The MaxO2ME oxygen monitor is intended for continuous monitoring of the concentration of oxygen being delivered to patients ranging from newborns to adults. It can be used in the pre-hospital, hospital and sub-acute settings. The MaxO2ME is not intended as a life supporting device.
Device Story
MaxO2ME is a handheld oxygen analyzer/monitor measuring 0-100% oxygen concentration. Device utilizes a MAX-550E galvanic fuel cell sensor to output voltage proportional to oxygen concentration; microprocessor processes signal for display and alarm functions. Operated in pre-hospital, hospital, and sub-acute settings by clinicians. Features include user-adjustable high/low oxygen alarms with visual (flashing LED) and audible indicators, self-diagnostic circuitry, and calibration reminder. Output is displayed on an LCD screen; healthcare providers use this data to monitor delivered oxygen levels. Device is not life-supporting; provides continuous monitoring to ensure oxygen delivery remains within set clinical parameters.
Clinical Evidence
No human clinical testing was performed. Bench testing included ISO 80601-2-55 (respiratory gas monitors), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), alarm performance (IEC 60601-1-8), sensor performance, gas leakage, interfering gas effects, temperature compensation, and measurement accuracy. All tests met acceptance criteria.
Technological Characteristics
Galvanic fuel cell (MAX-550F) oxygen sensor; 4 AA alkaline battery power; LCD display with backlight; 15mm T-adapter interface. Dimensions: 3.6" x 5.8" x 1.2". Operates 15°C–40°C. Microprocessor-based circuitry for signal processing, self-diagnostics, and alarm management. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-8.
Indications for Use
Indicated for continuous monitoring of delivered oxygen concentration in patients ranging from newborns to adults in pre-hospital, hospital, and sub-acute settings. Not for life support.
Regulatory Classification
Identification
An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 10, 2016
Maxtec, LLC C/O Paul Dryden Consultant 2305 South 1070 West Salt Lake City, Utah 84119
Re: K153659
Trade/Device Name: MaxO2ME Regulation Number: 21 CFR 868.1720 Regulation Name: Oxygen Gas Analyzer Regulatory Class: Class II Product Code: CCL Dated: May 12, 2016 Received: May 13, 2016
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K153659
#### Device Name
#### MaxO2ME
#### Indications for Use (Describe)
The MaxO2ME oxygen monitor is intended for continuous monitoring of the concentration of oxygen being delivered to patients ranging from newborns to adults. It can be used in the pre-hospital, and subacute settings. The MaxO2ME is not intended as a life supporting device.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740
**ELOW THIS LINE – CONTI**
NOT SEND YOUR COMPLETED FORM TO THE DDA STAFF EMAIL ADDRESS BELOW.
**ORM FDA 3881 (1/14)**
{3}------------------------------------------------
#### 510(k) Summary June 10, 2016 Page 1 of 5
| Official Contact: | Bruce Brierley<br>President<br>Maxtec LLC<br>2305 South 1070 West<br>Salt Lake City, Utah 84119<br>Tel – 385-549-8070 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Proprietary or Trade Name: | MaxO2ME |
| Common/Usual Name: | Oxygen analyzer |
| Classification Name: | 21CFR 868.1720<br>Class II<br>CCL |
| Predicate Device: | Precision PM5900 - K063096 |
| Reference Device: | MiniOX - K961644 |
#### Device Description:
The MaxO-ME is a handheld oxygen analyzer/monitor capable of measuring the oxygen concentration from 0% to 100% in a sample gas. A MAX-550E oxygen sensor outputs a voltage which is used by the Max O2ME to determine the concentration of oxygen based on a calibration at room air or 100% oxygen. The MaxO2ME contains alarms that can be controlled by the user to set a maximum or minimum allowable oxygen concentration.
Device Features:
- Oxygen sensor of approximately 1,500,000 O2 percent hours and / or 2 years.
- External probe with 10 ft., extendable cable and diverter fitting for standard 15 mm "T" adapter. ●
- Operation using 4 AA alkaline batteries (4 x 1.5 volts) for approximately 5,000 hours of ● performance with typical use.
- Oxygen-specific, galvanic sensor that achieves 90% of final value in approximately 15 seconds at ● room temperature.
- Self-diagnostic check of analog and microprocessor circuitry. ●
- Low battery indication.
- Calibration reminder timer that alerts the operator, using a calibration icon on the LCD display, to ● perform a unit calibration.
- . Adjustable high-level and low-level alarming capability with flashing LED and audible indication of alarm conditions.
- Smart high-low alarm setting to help adjust alarm settings quickly.
- . Back-light display with auto ambient light level detection.
