Intact Advance and Intact Excision XL, Intact BLES
K142477 · Intact Medical Corporation · KNW · Dec 15, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K142477
Device Name
Intact Advance and Intact Excision XL, Intact BLES
Applicant
Intact Medical Corporation
Product Code
KNW · Gastroenterology, Urology
Decision Date
Dec 15, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K142477 · Dec 15, 2014
Intact Advance and Intact Excision XL, Intact BLES
Intact Medical Corporation
Peer-reviewed clinical literature; Clinical study cohorts (e.g., IPET trial)
The sponsor provided a summary of medical literature documenting the performance of the Intact BLES and Excision XL wand in clinical practice to support an expanded Indications for Use claim regarding the complete excision of breast abnormalities.
Literature review; Complete excision rates; Breast biopsy; Clinical performance
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review of Intact BLES/Excision XL clinical use; Retrospective/Observational clinical literature review
Patients with breast abnormalities undergoing percutaneous biopsy; Sample Size: 224
Not applicable for this study
Complete Excision Rate
Indications for Use
The Intact® BLES with Intact® Excision XL wand and handle is indicated to provide tissue samples for diagnostic sampling of breast abnormalities. The Intact® BLES with Intact® Excision XL wand and handle is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality. The Intact® BLES with Intact® Excision XL wand and handle is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality. The Intact® BLES with Intact® Excision XL wand and handle is intended to preserve lesion architecture in samples with a diameter of 12-30mm. The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpable or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Intact BLES with Intact® Excision XL wand and handle may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Device Story
High-frequency, vacuum-assisted electrosurgical biopsy system; captures intact breast tissue samples to preserve architecture for histology. Components: sterile single-use probe/wand with distal cut/capture electrode; handle; Model 3000 Controller (RF generator); foot pedal; vacuum source; table mounts for stereotactic guidance. Operation: probe inserted to target lesion; monopolar RF current performs circumscribing incision; vacuum assists tissue capture. Used in clinical settings by physicians for percutaneous stereotactic or ultrasound-guided procedures. Output: intact tissue specimen for pathologist evaluation. Benefits: enables definitive diagnosis and partial/complete removal of lesions; reduces need for open surgical excision.
Clinical Evidence
Literature review of 5 studies (n=224 patients) evaluating complete excision rates as a surrogate for diagnostic capability. Reported complete excision rates ranged from 49% to 100%, with a total aggregate rate of 76%.
Indicated for patients with breast abnormalities requiring tissue sampling for histologic examination. Applicable to imaged abnormalities (partial or complete removal) and palpable abnormalities (partial removal), including those classified as benign (e.g., fibroadenoma, fibrocystic lesion).
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles merging to form a single image. The profiles are arranged in a way that suggests movement or progress.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 15, 2014
Intact Medical Incorporated % Dr. John J. Smith Hogan Lovells US LLP 555 Thirteenth Street, North West Washington, District of Columbia 20004
Re: K142477
Trade/Device Name: Intact® BLES with Intact® Excision XL wand and handle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: November 21, 2014 Received: November 21, 2014
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
510(k) Number (if known) K142477
Intact® BLES with Intact® Excision XL wand and handle Device Name:
#### Indications for Use (Describe)
- The Intact® BLES with Intact® Excision XL wand and handle is indicated to provide tissue sampling of breast . abnormalities
- . The Intact® BLES with Intact® Excision XL wand and handle is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality
- The Intact® BLES with Intact® Excision XL wand and handle is intended to histologic examination with partial removal of a palpable abnormality
- . The Intact® BLES with Intact® Excision XL wand and handle is intended to preserve lesion architecture in samples with a diameter of 12-30mm
The extent of a histologic abormality cannot always be readly determined from palpation or imaged appearance. Therefore of removal of the palpable or imaged evidence of an abnormality does not premoval of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essettial that the tissue margins be examined for completeness of removal usings standard surgical procedures.
