NICAL SMART MICRO CR system for intraoral dental images
K143703 · Nical Spa · MUH · Mar 18, 2015 · Dental
Device Facts
Record ID
K143703
Device Name
NICAL SMART MICRO CR system for intraoral dental images
Applicant
Nical Spa
Product Code
MUH · Dental
Decision Date
Mar 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Indications for Use
NICAL SMART MICRO CR system for intraoral dental images is indicated for capturing, digitization and processing of intra oral x-ray images stored in imaging plate recording media.
Device Story
NICAL/SMART MICRO is a Computed Radiography (CR) system for dental imaging. It uses reusable X-ray storage phosphor plates (BaSrFBrl:Eu) as input media. After X-ray exposure in the patient's mouth, the plate is inserted into the reader. A laser diode (620-650 nm) scans the plate, releasing latent image energy as light (420 nm peak). A light guide collects this light, directing it to a Hamamatsu photomultiplier tube (PMT). The PMT signal is digitized via a 16-bit A/D converter and processed (digital filtering, gain/offset correction, flat fielding). The system outputs digital images compatible with DICOM 3.0 for viewing on a computer workstation, archiving, or printing. Used in dental clinics by dental professionals. The device replaces traditional film, allowing for digital image adjustment and storage. Patient cross-contamination is managed via single-use barrier envelopes.
Clinical Evidence
Bench testing only. Performance verified via MTF analysis, spatial resolution (10.5 lp/mm), and image quality assessment. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2. Biocompatibility of hygienic envelopes verified. Clinical images were reviewed by a board-certified radiologist to confirm excellent quality, though not required for substantial equivalence.
Technological Characteristics
Computed Radiography scanner. Uses BaSrFBrl:Eu phosphor imaging plates. Laser diode (620-650 nm) scanning; Hamamatsu photomultiplier tube (PMT) sensor. 16-bit A/D conversion. Connectivity via USB 2.0 (FTDI 2232HL IC). DICOM 3.0 compatible. Dimensions: 110x154x270 mm. Power: 100-240V, 50/60Hz. Software performs digital filtering, gain/offset correction, and flat fielding.
Indications for Use
Indicated for capturing, digitization, and processing of intraoral x-ray images stored in imaging plate recording media for dental patients.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing design that resembles a wing or a wave.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 18, 2015
NICAL SPA % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Court NAPLES FL 34114
Re: K143703
Trade/Device Name: NICAL SMART MICRO CR System for Intraoral Dental Images Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: January 28, 2015 Received: March 6, 2015
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Exhibit 4.
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K143703
Device Name NICAL SMART MICRO CR system for intraoral dental images
Indications for Use (Describe)
NICAL SMART MICRO CR system for intraoral dental images is indicated for capturing, digitization and processing of intra oral x-ray images stored in imaging plate recording media.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### Exhibit 1. 510(k) SUMMARY
{5}------------------------------------------------
## 510(k) Summary, K143703 This 510(k) summary information is being submitted in accordance with requirements of 21 CFR Part 807.92.
## Date prepared: March 2, 2015
- 1. Company and Correspondent: NICAL SPA Via Soffredini, 43 -20126 Milano Italy Phone +39 022571110 -Fax +39 022572207 www.nical.com Prepared by: Roberto Niccolucci, President
- 2. Device
Trade/Proprietary Name: NICAL/SMART MICRO Common Name: Computed Radiography Scanner System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH
- 3. Predicate Device: FireCR Dental, made by 3D Imaging & Simulations Corp., K131442.
- 4. Indications for Use: The NICAL/SMART MICRO Dental Imaging System is indicated for capturing, digitization and processing of intra oral x-ray images stored in imaging plate recording media.
