The manufacturer conducts a post-marketing review of published clinical literature to evaluate the device's performance and safety in routine clinical environments, ensuring identified risks are covered and no new risks are identified.
Post-marketing review; Literature review; Clinical environment performance
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Post-marketing literature review; Literature review of clinical studies; Study Period: Post-marketing
Patients in clinical environments using the Myrian device
Not applicable for this study
Device performance and safety in clinical environment
Indications for Use
Myrian® is a software medical device aimed at reviewing images produced by all standard medical imaging devices. It also includes DICOM communication, media interchange (printing, CD burning, storing) and reporting features. Myrian® can be run from any standard client platform (such as PC) that might be purchased independently by the end user. Common Users are trained medical professionals, including surgeons, radiologists, clinicians and technicians. This device is not indicated for mammography use. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
Device Story
Myrian is a software-based medical image review and analysis platform. It ingests DICOM-compliant images from standard medical imaging modalities. The software provides tools for visualization (MPR, 3D rendering, endoscopic exploration), manual/interactive annotation (ROI, POI, paths), and longitudinal follow-up for oncology. It operates on standard off-the-shelf PC hardware, either as a standalone application or via a 'Remote Execution Mode' (REM) where a server streams screen content to client platforms over a network. Physicians use the software to perform measurements, quantify lesions, simulate surgical procedures (e.g., liver/lung excision), and generate reports. The device supports clinical decision-making by automating response calculations (e.g., RECIST 1.1, Cheson) and facilitating preoperative planning. Benefits include standardized workflow, improved lesion tracking, and remote access to diagnostic tools for radiologists and surgeons.
Clinical Evidence
Bench testing only. Evidence includes pre-integration, functional, integration, final integration, regression, and build testing to verify system requirements and specifications. Performance and safety were evaluated through internal performance evaluations of segmentation engines and clinical features. Post-marketing review of literature is used to confirm safety and performance in clinical environments.
Technological Characteristics
Software-based medical image processing system. Operates on standard off-the-shelf PC hardware. Supports DICOM communication, media interchange, and reporting. Visualization modes include MPR, 3D (MIP, MiniP, Average, Volume Rendering), and endoscopic exploration. Connectivity via LAN/WAN for remote execution. Security via encryption of screen streams. Software includes modules for lung, liver, vessel, and oncology analysis. No specific material standards applicable as it is a software-only device.
Indications for Use
Indicated for trained medical professionals (surgeons, radiologists, clinicians, technicians) to review, analyze, and report on medical images from standard imaging devices. Contraindicated for primary interpretation of mammography using lossy compressed or digitized film-screen images; mammography requires FDA-approved monitors with at least 5 MP resolution.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
JUL 12 2012
# 5. 510(K) SUMMARY
### [As Required by 21 CFR 807.92] Summary of Safety and Effectiveness
| Preparation date | July 13th, 2012 |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter<br>Name<br>Registration number | INTRASENSE<br>3006546169 |
| Address<br>Tel Fax | 1231 Avenue du Mondial 98 34900 Montpellier FR<br>(+33) 467 130 130 / (+33) 467 130 132 |
| Contact Persons | Mr Frédéric BANEGAS<br>Phone number : (+33) 467 130 130<br>Fax number : (+33) 467 130 132<br>Ms Colette MAURIN<br>Phone number : (+33) 467 130 130<br>Fax number : (+33) 467 130 132 |
| Device name | |
| Common Name<br>Trade Name<br>Model number | System, Image Processing<br>Myrian®<br>N/A |
| Device classification | |
| Classification name<br>Code product<br>Panel<br>Regulation number<br>Regulatory class | System, Image Processing, Radiological<br>LLZ<br>892<br>892.2050<br>II |
| Predicate Devices | |
[K081985] Cleared [September 25, 2008] [Advantage Workstation Server], manufactured by [General Electric Healthcare]; [K061624] Cleared [June 27, 2006] [Vital Images, Inc.] [Vitrea2 Version 3.9], manufactured by [Vital Images, Inc.]; [K082228] Cleared [July 31, 2008] [Planisight Linasys] [LSPS], manufactured by [Pathfinder Therapeutics Inc.], [K091001] Cleared [June 29, 2009] [intrasense] [Myrian v1.4], manufactured by [intrasense] [K093621] Cleared [February 23, 2010] [syngo.PET&CT Oncology], manufactured by [Siemens Medical Solutions USA, Incl (K073194) Cleared (November 10, 2007) (CMRtools and plug-ins VentricularTools), manufactured by (Cardiovascular Imaging Solutions, Ltd]
# Explanation of how the device operates
Myrian with its modules is designed to run on standare, through the installed operating system. The hardware is all 'off-the-shelf standard computer components and may be purchased independently by the end user.