In instances when a patient presents with a palpable absornality that has been classified as benign through criteria (e.g., fibroadenoma, fibrocystic lesion), the Intact® Excision XL wand and handle may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
#### 510(k) Summary
Intact® BLES with Intact® Excision XL wand and handle
| Applicant | Intact Medical Corporation<br>550 Cochituate Road - Suite 25<br>Framingham, MA 01701<br>Phone: 508-655-7820<br>Fax: 508-655-7822 | |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Establishment Registration Number | 1226766 | |
| Contact Person | John J. Smith, M.D., J.D.<br>Hogan Lovells US LLP<br>555 Thirteenth St. NW<br>Washington, DC 20004<br>Phone: 202-637-3638<br>Fax: 202-637-5910 | |
| Summary Date | December 15, 2014 | |
| Proprietary Name | Intact® BLES with Intact® Excision XL wand and handle | |
| Classification | Class II | |
| Classification Name | Biopsy Instrument and Electrosurgical Cutting &<br>Coagulation Device & Accessories | |
| Regulation Number | 21 C.F.R. § 876.1075 | |
| Classification | Product Code KNW | |
| Predicate Devices | 10mm, 12mm, 15mm and 20mm Intact® BLES<br>(K060413) manufactured by Intact Medical Corporation<br>Intact® Advance and Intact Excision XL (K081057<br>"Primary Predicate") manufactured by Intact Medical | |
#### Device Description
The intent of the device is to be a biopsy tool to capture a sample of breast tissue in one intact sample to preserve the tissue architecture for histologic examination. The device is a high frequency, vacuumassisted electrosurgical devices used to remove tissue by electro-surgical cutting and simultaneous capture of an incised tissue volume. The devices include the following components:
#### 1. Probe/Wand
- (1a) Intact BLES Probe/wand
- (1b) Intact Advance BLES Probe/wand
- (1c) Intact Excise XL Probe/wand
{4}------------------------------------------------
# 2. Handle
(2) The Intact BLES Handle
- (2a) Intact Excise XL Handle
- 3. Intact Medical ™ Model 3000 Controller
- 4. Controller Power Cord Footpedal
- 5. Table Mount
- (6a) Table Mount, BLES attaches the Intact BLES Handle to a stereotactic table (Fischer and Lorad mounts).
- (6b) Table Mount, Excise XL attaches the Intact Excise XL Handle to a stereotactic table (Fischer and Lorad mounts).
- 6. Vacuum Source
- 7. Cart
# Probe/Wand:
The probes/wands are sterile and intended for single use. They utilize a sharp blade to access the target lesion.
The Cut/Capture Electrode is located at the distal end of the wand. It consists of five (5) small diameter wire electrodes that cut tissue with a mono-polar electro-surgical cutting current and purse down to close the distal end of the cutting/capture element to make a circumscribing incision and capture of the target tissue.
# Handle:
The handles consist of a hand piece at the distal end of a cable that mates with the Probe/Wand, a connector at the proximal end of a cable that mates with the Controller, a grip provided for ultrasound hand held procedures. The grip is replaced with a cradle for stereotactically guided procedures.
# Controller:
Intact Medical Model 3000 Controller consists of a radiofrequency generator, which operates at a single, preset frequency to apply a high-frequency voltage during the biopsy specimen capture phase.
# Power Cord:
The Intact Controller Power Cord is a standard detachable power cord that connects the Controller to the line voltage output receptacle.
# Foot Pedal:
The foot pedal assembly is a floor-mounted, three-pedal device, which attaches to the Intact Model 3000 Controller. The three pedals provide probe and handle controls to allow the user an alternate interface (to the handle buttons).
# Cradle:
Multiple versions of the cradle are available to fit commercially available stereo-tactic tables.
# Vacuum Source:
An off the shelf unit that allows for suction and removal of fluids during the procedure
{5}------------------------------------------------
#### Intended Use/Indications for Use
- . The Intact® BLES with Intact® Excision XL wand and handle is indicated to provide tissue samples for diagnostic sampling of breast abnormalities
- . The Intact® BLES with Intact® Excision XL wand and handle is intended to provide breast tissue for histologic examination with partial or complete removal of an imaged abnormality
- The Intact® BLES with Intact® Excision XL wand and handle is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality
- . The Intact® BLES with Intact® Excision XL wand and handle is intended to preserve lesion architecture in samples with a diameter of 12-30mm
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpable or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Intact BLES with Intact® Excision XL wand and handle may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histologic evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
# Technical and Operational Characteristics
The proposed 510k is limited to a change in label claims. No changes have been made to the device. The Intact® BLES with Intact® Excision XL wand and handle is identical in design, operational and technological characteristics as the predicate devices (K060413; K081057) and supports that no new safety concerns are being raised by change in intended use/indications for use statement and thus raises no new issues of safety or effectiveness. ).