- 5. Description: The NICAL/SMART MICRO is a Computed Radiography System which produces the Xray diagnostic image in digital format instead of using traditional screens and film. This device utilizes reusable X-ray storage phosphor plate (IP) that is sensitive to X-ray and stores latent image when it is exposed to X-ray. After X-ray exposure to the X-ray storage phosphor plate, X-ray storage phosphor plate is scanned by means of laser in the device. Latent image in the X-ray storage phosphor plate is released in a form of light by laser scanning. Then the light is collect and converted into a form of digital image. The signal processing is made to the digital image data such as the digital filtering, the gain & offset correction and flat fielding. The image can then be viewed on a computer workstation, adjusted if necessary, then stored locally, sent to an archive, printed or sent to PACS system. After acquisition of the latent image from the X-ray storage phosphor plate, it is erased thoroughly to be reused. Details: A laser diode is focused on the plate with a diameter of 30 µm, total power is 35 mW (CW) at wavelength from 620 to 650 µm. The blue light (400 nm) is collected from a light guide (NICAL patent) and sent to a Hamamatzu photomultiplier (PMT). Wavelength (Peak): 420 nm. The signal output from PMT is digitized from a 16 bit A/D converter. The 16 bit digitized imaging signal is sent to an FTDI 2232HL IC that contains all the logic to interface to USB2 standard. The time to read a full image depends on the size of the plate: size 0: 5 seconds; size 3: 9 seconds.
- 6. Comparison with predicate device: The indications for use and the technologies employed are virtually identical to those of the predicate. Imaging plates are exposed to x-rays while in the patient's mouth. The plates are removed and scanned by the scanning device. The resulting images can be viewed by the dentist and stored on a personal computer. The plates are then
{6}------------------------------------------------
erased and can be reused. Patient cross contamination is prevented by the use of single use barrier envelopes placed over the plates. The imaging plates are identical to those used in the predicate device. Comparison of the test results: The MTF performance (see chart below) is superior on the Nical system as compared to the predicate. The effective pixel pitch is smaller, 30 um vs 35 um giving slightly better overall resolution. This shows up in a comparison of the spatial resolution: 9.0lp/mm @ 35um on the predicate vs. 10.5 lp/mm @ 30um on the Nical device. A detailed comparison table follows.
| | | FireCR Dental<br>3D Imaging & Simulations<br>Corp. | NICAL/SMART MICRO Dental Imaging<br>System |
|------------------------|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | | K131442 | K143703 |
| Intended Use | | The FireCR Dental imaging<br>system is indicated for<br>capturing, digitization and<br>processing of intra oral x-<br>ray images stored in<br>imaging plate recording<br>media. | The NICAL/SMART MICRO Dental<br>Imaging System is indicated for<br>capturing, digitization and processing of<br>intra oral x-ray images stored in imaging<br>plate recording media. |
| Physical<br>Char. | Overall<br>Dimensions | Reader<br>185 x 100 x 293mm | Reader<br>110x154x270 mm |
| | Imaging Area | Size 0 : 22 x 31mm<br>Size 1 : 24 x 40mm<br>Size 2 : 31 x 41mm<br>Size 3 : 27 x 54mm<br>Size 4c : 48 x 54mm | Size 0 (22mm x 31mm),<br>Size 1 (24mm x 40mm),<br>Size 2 (31mm x 41mm),<br>Size 3 (27mm x 54mm)<br>N/A |
| | Effective Pixel<br>Pitch | 35um, 64um | 30 um. |
| | Spatial<br>Resolution | 9.0lp/mm @ 35um | 10.5 lp/mm @ 30um |
| | Image Matrix | Size 0<br>343 x 484 @ 64um<br>628 x 885 @ 35um<br>Size 1<br>375 x 625 @ 64um<br>685 x 1143 @ 35um<br>Size 2<br>484 x 640 @ 64um<br>886 x 1171 @ 35um<br>Size 3<br>421 x 843 @ 64um<br>771 x 1542 @ 35um<br>Size 4c<br>750 x 843 @ 64um<br>1370 x 843 @ 35um | Size 0<br>726x1024 pixel 30 um.<br>Size 1<br>792x1321 pixel 30 um.<br>Size 2<br>1024x1354 pixel 30 um.<br>Size 3<br>891x1783 pixel 30 um.<br>Not applicable |
