The Puritan Bennett™ 980 Ventilator Systém is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support, delivered invasively or noninvasively, to patients who require the following types of ventilator support: • Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or noninvasively (via mask or nasal prongs) • Assist/ Control, SIMV or Spontaneous modes of ventilation.
Device Story
Dual-microprocessor-based, touchscreen-controlled critical care ventilator; provides invasive/non-invasive continuous positive pressure ventilation. Inputs include patient respiratory signals; system processes data to deliver precise, sensitive breaths; supports Assist/Control, SIMV, and Spontaneous modes. Used in hospitals/institutions by clinicians; features advanced technology to support patient-ventilator synchrony and amplify spontaneous breathing efforts. Output consists of controlled mechanical ventilation; clinicians monitor via touchscreen interface to adjust treatment. Benefits include improved patient comfort, precise breath delivery, and optimized ventilation outcomes for critically ill patients.
Clinical Evidence
No clinical testing was necessary to show substantial equivalence. Evidence consists of extensive non-clinical verification and validation, including reliability, human factors, usability, labeling, breathing system performance, alarms, software verification, hardware safety (IEC 60601-1, IEC 60601-2, ISO 80601), electromagnetic compliance, electrical verification, mechanical subsystems, environmental testing, packaging, and biocompatibility (ISO 10993-1).
Technological Characteristics
Dual-microprocessor-based architecture; touchscreen interface. Supports invasive/non-invasive ventilation for patients ≥0.3kg. Features include proximal flow sensing and standby modes. Compliance with IEC 60601-1, IEC 60601-2, ISO 80601, and ISO 10993-1. Software-controlled breath delivery and alarm management.
Indications for Use
Indicated for neonatal, pediatric, and adult patients (minimum weight 0.3kg) requiring invasive or non-invasive continuous positive pressure ventilatory support in hospital or intra-hospital transport settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
840 Ventilator System with Expanded NeoMode Option (K092847)
{0}------------------------------------------------
## K231252
# FEB 2 4 2014
## 510(k) Summary of Substantial Equivalence pursuant to 21 CFR 807.92
Date: May 1st, 2013
#### General Company Information
Submitted by:
Contact:
### Manufacturing Site:
Covidien 6135 Gunbarrel Avenue Boulder, CO 80301
Michael Aymami Director Regulatory Affairs michael.aymami(@covidien.com (303) 305-2481
Covidien, formerly NELLCOR PURITAN BENNETT Ireland, Ltd Mervue, Galway, Ireland
Registration number 8020893
#### Device Trade Name
Puritan Bennett™ 980 Series Ventilator System
## Device Classification Name
Ventilator, continuous, facility use.
#### Device Classification Regulation
The Device Regulation regarding the device is a class II per 21 CFR 868.5895.
#### Device Product Code, CFR Section. Panel
- > Product Code: CBK
- > CFR Section: 21 CFR§ 868.5895
- > Device Panel: Anesthesiology
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "COVIDIEN" in bold, black letters. To the left of the word is a black square with a white cross in the center. The cross is made up of three horizontal and three vertical bars.
# Predicate Devices:
The Puritan Bennett™ 980 Series Ventilator System is substantially equivalent to the previous 840 Ventilator System from Covidien and the AVEA Ventilator from CareFusion:
- K092847 840 Ventilator System with Expanded NeoMode Option .
- K103211 AVEA Ventilator .
# Device Description:
The Puritan Bennett 980 Series Ventilator System is a dual-microprocessor-based, touchscreen controlled, critical care ventilator intended to provide continuous ventilation for neonate to adult patients who require either invasive ventilation or non-invasive ventilation. It can be used in hospitals and institutions and for intra -hospital transport applications with access to the appropriate services.
The ventilator system offers features that improve patient comfort while delivering sensitive, precise breaths to critically ill patients of any age. The product ventilates Neonatal, Pediatric, and Adult patients with predicted body weights from 0.3 kg, and with tidal volumes for mandatory volume controlled breaths from 2 mL to 2500 mL.
The ventilator is designed to help clinicians improve ventilation outcomes and quality of care for patients, and can be customized with advanced technology focused on supporting patient ventilator synchrony. A full suite of software options, safety features and accessories to fit a variety of patient needs, from infant to adult are available. Software options assist clinicians in assessing the best treatment approach for their patients at any given time by amplifying the patient's own spontaneous effort to breathe.
