K152587 · Dermascalp, LLC · OAP · Feb 4, 2016 · Physical Medicine
Device Facts
Record ID
K152587
Device Name
DermaScalp Laser Cap
Applicant
Dermascalp, LLC
Product Code
OAP · Physical Medicine
Decision Date
Feb 4, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DermaScalp Laser Cap is indicated to promote hair growth in females with Androgenetic Alpopecia who have Ludwig-Savin Classifications I-II and who have Fitzpatrick Skin Types I to IV.
Device Story
DermaScalp MD and MDX Laser Caps are hands-free, portable, non-invasive low-level laser devices; intended to treat Androgenetic Alopecia and promote hair growth in females. Device consists of a concave scalp covering containing an array of red laser diodes (650 nm wavelength, <5 mW per diode, continuous wave output). User wears the cap for 30-minute sessions; device delivers distributed coherent laser light to the entire scalp via bio-stimulation. Device is intended for prescription use. Performance is verified through bench testing against electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and laser safety (IEC 60825-1) standards. Clinical benefit is derived from therapeutic bio-stimulation of hair follicles.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and laser safety (IEC 60825-1 Class 3R). Performance testing verified that the device meets design specifications and operates identically to the predicate devices regarding laser wavelength, power, energy type, and treatment parameters.
Technological Characteristics
Hands-free, portable concave scalp covering; 650 nm red laser diodes; continuous wave output; <5 mW per diode; 80 diodes (MD model) or 224 diodes (MDX model); bio-stimulation modality; ISO 10993-1:2009 compliant materials; IEC 60601-1/60601-1-2 electrical/EMC safety; IEC 60825-1 Class 3R laser classification.
Indications for Use
Indicated for females with Androgenetic Alopecia, Ludwig-Savin Classifications I-II, and Fitzpatrick Skin Types I-IV to promote hair growth.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 4, 2016
Dermascalp LLC Mr. John Carullo Managing Member 1700 Alma Drive Suite 400 Plano, Texas 75075
Re: K152587
Trade/Device Name: DermaScalp Laser Cap Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: November 20, 2015 Received: January 4, 2016
Dear Mr. Carullo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K152587
### Device Name
DermaScalp Laser Cap Models: DermaScalp MD Laser Cap & DermaScalp MDX Laser Cap
#### Indications for Use (Describe)
The DermaScalp Laser Cap is indicated to promote hair growth in females with Androgenetic Alpopecia who have Ludwig-Savin Classifications I-II and who have Fitzpatrick Skin Types I to IV.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and mount of response, including the and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### ATTACHMENT 2
# 510(k) SUMMARY
| 510(k) Owner: | DermaScalp LLC<br>1700 Alma Dr. Suite 400<br>Plano, TX 75075<br>Contact: John Carullo<br>Phone: 214-683-0724 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Summary Prepared: | Sebtember 04, 2015 |
| Device: | |
| | Trade Name: DermaScalp Laser Cap |
| | Common/Classification Name: Light Therapy Hair System<br>Product Code OAP |
| | 21 C.F.R. § 890.5500 (Infrared lamp) |
| | Classification: Class II |
| Predicate Devices: | LaserCap Elite - Transdermal Cap Inc - K150613<br>LaserCap PRO - Transdermal Cap Inc - K150613 |
| Device Description: | The DermaScalp MD Laser Cap and the DermaScalp MDX Laser Cap devices<br>are a hands-free, portable, non-invasive, low-level laser device intended to<br>treat Androgenetic Alopecia (Hair Loss) and to promote hair growth in<br>Females. The DermaScalp Laser Cap devices provide distributed red laser<br>light dispersing from a "Concave" scalp covering utilizing laser modules with<br>a 650 nm wavelength, <5 mW output power, producing a continuous wave<br>"CW" output beam. The "Concave" scalp covering is designed to maximize<br>the delivery of the coherent laser light to effectively cover the entire scalp of<br>the user during treatment. |
{4}------------------------------------------------
| Intended Use: | The DermaScalp Laser Cap is indicated to promote hair growth in females with<br>Androgenetic Alpopecia who have Ludwig (Savin) Classifications I-II and<br>who have Fitzpatrick Skin Types I to IV. |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics: | The DermaScalp MD Laser Cap and the DermaScalp MDX Laser Cap devices<br>are a hands-free portable device that produce red laser energy dispersed from a<br>"Concave" scalp covering as it rests upon a user's head and creates a laser field<br>that covers the users entire scalp area. The devices produce timed treatments of<br>equally distributed laser energy to the full scalp area. The treatment received<br>from the laser energy promotes hair growth in Females and treats Androgenetic<br>Alopecia (Hair Loss) by the therapeutic modality of bio-stimulation. |
| Biocompatibility<br>Data: | The construction of the DermaScalp MD Laser Cap and the DermaScalp MDX<br>Laser Cap devices do not raise any biocompatibility issues. All the materials<br>used in the construction of the DermaScalp devices adhere to the requirements<br>of ISO 10993-1:2009, Biological Evaluation of Medical Devices - Part 1:<br>Evaluation and Testing within a Risk Management Process. |
