The CorPath 200 System is intended for use in the remote delivery and manipulation of coronary guidewires and balloon/stent catheters during percutaneous coronary intervention (PCI) procedures.
Device Story
CorPath 200 System enables remote robotic-assisted percutaneous coronary intervention (PCI). System comprises bedside unit with robotic drive/articulated arm and remote workspace with interventional cockpit/control console. Physician operates joysticks/touch-screen at remote console to manipulate commercially available guidewires and balloon/stent catheters loaded into single-use cassette. System provides computer-controlled movement of devices through coronary vasculature under fluoroscopic guidance. Remote operation allows physician to remain seated away from radiation source, potentially reducing operator radiation exposure. Software monitors communication between console and robotic drive, providing alerts for errors. System facilitates precise device advancement, retraction, and rotation during PCI.
Clinical Evidence
PRECISE clinical study: prospective, single-arm, multi-center, nonrandomized study (n=164). Primary endpoints: clinical procedural success (97.6%), device technical success (98.8%), and absence of Major Adverse Cardiac Events (MACE) (97.6%). 100% of subjects achieved post-procedure stenosis <30%. Results support safety and effectiveness for robotic-assisted PCI.
Technological Characteristics
System consists of bedside unit (robotic drive, articulated arm, single-use cassette) and remote workspace (control console, radiation shield, angiographic monitors). Electromechanical interface for device manipulation. Connectivity via communication cable between console and drive. Software performs continuous monitoring of communication link. Biocompatibility, sterilization, and shelf-life testing performed.
Indications for Use
Indicated for patients undergoing percutaneous coronary intervention (PCI) procedures requiring the delivery and manipulation of coronary guidewires and balloon/stent catheters.
Regulatory Classification
Identification
A steerable catheter control system is a device that is connected to the proximal end of a steerable guide wire that controls the motion of the steerable catheter.
{0}------------------------------------------------
K12-0834
P 1 of 5
# 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807, Section 807.92(c).
Applicant: Corindus, Inc. 11 Erie Drive JUL 1 9 2012 Natick, MA 01760, USA Michail M. Pankratov, MD, PhD Vice President, Clinical and Regulatory Affairs Email: Michail.Pankratov(@Corindus.com Telephone: (508) 653-3335 x 206 Fax: (508) 589-4050 Cell: (617) 480-4543 Contact Person: Mona Advani Regulatory Affairs Consulting Agent for Corindus, Inc. CardioMed Device Consultants, LLC 5523 Research Park Drive, Suite 360
Baltimore, MD 21228 Email: madvani@cardiomedIlc.com (b): (650) 575-5819 (F): (617) 663-6052
Date Prepared: March 16, 2012
Device_Information:
Corindus CorPath® 200 System Trade Name: Common Name: CorPath System Product Code: DXX (Steerable catheter control system) Regulation Number: 21 CFR 870.1290 (Steerable catheter control system)
#### Predicate Device(s):
- Niobe® MNS w/Navigant ™ Navigation Workstation (K060967) l.
- 2. Sensei® Robotic Catheter System (K091808)
{1}------------------------------------------------
K12 08 sf
P. 2 of 5
| Device Name | CorPath 200 System | Sensei Robotic Catheter System | Niobe MNS w/Navigant Navigation Workstation (NWS05) |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison Criteria | | | |
| Manufacturer | Corindus, Inc. | Hansen Medical, Inc. | Stereotaxis, Inc. |
| 510(k) Number | TBD | K091808 | K060967 |
| Product Code | DXX | DXX | NDQ |
| Regulation Number | 21 CFR 870.1290 | 21 CFR 870.1290 | 21 CFR 870.1290 |
| Device Class | II | II | II |
| Intended Use | Remote delivery and manipulation of coronary guidewires and balloon/stent catheters. | Remote manipulation and positioning of robotically steerable catheters. | Remote navigation of magnetically-adapted coronary guidewires and catheters. |
| Indications for Use | Percutaneous Coronary Intervention (PCI) | Cardiac Mapping | Navigation through coronary vasculature<br>Cardiac Mapping |
| Device Design | The CorPath 200 System is composed of: a Remote Workspace with an operator interface and a Bedside Unit with a robotic drive. | The Sensei Robotic Catheter System is composed of a Remote Workstation with an operator interface, a robotic arm and a control catheter. | The Stereotaxis Niobe MNS with Navigant System is composed of a Remote Workstation with an operator interface and a magnetic navigation system. |
| Operational Principles | The physician, seated at the Remote Workspace, manipulates coronary guidewires and/or balloon/stent catheters using joysticks or touch-screen controls on the Control Console. | The physician, seated at the Remote Workstation, manipulates catheters using joysticks or track ball controls. | The physician, seated at the Remote Workstation, manipulates magnetic coronary devices through use of joysticks and/or touch screen technology. |
| Device Name | CorPath 200 System | Sensei Robotic Catheter System | Niobe MNS w/Navigant Navigation Workstation (NWS05) |
| Comparison Criteria | | | |
| Use of Fluoroscopy | Yes | Yes | Yes |
# Summary of Comparison of the CorPath 200 System to Predicate Devices
8
.
{2}------------------------------------------------
## Device Description
The CorPath 200 System is intended for use by physicians in the delivery and manipulation of coronary guidewires and balloon/stent catheters during percutaneous coronary intervention ("PCI") procedures. The CorPath 200 System allows the physician to deliver and manipulate guidewires and balloon/stent catheters through the coronary vasculature under angiography-assisted visual guidance using computer controlled movements while in a seated position and away from the radiation source.
