K152928 · Siemens Medical Solutions USA, Inc. · OWB · Nov 13, 2015 · Radiology
Device Facts
Record ID
K152928
Device Name
Multitom Rax
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OWB · Radiology
Decision Date
Nov 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Multitom Rax is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Multitom Rax is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from but not limited to gastrointestinal examinations, cranial, skeletal, thoracic and lung exposures as well as examination of the urogenital tract. The unit may also be used in lymphography, endoscopy, myelography, venography, pediatrics, arthrography, interventional radiography, digital angiography and digital subtraction angiography (DSA). The Multitom Rax may be used for outpatient and emergency treatment, as well as for mobile transport (wheelchair and bed) examinations. The Multitom Rax is not for mammography examinations.
Device Story
Multitom Rax is a stationary X-ray system for radiography and fluoroscopy; comprises floor-mounted patient table, ceiling-suspended X-ray tube, and ceiling-suspended solid-state X-ray imager (SSXI). System synchronizes tube and detector movements for rotation around a center; series imaging supports 3D post-processing. Operated by healthcare professionals in clinical settings (outpatient, emergency). Features 'RAXalign' for automated tube-to-detector/detector-to-tube alignment; includes organ programs with stored positions for rapid movement. Collision control software manages increased movement flexibility. Automatic exposure control utilizes 5 dose measurement fields. Output consists of digital X-ray images for visualization of anatomical structures; aids clinicians in diagnostic and interventional decision-making. Benefits include comprehensive imaging capabilities, pediatric-specific programs, and efficient positioning for diverse clinical examinations.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing performed to internal functional specifications, software requirements, and risk hazards. Compliance with IEC and ISO standards confirmed. Clinical image quality evaluation was not conducted as the device utilizes the same solid-state X-ray imager (SSXI) as the primary predicate.
Technological Characteristics
Stationary X-ray system; floor-mounted table, ceiling-suspended tube and SSXI. Energy source: X-ray generator. Connectivity: DICOM compliant. Software: Includes collision control, automated alignment (RAXalign), and automatic exposure control (5-field). Sterilization: Not applicable. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 62366, ISO 14971, IEC 62304, IEC 60601-2-28, IEC 60601-2-54, IEC 61910-1, NEMA PS 3.1-3.20, IEC 60825-1, ISO 10993-1, IEC 60601-2-43.
Indications for Use
Indicated for radiographic and fluoroscopic diagnostic imaging of anatomical structures, including gastrointestinal, cranial, skeletal, thoracic, lung, urogenital, lymphographic, endoscopic, myelographic, venographic, pediatric, arthrographic, and interventional procedures. Suitable for outpatient, emergency, and mobile transport settings. Contraindicated for mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal are three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 13, 2015
Siemens Medical Solutions USA, Inc. % Mr. Darren Dorman Regulatory Affairs Specialist 40 Liberty Blvd., 65-1A MALVERN PA 19355
Re: K152928
Trade/Device Name: Multitom Rax Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Codes: OWB, JAA Dated: October 5, 2015 Received: October 6, 2015
Dear Mr. Dorman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Oaks
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K152928
Device Name Multitom Rax
### Indications for Use (Describe)
The Multitom Rax is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations.
The Multitom Rax is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from but not limited to gastrointestinal examinations, cranial, skeletal. thoracic and lung exposures as well as examination of the unit may also be used in lymphography, endoscopy, myelography, pediatrics, arthrography, interventional radiography, digital angiography and digital subtraction angiography (DSA).
The Multitom Rax may be used for outpatient and emergency treatment, as well as for mobile transport (wheelchair and bed) examinations.
The Multitom Rax is not for mammography examinations
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in all capital letters. The color of the text is a light blue or turquoise color.
## 510(k) Summary: Multitom Rax
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355
October 2, 2015 Date Prepared:
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### 1. General Information:
Importer / Distributor: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869
Manufacturing Site: SIEMENS AG Sector Healthcare Siemensstraße 1 D-91301 Forchheim, Germany Establishment Registration Number: 3004977335
#### 2. Contact Person:
Mr. Darren Dorman Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard · Mail Code 65-1A Malvern, PA 19355, USA Phone: +1 610 448-6483 ▪ Fax: +1 610 640-4481 Email: darren.dorman@siemens.com
## Alternate Contact Person :
Ms. Patricia D. Jones Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, Mail code 65-1A Malvern, PA 19355, USA Phone : +1-610-448-6474. Fax : +1-610-640-4481 Email : patricia.d.iones@siemens.com
#### 3. Device Name and Classification:
{4}------------------------------------------------
4.