| Performance<br>Data: | The evaluation of the perfomance data presented confirms that the DermaScalp<br>MD Laser Cap and the DermaScalp MDX Laser Cap devices have the same or<br>similar Laser Wavelength, Laser Power, Laser Energy Type, Output Mode,<br>Treatment Time, Output Beam, Laser Field Treatment Area, Consumer Usage<br>Focal Length as the FDA Cleared predicate devices.<br><br>Testing to IEC 60601-1 and 60601-1-2 confirm the devices adherence to LVD<br>electrical and EMC safely requirements. Testing to IEC 60825-1 certifies the<br>laser system to classifaction 3R, which is the same as the predicate devices.<br><br>Performance Testing is conducted to confirm compliance to design<br>specifications; all functions were verified to operate as designed. The<br>DermaScalp devices met all acceptance criteria in the performance testing. |
| Conclusions: | The DermaScalp MD Laser Cap and the DermaScalp MDX Laser Cap devices<br>are as safe and effective as the FDA Cleared predicate devices for female<br>treatment and is thefore Substantial Equivalent to the FDA Cleared predicate<br>devices with respect to intended use, technological characteristics and safety<br>characteristics. |
### Substantial Equivalence Comparison Table
{5}------------------------------------------------
| | New Device | Predicate Device |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K152587 | K150613 |
| Company | DermaScalp LLC | Transdermal Inc |
| Name | DermaScalp MD Laser Cap | LaserCap Elite |
| Indications for Use | The DermaScalp MD Laser Cap is indicated<br>to treat Androgenetic Alpopecia and to<br>promote hair growth in Females who have<br>Ludwig (Savin) I-II patterns of hair loss and<br>who have a Fitzpatrick Skin Types I to IV. | LaserCap Elite is indicated to treat<br>Androgenetic Alopecia and to promote hair<br>growth in Females who have Ludwig<br>(Savin) Hair Loss Classification of I to II,<br>and who have a Fitzpatrick Skin Type of<br>I to IV. |
| Indicated Use On | Females | Females |
| Cleared Use for | Prescription Use | Prescription Use |
| Device Type | Scalp Covering Device | Scalp Covering Device |
| Laser Assembly | Red Laser Diodes mounted within a<br>"Concave" scalp covering | Red Laser Diodes mounted within a<br>"Concave" scalp covering |
| Wavelength | 650nm | 650nm |
| Output Beam | Continous Wave "CW" | Continous Wave "CW" |
| Laser Diode Power (each) | <5mW Per IEC 60825-1 ED. 3.0 B:2014 | <5mW Per IEC 60825-1 ED. 3.0 B:2014 |
| Number of Laser<br>Diodes/Optics | Eighty (80) for Females | Eighty (80) for Females |
| Total Laser Energy Output | <400mW for Female use | <400mW for Female use |
| Light Energy Type | Coherent Laser | Coherent Laser |
| Laser Field | Entire Scalp Area | Entire Scalp Area |
| Laser Diode Array | Linear Rows within a<br>"Concave" scalp covering | Linear Rows within a<br>"Concave" scalp covering |
| Total Treatment Time | 30 minutes | 30 minutes |
| Device Application | Portable | Portable |
| Consumer Usage<br>Focal Length | 0.5" - 1.5" from Emanating Diode Location | 0.5" - 1.5" from Emanating Diode Location |
| Ambient Operating<br>Temperature | 50° to 86°F (10° to 30°C) | 32° to 104°F (0° to 40°C) |
| Safety | CE Mark | CE Mark |
| User Manual | Yes | Yes |
{6}------------------------------------------------
| Substantial Equivalence Comparison Table | | |
|------------------------------------------|--|--|
|------------------------------------------|--|--|
| | New Device | Predicate Device |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K152587 | K150613 |
| Company | DermaScalp LLC | Transdermal Inc |
| Name | DermaScalp MDX Laser Cap | LaserCap PRO |
| Indications for Use | The DermaScalp MDX Laser Cap is<br>indicated to treat Androgenetic Alpopecia<br>and promote hair growth in Females who<br>have Ludwig (Savin) Hair Loss<br>Classification of I-II patterns of hair loss<br>and Fitzpatrick Skin Types I to IV. | LaserCap PRO is indicated to treat<br>Androgenetic Alopecia and promote hair<br>growth in Females who have Ludwig<br>(Savin) Hair Loss Classification of I to II,<br>and who have been determined to have a<br>Fitzpatrick Skin Type of I to IV. |
| Indicated Use On | Females | Females |
| Cleared Use for | Prescription Use | Prescription Use |
| Device Type | Scalp Covering Device | Scalp Covering Device |
| Laser Assembly | Red Laser Diodes mounted within a<br>"Concave" scalp covering | Red Laser Diodes mounted within a<br>"Concave" scalp covering |
| Wavelength | 650nm | 650nm |
| Output Beam | Continous Wave "CW" | Continous Wave "CW" |
| Laser Diode Power (each) | <5mW Per IEC 60825-1 ED. 3.0 B:2014 | <5mW Per IEC 60825-1 ED. 3.0 B:2014 |
| Number of Laser<br>Diodes/Optics | Two hundred twenty-four (224)<br>for Females | Two hundred twenty-four (224)<br>for Females |
| Total Laser Energy Output | <1.120mW for Female use | <1,120mW for Female use |
| Light Energy Type | Coherent Laser | Coherent Laser |
| Laser Field | Entire Scalp Area | Entire Scalp Area |
| Laser Diode Array | Linear Rows within a<br>"Concave" scalp covering | Linear Rows within a<br>"Concave" scalp covering |
| Total Treatment Time | 30 minutes | 30 minutes |
| Device Application | Portable | Portable |
| Consumer Usage<br>Focal Length | 0.5" - 1.5" from Emanating Diode Location | 0.5" - 1.5" from Emanating Diode Location |
| Ambient Operating<br>Temperature | 50° to 86°F (10° to 30°C) | 32° to 104°F (0° to 40°C) |
| Safety | CE Mark | CE Mark |
| User Manual | Yes | Yes |
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.