K153022 · Philips Medical Systems Nethrlands BV · LLZ · Nov 18, 2015 · Radiology
Device Facts
Record ID
K153022
Device Name
Philips IntelliSpace Cardiovascular
Applicant
Philips Medical Systems Nethrlands BV
Product Code
LLZ · Radiology
Decision Date
Nov 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
Philips IntelliSpace Cardiovascular software product is an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital medical images.
Device Story
IntelliSpace Cardiovascular (ISCV) is a multimodality image and information management system for cardiology departments. It ingests digital cardiovascular images, waveforms, and data from various modalities (ultrasound, X-ray, CT, MR, nuclear medicine). The system provides a centralized platform for clinicians to review, analyze, quantify, and report on patient studies. It supports remote deployment, thin-client browser-based access (HTML5), and integration with hospital information systems (HIS/EMR). Clinicians use the system to access patient cardiovascular history, perform measurements, and generate reports. The modular design allows for configurable workflows and data exchange via standard protocols (DICOM, HL7). By consolidating data and providing diagnostic tools, the device facilitates efficient clinical decision-making and patient management across the cardiovascular care continuum. It does not interact directly with the patient and does not apply radiation or contrast agents.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence is based on non-clinical performance testing, including system-level tests, performance tests, and safety testing derived from risk analysis, demonstrating that the device meets all defined functionality requirements.
Technological Characteristics
Software-based PACS; runs on standard IT hardware and server/desktop operating systems. Supports virtualization (VMWare) and thin-client (HTML5) browser access. Connectivity via TCP/IP, DICOM, DSR, HL7, NFS, FTP. Integrates with Philips TSM/Xper IM. Features include automated study placement, DICOM SR mapping, and consistency check mechanisms for diagnostic guidance. No patient-contacting components; no radiation/contrast application.
Indications for Use
Indicated for patients undergoing radiology procedures and users of the equipment. The system performs import, export, storage, archival, review, analysis, quantification, reporting, and database management of digital cardiovascular images, waveforms, and data.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2015
Philips Medical Systems Nederland BV % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street NW Buffalo MN 55313
Re: K153022
Trade/Device Name: Philips Intellispace Cardiovascular Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 5, 2015 Received: November 6, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K153022
Device Name Philips IntelliSpace Cardiovascular
#### Indications for Use (Describe)
Philips IntelliSpace Cardiovascular software product is an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital cardiovascular images, waveforms and data related to cardiology.
Philips IntelliSpace Cardiovascular offers support for third party applications in order to enable the use of commercially available tools and specified applications for analysis, quantification and reporting. It allows multiple users fast access to, and exchange of specific and/or multiple cardiology exams.
Philips IntelliSpace Cardiovascular software runs on standard information technology hardware, utilizing the standard information technology operating systems and user interface. Communication and data exchange are done using standard protocols. Philips IntelliSpace Cardiovascular will also be made available for use on specified Cardiovascular Monitoring Systems, which use suitable hardware components.
The modular design allows configurability to tailor the image import, archive and communications solution to one's particular budgetary and performance needs. The number of modalities and reporting and/or viewing sites can be configured per system.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white.
## 510(k) Summary
# IntelliSpace Cardiovascular
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92.
Date Prepared: September 30, 2015
#### I. Submitter's name and address
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best<br>The Netherlands<br>Establishment Registration Number: 3003768277 |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dave J.W. Bierhuizen<br>Quality and Regulatory Manager<br>Phone: +31 40 2764366<br>E-mail: dave.bierhuizen@philips.com |
#### II. Device information
| Device Name: | IntelliSpace Cardiovascular | |
|-----------------|--------------------------------------------------------|---------------------------------------------|
| Common Name: | Cardiovascular Image and Information Management System | |
| Classification: | Classification name: | Picture Archiving and Communications System |
| | Device class: | Class II |
| | Classification regulation: | 21 CFR 892.2050 |
| | Classification panel: | Radiology |
| | Product code: | LLZ |
#### III. Predicate device information
| Predicate Device: | Trade name: | Philips Xcelera |
|-------------------|----------------------------|---------------------------------------------|
| | Manufacturer: | Philips Medical Systems Nederland B.V. |
| | 510(k) clearance: | K061995 |
| | Classification name | Picture Archiving and Communications System |
| | Device class: | Class II |
| | Classification regulation: | 21 CFR 892.2050 |
| | Classification panel: | Radiology |
| | Product code: | LLZ |
#### IV. Device Description
IntelliSpace Cardiovascular is a comprehensive cardiac image and information management solution designed to provide clinicians with convenient access to the detailed records of all cardiac across their complete cardiovascular care continuum. The application also provides hospital administrators and department managers with detailed operational information, as well as productivity and outcomes reporting. Key components include a wide range of detailed clinical modules that capture data during diagnostic/therapeutic procedures and patient follow-up encounters. Interfaces to other systems and devices within the cardiology departments as well as across the enterprise system, such as HIS/EMR, are available.
