GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System.
Applicant
Gpc Medical, Ltd.
Product Code
HRS · Orthopedic
Decision Date
May 17, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System are provided nonsterile. GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System are indicated for treating fractures of various bones including the clavicle, pelvis, scapula, long bone ( humerus, ulna, radius, femur, tibia and fibula), and small bone (metacarpals, metatarsals, phalanges). The system is indicated for use in adult patients only. All implants are for single use only.
Device Story
System consists of metallic bone plates and screws for osteosynthesis; used for fracture fixation and reconstruction. Inputs: patient anatomy/fracture site; device transforms inputs via mechanical stabilization of bone fragments. Output: rigid fixation of bone. Used in clinical/surgical settings by orthopedic surgeons. Healthcare providers use output to stabilize fractures, promoting bone healing. Benefits: restoration of skeletal alignment and stability.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including engineering analysis of mechanical properties (bending moment, stiffness, pull-out strength, torque-to-failure) and material compliance with ASTM standards.
Technological Characteristics
Materials: Stainless Steel and Titanium (ASTM F136, F138, F139). Components: Locking/non-locking bone plates and screws. Dimensions: Plates (2-22 holes), Screws (2.4mm-6.5mm diameter, 20mm-150mm length). Principle: Mechanical fixation/osteosynthesis. Supplied non-sterile; requires sterilization by user.
Indications for Use
Indicated for treating fractures of the clavicle, pelvis, scapula, long bones (humerus, ulna, radius, femur, tibia, fibula), and small bones (metacarpals, metatarsals, phalanges) in adult patients.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes 3.5 and 4.5mm Locking Compression Plate System (K082807)
Synthes 4.5 mm LCP Straight Reconstruction Plate (K051986)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, representing the department's focus on health and human well-being. The profiles are depicted in a simple, abstract manner.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 17, 2016
GPC Medical Limited Mr. Vikas Narang Director-Exports M-Block, DDA LSC, Vikas Puri New Delhi, 110018 INDIA
Re: K153716 Trade/Device Name: GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: April 8, 2016 Received: April 11, 2016
Dear Mr. Narang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
### Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known) K153716
Device Name
GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System.
#### Indications for Use (Describe)
GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System are indicated for treating fractures of various bones including the clavic, scapula, long bone ( humerus, ulna, radius, femur, tibia and fibula), and small bone (metacarpals, metatarsals, phalanges).
The system is indicated for use in adult patients only. All implants are for single use only.
#### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Premarket Notification 510(k) Summary as required by Section 807.92
#### General Company Information as required by 807:92 (a)
(a.1) The submitter's name, address, telephone number, a contact person, and the date the summary was prepared
#### Submitter's Name:
GPC Medical Ltd.
Address:
#### Office:
M Block, DDA LSC, Vikas Puri, New Delhi 110018 India
Factory: A-82, Sector A-4, Tronica City, Loni, Ghaziabad, Uttar Pradesh 201102
| Contact Person Name: | Mr. Vikas Narang |
|----------------------|------------------|
| Title: | Director-Exports |
| Phone Number: | +91-9810638797 |
| Dated: | 21-12-2015 |
### This is a bundled submission.
Throughout the submission there is a mention of GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System that represents the range of products covered under this 510(k) submission.
{4}------------------------------------------------
a.2: The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
Proprietary Name:
- . GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System.
Common or Usual Name:
- Orthopaedic Bone Plates
- Orthopaedic Bone Screws
### Classification Name:
- PLATES, FIXATION, BONE
- · SCREWS, FIXATION, BONE
| Product Code: | HRS, HWC |
|--------------------|----------------------------------------------------------------------------------------------------------------------------|
| Device Class: | II |
| Review Pane: | Orthopaedic |
| Regulation Number: | 21 CFR 888.3030 and 21 CFR 888.3040 |
| Variants/Types: | GPC BRAND Locking Bone Plates and Screws Osteosynthesis<br>Plating System are further subdivided into following categories |
| S. No. | Category | Types |
|--------|-----------------------|------------------------------------|
| 01 | Large Fragment Plates | Locking Version |
| 02 | Small Fragment Plates | Locking Version |
| 03 | Mini Fragment Plates | Locking Version |
| 04 | Bone Screws | Locking and Non-Locking<br>Version |
{5}------------------------------------------------
Further Description of GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System.
