ReadeR is a dental imaging system indicated for capturing, digitization and processing of intra oral x-ray images stored in imaging plate recording media.
Device Story
Computer radiography system for dental imaging; replaces traditional film. Input: reusable x-ray storage phosphor plates (IP) exposed to x-rays in patient's mouth. Operation: device scans IP with laser; releases latent image as light; converts light to digital signal. Processing: digital filtering, gain/offset correction, flat fielding. Output: digital image viewed on workstation, stored locally, sent to PACS, or printed. Used in dental offices by dentists/staff. Benefits: improved diagnostic capability for tooth decay and periodontal disease; reusable plates with disposable covers prevent cross-contamination.
Clinical Evidence
Bench testing only. Includes electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software validation (EN 62304), risk analysis (ISO 14971), and usability (IEC 62366). Image quality/resolution testing performed. Clinical images provided as supplemental evidence but not required for equivalence determination.
Technological Characteristics
Computed radiography scanner using high sensitivity photo multiplier tube (s-PMT). Imaging plates: BaSrFBrI:Eu phosphor. Connectivity: DICOM 3 compatible. Power: 100-240V, 50/60Hz. Software: Deep-View (OBL of Archimed Suite). Dimensions/Weight: 4.6 kg. Sterilization: disposable single-use plate covers.
Indications for Use
Indicated for dentists and qualified dental professionals to process intraoral x-ray images captured on imaging plates from the skull complex.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
Predicate Devices
NICAL SMART MICRO CR System for Intraoral Dental Images (K143703)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 10, 2016
Trident s.r.l. % Mr. Claude Berthoin President Denterprise International, Inc. 100 East Granada Blvd., Suite 219 ORMOND BEACH FL 32176
Re: K160386 Trade/Device Name: ReadeR. VieweR. OuickScan PSP Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: February 08, 2016 Received: February 11, 2016
Dear Mr. Berthoin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Oels
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160386
Device Name ReadeR, VieweR, QuickScan PSP
#### Indications for Use (Describe)
ReadeR is a dental imaging system indicated for capturing, digitization and processing of intra oral x-ray images stored in imaging plate recording media.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white drawing of an eagle. The eagle is facing to the right, and its wings are spread out. The eagle's head is turned to the side, and its beak is open. The eagle's feathers are detailed, and the overall image is well-drawn. The eagle appears to be in flight.
# 510k FDA Consulting
Medical Device Clearance
100 East Granada Blvd., Suite 219
Ormond Beach, FL 32176
386-506-8711
# 510(k) Summary
#### Submitter/Applicant
Trident s.r.l. Via Verdi 20 Assago, Milan 20090, Italy Phone: +39-02-87072380 Fax: +39-02-87072381 Contact: Giorgio Rizzo, President (giorgio.rizzo33@gmail.com)
Date Prepared: February 08, 2016
#### Preparer/Consultant
Denterprise International, Inc. 100 East Granada Blvd., Suite 219 Ormond Beach, FL 32176
Phone: 386-506-8711 Fax: 855-235-7902 Contact: Joyce St. Germain, Regulatory Executive (Joyce@510kfda.com) Secondary Contact: Claude Berthoin, President (Claude@denterpriseintl.com)
#### Device Classification
| Trade/Model Names: | ReadeR, VieweR, QuickScan PSP |
|-------------------------------|------------------------------------------|
| Common Name: | Computed Radiography Scanner System |
| Classification Name: | System, X-Ray, Extraoral Source, Digital |
| Regulation Name: | Extraoral source x-ray system |
| Regulation Number: | 21 CFR 872.1800 |
| Product Code: | MUH |
| Regulatory Class: | 2 |
| 510k Review Panel: | Radiology |
| Regulation Medical Specialty: | Dental |
{4}------------------------------------------------
# Predicate Device
The subject device claims equivalence to the following legally marketed predicate:
| 510(k) Number: | K143703 |
|----------------------|------------------------------------------------------------|
| Date Cleared | March 18, 2015 |
| Trade Name: | NICAL SMART MICRO CR System for Intraoral Dental<br>Images |
| Common Name: | Computed Radiography Scanner System |
| Classification Name: | System, X-ray, Extraoral Source, Digital |
| Regulation Name: | Extraoral source x-ray system |
| Regulation Number: | 21 CFR 872.1800 |
| Product Code: | MUH |
| Regulatory Class: | 2 |
| Medical Specialty: | Dental |
| 510k Review Panel: | Radiology |
#### Indications for Use
ReadeR is a dental imaging system indicated for capturing, digitization and processing of intraoral x-ray images stored in imaging plate recording media.
