CLEAR-VU SYSTEM, MODELS: CVL-01-1000, CVL-01-1030, CVL-01-1045

K080613 · Minimally Invasive Devices, LLC · OCT · Nov 6, 2008 · Gastroenterology, Urology

Device Facts

Record IDK080613
Device NameCLEAR-VU SYSTEM, MODELS: CVL-01-1000, CVL-01-1030, CVL-01-1045
ApplicantMinimally Invasive Devices, LLC
Product CodeOCT · Gastroenterology, Urology
Decision DateNov 6, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Clear-Vu is a single-use, disposable laparoscopic accessory device that is intended to facilitate intra-operative defogging and cleaning of the lens of a laparoscope during minimally-invasive surgery while maintaining visualization of the surgical site.

Device Story

Clear-Vu System is a single-use, disposable laparoscopic accessory; consists of a multi-lumen sheath assembly mounted on the laparoscope shaft. Device integrates with existing OR infrastructure: connects to CO2 insufflation circuit for lens defogging and to surgical irrigation system for saline lens flushing. Operated by surgeons or surgical staff during minimally-invasive procedures. By removing debris and fog from the lens, the device maintains continuous visualization of the surgical site, reducing the need to remove the laparoscope for manual cleaning, thereby improving procedural efficiency and patient safety.

Clinical Evidence

Bench testing performed in a surgical environment simulation chamber; preclinical animal testing performed in a porcine model to demonstrate effectiveness.

Technological Characteristics

Multi-lumen sheath assembly; single-use, disposable; connects to existing CO2 insufflation and surgical irrigation systems. Mechanical cleaning/defogging via fluid/gas flow.

Indications for Use

Indicated for use during minimally-invasive laparoscopic surgery to maintain clear visualization by defogging and cleaning the laparoscope lens.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a stylized symbol and surrounding text. The symbol is a stylized representation of a human figure embracing a bird, conveying a sense of care and protection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Minimally Invasive Devices, LLC % Mr. Michael H. Southworth, RAC Principal & Senior Consultant Southworth & Associates, LLC 1035 Waldo Way Twinsburg, OH 44087 JUL 2 7 2015 Re: K080613 > Trade/Device Name: Clear-Vu™ System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCT, GCJ Dated (Date on orig SE ltr): October 28, 2008 Received (Date on orig SE Itr): October 30, 2008 Dear Mr. Southworth, This letter corrects our substantially equivalent letter of November 6, 2008. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {1}------------------------------------------------ Page 2 - found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Statement 510(k) Number: __K480613 Device Name: Clear Vu System Indications for Use: Clear-Vu is a single-use, disposable laparoscopic accessory device that is intended to facilitate intra-operative defogging and cleaning of the lens of a laparoscope during minimally-invasive surgery while maintaining visualization of the surgical site. Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) 16 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Norman M. Whan eproductive, Abdomin and Radiological Devic 510(k) Number {3}------------------------------------------------ - ## 510(k) Summary of Safety and Effectiveness Clear-Vu System Minimally Invasive Devices, LLC : . NOV - 6 2008 | 510(k) Submitter | Minimally Invasive Devices, LLC<br>1275 Kinnear Road<br>Columbus, Ohio 43212 | NOV - 6 2008 | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Contact Person | Wayne L. Poll, MD, Chairman<br>Phone: (614) 580-2022<br>Fax: (614) 889-7630<br>E-mail: waynepoll@hotmail.com | | | Date Prepared | Revised October 28, 2008 | | | Device Name | Proprietary Name: Clear-Vu System<br>Common Name: endoscope lens cleaning and defogging device<br>Classification Name: "accessory, endoscope," a class II device in<br>accordance with 21 CFR § 876.1500 | | | Intended Use | The Clear-Vu System is a single-use, disposable laparoscopic accessory<br>device intended to facilitate intra-operative defogging and cleaning of<br>the lens of a laparoscope during minimally-invasive surgery while<br>maintaining visualization of the surgical site. | | | Device Description | The Clear-Vu System consists of a multi-lumen sheath assembly<br>mounted on the shaft of the laparoscope, which connects to the<br>existing CO₂ insufflation circuit to defog the laparoscope lens and<br>divert surgical debris, and connects to the existing surgical irrigation<br>system to clean the lens via saline flush. | | | Substantial<br>Equivalence | The Clear-Vu System is substantially equivalent in terms of safety and<br>effectiveness to the following predicate devices:<br>1. Defogging Heated Endoscopic Lens Protector / K062779<br>New Wave Surgical Corp.<br>2. Endo-Scrub® 2 / K982594<br>Xomed, Inc.<br>3. SeeClear™ Laparoscopic Smoke Evacuation System/510(k) Exempt<br>JLJ Medical Devices International, LLC | | | Technological<br>Comparison | The Clear-Vu System is substantially equivalent to the cited predicate<br>devices in terms of intended use, and the technological differences are<br>not significant relative to the safety or effectiveness of the new device. | | | Performance Testing-<br>Bench | The device was subjected to simulated use testing to demonstrate its<br>effectiveness in a surgical environment simulation chamber. | | | Performance Testing-<br>Animal | The device was subjected to preclinical animal testing to demonstrate its<br>effectiveness in a porcine model. | |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...