Generally there are following types of bone plates used with cortical (cortex) and Locking Screws. These bone plates are generally designed on the basis of the boñe contour and anatomy.
| S. No. | Type | Subtype |
|--------|-----------------------------|-----------------------------------------------------------------------------------|
| 01 | Small and Large<br>Fragment | T-Shaped, Hook Angled Plates. Reconstructions plates |
| 02 | Dynamic compression | Small DCP, Narrow DCP, Lengthening Narrow, Broad<br>DCP, Lengthening Broad Plates |
| 03 | Mini Fragment | T-Shaped |
{6}------------------------------------------------
### A3) Identification of the Predicate Device:
Following are the predicate device 510(k) with which we are declaring substantial equivalence:
Following is the range of variants covered with their corresponding predicate devices.
| S. No. | Subject Device | Predicate Device |
|--------|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| 01 | 4.5mm Narrow Locking Plate | Synthes 3.5 and 4.5mm Locking<br>Compression Plate System (K082807) |
| 02 | 4.5mm Broad Locking Plate | Synthes 3.5 and 4.5mm Locking<br>Compression Plate System (K082807) |
| 03 | 4.5mm/5.0mm Broad Curved Locking Plate | Synthes 3.5 mm and 4.5 mm<br>Curved Narrow and Broad Locking<br>Compression Plates (K082807) |
| 04 | 4.5mm Reconstruction Locking Plate | Synthes 4.5 mm LCP Straight<br>Reconstruction Plate (K051986) |
| 05 | Proximal Femur Locking Plate | Synthes (USA) LCP Proximal Femur<br>Plate (K030858) |
| 06 | Locking Lateral Proximal Tibia Plate, Left<br>and Right | Synthes LCP Proximal Tibia<br>Plate (K023802, K030597, K983787) |
| 07 | 4.5mm Distal Femur Locking Plate, Left and<br>Right | Synthes LCP Distal Femur<br>Plates (K062564) |
| 08 | Locking T-Plate 3.5mm, Oblique Angled, Left<br>and Right | Synthes Small Fragment Dynamic<br>Compression Locking( DCL) System<br>(K000684) |
| 09 | Small Dynamic Compression Plate 3.5mm with<br>LC Under Cuts | Synthes 3.5 (USA) Locking<br>Compression Plate System (K082807) |
| 10 | 3.5mm Small 'T' Plate – 3 Head Holes | Synthes Small Fragment Dynamic<br>Compression Locking( DCL)<br>System(K000684) |
| 11 | 3.5mm Small 'T' Plate – 4 Head Holes | Synthes Small Fragment Dynamic<br>Compression Locking(DCL)<br>System(K000684) |
| 12 | Proximal Tibia Locking Plate, Left and Right | Synthes 3.5mm Titanium LCP<br>Proximal Tibia Plate (K030597) OR<br>Synthes LCP Proximal Tibia Plates<br>(K011978) |
| 13 | Locking Anterolateral Tibia Distal Plate 3.5mm | Synthes (USA) LCP Anterolateral<br>Distal Tibia Plate 3.5 (K092812) |
| 14 | Locking Clavicle Hook Plate 3.5mm, Left and Right | Synthes Clavicle Hook Plate<br>(K061753) |
| 15 | 3.5mm Proximal Humerus Locking Plate | Proximal humerus plate long<br>Synthes K041860<br><br>K062494 EBI Optilock Upper<br>Extremity Plating System |
| 16 | Distal Humerus Locking Plate, Size 2.7/3.5mm,<br>Left and Right | Synthes 3.5mm LCP Distal<br>Humerus System (K033995) |
| 17 | Distal Humerus Locking Plate, 2.7/3.5mm with<br>Lateral Support, Left and Right | Synthes 3.5mm LCP Distal<br>Humerus System (K033995) |
| 18 | Locking Medial Distal Tibia Plate without Tab,<br>Left and Right | Synthes Medial Distal Tibia Plate<br>(K001945) |
| 19 | Locking Philous Plate | Synthes (USA) LCP® Proximal<br>Humerus Plates, Long (K041860) |
| 20 | 'T' Distal Radius 2.4mm Locking Plate | K982732 or K012114 (Synthes locking<br>distal radius plating system) |
| 21 | Locking Distal Radius Buttress Plate 2.4mm | Synthes (USA) LCP Radial Head Plating<br>System (K040777) |
| 22 | Self Tapping Locking Screw 2.4mm | Synthes 2.4mm Titanium locking<br>screw (K033975)<br><br>K051567 APTUS Titanium Fixation<br>System |
| 23 | Self Tapping Locking Screw 2.7mm | Synthes Small Fragment Dynamic<br>Compression Locking (DCL) System<br>(K000684)<br><br>K051567 APTUS Titanium Fixation<br>System |
| 24 | Self Tapping Locking Screw 3.5mm | Synthes Small Fragment Dynamic<br>Compression Locking (DCL) System<br>(K000684)<br><br>K051567 APTUS Titanium Fixation<br>System |
| 25 | Self Tapping Locking Screw 5.0mm | Synthes Large Fragment Dynamic<br>Compression Locking (DCL) System<br>(K000682)<br>K051567 APTUS Titanium Fixation<br>System |
| 26 | 6.5mm Self Tapping Locking Cannulated<br>Screw | Synthes Cannulated Screw System<br>K021932<br>K051567 APTUS Titanium Fixation<br>System |
{7}------------------------------------------------
{8}------------------------------------------------
### a4). A description of the device that is the subject of the premarket notification submission, such as might be found in the labeling or promotional material for the device
### Device Description:
GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System consists of various shape and sizes of plates featuring compression and locking holes, full-threaded-cortical, locking self-tapping screws.