The exact same version of the Myrian® application Server Platform and executed in "Remote Execution Mode" (REM). In REM mode, multiple Users can remotely access all Myrian® features from any Client Platform (off the shelf computers, including laptops), through a LAN or WAN network. This access is granted through a Web browser (via a standard URL) or a specific Client Application installed on the Client Platform
# IFU (Indications For Use)
Myrian" is a software medical device aimed at reviewing images produced by all standard medical imaging devices. It also includes DICOM communication, media interchange (printing, storing) and reporting features. Myrian " can be run from any standard client platform (such as PC) that might be purchased independently by the end user.
Common Users are trained medical professionals, including surgeons, radiologists, clinicians.
This device is not indicaled for mammography use. Lossy compressed mammography images must not be used for primary mage interpretations. Mammographic images may only be interpreted using an least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
# Intended Use
Myrian® is a medical inage review and aided diagnosis software Medical Device as delined by the 93/42/EEC Directive. It provides Users with the following features:
Import and export of DICOM files from/to any DICOM-compliant modality, workstation or PACS.
Visualization of DICOM images in various standard visualization modes (e.g. MPR, 3D ... etc.) with optional image-alignment feature;
Creation of Objects Of Interest ("OOI") for analysis and measurement purposes;
Generation of medical reports;
Virtual Cutting surface tool for preoperative evaluation of surgery strategies;
Longitudinal Follow-up of patient, designed to support the oncological worklow by helping the user to confirm the absence of lesions, including evaluation, quantification, follow-up and documentation of such lesions.
Common Users are trained medical professionals, including surgeons, radiologists, clinicians.
Myrian® is designed to be run:
On standard Standalone Plaform, through the installed operating system. The hardware of such platform onsists in "off-the-shell standard PC computer components and may be purchased independently by the end user.
Remotely, through a nework connecting a Client Platform (standard desktop or laptop PC, Apple® Mac, etc.) to the Sever Plation on which Myrian® is installed. This mode of the Myrian® application is called "Remote Execution Mode".
{1}------------------------------------------------
# Device Description
ರ
0
Myrian includes toolsets which enable the reviewing physician to provide any selected relevant information for diagnosis, surgery and treatment planning to the referring physician.
These toolsets are categorized as follow:
Imaging tools:
- Multi-Planar Reformatting views (MPR) in orthogonal, oblique or curved planes as well as 3D Views in various rendening 0 modes including MIP, MiniP, Average and Volume Rendering
- Cross-sectional or Endoscopic Exploration Modes along a Centerline, e.g. of a vessel, a colon, etc.
- 0 "Filet Visualization Mode", to visualize any tubular, hollow organ, such as a colon, as a flat, unfolded surface
Manual or interactive Objects Of Interest (OOI):
- Annotations of Interest (AOI), for information or measurement purposes 0
- Paths (considered as Annotations of Interest) 0
- Regions of Interest (ROI), for anatomical and pathological structure isolation (such as liver, spleen, lungs, colon ... etc.) 0 through which any measurement can be performed
- Points of Interest (POI), for marking areas such as lesions, etc.
Reporting tools:
- Objects of Interest (OOI) generate Reports which may be viewed and exported to standard film and paper printers or sent 0 electronically to intranet web servers and any DICOM device.
Manual or assisted image alignment tools for multiphase or time-based image comparison. Virtual ROI Cutting Surface tool for preoperative evaluation of surgery strategies, e.g. for a liver
Longitudinal Follow-up tools for oncology workflow:
- A dedicated workflow and user interface, including a way to organize images in "Time Points", every Time Point being a O subset of Studies selected by the User:
- 0 A set of measurements collected by the User in each Time Point through the abovementioned Objects of Interest
- 0 A mechanism to link together these Objects of Interest through the Time Point, called "Lesion of Interest"
- Some Follow-up Response Evaluation Criteria (one per Longitudinal Follow-up, e.g. RECIST 1.1 or Cheson 2007) attached 0 to the Lesions of Interest:
- An Automatic or Semi-automatic Response Calculation using the attached Follow-up Response Evaluation Criteria, used to 0 quantify the analysis. This Calculation is based on the strict implementation of the related reference publications (e.g. http://www.eorc.be/recist/documents/RECIST Guidelines.pdf for RECIST 1.1), and can be controlled and overridden by the User:
- A validation mechanism to sign and lock every analyzed Time Point used in the Longitudinal Follow-up; 0
- o Specific Reports based on the chosen Follow-up Response Evaluation Criteria.
