Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with nondiseased eyes. Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity. The lenses are to be prescribed for single use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Device Story
Soft (hydrophilic) contact lenses made of nelfilcon A polymer; 69% water content. Designed for daily disposable wear; single use only. Manufactured via Lightstream® Technology (full mold cast with integrated print step). Available in spherical, toric, and multifocal designs. Used by patients for vision correction; prescribed by eye care professionals. Lenses supplied sterile in blister packs with buffered saline. Visibility tint (phthalocyanine blue) and print inks (phthalocyanine blue/green) added for handling. Device provides optical correction for refractive errors; improves visual acuity for patients with myopia, hyperopia, astigmatism, or presbyopia.
Clinical Evidence
No clinical data provided. Substantial equivalence established via non-clinical process validation and comparison of physical/chemical properties to the predicate device.
Technological Characteristics
Material: 69% water, 31% nelfilcon A (polyvinyl alcohol partially acetalized with N-formylmethyl acrylamide). FDA Group 2 (high water, nonionic). Refractive index 1.38. Oxygen permeability 26 barrer (Dk-Polarographic method). Sterilization: Steam (autoclave). Packaging: Blister pack with phosphate-acetate buffered saline and Poloxamer 108.
Indications for Use
Indicated for daily wear optical correction of refractive ametropia (myopia, hyperopia, astigmatism) in non-aphakic, non-diseased eyes. Progressives indicated for myopia/hyperopia and/or presbyopia with reading addition ≤ +3.00D and astigmatism ≤ 2.00D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
DAILIES® AquaComfort Plus® family of (nelfilcon A) soft contact lenses
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 20, 2016
Alcon Laboratories, Inc. Dr. Andreas Friese Principal Regulatory Specialist 6201 South Freeway Fort Worth, TX 76134-2099
Re: K153643
> Trade/Device Name: Focus® DAILIES®, Focus® DAILIES® Toric, Focus® DAILIES® Progressives Soft (hydrophilic) Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens. Regulatory Class: Class II Product Code: LPL/MVN Dated: December 18, 2015 Received: December 21, 2015
Dear Dr. Friese:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR
regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K153643
#### Device Name
Focus® DAILIES®, Focus® DAILIES® Toric, Focus® DAILIES® Progressives
#### Indications for Use (Describe)
Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with nondiseased eyes.
Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
The lenses are to be prescribed for single use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Ink Shelf Life Change
{3}------------------------------------------------
# 510(k) Summary
# 1. Submitter Information:
| Company: | Alcon Laboratories, Inc.<br>6201 South Freeway<br>Fort Worth, TX USA 76134-2099 |
|-------------------------|---------------------------------------------------------------------------------|
| Primary Contact Person: | Dr. Andreas Friese<br>Principal Regulatory Specialist |
| Telephone: | +49 6022 240 514 |
| Fax: | +49 6022 240 512 |
| Email: | andreas.friese@alcon.com |
| Back up Contact Person: | Sherri Lakota<br>Head Global Regulatory Affairs, Vision Care |
| Telephone: | 817-615-5472 |
| Fax: | 817-551-4630 |
| Email: | sherri.lakota@alcon.com |
| Date Prepared: | 12 October 2015 |
## 2. Device Name:
| Common Name: | Soft Contact Lens |
|-------------------------|-------------------------------------------------------------------------|
| Trade/Proprietary Name: | Focus® DAILIES®, Focus® DAILIES® Toric,<br>Focus® DAILIES® Progressives |
| Classification Name: | Daily Wear Soft (Hydrophilic) Contact Lens |
| Device Classification: | Class II [21 CFR 886.5925 (b) (1)] |
## 3. Predicate Device:
Focus® DAILIES®, Focus® DAILIES® Toric, Focus® DAILIES® Progressives (nelfilcon A) contact lenses (cleared under K083216) have been identified as predicate device.