The plates and screws are fabricated from Stainless Steel and Titanium.
GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System contains several models based on the size of the device and application site such as fixation/reconstruction of small fragment bones, forefoot, mid-foot, rear-foot, ankle, or other bones appropriate for the size of the device. The plate implants are in many models available: like, Reconstruction Plates, T-Plates, Anatomical Plates and Clavicle Hook Plates.
These all are mainly divided into
- Large Fragment Plates
- . Small Fragment Plates
- Mini Fragment Plates ●
The number of the holes varies from 2 to 22.
The Locking screw implants are in corresponding diameter ranges from 2.4mm, 3.5mm, 5.0mm and 6.5mm diameters with lengths varying as per the requirements and minimum length: 20 mm to maximum length 150mm
The Non-locking screw implants are in 3.5mm and 4.5mm diameter and length range from 10mm to 140mm. The non locking screws 3.5mm and 4.5mm were cleared previously in 510k
Rev: 02 SECTION 5 Date: 27.04.2016 PAGE 6 OF 10
{9}------------------------------------------------
number K092493
These implants are supplied non-sterile, the products have to be sterilized prior to use.
{10}------------------------------------------------
### A5). (5) A statement of the intended use of the device
### Indications for Use:
· GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System are provided nonsterile.
GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System are indicated for treating fractures of various bones including the clavicle, pelvis, scapula, long bone ( humerus, ulna, radius, femur, tibia and fibula), and small bone (metacarpals, metatarsals, phalanges).
The system is indicated for use in adult patients only. All implants are for single use only.
### a6). Summary of Technological Characteristics as compared to the predicate devices:
### Substantial equivalence including comparison with predicate devices
A comparison between the GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System and predicate devices has been performed which has resulted in demonstration of similarities in dimensional and performance criteria.
| S. No. | Characteristics | Predicate Device Versus New<br>Device (GPC Brand) | Remarks |
|--------|-----------------------------|----------------------------------------------------------------------------------------|------------|
| 01 | Indications for use | Similar intended use in New Device<br>and Predicate device | Equivalent |
| 02 | Material | Same material used in New Device and<br>Predicate device | Equivalent |
| 03 | Performance<br>Standards | Same performance standards used in both<br>New Device as well as predicate device | Equivalent |
| 04 | Sterilization | Same method of sterilization used in<br>both New Device as well as Predicate<br>device | Equivalent |
| 05 | Dimensional<br>Verification | Same dimensions found in both New<br>Device as well as Predicate device | Equivalent |
### Following is the summary of parameters in which the comparison has been verified:
{11}------------------------------------------------
### b1). Discussion on the non-clinical testing performed
Following are the applicable product standards considered for non-clinical standards
Material A: Standards
B: Performance Standards
#### Material A: Standards:
The material standards are the essential part to be complied to first, as it is the basis of manufacturing metallic surgical implants.
We have complied to following material standards
- 1. ASTM F 136: Standard specification for wrought Titanium-6Aluminium-4Vanadium ELI (Extra low interstitial) Alloy for surgical implant applications.
- 2. ASTM F 138: Standard Specification for Wrought 18 chromium-14Nickel-2.5Molybdenum stainless steel bar and wire for surgical implants.
- 3. ASTM F 139: Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Sheet and Strip for Surgical Implants
We have verified the purchased material compliance to these standards and copies of the relevant test results were provided in the submission.
The performance of the plates was demonstrated using an engineering analysis of bending moment and bending stiffness based on material properties and the plate shaft dimensions.
The performance of the screws was demonstrated using an engineering analysis of pull-out strength and torque-to-failure based on material properties and the screw dimensions.
{12}------------------------------------------------
### b2). Discussion on the clinical evaluation referenced and relied upon:
GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System are of similar design and pattern as well as similar intended use. Clinical information was not necessary to demonstrate substantial equivalence.
### CONCLUSION:
### General, Safety and Performance conclusion:
From the available data available we can justify that the GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System are as safe, as effective and perform as same indications for use as that of already marketed predicate devices identified in a3. of 510(k) summary.
Hence these devices can be considered safe and effective for their intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.