CIBA VISION (NELFILCON A) ONE-DAY SOFT CONTACT LENSES
Applicant
Ciba Vision Corporation
Product Code
LPL · Ophthalmic
Decision Date
Apr 14, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes. Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity. The lenses are to be prescribed for single-use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Device Story
Soft contact lenses composed of nelfilcon A polymer; 69% water content. Designed for daily wear; single-use disposable. Corrects refractive ametropia (myopia, hyperopia, astigmatism) and presbyopia. Manufactured via Lightstream Technology (full mold cast); includes integrated printing step for visibility tint or color additives (phthalocyanine blue/green). Supplied sterile in blister packs with buffered saline. Prescribed by eye care professionals; patient self-applied. Output is optical correction; improves visual acuity for patients. No cleaning or disinfection required; discarded after single use.
Clinical Evidence
No clinical testing performed. Safety and effectiveness established via non-clinical stability testing and comparison to predicate devices with identical material and design characteristics.
Technological Characteristics
Material: 31% nelfilcon A polymer, 69% water. FDA Group 2 (high water, nonionic). Oxygen permeability: 26 barrer. Refractive index: 1.38. Manufacturing: Lightstream Technology (full mold cast). Sterilization: Steam (autoclave). Packaging: Blister pack with phosphate-acetate buffered saline and Poloxamer 108.
Indications for Use
Indicated for daily wear optical correction of refractive ametropia (myopia, hyperopia, astigmatism) in non-aphakic persons with non-diseased eyes. Progressives indicated for presbyopia with reading add ≤ +3.00D and astigmatism ≤ 2.00D. Single-use daily disposable.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Focus® DAILIES®, Focus® DAILIES® Toric, and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses (K050065)
FreshLook®, FreshLook® Toric, and FreshLook® Progressives (nelfilcon A) One-Day Color Contact Lens (K050213)
Submission Summary (Full Text)
{0}------------------------------------------------
| <div> <img alt="CIBA VISION Logo" src="..."/> <br/> Shared Passion for Healthy Vision and Better Life </div> | | CIBA VISION Corporation<br>11460 Johns Creek Parkway<br>Duluth, Georgia USA 30097 | 04 February 2009<br>Page 1 of 4<br>v03 |
|------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------------------------------------------|----------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------------------------------------------|----------------------------------------|
CIBA VISION Nelfilcon A Daily Wear Soft Contact Lenses
510(k) Summary of Safety and Substantial Equivalence
510(k) Summary
K 083216
1. Submitter Information:
| Company: | CIBA VISION® Corporation<br>11460 Johns Creek Parkway<br>Duluth, Georgia USA 30097 | APR 14 2009 |
|-----------------|------------------------------------------------------------------------------------------------------------------|-------------|
| Contact Person: | Martina Heim, PhD, RAC<br>Senior Regulatory Specialist, Global Regulatory Affairs<br>martina.heim@cibavision.com | |
| Telephone: | 678-415-3565 | |
| FAX: | 678-415-3454 | |
| Date Prepared: | 27 October 2008 | |
2. Device Name:
| Common Name: | Soft Contact Lens |
|-------------------------|-------------------------------------------------------------------------|
| Trade/Proprietary Name: | Focus® DAILIES®, Focus® DAILIES® Toric,<br>Focus® DAILIES® Progressives |
| Classification Name: | Daily Wear Soft Contact Lens |
| Device Classification: | Class II [21 CFR 886.5925 (b) (1)] |
#### 3. Predicate Device:
Predicate devices are:
1) Focus® DAILIES®, Focus® DAILIES® Toric, and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses, as cleared under K050065, 2) FreshLook®, FreshLook® Toric, and FreshLook® Progressives (nelfilcon A) One-Day Color Contact Lens, as cleared under K050213.
#### 4. Description of Device:
The lens material is 69% water and 31% nelfilcon A polymer (polyviny) alcohol partially acetalized with N-formylmethyl acrylamide). For VISITINT® lenses, the color additive phthalocyanine blue (also known as copper phthalocyanine) is added to the lens material to create a light blue edge to edge color to make them easier to see when handling. The lenses may be printed with inks containing one or more of the following color additives: phthalocyanine blue, phthalocyanine green.
{1}------------------------------------------------
| CIBAOVISION. | CIBA VISION Corporation |
|---------------------------------------------------|---------------------------|
| Shared Passion for Healthy Vision and Better Life | 11460 Johns Creek Parkway |
| | Duluth, Georgia USA 30097 |
## CIBA VISION Nelfilcon A Daily Wear Soft Contact Lenses 510(k) Summary of Safety and Substantial Equivalence
Focus® DAILIES®, Focus® DAILIES® Toric, Focus® DAILIES® Progressives One-Day contact lenses include spherical, toric, and multifocal lens designs in the following parameter ranges:
- Power Range: -20.00D to +20.00D .
- 0.010 mm for -3.00D spherical (varies with power) Center Thickness: .
Lenses have the following properties:
- 1.38 Refractive index: .
- Light transmittance: approximately 96 %T .
- Water content : 69% by weight .