# 4. Description of Device:
The lens material is 69% water and 31% nelfilcon A polymer (polyviny) alcohol partially acetalized with N-formylmethyl acrylamide). For VISITINT® lenses, the color additive phthalocyanine blue (also known as copper phthalocyanine) is added to the lens material to create a light blue edge to edge color to make them easier to see
{4}------------------------------------------------
when handling. The lenses may be printed with inks containing one or more of the following color additives: phthalocyanine blue, phthalocyanine green.
Nelfilcon A lens designs include spherical, toric and multifocal lenses in the following parameter ranges:
| Power Range: | -20.00 D to +20.00 D |
|------------------|-----------------------------------------------------------------|
| Center Thickness | varies with design and power<br>(0.10 mm for -3.00 D spherical) |
Lenses have the following properties:
| Refractive index: | 1.38 |
|----------------------|----------------------------------------------------------------------|
| Light transmittance: | approximately 96 %T |
| Water content: | 69% by weight |
| Oxygen permeability | 26 barrer measured at 35°C<br>(single point Dk-Polarographic method) |
Lenses are supplied sterile in sealed blister packs containing buffered saline. The compatibility and package integrity of the blister-pack packaging system has been demonstrated and successfully used for other Alcon marketed lens products and packaged lenses are effectively steam sterilized in a validated autoclave. Blister-pack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility).
## 5. Indications for Use:
Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes.
Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
{5}------------------------------------------------
The lenses are to be prescribed for single use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
# 6. Substantial Equivalence:
The modification involves using the same print technology already established for the Focus® DAILIES® family of (nelfilcon A) soft contact lenses as well as for the DAILIES® AquaComfort Plus® family of (nelfilcon A) soft contact lenses. The following matrix summarizes the characteristics of the modified device as compared to the predicate device.
| | Modified Device | Predicate Device |
|----------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|
| | Focus DAILIES® Family<br>(nelfilcon A) One-Day Soft<br>Contact Lens | Focus DAILIES® Family<br>(nelfilcon A) One-Day Soft<br>Contact Lens |
| 510(k)<br>number: | TBD | K083216 |
| Intended Use: | Daily Wear,<br>Daily Disposable | Daily Wear,<br>Daily Disposable |
| Material<br>Classification: | FDA Group 2<br>(>50% H2O, nonionic polymer) | FDA Group 2<br>(>50% H2O, nonionic polymer) |
| Lens<br>Material: | nelfilcon A | nelfilcon A |
| Water<br>Content: | 69% | 69% |
| Power Range: | +20.00 to -20.00 D | +20.00 to -20.00 D |
| Visibility Tint: | With or without copper<br>phthalocyanine | With or without copper<br>phthalocyanine |
| Manufacturin<br>g Method: | Lightstream® Technology:<br>Full mold cast, integrated print<br>step | Lightstream® Technology:<br>Full mold cast, integrated print<br>step |
| Lens Designs: | Spherical, toric, multifocal | Spherical, toric, multifocal |
| Sterilization: | Steam sterilization,<br>validated autoclave | Steam sterilization,<br>validated autoclave |
| Packaging: | Blister pack | Blister pack |
| Package<br>Storage Saline<br>Solution: | Phosphate-acetate buffered saline<br>with up to 0.5% Poloxamer 108. | Phosphate-acetate buffered saline<br>with up to 0.5% Poloxamer 108. |
Table 1: Substantial Equivalence Comparison
{6}------------------------------------------------
# Non-clinical Testing:
A successful process validation was performed on the modified device to verify equivalence of the device to the predicate device. This resulted in the acceptance criteria being met.
## Clinical Testing:
The scope of the device modification did not require clinical testing to establish safety and effectiveness of the modified device.
## Substantial Equivalence:
Focus® DAILIES®, Focus® DAILIES® Toric, and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses are substantially equivalent to the predicate lenses and similar to other daily wear soft contact lenses in terms of water content (69% water) and ionic characteristics (FDA Group II: high water, nonionic), and indications for use.
Any differences which may exist between the Focus® DAILIES®, Focus® DAILIES® Toric, and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses and other Group II soft hydrophilic contact lenses do not adversely affect the safety and effectiveness of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.