K160121 · Corindus, Inc. · DXX · Oct 27, 2016 · Cardiovascular
Device Facts
Record ID
K160121
Device Name
CorPath GRX System
Applicant
Corindus, Inc.
Product Code
DXX · Cardiovascular
Decision Date
Oct 27, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1290
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CorPath GRX is intended for use in the remote delivery and manipulation of guidewires and rapid exchange balloon/ stent catheters, and remote manipulation of guide catheters during percutaneous coronary intervention (PCI) procedures.
Device Story
Robotic system for remote PCI; enables physician to manipulate guidewires, balloon/stent catheters, and guide catheters from a shielded cockpit. System consists of a bedside unit (articulated arm, robotic drive, single-use cassette) and a remote workspace (control console, monitors, foot pedal). Physician uses joysticks and touchscreens to control robotic drive; system provides linear and rotational movement of interventional devices. Operates under independent angiographic fluoroscopy. Used in cath labs; allows physician to remain seated away from radiation source, reducing occupational radiation exposure. Output is physical movement of catheters/wires; assists in precise device placement during coronary interventions.
Clinical Evidence
Bench testing included performance, functional, guide catheter particulate analysis, simulated procedure testing, biocompatibility, software V&V, and EMC testing. Pre-clinical in-vivo porcine study conducted with 8 pigs treated via CorPath GRX and 4 pigs treated manually (control) for PCI (balloon angioplasty and stent deployment) in the mid-LAD artery to verify safety.
Technological Characteristics
System comprises a bedside unit (articulated arm, robotic drive, single-use cassette) and a remote workspace (control console, monitors). Cassette materials include custom plastic parts, machined stainless steel, miniature magnets, roller bearings, and springs. Connectivity via communication cable between robotic drive and control console. Operates under fluoroscopic guidance.
Indications for Use
Indicated for patients undergoing percutaneous coronary intervention (PCI) procedures requiring the remote delivery and manipulation of guidewires, rapid exchange balloon/stent catheters, and guide catheters.
Regulatory Classification
Identification
A steerable catheter control system is a device that is connected to the proximal end of a steerable guide wire that controls the motion of the steerable catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2016
Corindus, Inc. % Mona Advani Senior Consultant Cardiomed Device Consultants, LLC 5523 Research Park Drive, Suite 205 Baltimore, Maryland 21228
Re: K160121
Trade/Device Name: Corpath GRX System Regulation Number: 21 CFR 870.1290 Regulation Name: Steerable Catheter Control System Regulatory Class: Class II Product Code: DXX Dated: October 21, 2016 Received: October 24, 2016
Dear Mona Advani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Brian D. Pullin -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160121
Device Name CorPath GRX System
Indications for Use (Describe)
The CorPath GRX is intended for use in the remote delivery and manipulation of guidewires and rapid exchange balloon/ stent catheters, and remote manipulation of guide catheters during percutaneous coronary intervention (PCI) procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 807.92 (c).
# 1. SUBMITTER/510(K) HOLDER
| Company Name and Address: | Corindus Inc.<br>309 Waverley Oaks Road<br>Suite 105<br>Waltham, MA 02452 |
|---------------------------|------------------------------------------------------------------------------------------------------|
| Company Contact: | Robert Lavado<br>Email:Rob.lavado@corindus.com<br>Phone: (508) 635-3335 x 211<br>Fax: (508) 653-3355 |
# Establishment Registration Number: 3007822508
| Contact Person: | Mona Advani |
|-----------------|--------------------------|
| Email: | madvani@cardiomedllc.com |
| Phone: | (650) 575-5819 |
| Fax: | (617) 663-6052 |
- October 21, 2016 Date Prepared:
# 2. DEVICE NAME
| Proprietary Name: | CorPath GRX |
|----------------------|-----------------------------------------|
| Common/Usual Name: | CorPath GRX System |
| Product Code: | DXX (Steerable catheter control system) |
| Classification Name: | Steerable Catheter Control System |
### 3. PREDICATE DEVICE
- CorPath 200 System manufactured by Corindus, Inc., K120834 and K150892. .
# 4. DEVICE DESCRIPTION
The CorPath GRX System is intended to allow physicians to deliver and manipulate commercially available coronary guidewires, rapid exchange balloon/stent catheters, and guide catheters during percutaneous coronary intervention (PCI) procedures. During the use
{4}------------------------------------------------
of the CorPath GRX System, the physician maneuvers interventional devices using intuitive controls under independent angiographic fluoroscopy visual guidance using computer controlled movements while in a seated position away from the radiation source.
# 5. INDICATION FOR USE
The CorPath GRX is intended for use in the remote delivery and manipulation of guidewires and rapid exchange balloon/stent catheters, and remote manipulation of guide catheters during percutaneous coronary intervention (PCI) procedures.
# 6. TECHNOLOGICAL CHARACTERISTICS
The CorPath GRX System is intended to allow physicians to deliver and manipulate commercially available coronary guidewires, balloon/stent catheters, and guide catheters during percutaneous coronary intervention (PCI) procedures. During the use of the CorPath GRX System, the physician maneuvers interventional devices using intuitive controls under independent angiographic fluoroscopy.
