WebPAX is intended for use in the communication and storage of medical images. WebPAX is also intended for use as a comprehensive solution to view, optimize, and post-process diagnostic medical images as an aid to physicians and other healthcare professionals in the evaluation of digital imaging examinations. Due to special customer requirements based on the imaging modality and clinical focus, WebPAX can be configured with different combinations of clinical applications which are intended to assist the physician in diagnosis or treatment planning. This includes commercially available post-processing techniques such as multi-planar reconstruction (MPR). Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using cleared monitors intended for mammography display. MPR is not intended for mammography use. Not intended for diagnostic use on mobile devices.
Device Story
WebPAX is a software-only PACS solution; server-side software installed on customer-supplied hardware/virtual machines; client-side accessed via standard web browsers (Chrome, Internet Explorer). Inputs: DICOM-compliant medical images. Processing: server manages/stores images; client-side modules perform image visualization, optimization, and post-processing (e.g., multi-planar reconstruction via WebGL, echocardiography/cardiovascular MRI measurements). Outputs: processed diagnostic images and quantitative measurements (volumes, mass, velocity, time-intensity curves). Used in clinical settings by physicians/technicians to assist in diagnosis and treatment planning. Benefits: provides comprehensive tools for image evaluation and post-processing across various modalities.
Clinical Evidence
Bench testing only. Software verification confirmed requirements met on Chrome and IE11. Software validation of image analysis (basic, echo, cardiovascular MR measurements) showed >98% accuracy compared to predicate measurements on identical DICOM datasets. Reader study (n=3 board-certified physicians) evaluated 3D post-processing using six clinical datasets; performance was comparable to predicate Siemens Leonardo workstation, with all scores rated 1 (Excellent) or 2 (Very good).
Technological Characteristics
Software-only PACS; server-client architecture. Connectivity: TCP/IP, DICOM standard (SCP/SCU). Visualization: WebGL-based 3D post-processing (MPR). Measurements: ROI-based area, grayscale, distance, angle, velocity, and time-intensity curve analysis. Platform: customer-supplied server/virtual machine; web browser client. Software: WebPAX 9.3.
Indications for Use
Indicated for physicians and healthcare professionals to view, optimize, and post-process diagnostic medical images for evaluation, diagnosis, or treatment planning. Not for primary interpretation of lossy compressed mammographic or digitized film screen images. Not for diagnostic use on mobile devices.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized representation of a human figure with three faces in profile, stacked on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Auqust 17, 2016
Heart Imaging Technologies, LLC % Robert M. Judd, Ph.D. President 5003 Southpark Drive, Suite 140 DURHAM NC 27713
Re: K152949 Trade/Device Name: WebPAX Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 27, 2016 Received: August 1, 2016
Dear Dr. Judd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152949
Device Name WebPAX
#### Indications for Use (Describe)
WebPAX is intended for use in the communication and storage of medical images. WebPAX is also intended for use as a comprehensive solution to view, optimize, and post-process diagnostic medical images as an aid to physicians and other healthcare professionals in the evaluation of digital imaging examinations.
Due to special customer requirements based on the imaging modality and clinical focus, WebPAX can be configured with different combinations of clinical applications which are intended to assist the physician in diagnosis or treatment planning. This includes commercially available post-processing techniques such as multi-planar reconstruction (MPR).
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using cleared monitors intended for mammography display. MPR is not intended for mammography use.
Not intended for diagnostic use on mobile devices.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> |
|----------------------------------------------|-------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <input type="checkbox"/> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K152949
## Identification of Submitter
| Address: | Heart Imaging Technologies, LLC<br>5003 Southpark Drive, Suite 140<br>Durham, NC 27701 |
|--------------------------|----------------------------------------------------------------------------------------|
| Registration Number: | 3005107869 |
| Contact Person: | Robert M. Judd, Ph.D.<br>President<br>Telephone: 919-384-5044<br>FAX: 866-457-3694 |
| Date of Preparation: | July 29, 2016 |
| Identification of Device | |
| Trade Name: | WebPAX |
| Common Name: | Picture archiving and communications system |
| Regulation Number: | 892.2050 |
| Device Class: | II |
LLZ
#### Predicate Devices
Product Code:
### PREDICATE DEVICE 1:
| Manufacturer: | Siemens Medical Solutions, Inc. |
|--------------------|----------------------------------------------------|
| Trade Name: | LEONARDO syngo Cardiology Workstation |
| Common Name: | Picture archiving and communications system (PACS) |
| Regulation Number: | 892.2050 |
| Device Class: | II |
| Product Code: | LLZ |
| 510k Number: | K042203 |
#### PREDICATE DEVICE 2:
| Manufacturer: | Heart Imaging Technologies, LLC |
|--------------------|----------------------------------------------------|
| Trade Name: | WebPAX |
| Common Name: | Picture archiving and communications system (PACS) |
| Regulation Number: | 892.2050 |
| Device Class: | II |
| Product Code: | LLZ |
| 510k Number: | K051325 |
## Description of Device
WebPAX consists of a server, typically located in a hospital data center, and one or more client systems used by technicians and physicians. On the server side, WebPAX is a software-only solution that is installed on customer-supplied hardware, such as dedicated computer server or virtual machine. The server-side software communicates with other DICOM-compliant systems and with end-users via
(PACS)
{4}------------------------------------------------
standard Intranet and/or Internet mechanisms. On the client side, technicians and physicians access the WebPAX server using standard Internet web browsers, such as Google Chrome and Internet explorer.
