K152596 · Intact Medical Corporation · KNW · Feb 15, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K152596
Device Name
Intact® Gen2 System
Applicant
Intact Medical Corporation
Product Code
KNW · Gastroenterology, Urology
Decision Date
Feb 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Intact® Gen2 System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities. The Intact® Gen2 System is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality. The Intact® Gen2 System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality. The Intact® Gen2 System is intended to preserve lesion architecture in samples with a diameter of 12-30mm. The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpable or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Intact® Gen2 System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Device Story
High-frequency, vacuum-assisted electrosurgical biopsy system; captures intact breast tissue samples for histologic examination. System components: sterile single-use probe/wand (12-30mm), handle, controller, foot pedal, table mount, cart. Operation: probe uses five wire electrodes to cut tissue via mono-polar electrosurgical current; electrodes purse to capture tissue volume. Controller uses microprocessor to identify wand size via unique resistor and optimize power delivery; minimizes thermal necrosis. Used in clinical settings (stereotactic or ultrasound-guided) by physicians. Output: intact tissue specimen. Benefits: preserves lesion architecture for accurate diagnosis; reduces thermal artifact compared to predicate.
Clinical Evidence
Bench testing only. Testing included safety/essential performance (IEC 60601-1, IEC 60601-2-2), EMC (IEC 60601-1-2), software verification (IEC 62304), usability (EN 60601-1-6), and bench performance verification. Bench performance demonstrated that the system does not deliver more total energy to the biopsy sample than the predicate. Thermal artifact testing confirmed equivalence to the predicate.
Indicated for patients requiring diagnostic tissue sampling of breast abnormalities (imaged or palpable). Used for partial or complete removal of imaged abnormalities or partial removal of palpable abnormalities (e.g., fibroadenoma, fibrocystic lesions).
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
Intact® BLES with Intact® Excision XL wand and handle (K142477)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 15, 2016
Intact Medical Corporation % Dr. John Smith, M.D., J.D. Hogan Lovells US LLP 555 Thirteenth St. NW Washington, District of Columbia 22204
Re: K152596
Trade/Device Name: Intact® Gen2 System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: January 19, 2016 Received: January 19, 2016
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K152596
#### Intact® Gen2 System Device Name:
#### Indications for Use (Describe)
- · The Intact® Gen2 System is indicated to provide tissue sampling of breast abnormalities.
- · The Intact® Gen2 System is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality.
- · The Intact · Gen2 System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
- The Intact Gen2 System is intended to preserve lesion architecture in samples with a diameter of 12–30 mm. .
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpable or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Intact Gen2 System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue marqins be examined for completeness of removal using standard surgical procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including
the time to review instructions, search existing data sources, gather and maintain the data needed and
complete and review the collection of information. Send comments regarding this burden estimate or any
other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human
Services Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA)
Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Intact® Gen2 System | |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Intact Medical, Inc. |
| Establishment Registration Number | 1226766 |
| Contact Person | John Vacha<br>Intact Medical Corporation<br>550 Cochituate Road - Suite 25<br>East Wing, Floor 4<br>Framingham, MA 01701<br>Phone: 508-655-7820<br>Fax: 508-655-7822 |
| Summary Date | February 12, 2016 |
| Proprietary Name | Intact® Gen2 System |
| Classification | Class II |
| Classification Name | Biopsy Instrument and Electrosurgical Cutting &<br>Coagulation Device & Accessories |
| Regulation Number | 21 C.F.R. § 876.1075 |
| Classification | Product Code KNW |
| Predicate Device | Intact® BLES with Intact® Excision XL wand and<br>handle (K142477) |
### Device Description
The intent of the device is to be a biopsy tool to capture a sample of breast tissue in one intact sample to preserve the tissue architecture for histologic examination. The device is a highfrequency, vacuum-assisted electrosurgical device used to remove tissue by electrosurgical cutting and simultaneous capture of an incised tissue volume. The device includes the following components:
# 1. Probe/Wand
Intact® Gen2 Probe/wand (Sizes: 12mm, 15mm, 20mm and 30mm)
- 2. Handle
- 3. Gen2 Controller with Power Cord and Foot Pedal
- 4. Table Mount
{4}------------------------------------------------
Available to fit stereotactic tables (Fischer and Lorad mounts).
# 5. Cart
# Probe/Wand:
The probes/wands are sterile and intended for single use. They utilize a sharp blade to access the target lesion.
The Cut/Capture Electrode is located at the distal end of the wand. It consists of five (5) small diameter wire electrodes that cut tissue with a mono-polar electrosurgical cutting current and purse down to close the distal end of the cutting/capture element to make a circumscribing incision and capture of the target tissue.
