K061054 · Siemens Medical Solutions USA, Inc. · MQB · May 9, 2006 · Radiology
Device Facts
Record ID
K061054
Device Name
AXIOM ARISTOS FX PLUS
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
MQB · Radiology
Decision Date
May 9, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The AXIOM Aristos FX Plus is a dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems. The MPRS allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremities. It is not designed for mammography use. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions. The intended use and indications for use of this modified device as described in its labeling have not changed from its predicate device.
Device Story
Stationary x-ray system with flat panel detector; ceiling-mounted telescoping arms support x-ray tube and detector. Replaces conventional film/screen systems. Includes wireless remote control using RF technology for clinician-operated system movements; short-range, low-power radio link; safety-oriented communication protocol. Used in clinical settings by healthcare professionals. System provides visual/audible warnings and error messages for safety monitoring. Output is digital radiographic image for diagnostic decision-making.
Clinical Evidence
No clinical data. Bench testing only. System performance verified through established industry practices and final performance testing.
Technological Characteristics
Stationary x-ray system; flat panel detector; ceiling-mounted telescoping arms. Includes generator, x-ray tube, collimator, and wireless RF remote control. Safety features include visual/audible warnings and system error monitoring. Connectivity via proprietary communication protocol for remote control. Sterilization not applicable.
Indications for Use
Indicated for radiographic imaging of the whole body, including skull, spinal column, chest, abdomen, and extremities in patients of all ages. Contraindicated for mammography. Patient may be in sitting, standing, or supine positions.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
осыны 1 ггодку заншнагу
K061054
FOR
# SIEMENS AXIOM ARISTOS FX PLUS
510(K) SUMMARY
### WIRELESS REMOTE CONTROL
MAY - 9 2006
Submitted by: Siemens Medical Solutions USA. Inc. 51 Valley Stream Parkway Malvern, PA 19355
## April 14, 2006
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### 1. Contact Person:
Mr. Gary Johnson Technical Specialist, Regulatory Affairs Submissions
Siemens Medical Solutions USA, Inc 51 Valley Stream Parkway E-50 Malvern, PA 19355-1406 Phone:(601) 448-1778 Fax: (610) 448-1787
#### 2. Device Name and Classification
Product Name: Classification Name: Classification Panel: CFR Section: Device Class: Product Code:
AXIOM Aristos FX Plus Solid State X-ray Imager (SSXI) Radiology 21 CFR §892.1680 Class II 90MQB
#### 3. Intended Use:
The AXIOM Aristos FX Plus is a dedicated x-ray system with a flat panel detector which allows the acquisition of x-ray exposures without the use of conventional film/screen systems. The MPRS allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, and extremities. It is not designed for mammography use. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions. The intended use and indications for use of this modified device as described in its labeling have not changed from its predicate device.
Siemens Medical Solutions USA, Inc. Special 510(k) for AXIOM Aristos FX Plus
Page 7
CONFIDENTIAL
{1}------------------------------------------------
#### 4. Device Description:
The AXIOM Aristos FX Plus is designed as a set of two ceiling mounted telescoping arms. One arm is supporting the x-ray tube with the collimator; the other arm is supporting the FD (Flat panel detector) with new detector housing and two hand grips. Different table tops, including a painted one for children are available. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions.
A Wireless Remote Control using RF technology has been designed to allow the clincian to remotely operate the system movements from virtually anywhere in the room through a short-range, low-power radio communication link between a hand held transmitter and the receiver. Additional hardware (independent of microcontroller) and Siemens own communication protocol layers will provide a safety-oriented signal transmission.
#### ·Substantial Equivalence: ಕ್ಕೆ.
The AXIOM Aristos FX Plus stationary x-ray system is designed with a Flat Panel Detector. The ceiling mounted configuration allows acquisition of radiographic exposures of various anatomical regions of the body. It is substantially equivalent to the following SIEMENS Medical Systems devices:
AXIOM Aristos FX K013826, cleared on 2/4/02
The Wireless Remote Control is substantially equivalent to the wireless footswitch for AXIOM Artis system.
AXIOM Artis with Wireless Footswitch K052202, cleared on 3/7/06
A detailed Substantial Equivalence Comparison is provided in Section 5.
### Summary of Technological Characteristics of the Principal Device as compared with the 6. Predicate Device:
Many of the components (Generator, X-ray tube, Imaging system, Collimator, FD detector, Wireless Remote Control) used in the AXIOM Aristos FX Plus are either commercially available with current Siemens AXIOM Aristos FX system or include minor modifications to existing components.
Siemens Medical Solutions USA, Inc. Special 510(k) for AXIOM Aristos FX Plus
Page 8
CONFIDENTIAL
{2}------------------------------------------------
### General Safety and Effectiveness Concerns: 7.
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition, the AXIOM Aristos FX Plus System is continually monitored, and if an error occurs an error message will be displayed.
Furthermore, the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
Siemens Medical Solutions USA, Inc. Special 510(k) for AXIOM Aristos FX Plus
CONFIDENTIAL
Page 9
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
MAY - 9 2006
Mr. Gary L. Johnson Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway MALVERN PA 19355
Re: K061054
Trade/Device Name: AXIOM Aristos FX Plus Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: April 14, 2006 Received: April 17, 2006
Dear Mr. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image shows the FDA Centennial logo. The logo is circular and contains the letters "FDA" in the center. The words "Centennial" are below the letters. The years "1906 - 2006" are at the top of the logo. Below the logo is the text "Protecting and Promoting Public Health".
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| (Gastroenterology/Renal/Urology) 240-276-0115 | |
|-----------------------------------------------|--------------|
| (Obstetrics/Gynecology) | 240-276-0115 |
| (Radiology); | 240-276-0120 |
| | 240-276-0100 |
| | |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# SECTION 3 INDICATION FOR USE
510(k) Number (if known):
K061054
Device Name:
AXIOM Aristos FX Plus
AXIOM Aristos FX Plus is a dedicated x-ray system with a flat panel detector, which allows the acquisition of x-ray exposures without the use of conventional film/screen systems. The AXIOM Aristos FX Plus allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities, excluding mammography. Radiographic exposures may be taken with the patient in the sitting, standing, or lying positions. The intended use and indications for use of this modified device as described in its labeling have not changed from its predicate device.
(Please do not write below this line - continue on another page if needed)
pen source with MIT license
Concurrence of the CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use
Over-The-Counter Use
(Per 21 CFR §801.109)
David B. Lagemann
(Division Sign-Off) Division of Reproductive, and Radiological Devic 510(k) Number
Siemens Medical Solutions USA, Inc. Special 510(k) for AXIOM Aristos FX Plus
CONFIDENTIAL
. . . .
Page 17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.