WebPAX provides users with the ability to manage, store, and interpret digital medical images, including the following application specific modules:
- . Echocardiography Workstation
- Cardiovascular MRI Workstation
- 3D Post Processing (Multi-Planar Reconstruction)
# Indications for Use
WebPAX is intended for use in the communication and storage of medical images. WebPAX is also intended for use as a comprehensive solution to view, optimize, and post-process diagnostic medical images as an aid to physicians and other healthcare professionals in the evaluation of digital imaging examinations.
Due to special customer requirements based on the imaging modality and clinical focus, WebPAX can be configured with different combinations of clinical applications which are intended to assist the physician in diagnosis or treatment planning. This includes commercially available post-processing techniques such as multi-planar reconstruction (MPR).
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using cleared monitors intended for mammography display. MPR is not intended for mammography use.
Not intended for diagnostic use on mobile devices.
# Performance Testing
The following test reports have been provided in support of substantial equivalence:
- 1. Software Verification Report
The purpose of this report is to demonstrate that the WebPAX software meets all of the requirements described in the Software Requirement Specification. Testing was conducted with clients running on the two supported web browsers: Google Chrome and Microsoft Internet Explorer 11. All tests passed.
- 2. Software Validation Report
The purpose of this report was to validate the performance of the following WebPAX image analysis functionalities: Basic measurements, Echocardiography measurements, and Cardiovascular MR measurements. Testing was conducted with clients running on the two supported web browsers: Google Chrome and Microsoft Internet Explorer 11. [Note: Some functionalities are only available with the Google Chrome browser, and thus were only tested with that browser].
For all Software Validation tests for which the Predicate Device offered the same functionality, the "WebPAX Measurement Accuracy" was assessed by direct comparison to measurements made using the Predicate Device on the exact same DICOM dataset, image,
{5}------------------------------------------------
and location(s) within the image. In all cases WebPAX met the Acceptance Criteria that accuracy be greater than 98%.
- 3. Reader Study Report
The purpose of the Reader Study was to demonstrate that the 3D post processing capability of WebPAX 9.3 is equivalent to that of the predicate Siemens Leonardo Workstation (K042203). Three board-certified physicians who routinely read cardiovascular MRI studies were asked to evaluate six DICOM datasets that represent typical clinical scenarios. All scores for WebPAX were at least a 1 (Excellent) or a 2 (Very good). The performance of WebPAX was comparable to that of the predicate Siemens device.
# Substantial Equivalence Discussion
The following table compares WebPAX to the predicate device:
| FEATURE | WebPAX | Predicate<br>Device<br>(Siemens<br>Leonardo<br>K042203) | Comments |
|-------------------------------------|--------|---------------------------------------------------------|----------------------------------|
| Basic PACS Functions | | | |
| TCP/IP network connectivity | Yes | Yes | DICOM standard |
| DICOM storage class provider (SCP) | Yes | Yes | DICOM standard |
| DICOM storage class user (SCU) | Yes | Yes | DICOM standard |
| Digital image storage | Yes | Yes | DICOM standard |
| Image visualization | Yes | Yes | DICOM standard |
| Generic Tools | | | |
| Distance Measurements | Yes | Yes | DICOM standard |
| ROI Area Measurements | Yes | Yes | DICOM standard |
| ROI Grayscale Measurements | Yes | Yes | DICOM standard |
| Angle Measurements | Yes | Yes | Basic Geometry |
| Specialty Tools: Echocardiography | | | |
| Doppler Velocity | Yes | No | DICOM standard |
| Deceleration Time | Yes | No | DICOM standard |
| Velocity-Time Integral | Yes | No | DICOM standard |
| Stress Echo Display | Yes | No | DICOM standard |
| Specialty Tools: Cardiovascular MRI | | | |
| Volumes/Mass | Yes | Yes | ROI Area times Slice Thickness |
| Hyperintensity | Yes | Yes | ROI Area times Slice Thickness |
| Velocity Encoding | Yes | Yes | ROI Grayscale times Scale Factor |
| Time-Intensity Curves | Yes | Yes | ROI Grayscale versus time |
| T2 and T2* Decay | Yes | No | ROI Grayscale versus time |
| T1 Recovery | Yes | No | ROI Grayscale versus time |
| 3D Post Processing | | | |
| Multi-Planar Reconstruction | Yes | Yes | For WebPAX, based on WebGL |
{6}------------------------------------------------
Both the WebPAX and Siemens Leonardo devices provide generic PACS functionalities with common measurement tools and basic image processing.
Both devices also provide specialty tools for Cardiovascular MRI that allow assessment of cardiac volumes, masses, blood flow velocities, and time-intensity curves. The performance of these functions for both devices were directly compared. WebPAX also provides users with the ability to fit a monoexponential curve to MRI T2(*) and/or T1 grayscale data based on well-established techniques. WebPAX additionally provides specialty tools for Echocardiography. Although the Siemens Leonardo does not provide Echocardiography functions, all of these are an integral part of the DICOM standard itself and testing demonstrated that performance met established acceptance criteria.
Both devices provide multi-planar reconstruction (MPR) techniques that allow the user to re-slice 3D data into directions other than the original imaging planes. The underlying technology used by WebPAX to achieve this, namely WebGL, is different than that used by the Siemens Leonardo, but a reader study demonstrated equivalent performance.
# Summary of Substantial Equivalence
No significant differences were detected when comparing measurements and post-processing results made using the candidate WebPAX system with those of the predicate devices. We conclude that the WebPAX system is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.