K160334 · Keosys Sas · LLZ · Jun 21, 2016 · Radiology
Device Facts
Record ID
K160334
Device Name
Advanced Medical Imaging Software Suite (KSWVWR)
Applicant
Keosys Sas
Product Code
LLZ · Radiology
Decision Date
Jun 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Keosys Medical Imaging Suite (KSWVWR) is intended to be used by trained medical professionals included, but not limited to, radiologists, nuclear medicine physicians and physicists. Keosys Medical Imaging Suite is a software application intended to aid in diagnostic and evaluation of medical image data. Although this device allows the visualization of mammography images, it is not intended as a tool for primary diagnosis in mammography. Keosys Medical Imaging Suite can be used for display, process, temporarily store, print and also create and print reports from 2D and 3D multimodal DICOM medical image data. The imaging data can be Computed Tomography (CT), Magnetic Resonance (MR), Radiography X (CR, DX, XA, XRF, MG), Nuclear Medecine (NM) including planar imaging (Static, Whole body, Dynamic, Gated) and tomographic imaging (SPECT), Positon Emission Tomography (PT), Ultrasound (US). Keosys Medical Imaging suite provides tools like rulers, markers or region of interests (e.g. it can be used in an oncology clinical workflow for tumor burden assessment or therapeutic response evaluation). It is the user responsibility to check that the ambient luminosity conditions, the images compression ratio and the interpretation monitor specifications are consistent with a clinical diagnostic use of the data.
Device Story
Software-based multimodality diagnostic workstation; processes 2D/3D DICOM images from CT, MR, X-ray, NM, SPECT, PET, and US. Provides visualization, 3D post-processing, and measurement tools (rulers, markers, ROIs). Used by radiologists and nuclear medicine physicians in clinical settings for oncology workflows, including tumor burden assessment and therapeutic response evaluation. Outputs include displayed images, processed data, and printed reports. Assists clinicians in diagnostic decision-making by facilitating image manipulation, fusion, and quantitative analysis (e.g., SUV calculations). Benefits patients through improved diagnostic evaluation and standardized therapeutic response monitoring.
Clinical Evidence
Bench testing only. No clinical data provided. Performance and functional testing conducted per internal software development process and documented in a Software Test Plan.
Technological Characteristics
Software-only device; no hardware energy delivery. Operates on standard computer hardware. Supports DICOM standard for multimodal image input (CT, MR, X-ray, NM, SPECT, PET, US). Features include 3D post-processing, image fusion, and measurement tools (ROI, markers).
Indications for Use
Indicated for trained medical professionals (radiologists, nuclear medicine physicians, physicists) to aid in diagnostic evaluation of 2D/3D multimodal DICOM medical images (CT, MR, X-ray, NM, SPECT, PET, US). Includes oncology workflow tools for tumor burden/therapeutic response assessment. Not for primary mammography diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 21, 2016
KEOSYS SAS % Ms. Laurence Boy-Machefer Regulatory Affairs and Ouality Assurance Manager 13 Impasse Serge Reggiani, B.P. 10222 Saint Herblain cedex, 44815 FRANCE
Re: K160334
Trade/Device Name: Advanced Medical Imaging Software Suite (KSWVWR) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 13, 2016 Received: May 16, 2016
Dear Ms. Boy-Machefer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Oolo
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160334
Device Name
Advanced Medical Imaging Software Suite (KSWVWR)
Keosys Medical Imaging Suite (KSWVWR) is intended to be used by trained medical professionals included, but not limited to, radiologists, nuclear medicine physicians and physicists.
Keosys Medical Imaging Suite is a software application intended to aid in diagnostic and evaluation of medical image data. Although this device allows the visualization of mammography images, it is not intended as a tool for primary diagnosis in mammography.
Keosys Medical Imaging Suite can be used for display, process, temporarily store, print and also create and print reports from 2D and 3D multimodal DICOM medical image data. The imaging data can be Computed Tomography (CT),
Magnetic Resonance (MR), Radiography X (CR, DX, XA, XRF, MG), Nuclear Medecine (NM) including planar imaging (Static, Whole body, Dynamic, Gated) and tomographic imaging (SPECT), Positon Emission Tomography (PT), Ultrasound (US).
Keosys Medical Imaging suite provides tools like rulers, markers or region of interests (e.g. it can be used in an oncology clinical workflow for tumor burden assessment or therapeutic response evaluation).
It is the user responsibility to check that the ambient luminosity conditions, the images compression ratio and the interpretation monitor specifications are consistent with a clinical diagnostic use of the data.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a logo for Keosys Medical Imaging. The logo consists of a large blue circle with a white "K" inside of it. To the right of the circle is the word "KEOSYS" in blue, with the words "MEDICAL IMAGING" underneath it in a smaller font.