- Multiple Longitudinal Follow-ups per Patient. O
All Myrian® functionalities can be packaged, licensed and marketed as individual modules:
#### XP Lung:
Myrian XP Lung is designed to study the human respiratory system to help clinicians identify and analyse such as emphysema
The XP-Lung Module Engine makes use of a set of dedicated tools such as:
- Dedicated generic Protocols to predefine the workspace and available tools
- A semi-automatic Lung ROI segmentation engine
- The CutSimulator tool to simulate surgical action
Used in conjunction with other tools in the density histogram, it is possible to visualise, analyse and quantify healty, and pathological lung tissues rapidly with precision as well as to plan and simulate excision or lobectory
#### XPLungnodule:
Myrian® XP LungNodule enables volumetric analysis and follow-up of lung nodules making use of such functionalilies as:
- . Automatic segmentation and volumetric analysis of solitary solid lung nodules
- Automatic measurement and characterization as well as 3D visualisation of the nodules in their context -
- Matching and follow-up tools
- Density Histogram -
- Auto-adjusting display protocols, optimising the workspace and available tools -
- . Advanced Surfacic mode rendering
- Dedicated report template
#### XP Liver:
Myrian XP Liver is designed for use on CT datasets with contrast enhancement for portal veins and slices below 2 millimetres in thickness. The available dedicated tools and features include the measurements and analysis of liver stuctures as well as 2d/3d hepatectory simulation enabling to plan single phase and multiphase Liver surgery.
#### XP Vessel:
Myrian XP Vessel is designed for the visualisation and analysis of medical image data derived from CT, MRI scans of selected human vessels, including but not limited to the carcid artenes, the peripheral atteries, the aota, arteries of the brain, and any opacified veins.
{2}------------------------------------------------
The aim is to help automate rouline inspection of human vessels to detect stenosis, and dissection in the vessel. It also supports the interactive segmentation of any vessel and enable the isolation of highlighting or hiding of che data set from display for critical evaluation of selected part(s) of vessel. It is intended for use by radiologists, clinicians to acquire, process, render, measure, evaluate, archive, print and distribute DICOM compliant vessel image studies.
#### XL Onco:
Myrian® XL-Onco is designed around 3 methods and review of various types of lesions of interest. Each method comprises three stages (baseline, time point and has its specific constraints according to the coresponding guidelines (RECIST 1.0, RECIST1.1 and CHESON).
The use of the following enhanced features allows you to detect, assess and ensure volumeric longitudinal follow-up of the lesions:
- Automatic segmentation engine -
- -Editing tools
- Viewport layout enabling to compare one time point to the other .
- Follow-up dashboard including lesion graph and table.
The Myrian" System allows OEM customization of both the Graphical User Interface (GU) and the without implying any impact on system performance or system intended use.
This device is not indicated for mammography use. Lossy compressed mammography images and digitized film screen images must not be used for primary interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
An identical version of this software can be installed on an Application Server in "Remote Execution Mode" (REM). In REM mode, multiple Users can remotely access all Myrian® features from any Client Platform (off-the shell computers including laptops), through a LAN or WAN network. This access is granted through a Web browser (via a standard URL) or through a specific Client Application installed on the Client Platform.
Medical Images themselves are not transmitted. Interaction, a stream of screen content which is compressed using either a lossless or lossy compression scheme, is transmitted from the Application Server to the Client Platform.
Access to Myrian® REM and the stream of screen content are both encrypted to ensure security and confidentiality,
For diagnosic purposes, the User must that the compression scheme and the Client Platform display device used for reading the images both comply with state-of-the-art diagnosic requirements and all current legislation, rules and regulations in force.
Once accessed in REM mode, Myrian® is intended to be used exactly as the Myrian® Medical Device.
# Performance data
The performance and safety of Myran" is evaluated in pre-marketing phase during the through ventication and validation including the internal performance evaluations.
Before release of the software following testing are performed to checined specifications and fullifis its intended use:
- -Pre-integration testing assures that all modules aimed at being integrated into the system behave integration to the system
- -Functional Testing ensures that each element of the application meets of the business as outlined in the System Requirement Specification (SRS) and in the System Functional Sheet
- Integration Testing proves that all areas of the system interface with each other correctly and that there are no gaps in the data flow. -Final Integration Test proves that system works as integrated unit when all the fixes are complete.
- -Regression Testing ensures that there is no impact on previously released software. The regression testing is automated using ISVCR Automated Testing Tool.
