The M2 Wrist 2 MRI System is indicated for use as a magnetic resonance imaging device for producing transverse, sagittal and coronal images of the internal structure of the wrist (in patients with an arm length > 320mm). When interpreted by a trained physician, the resultant MR images provide information that can be useful in determining a diagnosis.
Device Story
M2 Wrist 2 MRI System is a 1 Tesla compact permanent magnet MRI device for wrist imaging. System inputs include RF signals from wrist anatomy; transforms signals via pulse sequences (2D/3D Spin-Echo, Spoiled Gradient Echo, Fast Spin Echo) into transverse, sagittal, and coronal images. Operated by trained personnel in fixed installations; no RF-shielded room required due to self-shielded magnet. Output is MR images displayed on a workstation for physician interpretation to assist in clinical diagnosis. Benefits include high-performance imaging in a compact, low-noise, low-field footprint.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical/mechanical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), software validation, and NEMA performance standards (MS-1, MS-3, MS-4, MS-5, MS-8, MS-10, MS-11, MS-12) for image quality, SNR, geometric distortion, acoustic noise, slice thickness, and SAR.
Indicated for patients with arm length > 320mm requiring magnetic resonance imaging of the wrist for diagnostic purposes. To be interpreted by a trained physician.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K130692
Page 1 of 6
# 510(k) Summary of Safety & Effectiveness
JAN 1 4 2014
Pursuant to CFR 807.92, the following 510(k) Summary is provided:
- George J. Hattub Submitter 1. (a) Address: MedicSense, USA . 291 Hillside Avenue Somerset, MA 02726 www.medicsense.com 1. (b) Manufacturer Aspect Imaging, Ltd. Address: 27 Shaked Street Industrial Area Hevel Modi'in Shoham, Israel 60850 Mfg. Phone: Tel.: +972 73 2239000 Contact Uri Hoffer Person: December 1, 2013 Date: Magnetic Resonance Imaging System have been classified as Class 2 LNH, 2. Device & Classification: Regulation Number 21 CFR 892.1000 Magnetic resonance diagnostic device M2 Wrist 2 MRI System Name: Predicate M2 Wrist MRI System K120701 3. Device: The M2 Wrist 2 MRI System is a 1 Tesla, compact, high-performance solution, 4. Description: based on a permanent magnet. Due to its remarkable structure, the external magnetic field is very low, thus offering unique safety advantage. The system has very low Eddy currents and exhibits very low gradient-related acoustic noise. The magnet is self-shielded and thus no RF shielded room is required. The type of installation is fixed. The M2 Wrist 2 MRI System's main components are: . Magnet Sub-system Wrist Coil . Electronics Cabinet . Aspect Imaging Proprietary Software . Computer . . Isolation Transformer
Specific Technical Description: see below in Section 6
{1}------------------------------------------------
- The M2 Wrist 2 MRI System is indicated for use as a magnetic resonance ട. Intended Use: imaging device for producing transverse, sagittal and coronal images of the internal structure of the wrist (in patients with an arm length > 320mm). When interpreted by a trained physician, the resultant MR images provide information that can be useful in determining a diagnosis.
#### 6. Technology: Technology and Comparison of Characteristics
With respect to technology and intended use, the M2 Wrist 2 MRI System is substantially equivalent to its predicate device which is the M2 Wrist MRI System. The primary differences are the addition of plastic panels, modified wrist coil, a change in the spectrometer, additional pulse sequences and added MRI viewing features. Based upon the validation results, Aspect Imaging believes these changes do not raise additional safety of efficacy concerns.