# Indications for Use:
The Puritan Bennett™ 980 Series Ventilator System is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3kg (0.66lb). It is suitable for service in hospitals (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support, delivered invasively or noninvasively, to patients who require the following types of ventilator support:
- . Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs)
- Assist/ Control, SIMV or Spontaneous modes of ventilation o
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Covidien. The logo consists of a black and white cross symbol on the left, followed by the word "COVIDIEN" in bold, black letters. The cross symbol is made up of a white cross inside a black square.
## Substantial Equivalence:
The Puritan Bennett 980 Series Ventilator System is a continuous ventilator, and is substantially equivalent to the currently marketed and cleared Puritan Bennett 840 ventilator system (K092847) and the AVEA ventilator (K103211) from CareFusion.
The Puritan Bennett 840 Ventilator system has been marketed in United States for over 10 years and has been the subject of several previous premarket notifications.
Testing of the Puritan Bennett 980 Series Ventilator System has been performed to establish substantial equivalence to the listed predicate devices.
## Comparison of Technological Characteristics:
The Puritan Bennett 980 Series Ventilator System is substantially equivalent to the Puritan Bennett 840 Ventilator and the AVEA ventilator. All three systems are under the same regulation and FDA product code and there are no significant differences in the indications for use.
All three devices are intended for neonatal, pediatric and adult patient populations in hospital environments have similar features and meet industry standards and performance.
The proximal flow and stand by features of the PB980 are present in the predicate AVEA device. All other PB980 features are either included in the predicate PB840 or simply represent new data or alarm display or entry systems which do not affect indications for use.
## Performance Testing:
Extensive non-clinical testing was done to establish substantial equivalence of the new Puritan Bennett™ 980 Series Ventilator System to the noted predicate devices.
An extensive Verification and Validation test plan was executed to verify the design of the Puritan Bennett 980 Series Ventilator System. Over 50 units were built and made available for different aspects of Verification and Validation testing.
Verification and validation testing included:
- Reliability .
- Human Factors and Usability .
- Labeling validation .
- Breathing System .
- Alarms .
- . Controls Design Verification
- Software Verification .
- Hardware Safety Compliance (including IEC 60601-1, IEC 60601-2 and ISO .
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "COVIDIEN" in bold, black letters. To the left of the word is a black square with a white cross in the center. The cross is made up of two intersecting rectangles, and there are also some other white shapes within the square.
80601)
- Electromagnetic Compliance Testing ●
- Electrical Verification ●
- Mechanical Subsystems ●
- Environmental Testing (Temperature/Humidity, Altitude, 0 Vibration, Cleanability, etc.)
- Packaging Verification o
- Biocompatibility Risk Analysis in conformance with ISO 10993-1 o
The results of non-clinical testing demonstrated that the Puritan Bennett 980 Series Ventilator System is substantially equivalent to the predicate devices. No clinical testing was necessary to show substantial equivalence.
#### Conclusion:
Verification and validation activities were conducted to establish the performance characteristics of the Puritan Bennett 980 Ventilator System. All testing demonstrated that the Puritan Bennett 980 Ventilator System met required design verification criteria and has acceptable performance when used in accordance with its labeling; the device is therefore suitable for its intended use. The device's intended use, operating principles, ventilation modes and performance parameters are comparable to the referenced predicate devices. Therefore, the Puritan Bennett 980 Ventilator System is substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2014
Covidien Mr. Michael Aymami Director, Regulatory Affairs 6135 Gunbarrel Avenue Boulder, CO 80301
Re: K131252
Trade/Device Name: Puritan Bennett™ 980 Ventilator System Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: January 15, 2014 Received: January 16, 2014
Dear Mr. Aymami:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part
807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/5/Picture/7 description: The image shows a logo with the text "Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID". The logo is in black and white and has a stylized design. The word "FOR" is also present in the image.
Erin I. Keith. M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Section 7
# Indications for Use
510(k) Number (if known): K131252
Device Name: Puritan Bennett 980 Series Ventilator System
Indications for Use:
The Puritan Bennett™ 980 Ventilator Systém is designed for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilatory support, delivered invasively or noninvasively, to patients who require the following types of ventilator support:
· Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or noninvasively (via mask or nasal prongs)
· Assist/ Control, SIMV or Spontaneous modes of ventilation.
Prescription Use _ X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
> (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
K131252
Image /page/6/Picture/12 description: The image shows a logo with the letters "FDA" in a stylized, geometric design. The letters are bold and outlined, giving them a three-dimensional appearance. The logo is black and white, with the letters standing out against the white background.
Anya C. Harry -S 2014.02.21 17:40:04 -05,00،
18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.