#### Intended Use
ReadeR system is intended to be used only by dentist and other qualified dental professionals to process x-ray images exposed to the imaging plates from the intraoral complex of the skull.
It is possible for both tooth decay and periodontal disease to be missed during a clinical exam, and radiographic evaluation of the dental and periodontal tissues is a critical segment of the comprehensive oral examination. In some cases extensive decay has been overlooked by a number of dentists prior to radiographic evaluation. So, the sensors and the PSP plates are in the best interest of the patient for improved healthcare. The ReadeR is a necessity for the dental office that is using the PSP plates.
# Device Description
The ReadeR is a computer radiography system which produces the x-ray diagnostic image in digital format instead of using traditional screens and film. This device does not have a wireless transmission. The device utilizes reusable x-ray storage phosphor plate (IP) that is sensitive to x-ray and stores latent image when it is exposed to x-ray. After x-ray exposure to the x-ray storage phosphor plate, x-ray storage phosphor plate is scanned by means of laser in the device. Latent image in the x-ray storage phosphor plates is released in a form of light by laser scanning. Then the light is collected and converted into a form of digital image. The signal processing is made to the digital image data such as the digital filtering, the gain and offset correction and flat fielding. The image can then be viewed on a computer workstation, adjusted if necessary, the stored locally, sent to an archive, printed or sent to PACS system.
{5}------------------------------------------------
After acquisition of the latent image from the x-ray storage phosphor plate, it is erased thoroughly to be reused.
# Comparison of Technological Characteristics with Predicate
The indications for use of the subject and predicate devices are identical and the technologies are substantially equivalent.
The imaging plates are exposed to x-rays while in the patient's mouth. The plates are removed and scanned by the scanning device. The images can be viewed by the dentist and stored on a personal computer. The plates are then erased and reused.
Patient cross contamination is prevented by the use of disposable, single use plate covers.
The imaging plates are identical to those used in the predicate device, as both systems are sharing the main part (CR Reader), the only difference with the subject and predicate device is the outside cover of the devices for private labeling. (The predicate device is horizontal and the subject device is vertical.)
The following table compares technological and other characteristics of the subject and predicate device.
| | | Trident<br>ReadeR<br>510(k) number unknown | Nical Smart Microchip CR<br>K 143703 |
|-----------------------------|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | | The ReadeR system is indicated<br>for capturing, digitization and<br>processing of intra oral X-Ray<br>images stored in imaging plate<br>recording media. | Nical Smart Microchip CR system<br>is indicated for capturing,<br>digitization and processing of<br>intra oral X-Ray images stored in<br>imaging plate recording media. |
| Physical<br>characteristics | Imaging<br>plate size | Size 0 (22mm x 31mm),<br>Size 1 (24mm x 40mm),<br>Size 2 (31mm x 41mm),<br>Size 3 27mm x 54mm) | Size 0 : 22 x 31mm<br>Size 1 : 24 x 40mm<br>Size 2 : 31 x 41mm<br>Size 3 : 27 x 54mm |
# Table of Comparison
{6}------------------------------------------------
| | Trident<br>ReadeR<br>510(k) number unknown | Nical Smart Microchip CR<br>K 143703 |
|--------------------------|---------------------------------------------------|---------------------------------------------------|
| Effective<br>pixel pitch | 30 μ | 30 μ |
| Spatial<br>resolution | 10.5 lp/mm @ 30μm | 10.5 lp/mm @ 30μm |
| | Size 0: 343 x 484 @ 64um<br>628 x 885 @ 35um | Size 0: 343 x 484 @ 64um<br>628 x 885 @ 35um |
| Image<br>Matrix | Size 1: 375 x 625 @ 64um<br>685 x 1143 @ 35um | Size 1: 375 x 625 @ 64um<br>685 x 1143 @ 35um |
| | Size 2: 484 x 640 @ 64um<br>886 x 1171 @ 35um | Size 2: 484 x 640 @ 64um<br>886 x 1171 @ 35um |
| | Size 3: 891 x 1783 @ 30 um | Size 3: 891x 1783 pixel 30 um |
| Weight | 4.6 kg | 4.7 kg |
| Imaging device | High Sensitivity Photo Multiplier<br>Tube (s-PMT) | High Sensitivity Photo Multiplier<br>Tube (s-PMT) |
| Operational<br>characteristics | Operating | Temperature :15-30°C | RH Temperature: 10 - 40 °C |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| | Conditions | Humidity: 15%-95% | Humidity: 30 - 90% RH |
| | Power | 100-240V, 50/60Hz | 100-240V, 50/60Hz |
| | Method of<br>imaging | Register Patient -> X-ray<br>Exposure | Register Patient -> X-ray<br>Exposure |
