K042203 · Siemens Medical Solutions USA, Inc. · LLZ · Sep 24, 2004 · Radiology
Device Facts
Record ID
K042203
Device Name
LEONARDO SYNGO CARDIOLOGY WORKSTATION
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Sep 24, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The LEONARDO syngo Cardiology Workstation is a medical diagnostic workstation for viewing, manipulation, communication, and storage of medical images and data on exchange media. The LEONARDO syngo Cardiology Workstation can be configured as a stand-alone diagnostic review and post-processing workstation with a variety of syngo or Windows based software options, which are intended to assist the physician in diagnosis or treatment planning. This includes commercially available post-processing techniques.
Device Story
LEONARDO syngo Cardiology Workstation; universal imaging platform based on Windows XP; functions as medical diagnostic workstation for viewing, manipulation, communication, and storage of medical images/data. Configurable with various syngo or Windows-based software options/clinical focus packages for post-processing diagnostic information. Used by physicians in clinical settings to aid evaluation of cardiological and radiological examinations; assists in diagnosis and treatment planning. Marketed as software-only solution or complete hardware/software workstation. Supports DICOM formatted images. Healthcare providers use output to inform clinical decision-making; benefits patient through enhanced diagnostic review and post-processing capabilities.
Clinical Evidence
No clinical data. Safety and effectiveness are supported by risk analysis, software development, verification, and validation testing adhering to industry standards.
Technological Characteristics
Software-only or hardware/software workstation; Windows XP operating system; supports DICOM formatted images; installed by Siemens service engineers; risk management via hazard analysis and verification/validation testing.
Indications for Use
Indicated for use by physicians as a diagnostic workstation for viewing, manipulating, communicating, and storing medical images and data to assist in diagnosis or treatment planning for patients undergoing cardiological and radiological examinations.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K04 2203
Siemens Medical Solutions, Inc 510(k) for LEONARDO syngo Cardiology
## 510(k) Summary
# SEP 2 4 2004
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
## I. GENERAL INFORMATION
| Establishment:<br>Address: | Siemens Medical Systems, Inc.<br>51 Valley Stream Parkway<br>Malvern, PA 19355 |
|----------------------------|--------------------------------------------------------------------------------|
|----------------------------|--------------------------------------------------------------------------------|
- 2240869 Registration Number: .
- Ana Ladino . Contact Person: Technical Specialist Regulatory Submissions Telephone: (610) 448-1785 (610) 448-1787 Telefax: ---------------------------------------------------------------------------------------------------------------------------------------------------------------------
Device Name:
- LEONARDO syngo Cardiology Workstation Trade Name: .
- Picture Archiving and Communications System (PACS) Classification: ●
- Radiology Classification Panel: .
- 21 CFR 8892.2050 CFR Section: .
- Class II . Device Class:
- LLZ . Product Code:
Date of Preparation of Summary: August 5th, 2004
## II. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE SUBSTANTIAL EQUIVALENCE DETERMINATION
#### Device Description and Intended Use: .
This premarket notification covers Siemens LEONARDO syngo Cardiology Workstation syngo is a universal imaging platform based on Windows XP. LEONARDO syngo Cardiology offers a comprehensive cardiology solution to view, optimize, post-process diagnostic information and aid the doctors in the evaluation of digital cardiological and radiological examinations and patient information.
Due to special customer requirements based on the modality image type and the clinical focus, the LEONARDO syngo Cardiology Workstation can be configured with different combinations of clinical applications. syngo applications can be
{1}------------------------------------------------
added to the LEONARDO syngo Cardiology Workstation either individually or as clinical focus packages.
The LEONARDO syngo Cardiology Workstation is a medical diagnostic workstation for viewing, manipulation, communication, and storage of medical images and data on exchange media.
The LEONARDO syngo Cardiology Workstation can be configured with a variety of syngo- or Windows based software options, which are intended to assist the physician in diagnosis or treatment planning. This includes commercially available post-processing techniques.
#### . Technological Characteristics:
The LEONARDO syngo Cardiology Workstation will be marketed as a software only solution for the end-user (with recommended hardware requirements) or as a complete work station for the end-user (hardware and software package). It will be installed by Siemens service engineers. The LEONARDO syngo Cardiology Workstation described supports DICOM formatted images and information. The workstation is based on the Windows XP operating system.
#### General Safety and Effectiveness Concerns: .
The device labeling contains instructions for use and any necessary cautions and warning, to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
#### Substantial Equivalence: .
The LEONARDO syngo Cardiology Workstation, addressed in this premarket notification, is substantially equivalent to the following commercially available device:
## LEONARDO (K040970)
The LEONARDO syngo Cardiology Workstation described in this premarket notification has the same intended use and similar technical characteristics as the device listed above.
In summary, Siemens is of the opinion that LEONARDO syngo Cardiology Workstation does not introduce any new potential safety risks and is substantially equivalent to and performs as well as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that form the wings and body of the bird.
# SEP 2 4 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Ana Ladino Technical Specialist, Regulatory Submission Siemens Medical Systems, Inc. 51 Valley Stream Parkway MALVERN PA 19355 -
Re: K042203 Trade/Device Name: LEONARDO syngo Cardiology Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II
Product Code: 90 LLZ Dated: August 5, 2004 Received: August 16, 2004
Dear Ms. Ladino:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to ,devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ought haling of substantial equivalence of your device to a legally premarket nothcation: "The PPF annums to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on any -4639. Also, please note the regulation entitled, "Misbranding Other of Othphanos at ( notification" (21CFR Part 807.97) you may obtain. Other general of releveloc to premanted non-on-er the Act may be obtained from the Division of Small miormation on your respensional Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Intended Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: _________________________________________________________________________________________________________________________________________________________________
## Indications for Use
The LEONARDO syngo Cardiology Workstation is a medical diagnostic workstation for The LEONARDO bylgo conmunication, and storage of medical images and data on exchange media.
The LEONARDO syngo Cardiology Workstation can be configured as a stand-alone diagnostic review and post-processing workstation with a variety of syngo or Windows agnoste forfow and possible intended to assist the physician in diagnosis or treatment planning. This includes commercially available post-processing techniques.
(Please do not write below this line - continue on another page if needed) ( 1 loabo do 110 v 1120 v 11210 - 314
----------------------------------------------------------------------------------------------------------------------------------------Concurrence of the CDRH, Office of Device Evaluation (ODE)
V Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ OR Prescription Use _ (Per 21 CFR 801.109)
David R. Ingram
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.