- Sleep Mode operation to extend battery life. ●
#### Indications for Use:
The MaxO2ME oxygen monitor is intended for continuous monitoring of the concentration of oxygen being delivered to patients ranging from newborns to adults. It can be used in the pre-hospital, hospital and sub-acute settings. The MaxO2ME is not intended as a life supporting device.
{4}------------------------------------------------
#### Patient Population
The MaxO2ME may be used on equipment where one desires to measure and monitor the delivered oxygen concentration. This is independent of a patient population.
#### Contraindications
There are no contraindications.
#### Environments of Use
Pre-hospital, hospital and sub-acute settings
#### Substantial Equivalence
This discusses how one can find the MaxO2ME substantially equivalent to the predicate Precision Medical PM5900 (K063096).
#### Indications for Use
Table 1 outlines the indications for use for both devices and one can see that they are similar, namely oxygen monitor is intended for continuous monitoring of the concentration of oxygen being delivered to patients, not intended as a life supporting device.
Discussion: One can find the proposed device substantially equivalent to the predicate Precision Medical PM5900 (K063096). There are no differences which raise any new substantial equivalence concerns.
Environment of Use -Oxygen monitors have commonly been used in pre-hospital and healthcare settings. The reference to the types of equipment that the predicate discloses supports pre-hospital and healthcare setting.
Discussion: One can find the proposed device substantially equivalent to the predicate Precision Medical PM5900 (K063096). There are no differences which raise any new substantial equivalence concerns.
Population - The predicate submission does not disclose details on patient population but does include incubators which would suggest newborns.
Discussion: Based upon the available information one can find the proposed device substantially equivalent to the predicate Precision Medical PM5900 (K063096). There are no differences which raise any new substantial equivalence concerns.
Performance Specifications - The following is a list of the differences between the proposed device and the predicate.
- . Response time is 15 seconds vs. 12 seconds for the predicate
- o The reference device MiniOX (K961644) has a response time of 20-30 seconds
- . Warm-up time is 3 seconds less for the predicate (12 vs. 15 seconds)
- Operating temperature range is wide by 10 degrees for the predicate vs. the proposed. ●
- Battery life is about 50% less than the proposed device (2000 hours vs. 5000 hours) ●
- The Sensor Life is less for the predicate vs. the proposed ●
- High Oxygen Alarm range is smaller by 3 % points (19-99% vs. 16-100%)
- o The reference device MiniOX (K961644) has the identical high alarm range
- . Size and weight are slightly different
#### Discussion:
While the subject device has some differences when compared to the Response Time and High Alarm range, the reference device, MiniOX (K961644), which has similar intended use, population and environment has similar Response time and High alarm, which supports that the differences do not raise any new substantial equivalence concerns. The other differences, battery life, warm-up time, sensor
{5}------------------------------------------------
#### 510(k) Summary June 10, 2016
### Page 3 of 5
life and physical size are similar and the defenses are improvements vs. the predicate. As such they would not raise any new substantial equivalence concerns. One can find the proposed device substantially equivalent to the predicate Precision Medical PM5900 (K063096) and reference device MiniOX (K961644).
#### Non-clinical Testing
#### Performance Testing
We performed a number of tests to demonstrate that the proposed device performed as intended.
- . ISO 80601-2-55 Performance of respiratory gas monitors
- ISTA2A Shipping Validation Test Report ●
- Sensor performance Test Report ●
- Gas leakage Test Report ●
- . Interfering gas effects Test Report
- . Temperature compensation Test Report
- Drift of measurement accuracy Test Report
- MaxO2ME Operating and Storage Environment Report
- Device Cleaning Report and Disinfection Test
- . Measurement accuracy Test Report
- IEC 60601-1 Electrical safety ●
- IEC 60601-1-2 - EMC
- IEC 60601-1-8 - Alarms
- Shelf-life / Real-time ●
Discussion: In all cases the proposed device passed or meets the acceptance criteria. One can find the proposed device substantially equivalent to the predicate Precision Medical PM5900 (K063096).
Biocompatibility - The materials that are in the gas pathway for the oxygen sensor. "Tee" adapter and diverter, are identical to other Maxtec supplied or cleared products which has similar intended use, population, environment of use and type of patient contact. As such no biocompatibility testing was performed.
Per G95-1 and ISO 10993-1:2009, these materials would be considered as:
- Externally communicating, Tissue contact, and Duration of Use prolonged (> 24 hours, < 30 ● days)
Animal - No animal testing was performed.
Clinical - No human clinical testing was performed.
#### Substantial Equivalence Conclusion-
It is sponsor's opinion that the MaxO2ME oxygen analyzer based upon the comparative testing is substantially equivalent to the predicate device.