| Characteristic /<br>Feature | Predicate Device<br>(System) | Modified Device<br>(System) |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Magnet Subsystem | | |
| Field Strength | $1.05 (+/- 2%)$ Tesla<br>vertical field | $1.05 (+/- 2%)$ Tesla<br>horizontal field |
| Bore Opening<br>Size (H x W) | 76 x 200 mm | same |
| Field of View | 115 x 80 x 50 mm | 110 x 80 x 50 mm |
| Gradient System<br>Type | Special purpose<br>gradient system | same |
| Gradient Strength | 190 mT/m | same |
| Slew Rate | 400 T/m/sec | same |
| Rise Time | 475 $\mu$ sec | same |
| Gradient Amplifier<br>Gain Scale | $\pm$ 10 A output per $\pm$ 1 V input | same |
| Acoustic Noise | <75db | same |
| Cooling | Built-in cooling fans<br>for magnet and<br>gradient subsystem | same |
| Wrist RF Coil | | |
| Central Frequency | 45 MHz | same |
| Solenoid (9 turns) | yes | same |
| B 1 Direction | yes | same |
| Coil Housing<br>Material | Ultralloy 910-5 GLU | Ultralloy 304 |
| Balance Matching<br>Circuit | yes | yes |
| Tuning Capacitor | yes | yes |
| Maximum RF<br>peak handling | 500 W | same |
| Topology | Transmit/Receive<br>Coil | same |
| | | |
| Electrical | Solenoid<br>Central Frequency:<br>45 MHz typical<br><br>No decoupling<br>circuits | same |
| Dimensions of<br>Magnet sub-<br>system | Height = 114 cm<br>Width = 79 cm<br>Length = 79 cm<br>Weight = Approx. 930 kg | Height = 125 cm<br>Width = 87 cm<br>Length = 82 cm<br>Weight = Approx. 1050 kG |
| Wrist Coil<br>Dimensions &<br>Positions | The maximum hand<br>size that can be<br>scanned is:<br>Width: 108 mm<br>Length: 210 mm<br>Height: 50 mm | The maximum hand size<br>that can be scanned is:<br>Width: 108 mm<br>Length: 220 mm<br>Height: 50 mm |
| Electronics Cabinet | | |
| RF Power<br>Amplifier | Tomco | same |
| Control Unit | Aspect | same |
| Gradient Amplifier | Copley | same |
| Power Distribution<br>Unit | Aspect | same |
| Spectrometer | Tecmag-Apollo | Tecmag-Redstone |
| Dimensions of<br>Electronics<br>Cabinet | Height: 130 cm<br>Width: 60 cm<br>Length: 80 cm<br>Weight: Approx. 205 kg | same |
| Isolation Transformer | | |
| Input | 200-240 VAC 50/60<br>Hz | same |
| Output | 200-240 VAC 9A,<br>2.16 kVA | same |
| Operator<br>Emergency Stop<br>Switch | yes | same |
| Main Circuit<br>Breaker LED for<br>Power Indication | yes | same |
| Proprietary Software | | |
| Software | Windows 7<br>Professional & Wrist<br>MRI Software | same |
| Conventional MRI | 2D Spin-Echo | 2D Spin-Echo<br>2D Spoiled |
| | | 3D Spoiled<br>Gradient Echo |
| | | 2D Fast Spin Echo |
| | | 3D Fast Spin Echo |
| | | Fat Suppression |
| Software Features | Perform MRI Scan<br>Sequence parameter<br>variability<br>Multi-Slice 2D<br>Viewer<br>DICOM Export | same |
| Wrist MRI<br>Software | Version 1.1 | Version R.1.2.0.6 |
| TNMR Software | Version 2.11.2 | Version 2.11.22 |
| Processor | Computer<br>Intel Xeon W3520<br>(2.66GHz,4.8GT/s,8<br>MB) Memory runs at<br>1066MHz | same |
| Resource DVD | Precision T3500<br>Diagnostics and Drivers | same |
| Memory | 4GB<br>(4x1GB)1066MHz<br>DDR3 ECC-UDIMM | same |
| Hard Drive | 2x320GB (7200RPM)<br>Serial ATA II with<br>NCQ and 16MB<br>DataBurst Cache | same |
| Raid Controller | C1 All SATA Hard<br>Drives, NON-RAID for 2<br>Hard Drive | same |
| Optical Drive | 16X DVD+/-RW Drive;<br>Power DVD 8.1<br>Software and Media<br>included | same |
| Graphics | 768 MB Quadro<br>NVIDIA FX1800 - 2<br>DP, 1 DVI (1 DP-<br>DVI, 1 DVI-VGA<br>adapter) | same |
| Operating System | English Windows 7<br>Professional with SP1 | same |
{2}------------------------------------------------
.