# 510(k) Summary of Safety and Effectiveness
This summary of safety and effectiveness is provided as part of the Premarket Notification for KSWVWR in conformance with 21 CFR 807.92.
| Date Prepared: | January 2016. |
|-----------------------------|--------------------------------------------------|
| Submitter: | Keosys S.A.S. |
| | 13, impasse Serge Reggiani, B.P. 10222 |
| | 44815 Saint-Herblain cedex, FRANCE |
| Contact name: | Ms. Laurence Boy-Machefer |
| Contact Email: | laurence.boy-machefer@keosys.com |
| Contact Telephone: | 00 33 (0)2 53 59 12 90 |
| Contact Fax: | 00 33 (0)2 53 59 12 91 |
| Device Trade Name: | Advanced Medical Imaging Software Suite (KSWVWR) |
| Device Common Name: | Viewer |
| Device Classification Name: | Picture Archiving and Communication System |
| Classification Panel: | 21 CFR §892.2050 |
| Product code: | LLZ |
| Device Classification: | Class II |
| Predicate Devices: | |
| Name | Manufacturer | 510(k)<br>number |
|-----------------------------------------------|-----------------------------|------------------|
| Myrian | Intrasense | K113620 |
| MIM 5.2 | MIM Software | K103576 |
| XD | Mirada | K101228 |
| Hermes Medical Imaging<br>Software Suite v5.4 | Hermes Medical Solutions AB | K140269 |
| Imagys | Keosys S.A.S | K102000 |
#### I) Device Description:
Keosys' Advanced Medical Imaging Software Suite (aka Viewer, aka KSWVWR) is a multimodality diagnostic workstation for visualization and 3D post-processing of Radiological and Nuclear Medicine medical images. It includes dedicated applications for the therapeutic response evaluation process in a multi-vendor, multi-modal and multi time-points context. The solution also includes the latest recommendations for SUV calculation.
#### II) Indications for Use:
Keosys Medical Imaging Suite (KSWVWR) is intended to be used by trained medical professionals included, but not limited to, radiologists, nuclear medicine physicians and physicists.
Keosys Medical Imaging Suite is a software application intended to aid in diagnostic and evaluation of medical image data. Although this device allows the visualization of mammography images, it is not intended as a tool for primary diagnosis in mammography.
Keosys Medical Imaging Suite can be used for display, process, temporarily store, print and also create and print reports from 2D and 3D multimodal DICOM medical image data. The imaging data can be Computed Tomography (CT), Magnetic Resonance (MR), Radiography X (CR, DX, XRF, MG), Nuclear Medecine (NM) including planar imaging (Static, Whole body, Dynamic, Gated) and tomographic imaging (SPECT, Gated SPECT), Positon Emission Tomography (PT), Ultrasound (US).
Keosys Medical Imaging suite provides tools like rulers, markers or region of interests (e.g. it can be used in an oncology clinical workflow for tumor burden assessment or therapeutic response evaluation).
{4}------------------------------------------------
It is the user responsibility to check that the ambient luminosity conditions, the images compression ratio and the interpretation monitor specifications are consistent with a clinical diagnostic use of the data.
### III) Comparison of the technical characteristics between the device and its predicate devices:
The intended use of the Viewer and the Substantial Equivalent devices are similar with regard to the following points:
- all of the devices are providing tools to aid the physician in the diagnosis of patient,
- all of the devices support display, manipulation and fuse tools,
- all of the devices can use medical images from multiple modalities as SPECT, PET, etc.
The indications for use of the Viewer and SE devices are similar for the reasons expressed in the above point.
The target population is globally the same for all the devices.
All of the devices can be used inside the medical environment (within the "intended use" and the "indications for use" limits). Systems requirements differs only slightly.
The devices are not using any energy apart from the electrical needs for the computer power supply. Moreover, there's no energy delivered.
The Viewer and SE devices are all software components which rely on some human factors represented by the user of the device.
The general design of all the devices software appears to be identical, based on available data.
All of the devices have specific manufacturer's software methods and software to address the performance and functional testing.
#### IV) Testing:
Keosys Medical Imaging Advanced Software Suite is tested according to the specifications that are documented in a Software Test Plan. Performance and functional testing are an integral part of Keosys's software development process.
Image /page/4/Picture/15 description: The image contains the logo for Keosys Medical Imaging. The logo features a large, stylized letter "K" in blue, followed by the word "KEOSYS" in a similar blue font. Below "KEOSYS" are the words "MEDICAL IMAGING" in a smaller font, also in blue. The overall design is clean and professional, suggesting a company in the medical technology field.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.