The solution supports a remote deployment model.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
#### V. Intended use/ Indications for Use
#### Intended Use
Philips IntelliSpace Cardiovascular software product is an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital medical images.
#### Indication for Use
Philips IntelliSpace Cardiovascular software product is an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital cardiovascular images, waveforms and data related to cardiology.
Philips IntelliSpace Cardiovascular offers support for third party applications in order to enable the use of commercially available tools and specified applications for analysis, quantification and reporting. It allows multiple users fast access to, and exchange of specific and/or multiple cardiology exams.
Philips IntelliSpace Cardiovascular software runs on standard information technology hardware and software, utilizing the standard information technology operating systems and user interface. Communication and data exchange are done using standard protocols. Philips IntelliSpace Cardiovascular will also be made available for use on specified Cardiovascular Monitoring Systems, which use suitable hardware components.
The modular design allows configurability to tailor the image import, archive and communications solution to one's particular budgetary and performance needs. The number of modalities and reporting and/or viewing sites can be configured per system.
#### VI. Comparison of Technological Characteristics with the Predicate Device
Philips IntelliSpace Cardiovascular (ISCV) software is an integrated multimodality image and information system, which employs and further builds on the same fundamental scientific technology as the Philips Xcelera software.
IntelliSpace Cardiovascular software runs on standard information technology hardware and software.
IntelliSpace Cardiovascular does not interact with the patient, and does not apply any radiation or contrast agent to the patient.
A comparison matrix (please see Table 5-1 below) shows the similarities and differences. The ISCV software with the listed enhancements is comparable to the previously cleared Xcelera product as the changes do not impact the Intended Use or patient population. Enhancements relate to upgrades in current standard IT application and technology environment or clinical practices to improve customer workflows. All changes are implemented and tested utilizing released design and change control processes. No new issues of safety or effectiveness are raised as compared to the predicate device.
| # | Specification / Feature | IntelliSpace Cardiovascular<br>(Subject device) | Xcelera<br>(Predicate device - K061995) |
|-------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Intended Use / Target population | | |
| 1. | Intended Use | Philips IntelliSpace Cardiovascular software<br>product is an integrated multimodality image<br>and information system designed to perform<br>the necessary functions required for import,<br>export, storage, archival, review, analysis,<br>quantification, reporting and database<br>management of digital medical images. | Philips Xcelera software product is an<br>integrated multimodality image and<br>information system designed to perform the<br>necessary functions required for import,<br>export, storage, archival, review, analysis,<br>quantification, reporting and database<br>management of digital medical images. |
| 2. | Target population | Patients undergoing radiology procedures<br>and the users of the equipment. | Patients undergoing radiology procedures<br>and the users of the equipment. |
| # | Specification / Feature | IntelliSpace Cardiovascular<br>(Subject device) | Xcelera<br>(Predicate device - K061995) |
| | Technology | | |
| 3. | Hardware Platform | Standard IT hardware | Standard IT hardware |
| | requirements | | |
| 4. | Operating Platform<br>requirements | Current industry standard versions of server<br>and desktop operating systems. | Industry standard versions of server and<br>desktop operating systems. |
| 5. | Browser support | Thin clients are supported on HTML5<br>capable browsers. | Not supported |
| 6. | Virtualization | Server virtualization enabled with VMWare | Not supported |
| 7. | Imaging and Other<br>Communications<br>Protocols | TCP/IP, DICOM, DSR, HL7, NFS, FTP | TCP/IP, DICOM, DSR, HL7, NFS, FTP |
| 8. | Support for launching 3rd<br>party medical devices | Plug-in support for: ultrasound,<br>cardiovascular X-ray, nuclear medicine,<br>computed tomography, magnetic<br>resonance, and electrophysiology studies.<br>URL launch of 3rd party medical devices<br>(that allow URL launch) | Plug-in support for: ultrasound,<br>cardiovascular X-ray, nuclear medicine,<br>computed tomography, magnetic<br>resonance, and electrophysiology studies. |
| 9. | EMR/HIS Interface | A mechanism to launch from the EMR<br>system directly into ISCV, and into the EMR. | Not supported |
| 10. | Integration with Philips<br>information management<br>systems | TSM (Table Side Module) Xper IM Philips CVIS | TSM (Table Side Module) Xper IM |
