laser | | | |
| Classification Name: | Instrument, Surgical, Powered, laser<br>79-GEX, 21 CFR 878.4810 | | | |
| Legally Marketed<br>Predicate Devices: | Cutera enlighten Laser System (K133945 and K140727)<br>Cynosure PicoSure (K121346)<br>Quanta System Q-Plus T (K073549)<br>Cynosure RevLite (K133254) | | | |
| Device Description: | The enlighten III laser system is a multi-wavelength, pulsed laser<br>system. A key feature of the device is its ability to produce multiple<br>laser wavelengths and pulse widths (1064 and 532 nm: 750 ps and<br>2 ns; 670 nm: 660 ps).<br><br>The laser, power supplies, and control electronics are housed<br>inside a console equipped with a touchscreen control panel. The<br>laser treatment parameters are selected using the control panel.<br>Laser emission is activated by depressing a footswitch. An<br>articulated arm with rotational mirror knuckles delivers the laser<br>beam from a laser head inside the console to a handpiece. The<br>handpiece is equipped with an optical zoom assembly that<br>determines the spot size of the laser beam on the treatment<br>surface. The beam of a low-power diode laser is also transmitted<br>through the arm to provide an aiming beam. | | | |
| Indications for Use: | <b>1064 nm</b><br>The 1064 nm wavelength of the enlighten III laser system is<br>indicated for:<br>treatment of benign pigmented lesions on patients with all<br>skin types (Fitzpatrick I-VI) tattoo removal for dark colored tattoo inks and for<br>multicolored tattoos containing dark colored tattoo inks on | | | |
patients with all skin types (Fitzpatrick I-VI)
{4}------------------------------------------------
## 532 nm
The 532 nm wavelength of the enlighten III laser system is indicated for:
- treatment of benign pigmented lesions on patients with . Fitzpatrick skin types I-III
- tattoo removal for lighter colored tattoo inks, including red . and yellow inks, on patients with Fitzpatrick skin types I-III
### 670 nm
The 670 nm wavelength of the enlighten III laser system is indicated for treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III.
IEC 60601-1 Medical Electrical Equipment - Part 1: General Performance Data: Requirements for Safety (Edition 3.1. 2012), including:
- . IEC 60601-1-6 Medical Electrical Equipment - Part 1-6: General Requirements for Safety - Collateral Standard: Usability (Edition 3.1, 2013) - Test Report Attachment 4
- . IEC 62366 Medical Devices - Application of Usability Engineering to Medical Devices (Edition 1.1, 2014) - Test Report Attachment 5
- IEC 60601-2-22 Medical Electrical Equipment Part 2: . Particular Requirements for Basic Safety and Essential Performance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment (Edition 3.1, 2012) – Test Report Attachment 6
- . IEC 60825-1 Safety of Laser Products - Part 1: Equipment Classification and Requirements (Edition 2.0, 2007) - Test Report Attachment 7
The product also fulfills the requirements of AAMI/ANSI ES60601-1:2005+A2 (R2012) + A1.
IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility (Edition 3, 2007)
enlighten III Software Verification and Validation Testing Report (V0141 r2)
An IRB-approved prospective clinical study was conducted to Results of Clinical Study: assess the safety and efficacy of the 670 nm wavelength of the enlighten III laser system for the treatment of benign pigmented lesions.
> Forty-eight subjects with Fitzpatrick skin types I-III diagnosed with benign pigmented lesions of the hands, body, or face were enrolled and received a single picosecond 670 nm treatment. Subjects were enrolled in two primary cohorts: 25 subjects with 32 subjecttreatment-areas were enrolled in cohort 1; 3 additional subjects with
{5}------------------------------------------------
5 subject-treatment-areas were enrolled between the two primary cohorts; and 20 subjects with a single subject-treatment-area were enrolled in cohort 2. Forty-six subjects completed the study; one patient was lost to follow-up from the 3 subjects enrolled between the two primary cohorts; and one subject was lost to follow-up from the 2nd cohort.
The first cohort of 25 subjects and the 3 additional subjects were enrolled under protocol Ver. 1, and standardized photographs were taken at baseline and 12 weeks following treatment. The 200 cohort of 20 subjects was enrolled under protocol Ver. 2, and standardized photographs were taken at baseline and 6 weeks following treatment. Blinded assessment of improvement (clearing) in the benign pigmented lesions was completed by three independent dermatologists. Pain levels during treatment and adverse events were recorded during all visits.
