K121346 · Cynosure, Inc. · GEX · Nov 27, 2012 · General, Plastic Surgery
Device Facts
Record ID
K121346
Device Name
PICOSURE WORKSTATION
Applicant
Cynosure, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 27, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PicoSure™ workstation is indicated for tattoo and benign pigmented lesions removal.
Device Story
PicoSure™ is a high-powered, flashlamp-excited Q-switched alexandrite laser system (755-nm wavelength). It delivers laser energy to tattoo dye or pigment particles under the skin; the combination of wavelength, pulse duration (450ps–900ps), and energy fluence disrupts particles without harming surrounding tissue. Fragmented particles surface and fade as the epidermis renews. Used in clinical settings by trained professionals for tattoo and benign pigmented lesion removal. Output is controlled laser energy; healthcare providers use the device to achieve tattoo clearance. Benefits include effective pigment removal with a safety profile comparable to existing laser systems.
Clinical Evidence
Three clinical studies (two controlled, one uncontrolled) evaluated tattoo removal efficacy and safety. Sample size and specific metrics were not detailed, but effectiveness was measured via before-and-after photo evaluation and percentage of tattoo clearance. Results showed high levels of clearance and physician satisfaction, with the device demonstrating a higher percentage of clearance than the control device in controlled studies. Safety was assessed through incidence and severity of adverse events; no serious or unexpected events occurred. Common reactions included mild pain and blistering. All treatment arms showed comparable safety profiles.
Technological Characteristics
Flashlamp-excited Q-switched alexandrite laser; 755nm wavelength; pulse width 450ps-900ps; energy per pulse 0.2 J/cm2; max average fluence 6.37 J/cm2; spot sizes 2-10mm. Conforms to 21 CFR 1040 (Laser Performance Standard). System includes hardware and software components for laser control.
Indications for Use
Indicated for tattoo and benign pigmented lesions removal in patients requiring laser-based dermatological treatment.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked slightly offset from each other. Above the profiles is a design element resembling a wing or feather. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 23, 2016
Cynosure, Incorporated Ms. Irina Kulinets Vice President of Regulatory Affairs and Quality Systems 5 Carlisle Road, Westford, Massachusetts, 01886
Re: K121346 Trade/Device Name: Picosure™ workstation Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 8, 2012 Received: November 9, 2012
Dear Ms. Kulinets:
This letter corrects our substantially equivalent letter of November 27, 2012.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K121346
Device Name Picosure™ workstation
Indications for Use (Describe)
The PicoSure™ workstation is indicated for tattoo and benign pigmented lesions removal.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(K) Owner: | CYNOSURE, INC.<br>5 Carlisle Road<br>Westford, MA 01886<br>Telephone: 978-256-4200<br>Fax: 978 256 6556 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------|
| Contact: | Irina Kulinets<br>VP of Regulatory Affairs and Quality Systems<br>CYNOSURE, INC.<br>5 Carlisle Road<br>Westford, MA 01886 |
| Submitter: | Connie Hoy<br>SVP of Regulatory Affairs and Quality Systems<br>CYNOSURE, INC.<br>5 Carlisle Road<br>Westford, MA 01886 |
| Manufacturer: | CYNOSURE, INC.<br>5 Carlisle Road<br>Westford, MA 01886 |
| Date Prepared: | March 29, 2016 |
| Trade name: | PicoSureTM workstation |
| Common name: | Powered Laser Surgical Instrument (Laser<br>for Surgery and Dermatology) |
| Classification name: | 21 CFR 878.4810 |
| Product Code(s): | GEX (Laser Surgical Instrument for Use in General and<br>Plastic Surgery and in Dermatology) |
| Classification: | Class II |
4. . .... 网 | 时
ાં
.
.
g : ಿ ಕಾ
:
তে
.
网
.
网
at
{4}------------------------------------------------
Review Panel:
### General & Plastic Surgery
### Predicate Devices (Claiming Substantial Equivalence):
K992814 Cynosure Accolade/Accolade II Laser K103118 HOYA ConBio RevLite
#### Summary Description of the Device:
The PicoSure™ workstation is a high-powered, laser system that delivers energy in the 755-nm wavelength. Same as its predicates, the PicoSure™ workstation is intended for tattoos and benign pigmented lesions removal. The combination of wavelength, pulse duration and energy fluence are disrupting the tattoo dye or pigment particles under the skin without harming the surrounding tissue. The fragmented dye or pigment particles eventually surface and fade as the epidermal layer of the skin is renewed.
