K160190 · Sunny Medical Device (Shenzhen) Co., Ltd. · DXT · Nov 10, 2016 · Cardiovascular
Device Facts
Record ID
K160190
Device Name
Sunmed Control Syringes
Applicant
Sunny Medical Device (Shenzhen) Co., Ltd.
Product Code
DXT · Cardiovascular
Decision Date
Nov 10, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
The Sunmed Control Syringes is intended to inject contrast media or saline for angiographic procedures.
Device Story
Sunmed Control Syringes are single-use, manual angiographic syringes used to inject contrast media or saline. The device operates on a dynamics principle using an interference fit between a plunger/piston and a barrel to generate aspirating or driving force. Available in 10ml and 12ml volumes, the syringes feature either a standard or locking plunger design and are equipped with either rotating or fixed male luer connectors. Used in clinical settings, the device is operated manually by a clinician to deliver fluids during angiographic procedures. The clear, transparent barrel with printed scales allows for accurate volume measurement and monitoring during injection. The locking plunger variant provides enhanced security and stability during use.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (acute systemic toxicity, skin sensitization, in vitro hemolysis, intracutaneous reactivity, in vitro cytotoxicity), package integrity, asepsis, aging, and EtO/ECH residue testing.
Technological Characteristics
Single-use manual syringe; 10ml and 12ml volumes; standard or locking plunger types; rotating or fixed male luer connectors compliant with ISO 594-2:1998; pressure rating 300 kPa; Ethylene Oxide sterilization.
Indications for Use
Indicated for patients undergoing angiographic procedures requiring the injection of contrast media or saline.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 10, 2016
Sunny Medical Device (Shenzhen) Co., Ltd. James Qi Zhang General Manager 56 Lehigh Aisle Irvine, California 92612
Re: K160190
Trade/Device Name: Sunmed Control Syringes Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector and Syringe Regulatory Class: Class II Product Code: DXT Dated: September 28, 2016 Received: October 3, 2016
Dear James Qi Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Brian D. Pullin -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160190
Device Name Sunmed Control Syringes
Indications for Use (Describe)
The Sunmed Control Syringes is intended to inject contrast media or saline for angiographic procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(k) Summary(K160190)
- 1. Submitted by: Sunny Medical Device (Shenzhen) Co., Ltd. Registered Address: 4/F and 5/F, Dongpingxing Creative Science Park, Longgang District, Shenzhen, Guangdong, P.R. China 518172 56 Lehigh Aisle, Irvine, CA 92612 Contact Address:
| Telephone: | (949)216-8838 |
|------------|--------------------------------|
| Fax: | (949)423-0168 |
| Contact: | JamesQi Zhang, General Manager |
| E-mail: | jamesqizhang@gmail.com |
| Date: | November 9, 2016 |
### 2. Proposed Device:
| Trade/Proprietary Name: | Sunmed™ Control Syringes |
|-------------------------|-----------------------------------|
| Common/Usual Name: | Control Syringes |
| Classification: | II |
| Classification Name: | Angiographic injector and syringe |
| Regulation Number: | 870.1650 |
| Product Code: | DXT |
- 3. Predicate Device:
| 510(k)<br>Number | Trade Name | Manufacture |
|------------------|------------------------------------------|----------------------------------------------|
| K093830 | Medline Angiographic Control<br>Syringes | Medline Industries Inc. |
| K072696 | ANT Angiographic Syringes | Shenzhen Ant Hi-Tech<br>Industrial Co., Ltd. |
Note: In the case of other characteristics substantially the same, Mediline Angiographic Control Syringes and our company's product are more similar in the specification model and materials. Medline Angiographic Control Syringes is identified as the primary predicate, which manufactured by Medline Industries Inc.
{4}------------------------------------------------
## 4. Device description
The Sunmed™ Control Syringes using dynamics principle, by means of the interference fit between the plunger with piston and the barrel, using the aspirating or driving force generated by manual function to draw or inject angiography medicament or saline for clinical disposable use.
The Sunmed™ Control Syringes is a single-use device that consists of two different structures: one is the standard type; the other is locking plunger type. The properties of the standard plunger type and locking plunger type are in table below:
| Standard plunger Type | Locking Plunger Type |
|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Humanized design, easy and simple operation, can be single hand operation | Humanized design, easy and simple operation, can be single hand operation |
| The wall of the tube is clear and transparent, the inner surface is smooth, and the promotion is convenient | The wall of the tube is clear and transparent, the inner surface is smooth, and the promotion is convenient |
| Printing scale is clear and accurate | Printing scale is clear and accurate |
| Pressure rating: 300 kPa | Locking design improve the security and stability.<br>Pressure rating: 300 kPa |
The volume of two type control syringes that to be offered in 10ml, 12ml.
The sunmed™ Control Syringes are available with rotating male luer and fixed male luer. The rotating male luer can be rotated. When connected with the other device, they can rotate freely. While the fixed male luer cannot be rotated. The rotating male luer and fixed male luer both are in line with ISO 594-2:1998.
The sterilization method of the Sunmed™ Control Syringes is Ethylene Oxide sterilization.
- 5. Intended Use
The Sunmed™ Control Syringes is intended to inject contrast media or saline for angiographic procedures.
{5}------------------------------------------------
- 6. Technological Comparison to Predicate Device
The technological characteristics of the subject device, the Sunmed™ Control Syringes, are equivalent in intended use, materials, operating principle, sterility assurance level, and method of sterilization to the Medline Angiographic Control Syringes; and the Sunmed™ Control Syringes is also similar to the ANT Angiographic Syringes in that it consists of the intended use, operating principle, sterility assurance level, and method of sterilization.
- Summary of Non-Clinical Testing 7.
The following tests were performed; the results of the performance testing demonstrated the substantial equivalence of the Sunmed™ Control Syringes.
Biocompatibility Testing: Acute Systemic Toxicity Test (two kinds of solvent) Skin sensitization Test (two kinds of solvent) In Vitro Hemolysis Study Intracutaneous Reactivity Test (two kinds of solvent) In Vitro Cytotoxicity Test
Package Penetrate Testing Asepsis Testing Aging Testing EtO and ECH Residue Testing Bench Testing
- 8. Clinical Evaluation was not applicable.
- 9. Conclusions
Based on the information presented in this 510(k) premarket notification, the Sunmed™ Control Syringes is considered substantially equivalent to the Medline Angiographic Control Syringes and ANT Angiographic Syringes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.