The wand contains a resistor of unique value for each wand size (12mm, 15mm, 20mm, or 30mm) that is used by the Controller to identify the specific wand size installed in the handle.
### Handle:
The handles consist of a hand piece at the distal end of a cable that mates with the Probe/Wand, and a connector at the proximal end of a cable that mates with the Controller. The handle is either mated to a stereotactic table using a mount or handheld for ultrasound-guided procedures.
#### Controller:
The Intact Gen2 Controller consists of a radiofrequency generator, which operates at a single, preset frequency. The Intact Gen2 Controller is designed to apply a high-frequency voltage during the biopsy specimen capture phase (involving Capture Electrodes). The power applied for each wand size is optimized to (1) be sufficiently high to form the electrical arc essential to effect the highly localized vaporization of cellular fluid which, thereby, enables electrosurgical cutting and (2) be sufficiently low to minimize the depth of necrosis induced in the biopsy specimen and the surrounding tissue. The controller incorporates a microprocessor and software.
### Power Cord:
The Intact® Controller Power Cord is a standard detachable power cord that connects the Controller to the line voltage output receptacle.
### Foot Pedal:
The foot pedal assembly is a floor-mounted, two-pedal device, which attaches to the rear of the Intact" Gen2 Controller. The two pedals provide the Arm and Capture functions, analogous to those buttons on the handle. Either the footswitch or the handle controls can be used interchangeably.
### Cradle:
Multiple versions of the cradle are available to fit commercially available stereo-tactic tables. .
{5}------------------------------------------------
## Cart:
This is an off-the-shelf moveable cart that houses the components of the system.
Required for the procedure but not included in the system are: An off-the-shelf vacuum unit and a patient return electrode.
# Intended Use/Indications for Use
- · The Intact · Gen2 System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities.
- · The Intact® Gen2 System is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality.
- · The Intact® Gen2 System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
- · The Intact® Gen2 System is intended to preserve lesion architecture in samples with a diameter of 12-30mm.
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpable or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Intact" Gen2 System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
# Technical and Operational Characteristics
The Intact Gen2 System works on the same principle of operation as the predicate device. Both require setting up a high-frequency voltage between an electrosurgical electrode (i.e., active electrode) and the tissue to be incised. This voltage creates an arc between the electrode and the tissue. This arc, by virtue of the high temperature, induced at the point of impingement with tissue, vaporizes the tissue in its vicinity.
This 510(k) notice implements technological modifications to the wands (addition of a resistor) and controller (addition of a microprocessor and software) to individualize the amount of power required to capture a sample based on probe size. Thus, less power is required to capture a tissue sample
{6}------------------------------------------------
for each sized probe. The proposed device is capable of performing to its intended use similar to the predicate device.
{7}------------------------------------------------
# Summary of Testing
| Verification Testing Summary Table | | | |
|----------------------------------------------|------------------------------------|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Test Type | Proposed<br>Intact® Gen2<br>System | Predicate<br>Intact® BLES<br>System Intact<br>BLES<br>(K142477) | Comments |
| Safety and<br>Essential<br>Performance Tests | Pass | Pass | Equivalent<br>Both systems meet requirements of IEC<br>60601-1, 3rd Edition<br>and IEC 60601-2-2 |
| EMC Tests | Pass | Pass | Equivalent<br>Both systems meet requirements of IEC<br>60601-1-2 |
| Software<br>Verification Tests | Pass | Not applicable | Proposed software meets requirements of<br>IEC 62304.<br>Predicate contains no software, utilized<br>hardware for all functions. |
| Usability Tests | Pass | Pass | Equivalent<br>Both systems meet the requirements of<br>EN 60601-1-6. |
| Bench Performance<br>Verification Test | Pass | Pass | Equivalent<br>Demonstrated that the proposed system<br>does not deliver more total energy to the<br>biopsy tissue sample than the predicate. |
| Thermal Artifact<br>Verification Testing | Pass | Pass | Equivalent<br>Demonstrated that the thermal artifacts in<br>biopsy tissue samples are equivalent |
# Rationale for Substantial Equivalence
The proposed modifications reduce the power introduced to the patient to capture a similar intact sample (size, shape, weight, architectural integrity, thermal artifact) as that of the predicate device. The proposed device is capable of performing to its intended use similar to the predicate device.
The Intended Use/Indications for Use statement and technological characteristics of the Intact Gen2 System device and the predicate devices were compared. The Intended Use/Indications for Use statements of the devices are identical. The proposed changes to the technology are supported by extensive bench and animal testing described above and are similar to that of the predicate device with the same intended use.
# Conclusion
Based on similar technology and operation characteristics as well as the bench and animal performance testing, the Intact Gen2 System device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.