For cohort 1, to create a dataset with 25 observations, one from each subject, a single treatment area was randomly selected from each subject for whom multiple treatments were done. This dataset with 25 observations was then created 100 times using a different randomization each time. Greater than 50% BPL lightening was observed in 66% of treated BPL areas, and the median improvement score of 2.00 was statistically significant (one-sample Wilcoxon Signed Test, 95% Cl: 1.67 - 2.67, p<0.0001). Investigator and subject subjective assessments of improvement were consistent with the blinded review.
For cohort 2, greater than 50% BPL lightening was observed in 74% of treated subjects, and the median improvement score of 2.33 was statistically significant (one-sample Wilcoxon Signed Test, 95% CI: 1.67 - 2.67, p<0.001). The mean and standard deviation were 2.12 ± 0.61. Investigator and subject subjective assessments of improvement were consistent with the blinded review.
For cohort 1, the reviewers were consistent (kappa of 0.874) and accurate (kappa of 0.833) in identification of the baseline and posttreatment photographs. Seventy-eight percent of the subject photographs were scored identically or within one grade difference. For cohort 2, the reviewers were consistent (kappa of 1.0) and accurate (kappa of 1.0) in identification of the baseline and posttreatment photographs. Ninety-five percent of the subject photographs were scored identically or within one grade difference. The 3 blinded reviewers' scores were an exact match in 16% of BPL photographs, and in 79% of photographs the 3 scores were within one grade difference.
For cohort 1, 96% percent of subjects were satisfied or extremely satisfied with the results of a single treatment and would repeat the treatment or recommend it to others. For cohort 2, 85% of subjects were satisfied or extremely satisfied with the results of a single treatment, 90% percent of the subjects would repeat the treatment,
{6}------------------------------------------------
and 95% would recommend it to others.
All subjects tolerated treatments well (cohort 1, mean pain score of 1.0 ± 0.2 [1-2]; cohort 2, mean pain score of 0.96 ± 0.37 [0-2] on a 10-point scale). As expected, all subjects experienced minor erythema, edema, and lesion frosting immediately following laser treatment, and most experienced mild crusting of treated lesions. All adverse effects resolved or improved during the study period with no required intervention. No serious adverse events were noted.
Treatment of benign epidermal pigmented lesions with the 670 nm output wavelength of the enlighten III laser system was found to be safe and effective, with minimal discomfort and adverse effects, allowing the conclusion that the enlighten III laser system is substantially equivalent to the predicate devices for the requested indications.
Summarv of The Cutera enlighten III laser system has the same intended use Technological and indications for use, as well as the same or very similar Characteristics: technological characteristics and operating principles, as the Cutera enlighten laser system (primary predicate: K133945 and K140727), Cynosure PicoSure (K121346), Quanta System Q-Plus T (K073549), and Cynosure RevLite (K133254). The enlighten III device design and components are very similar to those of the previously cleared enlighten laser system, where the only key difference is the addition of components to produce the 670 nm wavelength. The device design and components are also very similar to those of the Cynosure PicoSure, Quanta System Q-Plus T, and Cynosure RevLite predicates. For each of these systems:
- the user interface is located at the front/top of the console; ●
- . the treatment handpiece is attached to an articulating arm that is connected to the main system console;
- the laser output at each wavelength is generated within the laser chassis and delivered to the skin through the handpiece attached to the articulated arm; and
- . the handpiece allows the spot size on the skin to be adjusted according to device specifications.
Each system thus consists of an articulating arm and attached handpiece, as well as an electrically powered system console that produces the laser energy and houses the software and user interface.
The enlighten III laser system provides similar key design aspects, including the same or similar spot sizes, laser wavelengths, pulse widths, and laser types as its predicate devices. The repetition rate of the enlighten III laser system is the same as or within the repetition rate range of the predicate devices. Further, each of the devices presents a range of spot sizes to allow the user to choose the most appropriate spot size
{7}------------------------------------------------
for each patient. Therefore, the minor differences do not raise any new safety or effectiveness questions, because the enlighten III parameters are the same as or within the range of the predicates.
See Table 5A below.
- Conclusion: The enlighten III laser system is substantially equivalent to the predicate devices in terms of indications for use, technical specifications, operating performance features, and general design.