- Build Testing is to be performed by the Build Engineer: no warning or compilation error should appear and binaries should be silently produces.
Tests are performed by refering to a Test Plan which contains all test procedures created to guaranty that all the functional and operational aspects of the software are under control. The tests runs were successfully completed in the Software.
Internal performance evaluations are performed to check that the software fulfills its intended aim particularly with complete safety for the patient.
Clinical features refer to functionality which:
- is closely related to patient risk
and/or - is intended to be used for diagnosis purpose
Typically it refers to segmentation engines.
According to the intended use of the feature the statistical analysis to apply is defined. Sample size and relevant inclusion criteria are specified in order to select the appropriate exams.
The performance and safety of Myrian" is also evaluated in post-marketing review. This review is performed for every considered paper evaluated as suitable the performances and/or safely of Myrian®. Every such document is analyzed to ensure that all identified risks are already covered and no new risk is identified. The literature review demonstrates Myrian's capability to be used in clinical environment as intended by the manufacturer.
Based on this evidence Myrian achieves its intended performance during normal conditions of use. Known and foreseeable risks and any adverse events are minimized and acceptable when weighed against the benefits of the intended performance.
{3}------------------------------------------------
# Technological characteristics
Myrian" is a multi modality medical diae for the review and analysis of anatomy and pathology in multi-dimages acquired from a variety of imaging devices. Myrian provides users with several visualization modes and includes DICOM communication, media interchange (printing, CD burning, stoning). This intended use is substantially equivalent to the one of the predicate devices Advantage Workstation Server® (K081985) and Vitrea2® (K061624).
i
Myrian orovides user with annotations, measurement, quantification and reporting tools. These features are substantially equivalent to hose of the predicate device Vitrea2 (K061624).Myian® contains an Endoscopic exploration mode to visualize any tubular hollow organ such as colon. This feature is substantially equivalent to the one of the predicate device Vitrea2® (K061624).
Myrian" includes tolsets which enable the review to provide any selected thismation for the destrement of the ment of the ment
Myrian provides users with longitudinal follow-up tools for oncology workflow. These features are substantially equivalent to those of the predicate device syngo.PET&CT Oncology (K093621).
The Iron Load Calculator available in Myrian® is substantial equivalent to the Thalassaematools included in the predicate device CMRtools (K073194), and which allows the calculation of a property called T2 that characterizes iron loading in the liver.
Myrian can be installed on an Application Server Platform and execution Mode". This mode of operation is substantialy equivalent to the one of the predicate device Advantage Workstation Server® (K081985).
## Conclusion
The 510(k) Pre-market Notification for Myrian adequate information and data to enable FDA-CDRH to determine substanial equivalence to the predicate devices listed above.
Myrian® has the same intication for use and rew type of safety and effectiveness questions. Performance data provided demonstrate that the device is safe, effective, and is substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
# Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Frederic Banegas CTO Intrasense 1231 Avenue du Mondail 98 MONPELLIER 34000 FRANCE
Re: K113620
Trade/Device Name: MYRIAN 1.11 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 2, 2012 Received: July 10, 2012
# Dear Mr. Banegas:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 use stated in the encrosure) to regars) to regard of the Medical Device Amendments, or to commerce pror to Pricy 20, 2017 11:12 in accordance with the provisions of the Federal Food, Drug, de neces that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The 1 ou may, merelere, maines of the Act include requirements for annual registration, listing of general controlo profitetvring practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such If your device is thassified (200 as regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I reast of advised that I Dr o recament at your device complies with other requirements of the Act that I Dri has made a actor regulations administered by other Federal agencies. You must of ally it catales and regarations and regulations, but not limited to: registration and listing (21 Comply with an the 110. or studies and 809); medical device reporting (reporting of
JUL 12 2012
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice mourements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket whitication. The FDA finding of substantial equivalence of your device to a legally marketed nouricate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you decire openits and office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket s 150. Fired the more 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou inay of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): _K113620
Device Name: MYRIAN 1.11
Indications for Use:
Myrian® is a software medical device aimed at reviewing images produced by all standard medical imaging devices. It also includes DICOM communication, media interchange (printing, CD burning, storing) and reporting features. Myrian® can be run from any standard client platform (such as PC) that might be purchased independently by the end user.
Common Users are trained medical professionals, including surgeons, radiologist clinicians and technicians.
ctinformic and coomisor of mammography use. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA.
> Over-The-Counter Use Prescription Use ਮ Prescription Use
> (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
WAH D OK
Division Sign Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K113620
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.