# Summary of Testing
The Intact® BLES with Intact® Excision XL wand and handle was previously subjected to a comprehensive test program that included electrical safety and electromagnetic compatibility testing, animal testing, clinical testing, biocompatibility and sterilization testing. Since there is no change to the technology of the Intact® BLES with Intact® Excision XL wand and handle and the only proposed change is limited to the Indications for Use, no additional bench or animal testing was conducted to support demonstration of substantial equivalence in performance from the Intact® Excision XL wand and handle to its predicate device.
However, additional clinical data was generated to support the proposed additional indications for use claim. Complete Excision Rates were evaluated as a surrogate measure to assess the ability to remove sufficient, undisrupted tissue to preserve tissue architecture and to permit a definitive diagnosis.
{6}------------------------------------------------
Provided in the table below is a tabular summary of representative medical literature documenting the use of the Intact® BLES with Intact® Excision XL wand and handle as a tool in the field of breast biopsy.
| Citation | Author | Patient Quantity | Complete Excision Rate<br>(%) |
|----------|-------------------------------------------------------------------------------------------|------------------|-------------------------------|
| 1 | Whitworth, P | 25 | 88 |
| 2 | Whitworth, P,<br>Simpson, Poller,<br>Schonholz, Turner,<br>Phillips, Johnson,<br>McEachin | 10 | 100 |
| 3 | Naglaa Abdel Razek | 76 | 76 |
| 4 | Schonholz, S | 37 | 49 |
| 5 | Steven D. Allen ,<br>Ashish Nerurkar ,<br>Guidabaldo U. Querci<br>Della Rovere | 76 | 83 |
| TOTAL | BLES | 224 | 76% |
Tabular summary of medical literature with Intact® Excision XL wand and handle
# Citation number:
- 1. Whitworth, P. Percutaneous Stereotactic Lumpectomy and Partial Breast Irradiation for Small Breast Cancers: Reducing Overtreatment Presented at: Miami Breast Meeting 2014 -Publication pending
- 2. Whitworth, Simpson, Poller, Schonholz, Turner, Phillips, Johnson, McEachin: Definitive Diagnosis for High-Risk Breast Lesions Without Open Surgical Excision: The Intact Percutaneous Excision Trial (IPET), Annals of Surgical Oncology (2011) 18:3047-3052
- Naglaa Abdel Razek. Percutaneous breast lesion system (BLES): a new tool for 3. complete closed excision of high risk lesions. European Congress of Radiology 2013.
- Schonholz, S. Complete Excision and Reexcision Rates Following Percutaneous biopsy using 4. the Intact Breast Lesion Excision System and the Mammotome. ASBS 2013 in publication for review with Journal of Surgical Oncology.
- 5. Steven D. Allen & Ashish Nerurkar &Guidabaldo U. Querci Della Rovere The breast lesion excision system (BLES): a novel technique in the diagnostic and therapeutic management of small indeterminate breast lesions? European Radiology 2011.
{7}------------------------------------------------
#### Rationale For Substantial Equivalence
The Intended Use/Indications for Use statement and technological characteristics of the Intact® BLES with Intact® Excision XL wand and handle and the predicate devices were compared. The Intended Use/Indications for Use statement of the devices have the same general claims and do not raise new questions of safety and performance. Both the predicate device and proposed 510k device an intact sample of tissue for histological assessment. This proposed change to the Indications for Use is well supported by a review of the peer reviewed clinical literature and society presentations and are similar to that of the predicate devices with the same general intended use.
#### Conclusion
Based on the similar label, identical technology and operation characteristics as well as the clinical performance testing, the Intact® BLES with Intact® Excision XL wand and handle device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.