The Syzygy Stabilization System is intended to provide immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and/or sacral spine, specifically as follows: • When intended for pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients the Syzygy Stabilization System is indicated for one of the following: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); fracture; dislocation; scoliosis; kyphosis; spinal tumor; and/or failed previous fusion (pseudarthrosis). · In addition, when used as a pedicle screw fixation system, the Syzygy Stabilization System is indicated for skeletally mature patients having degenerative spondylolisthesis with objective evidence of neurologic impairment and/or severe spondylolisthesis (grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; who are receiving fusions using autogenous bone graft only; who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and who are having the device removed after the development of a solid fusion mass.
Device Story
System comprises posted screws, couplers, and surgical instruments; used with Savannah-T pedicle screw and rod system. Screws are self-tapping, cancellous thread design, cannulated or non-cannulated. Couplers (medial/cranial) attach screws to longitudinal rods to create rigid construct. Used in thoracic, lumbar, and sacral spine. Operated by surgeons in clinical/OR settings. Provides spinal immobilization/stabilization as adjunct to fusion. Benefits patient by stabilizing spinal segments to facilitate bone fusion in cases of deformity or instability.
Clinical Evidence
Bench testing only. Mechanical testing performed on worst-case configurations per ASTM F1717-14, including static axial compression/tension bending, static torsion, and dynamic axial compression/tension bending. A cadaveric validation study was also conducted. All tests met protocol requirements and applicable ASTM standards.
Technological Characteristics
Materials: Ti-6Al-4V (ASTM F136). Components: Posted screws (self-tapping, cancellous thread), couplers, longitudinal rods. Configuration: Cannulated and non-cannulated options. Sterilization: Provided non-sterile for single-use. Mechanical performance validated per ASTM F1717-14.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization as an adjunct to fusion for degenerative disc disease, fracture, dislocation, scoliosis, kyphosis, spinal tumor, failed fusion, or degenerative spondylolisthesis with neurologic impairment/severe spondylolisthesis (grades 3-4) at L5-S1. Contraindicated for patients not meeting skeletal maturity or those not receiving autogenous bone graft when treating specific spondylolisthesis conditions.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
Savannah Lumbar Percutaneous Stabilization System (K072116)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping to create a sense of depth and connection. The profiles are rendered in a simple, line-art style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 27, 2016
Amendia, Incorporated Ms. Kristen Allen Sr. Regulatory Affairs Specialist 1755 West Oak Parkway Marietta, Georgia 30062
Re: K153152
Trade/Device Name: Syzygy Stabilization System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH Dated: March 24, 2016 Received: March 25, 2016
Dear Ms. Allen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K153152
Device Name Syzygy Stabilization System
#### Indications for Use (Describe)
The Syzygy Stabilization System is intended to provide immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and/or sacral spine, specifically as follows:
• When intended for pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients the Syzygy Stabilization System is indicated for one of the following: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies); fracture; dislocation; scoliosis; kyphosis; spinal tumor; and/or failed previous fusion (pseudarthrosis).
· In addition, when used as a pedicle screw fixation system, the Syzygy Stabilization System is indicated for skeletally mature patients having degenerative spondylolisthesis with objective evidence of neurologic impairment and/or severe spondylolisthesis (grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; who are receiving fusions using autogenous bone graft only; who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and who are having the device removed after the development of a solid fusion mass.
| Type of Use (Select one or both, as applicable) | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|--|----------------------------------------------|------------------------------------------------------------------------------------------------|
| <div> <span> <tr><td>Type of Use (Select one or both, as applicable)</td><td></td></tr> <tr><td>Prescription Use (Part 21 CFR 801 Subpart D)</td><td><span style="font-size: 20px;"><b>"></b></span> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr> </span></div> | Type of Use (Select one or both, as applicable) | | Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;"><b>"></b></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
| Type of Use (Select one or both, as applicable) | | | | |
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;"><b>"></b></span> Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AMENDIA. The logo consists of the word "AMENDIA" in blue, with a blue square around the letter "A". Below the word "AMENDIA" is the tagline "creating balanced solutions" in a smaller, lighter blue font. The logo is simple and modern, and the blue color gives it a sense of trust and reliability.
