K150613 · Transdermal Cap, Inc. · OAP · Jul 22, 2015 · Physical Medicine
Device Facts
Record ID
K150613
Device Name
LaserCap LC Pro and LC Elite
Applicant
Transdermal Cap, Inc.
Product Code
OAP · Physical Medicine
Decision Date
Jul 22, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LCPro and LCElite are indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I - II and Fitzpatrick Classification of Skin Phototypes I to IV.
Device Story
LaserCap LCPro and LCElite are helmet-based low-level laser therapy (LLLT) devices for home use. LCPro utilizes 224 pulsed-emission diode lasers; LCElite utilizes 80 continuous-emission diode lasers; both operate at 650 nm. Devices cover the upper 1/3 of the head. Integrated safety interlock pauses therapy if the head moves outside the radiation zone and resumes upon repositioning. Audible tones signal therapy start (2 beeps) and end (one long beep). Powered by rechargeable Lithium Ion battery packs. Prescribed by physicians to promote hair growth in females with androgenetic alopecia. Benefits include non-invasive hair growth stimulation.
Clinical Evidence
No clinical trial data submitted. Bench testing only, including conformance to IEC 60825-1, IEC 60601-1-11, IEC 60601-1-2, and IEC 62133.
Technological Characteristics
Helmet-based LLLT system. LCPro: 224 pulsed-emission diode lasers (650 nm). LCElite: 80 continuous-emission diode lasers (650 nm). Powered by rechargeable Lithium Ion battery packs. Safety interlock for head positioning. Audible tone feedback. Standards: IEC 60825-1 (laser safety), IEC 60601-1-11 (home healthcare), IEC 60601-1-2 (EMC), IEC 62133 (battery safety).
Indications for Use
Indicated for females with androgenetic alopecia, Ludwig-Savin Classifications I-II, and Fitzpatrick skin phototypes I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around a central emblem. The emblem consists of a stylized image of an eagle or bird with three human profiles nested within its wing or body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Transdermal Cap Incorporated % Mr. Raymond R. Blanche NST Consulting, LLC 641 Shunpike Road, Suite 311 Chatham, New Jersey 07928
July 22, 2015
Re: K150613
Trade/Device Name: LaserCap LCPro and LCElite Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP Dated: June 17, 2015 Received: June 19, 2015
Dear Mr. Blanche:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours.
Joshua C. Nipper -S
For
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150613
Device Name LaserCap LCPro and LCElite
Indications for Use (Describe)
The LC PRO and LC ELITE are indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications I-II and Fitzpatrick Classification of skin phototypes of I-IV.
Type of Use (Select one or both, as applicable)
| ☑ Export under an IDP with a DE1191 Statement. Please see instructions. | ☐ This is a one-time-only Export. |
|-------------------------------------------------------------------------|-----------------------------------|
|-------------------------------------------------------------------------|-----------------------------------|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Transdermal Cap, Inc.
# Submitter's Contact Information
| Name: | Raymond R. Blanche |
|--------------------------|--------------------------------------------------------------------------|
| Address | NST Consulting, LLC<br>641 Shunpike Road, Suite 311<br>Chatham, NJ 07928 |
| Telephone:<br>Facsimile: | (973) 539-7444<br>(973) 539-7445 |
## Name of Device and Name/Address of Sponsor
| Trade Name: | Lasercap, LCPro and LCElite |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor Contact<br>Information: | Michael Rabin, MD<br>Transdermal Cap, Inc.<br>3615 Superior Avenue, Suite 3104F<br>Cleveland, Ohio 44114<br>T. 877-711-4769<br>F. 860-201-1846 |
| Common or Usual Name: | Lamp, non-heating, for promotion of hair growth |
| Classification Name: | Infrared lamp per 21 CFR 890.5500 |
Classification Code: OAP (Laser, comb, hair)
# Predicate Devices:
| Device Trade Name | Manufacturer |
|-------------------------------|-----------------|
| Capillus272 Pro – K143199 | Capillus, LLC |
| Theradome LH 80 Pro – K113097 | Theradome, Inc. |
| Reference Devices: | |
| Hairmax Lasercomb - K142573 | Lexington Inter |
Date Prepared: July 16, 2015 Revised 。
『
rnational
{4}------------------------------------------------
### Intended Use / Indications for Use
The LCPro and LCElite are indicated to promote hair growth in females with androgenetic alopecia who have Ludwig-Savin Classifications of I - II and Fitzpatrick Classification of Skin Phototypes I to IV.
### Technological Characteristics
The LCPro consists of 224 red, visible-light, pulsed-emission diode lasers operating at 650 nanometers, that are configured within a protective inner liner and outer helmet. The LCElite is physically similar to the LCPro, except that it is configured with 80 red, visible-light, continuous-emission diode lasers, also operating at 650 nanometers and also configured within a protective inner liner and outer helmet. The use of these specific number of diode lasers provides for a full coverage of the upper 1/3 of the head; i.e., the area commonly covered with stylized hair. The helmet system will automatically pause therapy if the subject's head is moved outside of the zone of radiation and will resume therapy when the correct head position is reestablished. This is achieved by a safety interlock. At the beginning and end of a therapy session, audible tones are delivered to the user, indicating that therapy has begun (2 beeps) or ended (one long beep). The system is powered by rechargeable Lithium Ion battery cells assembled into a proprietary battery pack. Both the battery cells pack and charger are fully compliant to recognized, international standards.
{5}------------------------------------------------
### Performance Data:
No Clinical trial data for LCPro and LCElite were submitted for this 510(k). The LCPro and LCElite were tested for standards conformance with IEC 60825-1, IEC 60601-1-11, IEC 60601-1-2, and IEC 62133.
## Substantial Equivalence
The LCPro and LCElite utilize the same technological characteristics as the predicate devices. The LCPro uses the same pulsing characteristics for its laser diodes as the K143199 (Capillus272 Pro) device, and provides similar energy and irradiance output to the treatment area as this predicate. The LCElite uses continuous-output laser diodes as the K113097 (Theradome LH80 Pro) device does, and provides similar energy and irradiance output to the treatment area as this predicate The LCPro contains 224 lasers and the LCElite contains 80. The target audience of these devices is the same. The two offerings are presented to offer the prescribing physician an economic alternative for his patients.
The LCPro and LCElite are designed to provide low-level-laser therapy. Both predicate devices and the proposed devices use red light diode lasers. The sponsor believes that the differences between the subject devices of this 510(k) compared with the predicate devices, do not substantially affect the therapeutic value or the safety profile. There are differences in the physical appearances of the LCPro and LCElite, compared to the predicates.
### Conclusion
The LCPro and LCElite utilize the same technological characteristics as the predicate devices listed above, and are to be used for the same intended use. The sponsor believes that there are no new types of safety and effectiveness questions for the LCPro and LCElite when compared with
{6}------------------------------------------------
the predicate devices, and therefore that the devices should be considered substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.