Healthcare
Multitom Rax
| Trade Name: | Multitom Rax |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device: | Interventional fluoroscopic X-ray system |
| Regulation Description: | Image-intensified fluoroscopic x-ray system |
| Regulation Medical Specialty: | Radiology |
| Review Panel: | Radiology |
| Product Code: | OWB |
| Subsequent Product Code: | JAA |
| Submission Type: | Traditional 510(k) |
| Regulation Number: | 892.1650 |
| Device Class | 2 |
| Legally Marketed Predicate Devices:<br>Primary Predicate: | |
| Trade Name: | AXIOM Luminos dRF |
| 510(k) Number: | K062623 |
| Device: | Interventional Fluoro X-Ray System |
| Regulation Description: | Image-intensified fluoro X-ray System |
| Regulation Medical Specialty: | Radiology |
| Review Panel: | Radiology |
| Product Code: | OWB |
| Subsequent Product Codes: | JAA, OXO |
| Submission Type: | Traditional 510(k) |
| Regulation Number: | 892.1650 |
| Device Class | 2 |
| Recall Information: | Z-0326-2009; Z-0327-2009; Z-0717-<br>2011; Z-1159-2012; Z-1383-2012; Z-<br>1848-2013; Z-0016-2014; Z-1845-2014;<br>Z-0374-2014; Z-0140-2014; Z-1711-2014;<br>Z-2292-2014; Z-2649-2014 |
| Secondary Predicate: | |
| Trade Name: | AXIOM Aristos FX Plus |
| 510(k) Number: | K061054 |
| Device: | Solid State X-Ray Imager |
| Regulation Description: | Image-intensified fluoro X-ray System |
| Regulation Medical Specialty: | Radiology |
| Review Panel: | Radiology |
| Product Code: | MQB |
| Submission Type: | Special 510(k) |
| Regulation Number: | 892.1680 |
| Device Class | 2 |
| Recall Information: | Z-0984-03; Z-0163-2009; Z-1959-2011;<br>Z-0860-2012; Z-1583-2015 |
{5}------------------------------------------------
#### 5. Device Description:
The Multitom Rax is a stationary X-ray system for radiography and fluoroscopy. The Multitom Rax consists of a floor mounted patient table and ceiling suspended X-ray tube and a ceiling suspended solid state X-ray imager (SSXI). Together with an X-ray generator and a digital imaging system the Multitom Rax provides comprehensive image acquisition modes to support radiographic and fluoroscopic imaging procedures. X-ray tube and SSXI suspension movements are synchronized to provide rotation around a center. Series imaging acquired during the rotation are provided to 3D post-processing workstations.
#### 6. Indication for Use:
The Multitom Rax is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, anqiography and pediatric examinations.
The Multitom Rax is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from but not limited to gastrointestinal examinations, cranial, skeletal, thoracic and lung exposures as well as examination of the urogenital tract. The unit may also be used in lymphography, endoscopy, myelography, venography, pediatrics, arthrography, interventional radiography, digital angiography and digital subtraction angiography (DSA).
The Multitom Rax may be used for outpatient and emergency treatment, as well as for mobile transport (wheelchair and bed) examinations.
The Multitom Rax is not for mammography examinations.
#### 7. Substantial Equivalence:
The new device Multitom Rax is within the same classification regulation for the same indication for use as the Primary predicate AXIOM Luminos dRF (K062623, cleared 08/22/2007). The mechanical design is compared to the Secondary predicate the AXIOM Aristos FX Plus (K061054, cleared 05/09/2006). Documentation is provided to support a claim of substantial equivalence to these Siemens' predicate devices.
### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Devices:
Changes implemented for the subject device include hardware and software changes. The subject device, Multitom Rax, uses the same X-ray generator, the same X-ray tube and the same SSXI with the same digital imaging system and same image processing software as the Primary predicate AXIOM Luminos dRF. The system can be controlled in the room or remotely in the same fashion as the Primary predicate AXIOM Luminos dRF.
{6}------------------------------------------------
The mechanical design concept is the same as for the Secondary predicate the AXIOM Aristos FX Plus.
The subject device was changed in reqard to movement flexibility of the suspension arms and increased number of predefined automatic positions of tube and SSXI. Dedicated positions for certain examinations can be stored with the organ programs. When such an organ program is selected the Multitom Rax automatically moves to that position. This feature was also available with the secondary predicate AXIOM Aristos FX Plus.
The software algorithm for this movement has been optimized to find the fastest way to that position. There are hardware and software collision control features in place as well. The scanner offers both automated tube-todetector and detector-to-tube alignment. This feature is called: "RAXalign." Due to the increase in movement speed and the addition of circular movement a more sophisticated collisions software was implemented.