### 510(k) Summary
### Syzyqy Stabilization System
| Submitter: | Amendia, Inc.<br>1755 W. Oak Parkway<br>Marietta, GA 30062 |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Kristen Allen<br>Sr. Regulatory Affairs Specialist<br>910-612-4153 (P), 877-420-1213 (F)<br>kallen@amendia.com (e-mail) |
| Date Prepared: | April 26, 2016 |
| Trade Name: | Syzygy Stabilization System |
| Common Name: | Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease |
| Device Product Code<br>and Classification: | Regulation Number: 21 CFR 888.3070<br>NKB, Class III, Pedicle Screw Spinal System, For Degenerative<br>Disc Disease<br>MNI, Class II, Pedicle Screw Spinal System<br>MNH, Class II, Spondylolisthesis Spinal Fixation Device System |
| Primary Predicate: | Savannah Lumbar Percutaneous Stabilization System (K072116) |
| Additional Predicate: | Savannah-T Pedicle Screw System (K132925)<br>Pangea (K052123)<br>Talon Spinal System (K102995) |
#### Device Description:
The Syzygy Stabilization System consists of posted screws, couplers, and associated surgical instruments. The system is used in conjunction with the Amendia Savannah-T pedicle screw and rod system. The Syzygy posted screws are self-tapping with a cancellous thread design. They are available in cannulated and non-cannulated configurations, in a variety of diameters and lengths. The Syzygy posted screws are attached to the pedicles via couplers (medial and cranial) which accept and secure the longitudinal rods to build a rigid construct. The screws and couplers are manufactured from Ti-6AI-4V (ASTM F136) and are provided non-sterile for singleuse.
#### Indications and Intended use:
The Syzygy Stabilization System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and/or sacral spine, specifically as follows:
- When intended for pedicle screw fixation of the non-cervical posterior spine in skeletally . mature patients the Syzygy Stabilization System is indicated for one or more of the following: degenerative disc disease (defined as discogenic back pain with degeneration of
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Amendia. The logo is blue and features the word "AMENDIA" in a sans-serif font. The "A" is inside of a blue square. Below the word "AMENDIA" is the tagline "creating balanced solutions" in a smaller font.
the disc confirmed by history and radiographic studies); spondylolisthesis; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and/or failed previous fusion (pseudarthrosis).
- . In addition, when used as a pedicle screw fixation system, the Syzygy Stabilization System is indicated for skeletally mature patients having degenerative spondylolisthesis with objective evidence of neurologic impairment and/or severe spondylolisthesis (grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; who are receiving fusions using autogenous bone graft only: who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and who are having the device removed after the development of a solid fusion mass.
#### Summary of Technological Characteristics:
The subject device is substantially equivalent to the predicate device as well as other similar devices cleared by FDA for commercial distribution in the United States. The Subject Device is equivalent to the predicate in regards to technological characteristics including design (rod, screw and coupler configuration), intended use (as described above), material composition (Titanium Alloy), sizes (dimensions are comparable to those offered by the predicate systems) and fundamental scientific technology (same as previously cleared devices).
#### Summary of Performance Testing:
Mechanical testing for the Syzygy Stabilization System was performed on the worst case subject device in accordance with ASTM standards.
| Test | Standard |
|-----------------------------------|---------------|
| Static Axial Compression Bending | ASTM F1717-14 |
| Static Axial Tension Bending | ASTM F1717-14 |
| Static Torsion | ASTM F1717-14 |
| Dynamic Axial Compression Bending | ASTM F1717-14 |
| Dynamic Axial Tension Bending | ASTM F1717-14 |
A cadaveric validation study was also performed.
For all test methods, the subject devices met the requirements as established by the test protocol and applicable ASTM standards. The results demonstrated that the Subject device is substantially equivalent to the Predicate.
#### Conclusion:
Based on the comparison to predicate device, the Syzygy Stabilization System has been shown to be substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.