The CorPath GRX System is composed of the following two functional sub-units:
- 1. Bedside Unit: Which consists of the Articulated Arm, Robotic Drive and Single-use Cassette, and the
- 2. Remote Workspace: Which consists of the Interventional Cockpit (radiation shielded) which houses the Control Console, angiographic monitor(s), hemodynamic monitors and X-ray foot pedal.
Commercially available guidewires, rapid exchange balloon/stent catheters, and guide catheters are loaded into the Single Use Cassette. By using the joysticks or the Control Console touch screens the physician can control the Robotic Drive to advance, retract, and rotate the guide wire, advance and retract the balloon/stent catheter, and advance, retract and rotate the guide catheter. The Robotic Drive and Control Console communicate via a single communication cable.
# 7. Substantial Equivalence
The product subject of this premarket notification is substantially equivalent in design and functionality to the CorPath 200 System (K120834, K150892).
The table below compares the technical characteristics of the CorPath GRX System to the predicate CorPath 200 System.
{5}------------------------------------------------
| Characteristics | CorPath GRX System<br>(Subject Device, previously described<br>as the CorPath 300 System) | CorPath 200 System<br>K150892 and K120834<br>(Predicate Device) |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Corindus, Inc. | Corindus, Inc. |
| Indication for use | The CorPath GRX System is intended<br>for use in the remote delivery and<br>manipulation of coronary guidewires<br>and rapid exchange balloon/stent<br>catheters, and remote manipulation of<br>guide catheters during percutaneous<br>coronary intervention (PCI) procedures. | The CorPath 200 System is<br>intended for use in the remote<br>delivery and manipulation of<br>coronary guidewires and rapid<br>exchange balloon/stent catheters<br>during percutaneous coronary<br>intervention (PCI) procedures. |
| Intended use | Percutaneous coronary intervention<br>(PCI) procedures | Percutaneous coronary intervention<br>(PCI) procedures |
| Device Design | The CorPath GRX System is composed<br>of:<br>Control Console<br>Remote Workspace<br>Articulated Arm<br>Robotic Drive<br>Cassette (single-use) | The CorPath 200 System is<br>composed of:<br>Control Console<br>Remote Workspace<br>Articulated Arm<br>Robotic Drive<br>Cassette (single-use) |
| Operational principles | The physician, seated at the Remote<br>Workspace manipulates coronary<br>guidewires, balloon/stent catheters and<br>guide catheters using joysticks or touch-<br>screen controls on the Control Console | The physician, seated at the<br>Remote Workspace, manipulates<br>coronary guidewires and/or<br>balloon/stent catheters using<br>joysticks or touch-screen controls<br>on the Control Console. |
| Visualization during<br>Procedure | Fluoroscopy | Fluoroscopy |
| Compatible with<br>Commercially available<br>PCI Devices | Yes, compatible with commercially<br>available:<br>Guide cathetersGuidewiresRapid Exchange balloon/stent<br>catheters | Yes, compatible with commercially<br>available:<br>GuidewiresRapid Exchange balloon/stent<br>catheters |
| Linear movement of PCI<br>devices | Yes, allows for linear movement of:<br>GuidewiresRapid exchange balloon/stent<br>cathetersGuide Catheters | Yes, allows for linear movement<br>of:<br>GuidewiresRapid Exchange balloon/stent<br>catheters |
| Rotational movement of<br>PCI devices | Yes, allows for rotational movement of:<br>GuidewiresGuide Catheters | Yes, allows for rotational<br>movement of:<br>Guidewires |
| Materials (single-use<br>Cassette) | The single-use sterile Cassette is<br>assembled from a variety of custom<br>plastic parts and machined stainless<br>steel components, as well as off-the-<br>shelf miniature magnets, roller bearings<br>and springs. | The single-use sterile Cassette is<br>assembled from a variety of custom<br>plastic parts and machined stainless<br>steel components, as well as off-<br>the-shelf miniature magnets, roller<br>bearings and springs. |
{6}------------------------------------------------
# Bench Testing
Bench testing and in-vivo animal studies were performed to determine substantial equivalence. Specifically, the following bench tests were performed.
- Performance testing of the CorPath GRX System .
- Functional testing of the CorPath GRX System .
- Guide Catheter Particulate Analysis
- Simulated Procedure Testing ●
- Biocompatibility Testing of the Single-Use Cassette ●
- Software verification and validation testing ●
- EMC Testing of the CorPath GRX System
# Pre-Clinical Study
To further verify the safety of the CorPath GRX System, an in-vivo porcine study was conducted. Eight (8) pigs underwent PCI (balloon angioplasty and stent deployment) of the mid left anterior descending artery (LAD) performed using the CorPath GRX System and the results were compared to a four (4) pig control group which was treated manually.
# 8. Conclusion
Based on the testing conducted, it is concluded that the CorPath GRX System is substantially equivalent to the predicate CorPath 200 system (K150892 cleared October 2, 2015 and K120834 cleared July 19, 2012).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.