Disposal Lens Cleaning Sheath

K153773 · Olympus Medical Systems Corp. · GCJ · Nov 22, 2016 · Gastroenterology, Urology

Device Facts

Record IDK153773
Device NameDisposal Lens Cleaning Sheath
ApplicantOlympus Medical Systems Corp.
Product CodeGCJ · Gastroenterology, Urology
Decision DateNov 22, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Disposable Lens Cleaning Sheath is an endoscope accessory that uses saline solution and carbon dioxide to remove debris from the lens surface of endoscopes used within the thoracic and abdominal cavities, including female reproductive organs.

Device Story

Disposable Lens Cleaning Sheath; single-use sterile accessory for flexible and rigid endoscopes. Device attaches to endoscope, saline source, and CO2 gas source. Operation involves manual activation to spray saline and CO2 gas onto the objective lens surface to clear debris. Used in thoracic and abdominal cavities, including female reproductive organs. Operated by clinicians during endoscopic procedures. Output is a cleared lens surface, improving visualization for the surgeon. Benefits include reduced need to remove the endoscope from the body to clean the lens, thereby shortening procedure time and maintaining surgical field visibility.

Clinical Evidence

Bench testing only. Performance testing included sterilization validation (ISO 11137-1), shelf life (ASTM F1980-07), and mechanical performance (insertion/removal, trocar insertion, gas/spray feeding, debris removal efficacy, durability, and package integrity). Biocompatibility testing conducted per ISO 10993 standards (cytotoxicity, sensitization, irritation, systemic toxicity).

Technological Characteristics

Single-use sterile sheath; includes insertion section, main body, and water/CO2 tube. Materials evaluated for biocompatibility per ISO 10993. Sterilization via radiation (ISO 11137). Mechanical performance tested for gas/spray delivery and durability. No software or electrical components.

