O-pel Laser and families of probe delivery devices
K160012 · Precise Light Surgical · GEX · Jun 2, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160012
Device Name
O-pel Laser and families of probe delivery devices
Applicant
Precise Light Surgical
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 2, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The O-pel laser and families of probe delivery devices are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including Arthroscopy, Urology, ENT, Dermatology, Plastic Surgery, General Surgery, Gastroenterology, Thoracic and Pulmonary and Gynecology.
Device Story
O-pel laser is an infrared laser system (approx. 2μm wavelength) for soft tissue surgery; includes laser console, control/display panel, fiber port, footswitch, and fiber optic probe delivery devices. Used in OR or clinical settings by physicians for open, laparoscopic, or endoscopic procedures. Laser energy delivered via fiber optics to target tissue; enables precise cutting, ablation, and coagulation. Benefits include minimally invasive access and controlled tissue interaction. Output controlled by surgeon via footswitch and console interface; provides thermal energy for surgical effect.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing, including electromagnetic compatibility (IEC 60601-1-2:2007), electrical safety (IEC 60601-1:2007), laser safety (IEC 60601-2-22, IEC 60825-1), wireless coexistence, and biocompatibility (ISO 10993-5, ISO 10993-7).
Indicated for patients requiring soft tissue incision, excision, resection, ablation, vaporization, coagulation, or hemostasis across multiple surgical specialties including ENT, Dermatology, Plastic Surgery, General Surgery, Thoracic/Pulmonary, Urology, Gastroenterology, Gynecology, and Arthroscopy/Orthopedics.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Lumenis Versapulse PowerSuite Holmium (Ho:YAG) and Dual Wavelength (Ho:YAG/Nd:YAG) Surgical Lasers & Delivery Devices with Accessories (K011703)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three wavy lines that also suggest human profiles.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2016
Precise Light Surgical Mr. Ken Arnold President And CEO 310 W. Hamilton Ave., Suite 210 Campbell, California 95008
Re: K160012
Trade/Device Name: O-pel Laser And Families Of Probe Delivery Devices Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: Mav 2, 2016 Received: May 4, 2016
Dear Mr. Arnold:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K160012
#### Device Name
O-pel Laser And Families Of Probe Delivery Devices
Indications for Use (Describe)
The O-pel laser and families of probe delivery devices are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including Arthroscopy, Urology, ENT, Dermatology, Plastic Surgery, General Surgery, Gastroenterology, Thoracic and Pulmonary and Gynecology.
ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including:
- · Endonasal/sinus Surgery
- · Partial Turbinectomy
- · Polypectomy
- · Dacryocystorhinostomy
- · Frontal Sinusotomy
- · Ethmoidectomy
- Maxillary antrostomy
- · Functional endoscopic sinus surgery
- · Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal
- · Tonsillectomy
- · Adenoidectomy
Dermatology and Plastic Surgery
Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including:
- · Basal cell carcinomas
- · Lesions of skin and subcutaneous tissue
- · Skin tags
- Plantar warts
#### General Surgery
Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- · Cholecystectomy
- · Lysis of adhesion
- · Appendectomy
- Biopsy
- · Skin incision
- · Tissue dissection
- · Excision of external tumors and lesions
- · Complete or partial resection of internal organs, tumors and lesions
- · Mastectomy
- · Hepatectomy
- · Pancreatectomy
{3}------------------------------------------------
- · Splenectomy
- · Thyroidectomy
- · Parathyroidectomy
- · Herniorrhaphy
- · Tonsillectomy
- Lymphadenectomy
- · Partial Nephrectomy
- · Pilonidal Cystectomy
- · Resection of lipoma
- Debridement of Decubitus Ulcer
- Hemorrhoids
- · Debridement of Statis Ulcer
- · Biopsy
- Thoracic and Pulmonary
Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
- · Laryngeallesions
- · Airway obstructions including carcinoma
- Polyps and Granulomas
- · Palliation of obstructing carcinomas of the tracheobronchial tree
### Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- · Urethral strictures
- · Bladder neck incisions(BNI)
- · Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral Tumors
- · Ablation of Benign Prostatic Hypertrophy (BPH)
- · Transurethral incision of the prostrate (TUIP)
- · Laser resection of the prostate (HoLRP)
- · Laser enucleation of the prostate (HoLEP)
- · Laser ablation of the prostate (HoLAP)
- · Condylomas
- · Lesions of external genitalia
### Gastroenterology
Open and endoscopic gastroenterology surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) including:
- · Appendectomy
- Polyps
- Biopsy
- Gall Bladder calculi
- Biliary/Bile duct calculi
- Ulcers
- · Gastric ulcers
- · Duodenal ulcers
- · Non bleeding ulcers
- · Pancreatitas
- · Hemorrhoids
- · Cholecystectomy
- · Benign and Malignant Neoplasm
- · Angiodysplasia
- · Colorectal cancer
- · Telangiectasias
{4}------------------------------------------------
· Telangiectasias of the Osler-Weber-Renu disease
- Vascular Malformation
- Gastritis
- · Esophagitis
- · Esophageal ulcers
- · Varices
- · Colitis
- · Mallory-Weiss tear
- · Gastric erosions
Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis)
- · Intra-uterine treatment of submucous fibroids, benign endometrial polyps, and uterine septum by incision, excision, ablation, and or vessel coagulation
- · Soft tissue excision procedures such as excisional conization of the cervix
Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue)
- · Ablation of soft and cartilaginous tissue in Minimal Invasive Spinal Surgery including:
- · Percutaneous Laser Disc Decompression/Discectomy
- · Foraminoplasty
- · Ablation and coagulation of soft vascular and non vascular tissue in minimally invasive
spinal surgery
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
#### 510(k) Summary
#### Precise Light Surgical O-pel Laser and families of probe delivery devices
#### I SUBMITTER
Precise Light Surgical 310 W. Hamilton Ave Suite 210 Campbell. CA 95008 Phone: 844-669-1845 e-mail: karnold@preciselightsurgical.com
Contact person: Ken Arnold, President Date prepared: December 30, 2015
Establishment Registration Number: TBD
II DEVICE
Name of Device: O-pel laser and families of Probe delivery devices
Common or Usual Name: 2 micron Laser System
Classification Name: Powered Laser Surgical Instrument and Accessories (21CFR 878.4810)
Regulatory Code: II
Product Code: GEX
III PREDICATE DEVICE(S)
K051167 Allmed RevoLix and RevoLix Jr K011703 Modified Lumenis Versapulse PowerSuite Holmium (Ho:YAG) and Dual Wavelength (Ho:YAG/Nd:YAG) Surgical Lasers & Delivery Devices with Accessories
IV DEVICE DESCRIPTION
The O-pel laser is an infrared laser system with an output wavelength of approximately 2μm. The O-pel laser and families of probe delivery devices (O-pel laser system) have an Intended Use of cutting and coagulating tissue, specifically incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue. The O-pel laser system is comprised of the following components:
-a laser console -control and display panel; -a fiber port for delivery systems;
{6}------------------------------------------------
-a footswitch:
-a variety of fiber optic (probe) delivery devices
#### V INDICATIONS FOR USE
The O-pel laser and families of probe delivery devices are intended for use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including Arthroscopy, Urology, ENT, Dermatology, Plastic Surgery, General Surgery, Gastroenterology, Thoracic and Pulmonary and Gynecology.
### ENT
Endoscopic endonasal surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue) including:
- . Endonasal/sinus Surgery
- . Partial Turbinectomy
- . Polypectomy
- . Dacryocystorhinostomy
- . Frontal Sinusotomy
- . Ethmoidectomy
- . Maxillary antrostomy
- . Functional endoscopic sinus surgery
- . Lesions or tumors of the oral, nasal, glossal, pharyngeal and laryngeal
- Tonsillectomy
- . Adenoidectomy
### Dermatology and Plastic Surgery
Incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft, mucosal, fatty and cartilaginous tissue, in therapeutic plastic, dermatologic and aesthetic surgical procedures including:
- . Basal cell carcinomas
- . Lesions of skin and subcutaneous tissue
- . Skin tags
- . Plantar warts
### General Surgerv
Open laparoscopic and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Cholecystectomy
- . Lysis of adhesion
- . Appendectomy
- . Biopsy
- . Skin incision
- . Tissue dissection
- . Excision of external tumors and lesions
{7}------------------------------------------------
- . Complete or partial resection of internal organs, tumors and lesions
- Mastectomy .