- Oxygen permeability 26 barrer
- measured at 35°C (single point Dk-Polarographic method)
Lenses are supplied sterile in sealed blister-packs containing buffered saline. The compatibility and package integrity of the blister-pack packaging system has been demonstrated and successfully used for other CIBA VISION marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blisterpack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to establish and extend the labeled expiration date.
#### 5. Indications for Use:
◆
Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes.
Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eves who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
The lenses are to be prescribed for single-use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the CIBA Vision logo. The logo is black and white and features the words "CIBA VISION" in a bold, sans-serif font. The "O" in "VISION" is stylized to look like an eye. Below the logo is the tagline "Shared Passion for Healthy Vision and Better Life" in a smaller font.
CIBA VISION Corporation 11460 Johns Creek Parkway Duluth, Georgia USA 30097
## CIBA VISION Nelfilcon A Daily Wear Soft Contact Lenses 510(k) Summary of Safety and Substantial Equivalence
#### દ. Substantial Equivalence:
The following matrix summarizes the characteristics of the modified device as compared to the predicate devices:
| | Modified Device | K050065 | K050213 |
|------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|
| | Focus DAILIES Family<br>(nelfilcon A) One-Day Soft<br>Contact Lens | Focus DAILIES Family<br>(nelfilcon A) One-Day<br>Contact Lens | FreshLook Family<br>(nelfilcon A) One-Day Color<br>Contact Lens |
| Lens Material: | nelfilcon A | nelfilcon A | nelfilcon A |
| Material<br>Classification: | FDA Group 2<br>(> 50% H2O, nonionic<br>polymer) | FDA Group 2<br>(> 50% H2O, nonionic<br>polymer) | FDA Group 2<br>(> 50% H2O, nonionic<br>polymer) |
| Water Content: | 69% | 69% | 69% |
| Power Range: | +20.00 to -20.00D | +20.00 to -20.00D | +20.00 to -20.00D |
| Visibility Tint: | With or without copper<br>phthalocyanine | With or without copper<br>phthalocyanine | With or without copper<br>phthalocyanine |
| Manufacturing<br>Method: | Lightstream Technology™:<br>Full mold cast,<br>integrated print step | Lightstream Technology™:<br>Full mold cast | Lightstream Technology™:<br>Full mold cast,<br>integrated print step |
| Lens Designs: | Spherical, toric, multifocal | Spherical, toric, multifocal | Spherical, toric, multifocal |
| Sterilization: | Steam sterilization, validated<br>autoclave | Steam sterilization, validated<br>autoclave | Steam sterilization, validated<br>autoclave |
| Packaging: | Blister pack | Blister pack | Blister pack |
| Package Storage<br>saline solution | Phosphate-acetate buffered<br>saline with up to 0.02%<br>Poloxamer 108 | Phosphate-acetate buffered<br>saline with up to 0.02%<br>Poloxamer 108 | Phosphate-acetate buffered<br>saline with up to 0.02%<br>Poloxamer 108 |
#### Non-clinical Testing:
Successful stability testing supports the labeled expiration date. Additional chemical, physical, and toxicological testing results can be referenced from the predicate devices.
#### Clinical Testing:
The scope of the device modification did not require clinical testing to establish safety and effectiveness of the modified device.
### Substantial Equivalence:
The Focus® DAILIES®, Focus® DAILIES® Toric, and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses are substantially equivalent to the predicate
{3}------------------------------------------------
| 110100<br>CIBAC<br>Colli | |
|---------------------------------------------------|--|
| Shared Passion for Healthy Vision and Better Life | |
CIBA VISION Corporation 11460 Johns Creek Parkway Duluth, Georgia USA 30097
# CIBA VISION Nelfilcon A Daily Wear Soft Contact Lenses 510(k) Summary of Safety and Substantial Equivalence
lenses and similar to other daily wear soft contact lenses in terms of water content (69% water) and ionic characteristics (FDA Group II: high water, nonionic), clinical performance, and indications for use.
Any differences which may exist between the Focus® DAILIES®, Focus® DAILIES® Toric, and Focus® DAILIES® Progressives (nelfilcon A) One-Day Contact Lenses and other Group II soft hydrophilic contact lenses do not adversely affect the safety and effectiveness of the device.
:)
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
CIBA VISION Corporation c/o Martina Heim, Ph.D., RAC Senior Regulatory Specialist 11460 Johns Creek Parkway Duluth, Georgia 30097
APR 1 4-2009
Re: K083216
Trade/Device Name: Focus® DAILIES®, Focus® DAILIES® Toric, Focus® DAILIES® Progressives Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL, MVN Dated: March 24, 2009 Received: March 25, 2009
Dear Dr. Heim;
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Egeland, MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT PART II.
: : : : : : : : : : : : : : :
K083216 510(k) Number:
Device Name:
Focus® DAILIES®, Focus® DAILIES® Toric,
Focus® DAILIES® Progressives
Indications For Use:
Focus® DAILIES® and Focus® DAILIES® Toric (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia, hyperopia and astigmatism) in not-aphakic persons with non-diseased eyes.
Section and States and 122
Focus® DAILIES® Progressives (nelfilcon A) One-Day soft contact lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
The lenses are to be prescribed for single-use daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use:
Over the Counter Use
I
ing-chun Shi
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices
P
or
510(k) Number K083216
CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.