{8}------------------------------------------------
## Section 5 510(K) Summary
# Table 5A—Technical Specification Comparison
| | enlighten III Laser System<br>(K160488 current submission) | Cutera enlighten Laser System<br>(K133945 and K140727) | Quanta System Q-Plus T<br>(K073549) | Cynosure PicoSure<br>(K121346) | Cynosure RevLite<br>(K133254) |
|------------------------------|-----------------------------------------------------------------------------------|--------------------------------------------------------|-----------------------------------------------------------|-------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Wavelength | 1064 nm<br>532 nm<br>670 nm | 1064 nm<br>532 nm | 1064 nm<br>532 nm<br>694 nm | 755 nm | 1064 nm<br>532 nm<br>585 nm (with optional dye handpiece)<br>650 nm (with optional dye handpiece) |
| Max Pulse Energy | 1064 nm: 600 mJ<br>532 nm: 300 mJ<br>670 nm: 125 mJ | 1064 nm: 600 mJ<br>532 nm: 300 mJ | 1064 nm: 1000 mJ<br>532 nm: 500 mJ<br>694 nm: 1000 mJ | 200 mJJ | 1064 nm: 1.6 J<br>532 nm: 500 mJ<br>585 nm: unknown<br>650 nm: 170 mJ |
| Max Fluence | 1064 nm: 10 J/cm²<br>532 nm: 2.5 J/cm²<br>670 nm: 4.0 J/cm² | 1064 nm: 10 J/cm²<br>532 nm: 2.5 J/cm² | 1064 nm: 22 J/cm²<br>532 nm: 11 J/cm²<br>694 nm: 25 J/cm² | 6.37 J/cm² | 1064 nm: 12 J/cm²<br>532 nm: 5 J/cm²<br>585 nm: 10 J/cm²<br>650 nm: 6 J/cm² |
| Pulse Duration<br>(nominal) | 1064 nm: 750 ps or 2 ns<br>532 nm: 750 ps or 2 ns<br>670 nm: 660 ps | 750 ps or 2 ns | 1064 nm: 6 ns<br>532 nm: 6 ns<br>694 nm: 30 ns | 450 - 900 ps | 1064 nm: 7-20 ns<br>532 nm: 7-20 ns<br>585 nm: <7 ns<br>650 nm: <7 ns |
| Spot Size | 1064 nm and 532 nm: 2, 3, 4,<br>5, 6, 7, and 8 mm<br>670 nm: 2, 3, 4, 5, and 6 mm | 2, 3, 4, 6, or 8 mm | Up to 6 mm | Zoom 2-6 mm; fixed 2, 3, 4, 6, 8<br>or 10 mm | Zoom 2-8.5 mm |
| Output Mode | Pulsed | Pulsed | Pulsed | Pulsed | Pulsed |
| Repetition Rate | 1, 2, 3.3, 5, and 10 Hz or<br>single shot | ≤10 Hz or single shot | ≤10 Hz or single shot | ≤10 Hz or single shot | ≤10 Hz or single shot |
| Laser Media | Q-switched Nd:YAG laser | Q-switched Nd:YAG laser | Q-switched Nd:YAG laser and<br>Q-switched Ruby laser | Q-switched Alexandrite laser | Q-switched Nd:YAG laser |
| User Interface | Push-button control or LCD color<br>touchscreen | Push-button control or LCD color<br>touchscreen | Push-button control or LCD color<br>touchscreen | Push-button control or LCD color<br>touchscreen | Push-button control or LCD color<br>touchscreen |
| Treatment Beam<br>Activation | Footswitch | Footswitch | Footswitch | Footswitch | Footswitch |
| Delivery System | Articulated arm with laser<br>handpiece | Articulated arm with laser<br>handpiece | Articulated arm with laser<br>handpiece | Articulated arm with laser<br>handpiece | Articulated arm with laser handpiece |
| Aiming Beam | ≤1 mW Max at 450 nm | ≤1 mW Max at 635 nm | 635 nm | 630-690 nm | 650 nm |
| Handpiece (How<br>Supplied) | Non-sterile, reusable, cleanable | Non-sterile, reusable, cleanable | Non-sterile, reusable, cleanable | Non-sterile, reusable, cleanable | Non-sterile, reusable, cleanable |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.