| | | FireCR Dental<br>3D Imaging & Simulations<br>Corp. | NICAL/SMART MICRO Dental Imaging<br>System |
| | Photo | Image: FireCR Dental 3D Imaging | Image: NICAL/SMART MICRO Dental Imaging System |
| | Weight | 4.7kg | 4.6kg |
| | Imaging<br>Device | High Sensitivity Photo<br>Multiplier Tube (s-PMT) | High Sensitivity Photo Multiplier Tube<br>(s-PMT) |
| Operational<br>Char | Operating<br>Condition | Temperature :15-30°C<br>Humidity: 15%-95% RH | Temperature: 10 - 40 °C<br>Humidity: 30 - 90% RH |
| | Power | 100 – 240V, 50/60Hz | 100 – 240V, 50/60Hz |
| | Methods of<br>Exposure | Register Patient -><br>X-ray Exposure | Register Patient →<br>X-ray Exposure |
| | X-ray<br>Absorber | Imaging plate | Imaging plate |
| Functional<br>Char. | Output Data | Dicom3.0 Compatible | Dicom3.0 Compatible |
| | Performance | MTF: 70 kV 0.08sec dose.<br>55% 1 lp/mm<br>11% 3 lp/mm | MTF: 65Kv 0,1 sec dose.<br>100%. 1 lp/mm<br>95%. 3 lp/mm<br>80%. 5 lp/mm<br>30%. 10 lp/mm |
| | Defect<br>Compensation | By Calibration | By Calibration |
| | Dynamic<br>Range | 16 bit | 16 bit |
| | Image<br>Processing | Single image processing<br>parameter is used | Single image processing parameter is<br>used |
| | FireCR Dental<br>3D Imaging & Simulations<br>Corp. | NICAL/SMART MICRO Dental Imaging<br>System | |
| DICOM<br>Compatibility | DICOM 3.0 Compliant | DICOM 3.0 Compliant | |
{7}------------------------------------------------
{8}------------------------------------------------
| Imaging Plate Comparison | | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | FireCR Dental<br>3D Imaging & Simulations Corp. | NICAL/SMART MICRO Dental Imaging System |
| 510(k) number | K131442 | K143703 |
| Imaging Plate Composition | BaSrFBrl:Eu phosphor | BaSrFBrl:Eu phosphor |
| Plate Sizes | Size 0 : 22 x 31mm | Size 0 (22mm x 31mm) |
| | Size 1 : 24 x 40mm | Size 1 (24mm x 40mm) |
| | Size 2 : 31 x 41mm | Size 2 (31mm x 41mm) |
| | Size 3 : 27 x 54mm | Size 3 (27mm x 54mm) |
| | Size 4c : 48 x 54mm | N/A |
| Thickness | Thickness: approx. 350 mm | Thickness: approx. 350 mm |
| Typical luminescence | 400 nm | 400 nm |
| Image Retention | Recommended within 1 hour of exposure. Two hours after exposure 70% of the stored energy still is present with no visible loss of information upon readout. Image retention still exceeds 45% after 24h | Recommended within 1 hour of exposure. Two hours after exposure 70% of the stored energy still is present with no visible loss of information upon readout. Image retention still exceeds 45% after 24h |
- 7. Description of non-clinical testing: Software validation and risk analysis was performed. Electrical safety and EMC testing was performed by a qualified testing laboratory according to IEC 60601-1 and IEC 60601-1-2. A verification report on image testing was performed to check image quality and resolution. A quality assurance procedure has been established for the incoming inspection of the imaging plates. Biocompatibility of the hygienic envelopes has been verified.
- 8. Description of clinical testing: Clinical Images were reviewed by a Board Certified US based radiologist and were found to be of excellent quality, suitable for the intended use. These images were not necessary to establish substantial equivalence based on the comparison to the predicate device (Note: X-ray storage phosphor plate that is identical to the predicate image plate) but they provide further evidence in addition to the laboratory performance data to show that the complete system works as intended."
- 9. Conclusion: Based on the similarity to the predicate device in terms of technology, performance, and indications for use, we conclude that the Nicro Dental Imaging System is substantially equivalent to the predicate device named above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.