- . Hepatectomy
- . Pancreatectomy
- . Splenectomv
- . Thyroidectomy
- . Parathyroidectomy
- . Herniorrhaphy
- . Tonsillectomy
- . Lymphadenectomy
- . Partial Nephrectomy
- Pilonidal Cystectomy
- . Resection of lipoma
- . Debridement of Decubitus Ulcer
- . Hemorrhoids
- . Debridement of Statis Ulcer
- . Biopsy
#### Thoracic and Pulmonary
Open and endoscopic thoracic and pulmonary surgery (incision, excision, resection. ablation, vaporization, coagulation and hemostasis) of soft tissue
- . Laryngeal lesions
- . Airway obstructions including carcinoma
- . Polyps and Granulomas
- . Palliation of obstructing carcinomas of the tracheobronchial tree
#### Urology
Open and endoscopic surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
- Urethral strictures
- Bladder neck incisions(BNI)
- Ablation and resection of Bladder Tumors, Uretheral Tumors and Ureteral . Tumors
- . Ablation of Benign Prostatic Hypertrophy (BPH)
- . Transurethral incision of the prostrate (TUIP)
- . Laser resection of the prostate (HoLRP)
- . Laser enucleation of the prostate (HoLEP)
- Laser ablation of the prostate (HoLAP) .
- . Condylomas
- . Lesions of external genitalia
#### Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) including:
{8}------------------------------------------------
- Appendectomy .
- . Polyps
- . Biopsv
- . Gall Bladder calculi
- . Biliary/Bile duct calculi
- . Ulcers
- . Gastric ulcers
- . Duodenal ulcers
- . Non bleeding ulcers
- . Pancreatitas
- . Hemorrhoids
- . Cholecystectomy
- . Benign and Malignant Neoplasm
- . Angiodysplasia
- . Colorectal cancer
- . Telangiectasias
- . Telangiectasias of the Osler-Weber-Renu disease
- Vascular Malformation
- . Gastritis
- . Esophagitis
- . Esophageal ulcers
- . Varices
- . Colitis
- Mallory-Weiss tear ●
- . Gastric erosions
#### Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis)
- Intra-uterine treatment of submucous fibroids, benign endometrial polyps, and uterine septum by incision, excision, ablation, and or vessel coagulation
- . Soft tissue excision procedures such as excisional conization of the cervix
#### Arthroscopy
Arthroscopy/Orthopedic surgery (excision, ablation and coagulation of soft and cartilaginous tissue)
- Ablation of soft and cartilaginous tissue in Minimal Invasive Spinal Surgery ● including:
- Percutaneous Laser Disc Decompression/Discectomy
- . Foraminoplasty
- . Ablation and coagulation of soft vascular and non vascular tissue in minimally invasive spinal surgery
8-4
{9}------------------------------------------------
#### VI RATIONAL FOR SUBSTANTIAL EQUIVALENCE
The O-pel laser system with fiber optic delivery devices share the same intended use, indications for use, similar design features and functional features and therefore are substantially equivalent to the Allmed Revolix and the Lumenis VersaPulse systems.
#### PERFORMANCE DATA VII
Electromagnetic Compatibility. The laser console was tested to IEC 60601-1-2: 2007 by a third party and was found to comply.
Electrical Safety: The laser console was tested to IEC 60601-1: 2007 by a third party and was found to comply.
Laser Safety: The laser console was tested to IEC 60601-2-22 and IEC 60825-1: by a third party and was found to comply.
Coexistence: The laser console was tested to the FDA Guidance, "Radio Frequency Wireless Technology in Medical Devices" issued August 14, 2013 and was found to comply.
Biocompatibility. The biocompatibility evaluation of the Probe delivery device was conducted per FDA draft guidance document, "Use of International Standard ISO 10993. "Biological Evaluation of medical devices Part 1: Evaluation and Testing "" dated April 23, 2013. The following tests were performed:
ISO 10993-5: Biological Evaluation of medical devices Part 5- Tests for in vitro cytotoxicity ANSI /AAMI/ISO 10993-7:2008 (R) 2012 Biological evaluation of medical device - Part 7 Ethylene Oxide sterilization residues.
#### VIII CONCLUSIONS
The O-pel laser and families of Probe delivery devices share indications for use, similar design features, and functional features and therefore are substantially equivalent to the predicate devices. Any differences in technological characteristics do not raise any new issues of safety or efficacy. Substantial equivalence was not based upon clinical performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.