Indications for Use

Indicated for use as an endoscope accessory to remove debris from the lens surface of endoscopes during procedures within the thoracic and abdominal cavities, including female reproductive organs.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around a symbol. The symbol is a stylized representation of a human figure in profile, with three overlapping heads suggesting a sense of community or interconnectedness. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 2, 2016 Olympus Medical Systems Corp % Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034-0610 Re: K153773 > Trade/Device Name: Disposable Lens Cleaning Sheath Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: November 11, 2016 Received: November 14, 2016 Dear Sheri Musgnung. This letter corrects our substantially equivalent letter of November 22, 2016. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {1}------------------------------------------------ or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K153773 Device Name Disposable Lens Cleaning Sheath ### Indications for Use (Describe) The Disposable Lens Cleaning Sheath is an endoscope accessory that uses saline solution and carbon dioxide to remove debris from the lens surface of endoscopes used within the thoracic and abdominal cavities, including female reproductive organs | Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> <span style="vertical-align: middle;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="vertical-align: middle;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> | |-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff [PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov) *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* {3}------------------------------------------------ # 510(k) SUMMARY K153773 Disposable Lens Cleaning Sheath November 22, 2016 # 5.1 General Information | ■ Applicant: | OLYMPUS MEDICAL SYSTEMS CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan<br>192-8507<br>Establishment Registration No: 8010047 | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ■ Official Correspondent: | Sheri L. Musgnung<br>Olympus Corporation of the Americas<br>3500 Corporate Parkway PO Box 610<br>Center Valley, PA 18034-0610, USA<br>Phone: 484-896-3147<br>FAX: 484-896-7128<br>Email: sheri.musgnung@olympus.com | | ■ Manufacturer: | OLYMPUS MEDICAL SYSTEMS CORP.<br>34-3 Hirai, Hinode-machi,<br>Tokyo, Japan 190-0182<br>Establishment Registration No.: 3003637092 | # 5.2 Device Identification | ■ Device Trade Name: | Disposable Lens Cleaning Sheath | |-------------------------|---------------------------------| | ■ Common Name: | Lens Cleaning Sheath | | ■ Regulation Number: | 876.1500 | | ■ Regulation Name: | Endoscope and accessories | | ■ Regulatory Class: | II | | ■ Classification Panel: | General & Plastic Surgery | | ■ Product Code: | GCI | {4}------------------------------------------------ | Device Trade<br>Name | Common Name | Applicant | 510(k) No. | |-------------------------------------------------------|-------------------------------------------------------|------------------------------------|------------| | Clear-VuTM<br>System<br>(Primary<br>predicate device) | endoscope lens<br>cleaning and<br>defogging<br>device | Minimally Invasive Devices,<br>LLC | K080613 | #### Predicate Device Information 5.3 The predicate device has not been subject to a design related recall. #### 5.4 Device Description The Disposable Lens Cleaning Sheath is an endoscope accessory that uses saline solution and carbon dioxide to remove debris from the lens surface of endoscopes used within the thoracic and abdominal cavities, including female reproductive organs. The subject device is a single use disposable device and is provided to the user sterile. The subject devices consist of an insertion section, main body section, and water/CO2 gas tube section. The subject devices are used in combination with flexible endoscope and rigid endoscope. Upon use the device is attached to the compatible endoscope, water source, and gas source to remove debris from the objective lens surface of the endoscope by spraying saline solution and CO2 gas. #### ર્સ ર ર્ Indications for Use The Disposable Lens Cleaning Sheath is an endoscope accessory that uses saline solution and carbon dioxide to remove debris from the lens surface of endoscopes used within the thoracic and abdominal cavities, including female reproductive organs. {5}------------------------------------------------ #### 5.6 Comparison of Technological Characteristics The subject device and predicate device do not have the same technological characteristics. The construction of the device (thread), component materials (sheath and tip material), water supply, gas supply, and method of sterilization are different. The different technological characteristics of the subject device do not raise different safety or effectiveness questions as comparable device construction, materials, water supply, gas supply, and methods of sterilization have been evaluated in previous 510(k)s for this device type. Accepted scientific methods exist to assess the effects of the different technological characteristics including electrical, software, and performance testing. #### 5.7 Summary of non-clinical testing · Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment. ·Performance testing including sterilization validation, shelf life, and mechanical performance/bench testing was conducted to demonstrate the basic performance of the subject device and confirmed that the subject device performs as intended. ### Sterilization validation Sterilization validation was carried out with Methods VDmax25 in accordance with ISO11137-1:2006. # Shelf life testing Shelf life testing was conducted based on an accelerated aging test in accordance with ASTM F1980-07(Reapproved 2011), the standard guide for accelerated aging of sterile barrier systems for medical devices. Three years real-time aging test will be performed to demonstrate longer stability and support the results of the accelerated aging test. # Mechanical performance testing Mechanical performance testing was conducted on the following items to demonstrate the basic performance of the subiect device and confirmed that the subject device performs as intended. - Inserted to the endoscope and remove from the endoscope - Inserting into the trocar - Gas feeding operation - Spray feeding operation - Inspection of removing the debris - Insertion section durability - Fixing strength of the main body - Inspection of the package {6}------------------------------------------------ ### Biocompatibility testing Biocompatibility testing is performed in accordance with the FDA Guidance," Required Biocompatibility Training and Toxicology Profiles for Evaluation of Medical Devices, Blue Book Memo, G95-1" including cytotoxicity, sensitization, and irritation testing. · The following standards have been applied to the Disposable Lens Cleaning Sheath. | Standard No. | Standard Title | |-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ISO 10993-1 Fourth Edition:2009-10-15 | Biological Evaluation Of Medical Devices - Part1:<br>Evaluation And Testing Within A Risk Management<br>Process [Including: Technical Corrigendum 1 (2010)] | | AAMI ANSI ISO 10993-<br>3:2003/(R)2013 | Biological Evaluation Of Medical Devices Part 3:<br>Tests For Genotoxicity, Carcinogenicity And<br>Reproductive Toxicity | | AAMI ANSI ISO 10993-<br>5:2009/(R)2014 | Biological Evaluation Of Medical Devices – Part5:<br>Tests For In Vitro Cytotoxicity | | ISO 10993-10 Third Edition:<br>2010-08-01 | Biological Evaluation Of Medical Devices - Part 10:<br>Tests For Irritation And Skin Sensitization | | AAMI/ANSI/ISO 10993-11:<br>2006/(R)2010 | Biological Evaluation Of Medical Devices Pat11 :<br>Tests For Systemic Toxicity | | ISO 11137-1 First Edition:<br>2006-04-15 | Sterilization Of Health Care Products - Radiation -<br>Part 1: Requirements For Development, Validation<br>And Routine Control Of A Sterilization Process For<br>Medical Devices [Including: Amendment 1 (2013)] | | ISO 11137-2 Third Edition:<br>2013-06-01 | Sterilization Of Health Care Products - Radiation -<br>Part 2: Establishing The Sterilization Dose | | ASTM F1980-07<br>(Reapproved 2011) | Standard Guide For Accelerated Aging Of Sterile<br>Barrier Systems For Medical Devices | | ISO 14971 Second Edition:<br>2007-03-01 | Medical Devices - Application Of Risk Management<br>To Medical Devices | # 5.8 Conclusion The Disposable Lens Cleaning Sheath is substantially equivalent to the proposed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...