| | Image<br>receptor | Imaging plate | Imaging plate |
| | | | |
| Functional<br>characteristics | Output data | DICOM 3 compatible | DICOM 3 compatible |
| | Performance | MTF: 65Kv 0,1 sec exposure.<br>100%. 1 lp/mm<br>95%. 3 lp/mm<br>80%. 5 lp/mm<br>30%. 10 lp/mm | MTF: 70 kV 0.08 sec exposure.<br>55% 1 lp/mm<br>11% 3 lp/mm |
| | Defect<br>compensation | By Imaging plate calibration | By Imaging plate calibration |
| | Dynamic<br>range | 16 bit | 16 bit |
| | Image<br>Processing | Single image processing<br>parameter is used | Single image processing<br>parameter is used |
| | Imaging plate<br>characteristics | Composition | BaSrFBrl:Eu phosphor |
| | | Thickness | approx. 350 mm |
| | | Typical<br>luminescence | 400 nm |
| | | Image<br>Retention | Recommended within 1 hour of<br>exposure. |
| Imaging<br>software | | Deep-View | Nical Image Plus |
| | The two acquisition and imaging software are two Own Brand Labeling (OBL) of<br>the same original software Archimed Suite, that is CE certified with certificate<br>1575/MDD issued by IMQ (Italy).<br>Deep-View holds the CE certificate number 1749/MDD, while Nical Image Plus<br>holds certificate number 1748/MDD, both release by IMQ (Italy). | | |
| | The two software packages have the same functionalities, and differ only from the<br>aesthetic screens. | | |
{7}------------------------------------------------
{8}------------------------------------------------
The above comparison shows the subject and predicate devices are substantially equivalent in technology characteristics. The differences are highlighted and none of these differences make the subject device any less safe and effective as the predicate device. The devices vary in weight by 0.1 kg. (ReadeR being lighter in weight) due to the outer covering of the devices. The operating temperatures vary, but these devices are used in an office setting and the temperature will be controlled to a comfortable median temperature of the operating temperature range. This is also due to the outer covering of the two devices being different. The MTF on performance is listed above and the variation is that the subject device is at 0.1 sec dose and the predicate is at 0.08 sec dose, so the percentages are different. All other technological characteristics are identical. The subject and predicate devices are identical except for the outer cover and the subject device stands vertically and the predicate device is horizontal.
# Non-Clinical Performance Data
The following performance data is provided in support of the substantial equivalence determination. All tests performed are included in this submission.
Biocompatibility ... This device does not require a biocompatibility evaluation. The plate covers are commonly used and have previous approval.
Electrical Safety and EMC ... testing was performed by a qualified testing laboratory according to IEC 60601-1 and IEC 60601-1-2. The device met the requirements according to these standards and was performed by Prima Ricercaq & Sviluppo S.r.1. of Italy.
Software ... validation of software (EN 62304) and risk analysis (ISO 14971:2012) were performed and were met according to these standards. IEC 62366:2007 for use in conjunction with 60601-1-6:2010 was performed by TUV Rheinland North America and the standards were met.
Performance Testing ... Image testing was performed to check image quality and resolution. Some of these test results are shown in the comparison table above. A quality assurance procedure and inspection was performed on the imaging plates and all testing was approved and standards were met since the predicate has CE mark and the same accessories are supplied for the subject device. Clinical images were provided; these images were not necessary to establish substantial equivalence based on the modification of the device (note CR reader technology that is identical to the predicate) but they provide further evidence in addition to the bench performance data to show that the CR system works as intended.
#### Conclusion
The subject and predicate devices have the same intended use and are substantially equivalent in technological characteristics and performance (since they are, in fact, the same device, with different branding for marketing purposes). The ReadeR does not raise any
{9}------------------------------------------------
questions regarding safety and effectiveness since it is the same as the predicate device. The non-clinical data supports and demonstrates the safety of the device.
The conclusion is that ReadeR (VieweR and QuickScan PSP) warrants a finding of substantial equivalence to the legally marketed original Nical Smart Micro CR System, and therefore, should have clearance for premarket activities in the United States.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.