The Synthes USS (including the Click'X®, and USS VAS variable axis components, and Pangea™), Click'X® Monoaxial, Dual-Opening and the Small Stature USS (which includes small stature and pediatric patients) are non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2), a posterior hook fixation system (T1-L5), or as a anterolateral fixation system (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion. When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct. When used with the 3.5/6.0 mm parallel connectors, the Synthes USS (including the Click'X® and, USS VAS variable axis components, and Pangea™), Click'X® Monoaxial and Dual-Opening USS can be linked to the CerviFix® System. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including the Click'X®, and USS VAS variable axis components, and Pangea™), the Click' X® Monoaxial and Dual-Opening USS Systems. In addition, Synthes USS (including the Click'X®, and USS VAS variable axis components, and Pangea™), Click'X® Monoaxial and the Dual-Opening USS can be interchanged with all USS 6.0 mm rods and transconnectors
Device Story
Pangea System is a spinal fixation orthosis; component of Universal Spinal System (USS). Device consists of polyaxial dual-core screws and non-threaded quarter-turn locking caps. Used by surgeons in clinical settings to provide stabilization of the spine. System integrates with existing USS rods and transconnectors; allows linkage to CerviFix System via parallel connectors. Provides mechanical support to spinal segments to facilitate fusion; benefits patients by correcting deformities and stabilizing unstable spinal segments.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: Commercially pure titanium (ASTM F67) and titanium alloy (ASTM F1295). Design: Polyaxial dual-core screws, non-threaded quarter-turn locking cap. Connectivity: Mechanical linkage to USS rods, transconnectors, and CerviFix system via parallel connectors.
Indications for Use
Indicated for non-cervical spinal fixation (T1-S2 pedicle, T1-L5 hook, T8-L5 anterolateral) in patients with degenerative disc disease, spondylolisthesis, trauma, deformities (scoliosis, kyphosis, lordosis, Scheuermann's), tumor, stenosis, pseudoarthrosis, or failed fusion. Pedicle screw fixation limited to skeletally mature patients, except for Small Stature USS which includes pediatric patients.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
## 510(k) Summary
| Name of Firm: | Synthes Spine<br>1302 Wrights Lane East<br>West Chester, PA 19380 |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Angela Mikroulis<br>Spine Regulatory Affairs Specialist<br>Telephone: 610-719-5718<br>Facsimile: 610-719-5102 |
| Trade Name: | PangeaTMSystem |
| Common Name/<br>Classification Name: | Spinal interlaminal fixation orthosis;<br>Spinal intervertebral body fixation orthosis;<br>Pedicle screw spinal system |
| Device Product<br>Code and<br>Classification: | KWP, KWQ, MNH, MNI, NKB<br>21 CFR 888.3050, 21 CFR 888.3060, 21 CFR 888.3070<br>CLASS II and Class III |
| Predicate: | Click'X (K992739) |
| Device Description: | The PangeaTM System is similar to the cleared Click'X (K992739). Both are<br>components of the Universal Spinal System (USS).<br><br>The Pangea System includes polyaxial dual-core screws and a non threaded<br>quarter turn in design locking cap.<br><br>The Synthes Pangea polyaxial screws and locking cap are fabricated from<br>commercially pure titanium and titanium alloy, conforming to ASTM F67<br>and ASTM F-1295 respectively. |
{1}------------------------------------------------
| Intended Use /<br>Indications for Use: | The Synthes USS (including the Click'X®, and USS VAS variable axis<br>components, and Pangea™), Click'X® Monoaxial, Dual-Opening and the<br>Small Stature USS (which includes small stature and pediatric patients) are<br>non-cervical spinal fixation devices intended for use as posterior pedicle<br>screw fixation systems (T1-S2), a posterior hook fixation system (T1-L5), or<br>as a anterolateral fixation system (T8-L5). Pedicle screw fixation is limited<br>to skeletally mature patients with the exception of the Small Stature USS.<br>These devices are indicated for all of the following indications regardless of<br>the intended use: degenerative disc disease (defined as discogenic back pain<br>with degeneration of the disc confirmed by history and radiographic studies),<br>spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or<br>curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's<br>Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.<br><br>When treating patients with Degenerative Disc Disease (DDD), transverse<br>bars are not cleared for use as part of the posterior pedicle screw construct.<br><br>When used with the 3.5/6.0 mm parallel connectors, the Synthes USS<br>(including the Click'X® and, USS VAS variable axis components, and<br>Pangea™), Click'X® Monoaxial and Dual-Opening USS can be linked to<br>the CerviFix® System. In addition, when used with 3.5/5.0 mm parallel<br>connectors, the Synthes Small Stature USS can be linked to the Synthes USS<br>(including the Click'X®, and USS VAS variable axis components, and<br>Pangea™), the Click' X® Monoaxial and Dual-Opening USS Systems.<br><br>In addition, Synthes USS (including the Click'X®, and USS VAS variable<br>axis components, and Pangea™), Click'X® Monoaxial and the Dual-<br>Opening USS can be interchanged with all USS 6.0 mm rods and<br>transconnectors |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{2}------------------------------------------------
## Page 2 - Ms. Angela Mikroulis
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use Statement
K052123 510(k) Number:
Pangea TM Device Name:
The Synthes USS (including the Click'X®, USS VAS variable axis Indications for use: components, and Pangea™), Click'X® Monoaxial, Dual-Opening and the Small Stature USS (which includes small stature and pediatric patients) are non-cervical spinal fixation devices intended for use as posterior pedicle screw fixation systems (T1-S2), a posterior hook fixation system (T1-L5), or as a anterolateral fixation system (T8-L5). Pedicle screw fixation is limited to skeletally mature patients with the exception of the Small Stature USS. These devices are indicated for all of the following indications regardless of the intended use: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis, Scheuermann's Disease), tumor, stenosis, pseudoarthrosis, and failed previous fusion.
When treating patients with Degenerative Disc Disease (DDD), transverse bars are not cleared for use as part of the posterior pedicle screw construct.
When used with the 3.5/6.0 mm parallel connectors, the Synthes USS (including the Click' X®, USS VAS variable axis components, and Pangea™), Click'X® Monoaxial and Dual-Opening USS can be linked to the CerviFix® System. In addition, when used with 3.5/5.0 mm parallel connectors, the Synthes Small Stature USS can be linked to the Synthes USS (including the Click' X®, USS VAS variable axis components, and Pangea™), the Click'X® Monoaxial and Dual-Opening USS Systems.
In addition. Synthes USS (including the Click'X®, USS VAS variable axis components, and Pangea™), Click' X® Monoaxial and the Dual-Opening USS can be interchanged with all USS 6.0 mm rods and transconnectors.
| Prescription Use | X (Part 21 CFR 801 Subpart D) |
|----------------------|-------------------------------|
| | AND/OR |
| Over-The-Counter Use | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Page 1 of 1
Division of General, Restorative, and Neurological Devices
510(k) Number K052123
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.