The PointerShell Universal and PointerShell LS are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation. It is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to: ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. ENT related anterior skull base procedures.
Device Story
PointerShell Universal and PointerShell LS are reusable electromagnetic navigation instrument adapters; designed to attach to surgical shavers or instruments. Devices incorporate an integrated sensor tracked within a low-energy magnetic field generated by the Fiagon Navigation system. Navigation software maps instrument position onto preoperative CT/MRI/fluoroscopy scans. Used in ENT surgery (e.g., sinus, skull base) by surgeons to precisely locate anatomical structures. Output is visual representation of instrument position relative to patient anatomy on navigation display; assists surgeon in real-time guidance during procedures. Benefits include improved spatial orientation and precision during complex ENT interventions.
Clinical Evidence
Bench testing only. Mean target registration error (accuracy) measured at 1.1 mm for both PointerShell Universal and PointerShell LS, compared to < 1.5 mm for predicate devices. Testing confirmed functionality and compatibility with the Fiagon Navigation system.
Technological Characteristics
Electromagnetic tracking sensor integrated into reusable instrument adapters. Compatible with Fiagon Navigation system. PointerShell Universal accommodates shaft diameters from 2.5mm; PointerShell LS accommodates octagonal instruments up to 6mm diameter. User-sterilized. No specific material standards cited.
Indications for Use
Indicated for patients requiring stereotactic surgery in ENT procedures (e.g., sinus, transphenoidal, intranasal, anterior skull base) where rigid anatomical structures can be referenced to CT/MR models. No specific age or gender contraindications listed.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
Fiagon Navigation Extended Instrument Set ENT Component PointerShell 3mm, 4mm and 5mm (K141456)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: a circular text element and a symbolic graphic. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The graphic element features a stylized depiction of what appears to be an abstract human form or a symbol representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 23, 2016
Fiagon Gmbh Dr. Dirk Mucha CTO Neuendorfstrasse 23b 16761 Hennigsdorf Germany
Re: K160479
Trade/Device Name: Pointershell Universal, Pointershell LS Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: February 16, 2016 Received: February 22, 2016
Dear Dr. Mucha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known): K160479
Device Name: Fiagon Navigation - PointerShell Universal , PointerShell LS
Indications for Use:
The PointerShell Universal and PointerShell LS are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation.
Both the instruments are indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. ENT related anterior skull base procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
And / Or
Over-The-Counter-Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) Summary
## February 16, 2016
#### 1. Submitter Information/ 510(k) Holder
| Submitter: | Fiagon GmbH |
|------------|----------------------------------------------------|
| Address: | Neuendorfstrasse 23b<br>16761 Hennigsdorf, Germany |
| Telephone: | +49 3302 201 21 10 |
| Telefax: | +49 3302 201 21 15 |
Mr. Dirk Mucha, CTO Contact:
#### 2. Device Information
Trade Name: Fiagon Navigation - PointerShell Universal, PointerShell LS Common Name: Image guided surgery system Classification: Class II per 21 CFR 882.4560 Device: Ear, Nose, and Throat Stereotaxic Instrument Product Code: PGW
#### 3. Purpose of Submission
The purpose of this submission is to gain clearance for a geometrically modified instrument for the Image Guided Surgery. The instruments involved in this submission are a modified version of instruments that have been approved within Fiagon Navigation extended instrument Set ENT (K141456).
#### 4. Predicate Device Information
The device described in this submission is substantially equivalent to the following comparable unmodified devices:
| Predicate Device / legally marketed<br>device | Manufacturer | 510(k) No. | |
|-----------------------------------------------|------------------------------------------------------------------------------------------------|-------------|---------|
| 1 | Fiagon Navigation Extended Instrument<br>Set ENT<br>Component<br>PointerShell 3mm, 4mm and 5mm | Fiagon GmbH | K141456 |
{4}------------------------------------------------
#### 5. Device Description
The Fiagon Navigation - PointerShell Universal and PointerShell LS are reusable instruments intended to be used with the Fiagon Navigation system. The instruments are electromagnetically navigated devices that are
- Instrument adapters (designed for mechanical connection to shavers or surgical instruments a. with different diameters without motor).
Each device incorporates a sensor device, which is tracked by the navigation system with in the lowenergy magnetic field of a field generator (part of the navigation system)
The navigation software (part of the navigation system) displays the position of the instruments in preoperative scans (e.g., CT, MRI, fluoroscopy)
#### 6. List of components
| Components | Grouping | Sterility State |
|------------------------|--------------------|-----------------------------------------------------------|
| PointerShell Universal | Instrument Adapter | user sterilized, reusable, delivered in non-sterile state |
| PointerShell LS | Instrument Adapter | user sterilized, reusable, delivered in non-sterile state |
#### 7. Intended Use
The PointerShell Universal and PointerShell LS are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is indicated for use with the Fiagon Navigation system using electromagnetic navigation.
It is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
- -ENT Procedures:
- Transphenoidal access procedures. -
- -Intranasal procedures.
- -Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies,
Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies.
- -ENT related anterior skull base procedures.
{5}------------------------------------------------
#### 8. Comparison of Technological Characteristics
The substantial equivalence of the PointerShell Universal and PointerShell LS to the PointerShell 3mm, 4mm and 5mm are shown by similarities in their intended use, indications for use, materials, and performance.
All devices utilize:
- · Electromagnetic tracking technology for navigation included in surgical instruments
- · Are adapters for instruments for enabling image guidance
The primary difference between the unmodified devices and the new devices is change in the shaft diameter of the devices. While the PointerShell 3mm, 4mm and 5mm incorporate devices of their respective diameters, the PointerShell Universal is a combination of all these 3 devices in one. It incorporates instruments of shaft diameters ranging from 2.5mm.The PointerShell LS is used for navigation of octagonal shavers or octagonal surgical instruments with a maximum shaft diameter of 6mm. These changes do not raise new issues of safety and effectiveness.
#### 9. Performance Data
Testing was performed in order to determine device precision and accuracy. The following nonclinical tests were performed to determine substantial equivalence:
A mean bench accuracy as target registration error of 1.1 mm for Pointershell Universal and 1.1 mm for PointerShell LS was measured for the new devices, which compares to the values mean < 1.5 mm for the predicate devices in an identical test setup.
The results support the claim of substantial equivalence to the unmodified devices.
Testing was completed to ensure functionality and compatibility with the Fiagon Navigation system.
#### 10. Conclusion
Based on the indications for use, technological characteristics, performance testing and comparison to the predicates, the PointerShell Universal and the PointerShell LS have been shown to be substantially equivalent to the comparable devices PointerShell 3mm, 4mm and 5mm and does not present any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.