## Indication for Use
The CorPath 200 System is intended for use in the remote delivery and manipulation of coronary guidewires and balloon/stent catheters during percutaneous coronary intervention (PCI) procedures.
# Technological Characteristics
The CorPath 200 System is composed of two functional sub-units; the Bedside Unit and the Remote Workspace. The Bedside Unit consists of the Articulated Arm, the Robotic Drive and the single-use Cassette. The Remote Workspace consists of the Interventional Cockpit (radiation shield) which houses the Control Console, as well as angiographic monitor(s). Commercially available guidewires and balloon/stent catheters are loaded into the single-use Cassette. By using the joysticks or touch screen of the Control Console, the physician can send commands to the Robotic Drive via a communication cable that advances, retracts or rotates the guidewire, and/or advances or retracts the balloon/stent catheters. The CorPath 200 System's software continuously monitors the communication between the Control Console and the Robotic Drive and alerts the physician if any communication error occurs.
A comparison of the device design shows that there is substantial equivalence between the CorPath 200 System and the predicate devices. All products are designed for the remote navigation of devices through the heart vasculature. All robotic systems are composed of a bedside unit and a remote workstation. In all robotic system the navigation of devices is accomplished by use of joysticks or track balls and touch-screen controls by the physician seated in a remote seated positioned.
The CorPath 200 System uses an electromechanical interface to manipulate commercially available PCI devices. In contrast, in the case of the Niobe MNS system, navigation is achieved through the use of magnetic fields. Also in contrast to the CorPath 200 System, in the case of the Sensei Robotic system, navigation is achieved through the use of a mechanical interface.
# Performance Data
{3}------------------------------------------------
Non-clinical testing of the CorPath 200 System Device consisted of performance testing, biocompatibility, sterilization, packaging, and product shelf life testing. These tests demonstrated that the technological characteristics such as product performance, design and intended use are substantially equivalent to the currently marketed predicate device.
Biocompatibility testing conducted on the system is summarized in the table below:
| Test Description |
|-------------------------------------------------------|
| Cytotoxicity |
| Sensitization (Guinea Pig Maximization) |
| Intracutaneous Reactivity (Irritation) |
| Systemic / Acute Toxicity |
| Hemocompatibility (Hemolysis - direct contact method) |
| Hemolysis - Partial Thromboplastin Time (PTT) |
| Hemolysis-Platelet & Leukocyte (P&L) |
| Material-Mediated Pyrogenicity |
In vitro bench testing conducted on the system is summarized in the table below:
| Test Description |
|---------------------------------------------------------|
| PCI Device Advancement Force Test |
| PCI Device Advancement Force at High Speed Test |
| PCI Device Velocity Test |
| PCI Device Continuous Move Positional Accuracy Test |
| PCI Device Placement Accuracy Test |
| PCI Device Discrete Movement Positional Accuracy Test |
| PCI Device Torque Test |
| PCI Device Rotational Velocity Test |
| PCI Device Wear Test |
| Particulate Analysis Test |
| PCI Device Dimensional and Functional Performance Tests |
Additionally, GLP animal studies were conducted to support the safety and performance of the device prior to the pivotal clinical study. The purpose of the GLP Studies was to evaluate the performance of the CorPath 200 System for Delivery and Deployment of Standard PTCA Devices in a porcine model acutely and at 30 days. The study was titled "Robotically-Assisted PCI: Evaluation of CorPath 200 System for
{4}------------------------------------------------
Delivery and Deployment of Standard PTCA Devices in Acute and 30-day Chronic Swine" to support the design and development of the CorPath 200 System. This study was conducted in compliance with Good Laboratory Practice (GLP) regulations (21 CFR Part 58).
#### Clinical Data
The CorPath 200 System was evaluated in the PRECISE Clinical Study. The PRECISE Clinical Study was a prospective, single-arm, multi-center, nonrandomized'study of the CorPath 200 System. The objective of the study was to evaluate the safety and effectiveness of the clinical and technical performance of the CorPath 200 System in the delivery and manipulation of coronary guidewires and stent/balloon devices for use in PC1 procedures. One hundred and sixty-four (164) subjects were enrolled and evaluated in the PRECISE Clinical Study at nine (9) clinical sites. The overall rate of clinical procedural success was 97.6%. One hundred percent of subjects achieved post-procedure stenosis of less than 30% (as evaluated by a Core Laboratory) and 97.6% of subjects had an absence of Major Adverse Cardiac Events (MACE). The overall device technical success rate was 98.8%. Although there were some limitations with the collection of operator radiation exposure data, the PRECISE trial results show a reduction in radiation exposure to the primary operator.
#### Conclusion
Based on similar intended use, technological characteristics, and performance characteristics, the CorPath 200 System is substantially equivalent to the predicate devices.
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 19 2012
Corindus, Inc. c/o Mona Advani Regulatory Affairs Consulting Agent CardioMed Device Consultants, LLC 5523 Research Park Drive, Suite 360 Baltimore, MD 21228
Re: K120834
Trade/Device Name: Corindus CorPath® 200 System Regulation Number: 21 CFR 870.1290 Regulation Name: Steerable catheter control system Regulatory Class: Class II Product Code: DXX Dated: May 25, 2012 Received: May 29, 2012
Dear Ms. Advani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for are case of to May 28, 1976, the enactment date of the Medical Device Amendments, or to conniner to may been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, mereleve, issues of the Act include requirements for annual registration, listing of general controls provision practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2 - Ms. Mona Advani
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): ________________
Device Name: CorPath® 200 System
Indications for Use:
The CorPath 200 System is intended for use in the remote delivery and manipulation of coronary guidewires and balloon/stent catheters during percutaneous coronary intervention (PCI) procedures.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K120834 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.