. .
{3}------------------------------------------------
{4}------------------------------------------------
Though there are some minor differences in the characteristics of the two systems, these differences do not raise new questions of safety or efficacy. Furthermore, the M2 Wrist 2 MRI System has passed all the required tests and standards for MRI devices, as did the predicate M2 Wrist MRI System.
- 7. Performance Data Non-
Clinical:
Performance Standards:
The following performance tests were performed on the M2 Wrist 2 MRI System or its components:
- Electrical & Mechanical Safety (IEC 60601 -1) .
- Electromagnetic Compatibility (IEC 60601-1-2) .
- Software Validation .
- MR Image Quality Testing .
- NEMA MS- 1-2008 Determination of Signal-to-Noise Ratio (SNR) in . Diagnostic Magnetic Resonance Imaging
- NEMA MS 3-2008 Determination of Two-Dimensional Geometric . Distortion in Diagnostic Magnetic Resonance Images (Image Uniformity Test)
- . NEMA MIS 4 (2006) Acoustic Noise Measurement Procedure for Diagnosing Magnetic Resonance Imaging Devices
- NEMA MS 5-2010 Determination of Slice Thickness in Diagnostic . Magnetic Resonance Imaging
- NEMA MS 8 (2006) Characterization of the Specific Absorption Rate . (SAR) for MRI Systems
- NEMA MS 10-2006 Determination of Local Specific Absorption Rate . (SAR) in Diagnostic Magnetic Resonance Imaging
- NEMA MS 11 -2006 Determination of Gradient-Induced Electric Fields in . Diagnostic Magnetic Resonance Imaging
- NEMA MS 12-2006 Quantification and Mapping of Geometric Distortion . for Special Applications
- IEC 60601-2-33 Medical electrical equipment Part 2-33: Particular . requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnostic (2007 (Second Edition) + Al 1:2005 ± A2:2007)
- High Contrast Spatial Resolution Testing .
In all instances, the M2 Wrist 2 MRI System functioned as intended and/or met the requirements of the standard
- 7. Clinical Performance Not applicable Data: 8. Conclusions: Conclusions Drawn from Non-Clinical and Clinical Tests The performance tests demonstrate that M2 Wrist 2 MRI System may be safely
and effectively used in acquiring wrist MR images. The software validation and performance tests demonstrate that the M2 Wrist 2 MRI System meets its design and performance specifications and is substantially equivalent to the cleared M2 Wrist MRI System.
{5}------------------------------------------------
K130692
Page 6 of 6
### Substantial Equivalence:
8.
In summary, the indications for use of the M2 Wrist 2 MRI System are the same and thus substantially equivalent to the M2 Wrist MRI System. Furthermore, the basic technological characteristics of the M2 Wrist 2 MR1 System are similar to the predicate M2 Wrist MRI System. The differences in the technological characteristics do not raise new questions of safety and effectiveness. Consequently, the M2 Wrist 2 MRI System is substantially equivalent it's predicate device.
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, with three wavy lines representing the branches of government.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring. MD 20993-0002
January 14, 2014
Aspect Imaging Ltd. % Mr. George Hattub Senior Staff Consultant MedicSence, USA 291 Hillside Avenue SOMERSET MA 02726
Re: KI30692
Trade/Device Name: M2 Wrist 2 MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: December 1, 2013 Received: December 13, 2013
Dear Mr. Hattub:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The . general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2-Mr. Hattub
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safetw/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Andra D. O'Hara
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known): K130692
Device Name: M2 Wrist 2 MRI System
Indications For Use: The M2 Wrist 2 MRI System is indicated for use as a magnetic resonance imaging device for producing transverse, sagittal and coronal images of the internal structure of the wrist (in patients with an arm length > 320mm). When interpreted by a trained physician, the resultant MR images provide information that can be useful in determining a diagnosis.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .
Concurrence of CDRH, Office of Device Evaluation (ODE)
Michael D. O'Hara
(Division Sign-Off) Division of Radiological Health/OIR K130692 510(k)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.