{6}------------------------------------------------
# 510(k) Summary
| Attributes | Proposed<br>MaxO2ME | Predicate<br>Precision Medical – PM5900 (K063096) |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The MaxO2ME oxygen monitor is intended for continuous<br>monitoring of the concentration of oxygen being delivered to<br>patients ranging from newborns to adults. It can be used in the<br>pre-hospital, hospital and sub-acute settings. The MaxO2ME is<br>not intended as a life supporting device. | Intended to measure the concentration of oxygen being<br>delivered to the patient. The oxygen monitor is not intended as<br>a life supporting device |
| Environments of Use | Pre-hospital, hospital and sub-acute settings | Variety of medical applications such as anesthesiology (e.g.,<br>anesthesia machines), Respiratory devices (e.g., respirators,<br>ventilators, pediatric incubators), and oxygen therapy (e.g.,<br>oxygen tents). Can be considered Pre-hospital, hospital and<br>sub-acute settings |
| Patient Population | The MaxO2ME may be used on equipment where one desires to<br>measure and monitor the delivered oxygen concentration. This<br>is independent of a patient population. | Not specific but includes reference to pediatrics and incubators<br>which would imply newborns and older |
| Measurement Range | 0.0 to 100% | 0.0 to 100% |
| Resolution | 0.1% | 0.1% |
| Accuracy and Linearity | ±1% of full scale at constant temperature, RH and pressure<br>when calibrated at fill scale | ±1% of full scale at constant temperature, RH and pressure<br>when calibrated at fill scale |
| Total Accuracy | ±3% Actual oxygen level over full operating temperature range | ±3% Actual oxygen level over full operating temperature range |
| Response Time | 90% of final value in approx. 15 seconds at 23°C | 90% of final value in 12 seconds at 25°C<br>Reference device - MiniOX – K961644<br>90% in 20 to 30 seconds |
| Warm-up Time | None required | None required |
| Operating Temperature | 15°C – 40°C (59°F – 104°F) | 10°C – 45°C (50°F – 113°F) |
| Storage Temperature | -15°C – 50°C (5°F – 122°F) | -15°C – 50°C (5°F – 122°F) |
| Atmospheric Pressure | 800 – 1012 mBars | Up to 8,000 ft. |
| Humidity | 0-95% (non-condensing) | 0-95% (non-condensing) |
| Power requirements | 4 – AA Alkaline batteries | 4 – AA Alkaline batteries |
| Battery Life | Approx. 5000 hours, typical use | Approx. 1,500 – 2,000 hours, typical use |
| Low Battery Indications | “LOWBAT” icon on LCD display | Icon on LCD display |
| Sensor Type | Maxtec MAX-550F galvanic fuel cell | Galvanic fuel cell |
| Attributes | Proposed<br>MaxO2ME | Predicate<br>Precision Medical – PM5900 (K063096) |
| Expected Sensor Life | > 1,500,000 % O2 Hours, over 2 years typical application | > 1,000,000% O2 Hours |
| Alarm Systems | High/Low alarms, flashing yellow LEDs<br>Nominal 975 Hz audio buzzer (IEC 60601-1-8) | High/Low alarms, flashing yellow LEDs |
| Low Oxygen Alarm Range | 15% - 99% (>1% lower than high alarm) | 15% - 99% (>1% lower than high alarm)<br>Reference device - MiniOX – K961644<br>15 – 99% |
| High Oxygen Alarm Range | 16% - 100% (>1% higher than low alarm) | 18% - 99% (>1% higher than low alarm) inconsistent<br>Summary shows 19-99%, literature shows 18-100%<br>Reference device - MiniOX – K961644<br>16 - 100% |
| Accuracy | Exact to display alarm value | Exact to display alarm value |
| Materials | The materials in the gas pathway are considered as Externally<br>communicating, Tissue contact, and Duration of Use –<br>prolonged (> 24 hours, < 30 days)<br><br>The components of the sensor and diffuser are identical to the<br>Maxtec components cleared under K131252 and Tee adapter is<br>identical to Envitec K122290.<br>These components have similar intended use, population,<br>environment of use and type of patient contact | |
| Dimensions | 3.6"(W) x 5.8"(H)x1.2"(D) | 3.6"(W) x 5.4"(H)x1.7"(D) |
| Weight | Approx. 0.89 lbs. | Approx. 1.11 lbs. |
| Cable length | 9 ft. | 10 ft. |
| Accessories | Diverter<br>Tee adapter (15 mmm x 22 mm fittings)<br>Mounting brackets<br>DC power adapter | Diverter<br>Tee adapter (15 mmm x 22 mm fittings)<br>Mounting brackets<br>DC power adapter |
{7}------------------------------------------------
## 510(k) Summary June 10, 2016
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.