The Wrist 3 MRI System is intended for use as a magnetic resonance imaging device for producing axial, sagittal, coronal, and oblique images of the internal structure of the Wrist and Hand. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Device Story
1 Tesla permanent MRI system; produces axial, sagittal, coronal, and oblique images of wrist and hand. Device holds body part in uniform static magnetic field; applies sequences of RF pulses and gradient magnetic fields. Precession of magnetization sampled and processed into image data. Used in hospital or clinical settings; operated by trained personnel. Physician interprets images to assist in diagnosis. Benefits include non-invasive visualization of internal wrist/hand structures.
Clinical Evidence
No clinical studies were required. Substantial equivalence supported by bench testing, system/software verification and validation, and sample clinical images.
Technological Characteristics
1 Tesla permanent magnet MRI; gradient strength 215 mT/m; slew rate 1074 T/m/Sec. Includes two TX/RX linear volume RF coils (small/large). 15" touch display. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, and NEMA MS-1, MS-2, MS-3, MS-5, MS-12 standards.
Indications for Use
Indicated for patients requiring MR images of the wrist and hand for diagnostic purposes.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 1, 2016
Aspect Imaging Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW BUFFALO MN 55313
Re: K160185
Trade/Device Name: Wrist 3 MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH, MOS Dated: February 22, 2016 Received: February 23, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160185
Device Name Wrist 3 MRI System
Indications for Use (Describe)
The Wrist 3 MRI System is intended for use as a magnetic resonance imaging device for producing axial, sagittal, coronal, and oblique images of the internal structure of the Wrist and Hand. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Aspect Imaging. The logo consists of a stylized cube made up of three layers in shades of blue, green, and yellow. To the right of the cube is the company name, "aspectimaging," written in a sans-serif font, with the "t" in "aspect" stylized to resemble a plus sign.
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | February 16, 2016 | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Aspect Imaging Ltd. | |
| | 27 Shaked St. | |
| | Industrial Area Hevel Modi'in | |
| | P.O. Box 926 | |
| | Shoham, 6085001 | |
| | ISRAEL | |
| Primary Contact Person: | Israel A. Citron | |
| | VP Quality Assurance and Regulatory Affairs | |
| | Aspect Imaging LTD | |
| | Tel: +972-73-223-9000 Ext 90 | |
| | Fax: +972-3-716-5606 | |
| | E mail: icitron@aspectimaging.com | |
| Device Trade Name: | Wrist 3 MRI System | |
| Common/Usual Name: | MRI System | |
| Product Classification | II | |
| Classification Name: | Magnetic Resonance Diagnostic Device (21 CFR 892.1000) | |
| Product Code: | LNH, MOS | |
| Predicate Device(s): | K130692 M2 Wrist 2 MRI System | |
| Device Description: | The Wrist 3 MRI System is a 1Tesla Permanent MRI system<br>producing MR images of the wrists and hands. During MRI scan,<br>body parts to be imaged are held within a uniform static magnetic<br>field, and are subject to sequences of RF pulses and gradient<br>magnetic fields. The signal from the precession of the<br>magnetization created by these fields is sampled and processed to<br>produce image data. | |
| Intended Use: | The Wrist 3 MRI System is intended for use as a magnetic<br>resonance imaging device for producing axial, sagittal, coronal,<br>and oblique images of the internal structure of the Wrist and Hand.<br>When interpreted by a trained physician, these images provide<br>information that can be useful in determining a diagnosis | |
| Technology: | The Wrist 3 MRI System employs the same fundamental<br>scientific technology as its predicate device (M2 Wrist 2 MRI<br>System K130692). | |
| Comparison of<br>Specifications | Predicate Device<br>M2 Wrist 2 MRI System<br>(K130692) | Proposed Device<br>Wrist 3 MRI System |