| 11. | System entry screen | Browser-based workspace with applets,<br>system extensions and workflow modules.<br>Provides two layers: user-centric, for search<br>and configurable worklist functionality;<br>patient-centric: for detailed cardiovascular<br>history. | Main window including configurable worklist. |
| | Archiving | | |
| 12. | Automatic Study<br>placement and folder<br>creation | Yes | Yes |
| 13. | Study and Patient<br>Management, incl. delete<br>images from study, study<br>split, merge, and<br>enterprise master patient<br>index | Yes | Not supported |
| 14. | Study Type | Ultrasound, Nuclear medicine, X-ray (cath<br>and invasive vascular), all other modality<br>types related to cardiology. | Ultrasound, Nuclear medicine, X-ray (cath) |
| 15. | DICOM SR Mapping<br>Tooling | Yes | Not supported |
| 16. | Customized<br>measurements and<br>calculations tool | Yes | Yes |
| 17. | PDF Import | Import reports that are in PDF format from<br>external sources. | Not supported |
| 18. | Data Mining | Supported | Supported |
| | Viewing | | |
| 19. | Supported Data and<br>Modalities | All DICOM formats as stated in the<br>DICOM, including but not limited to:<br>Ultrasound, Cath, CT and MR. Philips' proprietary ultrasound image<br>format (DSR-TIFF). | All DICOM formats as stated in the<br>DICOM, including but not limited to:<br>Ultrasound, Cath, CT and MR. Philips' proprietary ultrasound image<br>format (DSR-TIFF). |
| 20. | Cath Viewer | Yes | Yes |
| 21. | Echo Viewer | Yes | Yes |
| 22. | Remote Viewing (thin<br>client) | Zero-install (thin) client, browser based<br>technology (HTML5) to perform review-only<br>of series, runs, loops and images. | Not supported |
| | Quantification and | | |
| | Reporting | | |
| 23. | Cath Viewer | Simple distance measurements (non- | Simple distance measurements (non- |
| | Measurements | persistent) | persistent) |
| # | Specification / Feature | IntelliSpace Cardiovascular (Subject device) | Xcelera (Predicate device - K061995) |
| 24. | Cath Quantitative Analysis | Yes | Yes |
| 25. | Echo Measurements and Calculations | 2D Doppler MMode Trending graph of measurements and z-scores Z-scores: Michigan data | 2D Doppler MMode Trending graph of measurements and z-scores Z-scores: Michigan data |
| 26. | | Z-scores: extensions of z-scores data (Boston) | Not supported |
| 27. | | Blood Pressure import from DICOM SR files from vendors (Philips, GE, Siemens, Toshiba) | Not supported |
| 28. | Diagnostic Guidance | Consistency check mechanism that will allow clinicians to predefine a set of rules. | Not supported |
| 29. | Remote reporting (tele-cardiology) | View the study from a remote location and perform reporting tasks on the study. | Not supported |
| 30. | Remote Echo Reporting (thin client) | Thin client, browser based technology (HTML5) to perform diagnostic review and 2D measurements on echo images. Supported on standard PC hardware; not supported on mobile devices. | Not supported |
| Other | | | |
| 31. | System access | Define user rights based on institute level. | Global user rights |
| 32. | Single access to configuration | Configure ISCV in a singular place | Not supported |
Table 5-1 Comparison Table of the modified ISCV to the predicate Xcelera product.
{5}------------------------------------------------
## Philips Medical Systems Nederland B.V.
Image /page/5/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white.
{6}------------------------------------------------
## Philips Medical Systems Nederland B.V.
Image /page/6/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
#### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### Summary of Non-clinical testing
No performance standards for PACS systems or components have been issued under the authority of Section 514. IntelliSpace Cardiovascular was tested in accordance with Philips verification and validation processes. Quality Assurance measures were applied to the system design and development, including:
- Risk Analysis
- Product Specifications
- . Design Reviews
- Verification & Validations .
#### Summary of Clinical Testing
The subject of this premarket submission, IntelliSpace Cardiovascular software did not require clinical studies to support equivalence.
#### Conclusions drawn from the Non-clinical and Clinical testing
Verification and Validation activities required to establish performance and functionality of IntelliSpace Cardiovascular were performed. Testing involved system level tests, performance tests, and safety testing from Risk Analysis. Testing performed demonstrated the IntelliSpace Cardiovascular meets all defined functionality requirements and performance claims.
#### VIII.Conclusion
The IntelliSpace Cardiovascular is substantially equivalent to the predicate device Xcelera (K061995) in terms of design features, fundamental scientific technology, intended use, and safety and effectiveness. Additionally, substantial equivalence was demonstrated with non-clinical performance testing, which complied with the requirements specified in the international and FDA-recognized consensus standards. The non-clinical performance tests provided in this 510(k) premarket notification demonstrate that the subject IntelliSpace Cardiovascular is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.