### Intended Use / Indications for Use:
The PicoSure™ workstation is indicated for tattoo and benign pigmented lesions removal.
### Technological Characteristics:
The PicoSure™ workstation contains equivalent basic technology, components and patient-contact materials as the FDA cleared Cynosure Accolade/Accolade II laser (K992814 and K072868). The fundamental scientific technology of the PicoSure™ workstation is substantially equivalent to the FDA cleared Cynosure Accolade/Accolade II laser (K992814 and K072868).
#### Performance Standards:
This device conforms to the Laser Performance Standard (21 CFR 1040). No additional performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for these devices.
{5}------------------------------------------------
## Performance Data: Non-clinical testing
Performance (bench and animal) testing summaries are included in this submission. Bench testing included:
- . Design verification which includes software and hardware verification
- Design validation which includes software and hardware validation .
- . Production verification which includes in-process inspection and test
Animal data has been used to establish the device safety. A study examined the effect of the PicoSure™ workstation in an animal model (porcine study).
## Performance Data: Clinical testing
Three clinical studies (two controlled and one uncontrolled) were presented in the submission to confirm safety and effectiveness of the device in the tattoo removal. Number of treatments varied in the studies from 4 to 10 sessions. The device effectiveness in removal tattoos was assessed by the evaluation before-and-after photos and calculation the % of tattoo clearance at the end of study for each treatment area. Subject and physician satisfaction with the treatment results was also surveyed and analyzed.
The device safety was evaluated by the incidence, timing, severity of treatment-emergent adverse events were categorized and summarized.
The device was found to be safe and effective, with high level of overall tattoo clearance and "extreme," physician satisfaction (physician evaluation) and, additionally, demonstrated more significant % of highest clearance score than control device in controlled studies.
Treatment-emergent adverse events were recorded and evaluated in all studies. All treatment arms had a comparable safety profile. None of the events were classified as adverse device reaction. No serious or unexpected adverse events were reported. The majority of reported adverse events across the studies were of mild severity; no severe adverse events were reported in any group. The most common reactions were pain and blistering.
{6}------------------------------------------------
## Substantial Equivalence:
The Cynosure PicoSure™ workstation is as safe and effective as its predicate devices currently cleared for the same indications. The Cynosure PicoSure™ workstation is substantially equivalent to the predicate device(s) as shown in the table below and does not does not raise any additional questions of safety and effectiveness. Substantial Equivalence Table is provided below.
| | Proposed Device | Predicate Device | Predicate Device |
|----------------------------|------------------------------------------------------------------|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| 510(k) # | | K992814 | K103118 |
| Company | CYNOSURE, INC. | CYNOSURE, INC. | HOYA ConBio |
| Name | Cynosure PicoSure™<br>workstation | Cynosure<br>Accolade | RevLite |
| Laser Type | Flashlamp excited Q-<br>switched alexandrite | Flashlamp excited Q-<br>switched alexandrite | Flashlamp excited Q-<br>switched alexandrite |
| Nominal Wavelength | 755nm | 755nm | 532 nm, 585 nm 650 nm,<br>& 1064 nm |
| Energy per pulse | 0.2 J/cm2 | 0.57 J/cm2 | 0.85 J/cm2 |
| Maximum Average<br>Fluence | 6.37 J/cm2 | 7.5 J/cm2 | 12 J/ cm² (1064nm)<br>5 J/ cm² (532nm)<br>10 J/ cm² (585nm)<br>6 J/ cm²<br>(650 nm)<br>1.2 J/ cm² (532Lite) |
| Rate | Single pulse, or 1, 2.5,<br>5, or 10 pulse(s) per<br>second (Hz) | Single pulse, or 1,2,5,<br>or 10 pulse(s) per<br>second (Hz) | Single & double pulse, 1,<br>2, 5, & 10 Hz pulses per<br>second |
| Pulse Width | 450ps - 900ps | 45-75 ns | 5-20 ns |
| Spot Sizes | Zoom 2-6 mm,<br>Fixed 2, 3, 4, 6, 8, 10<br>mm | Fixed<br>2.4, 3, 4 mm | Fixed 2 - 8 mm (varies by<br>wavelength) |
# Device Comparison Table - Substantial Equivalence
Conclusion:
PicoSure™ workstation is as safe and efficient as its predicate devices.
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.