| Intended Use/Indication for Use | The M2 Wrist II MRI System<br>is indicated for use as a<br>magnetic resonance imaging<br>device for producing transverse,<br>sagittal and coronal images of<br>the internal structure of the wrist<br>(in patients with an arm length ><br>320 mm). When interpreted by a<br>trained physician, the resultant<br>MR images provide information<br>that can be useful in determining<br>a diagnosis. | The Wrist 3 MRI System is<br>intended for use as a magnetic<br>resonance imaging device for<br>producing axial, sagittal,<br>coronal, and oblique images of<br>the internal structure of the<br>Wrist and Hand. When<br>interpreted by a trained<br>physician, these images provide<br>information that can be useful in<br>determining a diagnosis. |
| Patient Population | Patients requiring MR images of the<br>wrist | Patients requiring MR images of the<br>wrist and Hand |
| Anatomical Sites | Wrist | Wrist and Hand |
| Environment of Use | Hospital or clinical setting | Hospital or clinical setting |
| Energy Used and/or delivered | Magnetic Resonance | Magnetic Resonance |
| Human Factors | The M2 Wrist 2 MRI System is<br>designed similar to other<br>commercially available MRI<br>Systems and therefore is familiar<br>and easy for use for the user.<br>Furthermore, the device contains a<br>user-friendly software interface<br>through which the user may easily<br>access all device functions. | The Wrist 3 MRI System is<br>designed similar to other<br>commercially available MRI<br>Systems and therefore is familiar<br>and easy for use for the user.<br>Furthermore, the device contains a<br>user-friendly software interface<br>through which the user may easily<br>access all device functions. |
| Hardware Changes | | |
| Magnet: | | |
| - Physical Dimensions | 125x87x82 cm | 138x87x82 cm |
| - Bore opening | 76x200 mm | 85x220 mm |
| - Weight<br>Field Strength<br>Gradient: | 1050 Kg<br>1 Tesla Permanent Magnet | 1300 (1700 with Trolley) Kg<br>1 Tesla Permanent Magnet |
| -Strength | 190 mT/m | 215 mT/m |
| -Rise Time | 0.475mSec | 0.200mSec |
| -Slew Rate | 400 T/m/Sec | 1074 T/m/Sec |
| Computer: | | |
| -Display: | 24" LCD Desktop Display | 15" Touch Display |
| RF Coils | 1 Coil | 2 Coils: Small/Large |
| Coil Type | TX/RX | TX/RX |
| Coil Geometry | Oval | Oval |
| Inner dimensions (mm) | 50X108X86 | 60X100X169 / 75X135X192 |
| Coil Design | Linear Volume | Linear Volume |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Aspect Imaging. The logo consists of a teal and green geometric shape on the left, followed by the word "aspectimaging" in a dark purple sans-serif font. The geometric shape appears to be a stylized representation of a cube or layered structure.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Aspect Imaging. The logo consists of a geometric shape on the left and the text "aspectimaging" on the right. The geometric shape is made up of four stacked squares, each a different color: blue, green, yellow, and purple. The text "aspectimaging" is written in a sans-serif font, with "aspect" in purple and "imaging" in gray.
### Determination of Substantial Equivalence:
Summary of Nonclinical Testing:
The Wrist 3 MRI System and its applications were tested to comply with the below voluntary consensus standards.
- IEC/ES 60601-1
- IEC 60601-1-2
- 트 IEC 60601-2-33
- 트 NEMA MS-1
- NEMA MS-3
- 트 NEMA MS-2
- NEMA MS-12
- NEMA MS-5
In addition System and Software verification and Validation testing were performed to demonstrate performance of the device as part of design controls activity.
The following design control activities were applied to the development of the system:
- Risk Management
- Requirements Management
- Design Reviews
- 트 Unit level module verification
- 트 System Integration verification
- I Performance verification
- Safety and EMC testing
- I Simulated use validation testing
- Sample clinical images
### Summary of Clinical Tests:
The subject of this premarket submission, Wrist 3 MRI System, did not require clinical studies to support substantial equivalence.
- Conclusion: Based on bench testing and sample clinical images. Aspect Imaging LTD believes that the Wrist 3 MRI System is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.