The automatic exposure control processes the data from 5 dose measurement fields instead of 3 compared with the predicate devices. The number of organ programs was increased and additional programs are provided for pediatric applications.
## Summary of Non-Clinical Test Data:
Multitom Rax is a new device based on components used with both predicate devices AXIOM Luminos dRF(K062623) and AXIOM Aristos FX Plus (K061054) The Multitom Rax design was completed in accordance with Siemens Quality Management System Design Controls and Engineering, standards compliance, and Verification and Validation testing were successfully conducted. Tests were performed on the Multitom Rax which demonstrated that the device is safe and effective, performs comparably to the predicate devices, and is substantially equivalent to the predicate devices. Tests included verification/validation testing to internal functional specifications (including software). The Multitom Rax uses the same solid state X-ray imager (SSXI) as the Primary predicate. X-ray geometry and techniques are the same so that clinical image comparisons involving SSXI were not conducted. Documentation provided demonstrates compliance of the new device to all FDA requirements stated in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. including results of verification/validation tests to software requirements and software risk hazards.
Performance testing confirmed that the Multitom Rax complies with 21 CFR 1020.30-32 Federal Performance Standards for X-Ray Fluoroscopic equipment and with relevant voluntary safety standards for Electrical Safety
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in all capital letters. The color of the text is a light blue or teal color.
and Electromagnetic Compatibility testing, specifically IEC standards listed in the table
below. Together, the verification/validation activities successfully confirmed device requirements has been fulfilled and that system functionality is consistent with the user needs and intended uses. The Multitom Rax device correctly performs as designed and raises no new questions regarding safety or effectiveness. Therefore, when compared to the predicate devices the Multitom Rax supports a determination of substantial equivalence to the predicate devices.
#### 9. Summary of Non-Clinical Tests
The Siemens Multitom Rax complies with the voluntary standards as listed in the following table:
| Development<br>Organization and<br>Reference Number | Title of Standard |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 3rd, 2005 | Medical Electrical Equipment - Part 1: General Requirements<br>for Safety |
| IEC 60601-1-2 3rd<br>2007 | Medical Electrical Equipment - Part 1-2: General requirements<br>for basic safety and essential performance - Collateral<br>Standard: Electromagnetic Compatibility -- Requirements and<br>Tests |
| IEC 60601-1-3: 3rd<br>2008 | Medical electrical equipment - Part 1-3: General requirements<br>for basic safety and essential performance - Collateral<br>Standard: Radiation protection in diagnostic X-ray equipment |
| IEC 62366, 2007 | Medical devices - Application of usability engineering to<br>medical devices |
| ISO 14971, 2010 | medical devices - application of risk management to medical<br>devices |
| IEC 62304 Ed. 1.0,<br>2006 | Medical device software - Software life cycle processes |
| IEC 60601-2-28 Ed<br>2.0, 2010 | Medical electrical equipment - Part 2-28: Particular<br>requirements for the basic safety and essential performance of<br>X-ray tube assemblies for medical diagnosis |
| IEC 60601-2-54, 2009 | Medical electrical equipment - Part 2-54: Particular<br>requirements for the basic<br>safety and essential performance of X-ray equipment for<br>radiography and<br>radioscopy |
| IEC 61910-1, 2014 | Medical electrical equipment - Radiation dose documentation -<br>Part 1: Radiation dose structured reports for radiography and<br>radioscopy (IEC 61910-1:2014) |
| NEMA PS 3.1 - 3.20,<br>2011 | Digital Imaging and Communications in Medicine (DICOM) Set |
| IEC 60825-1, 2007 | Safety of laser products - Part 1: Equipment classification, and<br>requirements |
{8}------------------------------------------------
| ISO 10993-1. 2009 | Biological evaluation of medical devices -- Part 1: Evaluation<br>and testing within a risk management process |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-2-43, 2010 | Medical electrical equipment - Part 2-43: Particular<br>requirements for the safety of X-ray equipment for<br>interventional procedures |
#### 10 Summary of Clinical Tests:
For the subject of this premarket submission, Siemens did not do an evaluation of the clinical image quality as required by the "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices." because the subject device uses the same SSXI as the Primary predicate device.
#### 11. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Multitom Rax is continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed.
Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
#### 12. Conclusion as to Substantial Equivalence:
The Multitom Rax is intended for the same indications for use as the Primary predicate AXIOM Luminos dRF. The operating environment is the same and mechanical design the same as the Secondary predicate the AXIOM Aristos FX Plus. Siemens concludes via the documentation provided in this 510(k) submission that the Multitom Rax is substantially equivalent to the AXIOM Luminos dRF (K051133 cleared 06/01/2005) and the AXIOM Aristos FX Plus (K061045 cleared 05/09/2006).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.